| Literature DB >> 33580853 |
Barbara Ladisich1, Lukas Machegger2, Alexander Romagna1,3, Herbert Krainz1, Jürgen Steinbacher2, Markus Leitinger4, Gudrun Kalss4, Niklas Thon5, Eugen Trinka4, Peter A Winkler1, Christoph Schwartz6.
Abstract
BACKGROUND: Stereoelectroencephalography (SEEG) allows the identification of deep-seated seizure foci and determination of the epileptogenic zone (EZ) in drug-resistant epilepsy (DRE) patients. We evaluated the accuracy and treatment-associated morbidity of frameless VarioGuide® (VG) neuronavigation-guided depth electrode (DE) implantations.Entities:
Keywords: Accuracy; Clinical outcome; Depth electrode; Drug-resistant epilepsy; Epilepsy surgery; Pre-surgical evaluation; SEEG
Mesh:
Year: 2021 PMID: 33580853 PMCID: PMC8053662 DOI: 10.1007/s00701-021-04755-w
Source DB: PubMed Journal: Acta Neurochir (Wien) ISSN: 0001-6268 Impact factor: 2.216
Fig. 1Exemplary accuracy measurement of centre point deviation. The Euclidean distance from the planned preoperative trajectory to the placed electrode in postoperative MRI is demonstrated for the centre point (CP) of a single electrode
Patient population and surgery
| Patients ( | |
| Male:female ratio | 1:1.4 |
| Mean age (SD) (years) | 31.9 (9.3) |
| Median age (range) (years) | 32.0 (20.0–54.0) |
| Implanted lobes/regions | |
| Frontal ( | 8 |
| Central ( | 1 |
| Parietal ( | 14 |
| Temporal ( | 11 |
| Insular ( | 9 |
| Limbic ( | 12 |
| Occipital ( | 8 |
| Surgeries ( | |
| Dominant hemisphere implantation (%) | 9 (42.9) |
| Median number of DE/surgery (range) | 12.0 (3.0–16.0) |
| Mean number of DE/surgery (SD) | 10.5 (4.1) |
| Median duration of surgery (range) (min) | 196.0 (50.0–360.0) |
| Mean duration of surgery (SD) (min) | 197.3 (59.7) |
DE depth electrode, EA entry angle, SD standard deviation
Accuracy assessment
| Implantation and electrode characteristics ( | |
| Mean DE length (SD) (mm) | 49.6 (11.9) |
| Median DE length (range) (mm) | 49.3 (25.1–93.8) |
| Mean EA (SD) (°) | 24.6 (13.3) |
| Median EA (range) (°) | 23.6 (1.0–74.8) |
| Mean bone thickness (SD) (mm) | 9.0 (3.1) |
| Median bone thickness (range) (mm) | 9.0 (2.0–21.0) |
| Deviation analyses | |
| Mean EP deviation (SD) mm | 3.2 (2.4) |
| Median EP deviation (range) mm | 2.8 (0.1–27.3) |
| Mean CP deviation (SD) mm | 3.0 (2.2) |
| Median CP deviation (range) mm | 2.5 (0.1–25.1) |
| Mean TP deviation (SD) mm | 2.7 (2.0) |
| Median TP deviation (range) mm | 2.4 (0.0–19.4) |
| Mean ED length deviation (SD) mm | 1.9 (1.8) |
| Median ED length deviation (range) mm | 1.6 (0.0–18.4) |
CP centre point, DE depth electrode, EA entry angle, SD standard deviation, TP target point
Fig. 2Target point deviations of all performed SEEG procedures
Fig. 3Measured EP, CP and TP deviations for an exemplary patient
Statistical analysis
| Correlations— | |||
|---|---|---|---|
| Analysed factor | EP deviation (continuous variable) | CP deviation (continuous variable) | TP deviation (continuous variable) |
| DE length (continuous variable, Spearman’s correlation) | 0.44 | 0.22 | 0.01 |
| Entry angle (continuous variable, Spearman’s correlation) | 0.84 | 0.93 | 0.51 |
| Entry angle (categorized variable, Mann-Whitney | 0.74 | 0.46 | 0.47 |
| Bone thickness (continuous variable, Spearman’s correlation) | 0.35 | 0.68 | 0.36 |
| Bone thickness (categorised variable, Mann-Whitney- | 0.77 | 0.27 | 0.17 |
| Lobe (categorised variable, Kruskal-Wallis test) | 0.20 | 0.15 | 0.25 |
CP centre point, DE depth electrode, EP entry point, TP target point