BACKGROUND: Electrode placement in epilepsy surgery seeks to locate the sites of ictal onset and early propagation. An invasive diagnostic procedure, stereoelectroencephalography (SEEG) is usually implemented with frame-based methods that can be especially problematic in young children.
OBJECTIVE: To evaluate the feasibility and accuracy of a new technique for frameless SEEG in children using the VarioGuide® system (Brainlab AG, München, Germany).
METHODS: A frameless stereotactic navigation system was used to implant depth electrodes with percutaneous drilling and bolt insertion in pediatric patients with medically refractory epilepsy. Data on general demographic information of electrode implantation, duration, number, and complications were retrospectively collected. To determine the placement accuracy of the VarioGuide® frameless system, the mean Euclidean distances were calculated by comparing the preoperatively planned trajectories with the final electrode position observed on postoperative computed tomography scans.
RESULTS: From May 2011 to December 2015, 15 patients (8 males, 7 females; mean age: 8 yr, range: 3-16 yr) underwent SEEG depth electrode implantation of a total of 111 electrodes. The mean error measured by the Euclidean distance from the center of the entry point to the intended entry point was 3.64 ± 1.78 mm (range: 0.58-7.59 mm) and the tip of the electrode to the intended target was 2.96 ± 1.49 mm (range: 0.58-7.82 mm). There were no significant complications.
CONCLUSION: Depth electrodes can be placed safely and accurately in children using the VarioGuide® frameless stereotactic navigation system.

Entities:  

Mesh:

Year:  2018        PMID: 28961938     DOI: 10.1093/ons/opx192

Source DB:  PubMed          Journal:  Oper Neurosurg (Hagerstown)        ISSN: 2332-4252            Impact factor:   2.703


  3 in total

1.  Stereoelectroencephalography in the very young: Case report.

Authors:  Joshua Katz; Caren Armstrong; Svetlana Kvint; Benjamin C Kennedy
Journal:  Epilepsy Behav Rep       Date:  2022-05-18

2.  Technical Aspects of SEEG and Its Interpretation in the Delineation of the Epileptogenic Zone.

Authors:  Hui Ming Khoo; Jeffery A Hall; Francois Dubeau; Naoki Tani; Satoru Oshino; Yuya Fujita; Jean Gotman; Haruhiko Kishima
Journal:  Neurol Med Chir (Tokyo)       Date:  2020-11-06       Impact factor: 1.742

3.  VarioGuide® frameless neuronavigation-guided stereoelectroencephalography in adult epilepsy patients: technique, accuracy and clinical experience.

Authors:  Barbara Ladisich; Lukas Machegger; Alexander Romagna; Herbert Krainz; Jürgen Steinbacher; Markus Leitinger; Gudrun Kalss; Niklas Thon; Eugen Trinka; Peter A Winkler; Christoph Schwartz
Journal:  Acta Neurochir (Wien)       Date:  2021-02-13       Impact factor: 2.216

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.