Head-to-head comparison of SARS-CoV-2 antigen-detecting rapid test with professional-collected nasal versus nasopharyngeal swab.

Andreas K Lindner1,2, Olga Nikolai1,2, Chiara Rohardt1, Susen Burock3, Claudia Hülso1, Alisa Bölke1, Maximilian Gertler1, Lisa J Krüger4, Mary Gaeddert4, Frank Tobian4, Federica Lainati4, Joachim Seybold5, Terry C Jones6,7,8, Jörg Hofmann9, Jilian A Sacks10, Frank P Mockenhaupt1,2, Claudia M Denkinger4,11,2.   

Abstract

Entities:  

Year:  2021        PMID: 33574072      PMCID: PMC7877326          DOI: 10.1183/13993003.04430-2020

Source DB:  PubMed          Journal:  Eur Respir J        ISSN: 0903-1936            Impact factor:   16.671


× No keyword cloud information.
To the Editor: Antigen-detecting rapid diagnostic tests (Ag-RDTs) are likely to play a substantial role in innovative testing strategies for severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2) [1, 2]. Currently, most Ag-RDTs require nasopharyngeal (NP) sampling performed by qualified healthcare professionals. Nasal sampling would enable scaling of antigen testing strategies. The term nasal sampling is often not used uniformly, but can be differentiated as either anterior nasal sampling (entire absorbent tip of the swab, usually 1 to 1.5 cm, inserted into nostril), and nasal mid-turbinate (as described below) [3]. We conducted a prospective diagnostic accuracy study with the objective to directly compare the performance of professional-collected nasal mid-turbinate (NMT) versus NP swab, using a World Health Organization (WHO)-listed SARS-CoV-2 Ag-RDT. The reference standard was RT-PCR collected from a combined NP/oropharyngeal (OP) swab. The study was continued until 30 positive NP swab samples according to Ag-RDT were obtained, which is the minimum recommended by the WHO Emergency Use Listing Procedure to demonstrate sample type equivalency [4]. This manufacturer-independent study was conducted in partnership with the Foundation of Innovative New Diagnostics, the WHO collaborating centre for coronavirus disease 2019 (COVID-19) diagnostics. Adults at high risk for SARS-CoV-2 infection according to clinical suspicion who attended the ambulatory SARS-CoV-2 testing facility of Charité University Hospital Berlin, Germany, were enrolled from 11 to 18 November 2020. Participants were excluded if either of the swabs for the Ag-RDT or the RT-PCR reference standard could not be collected. Participants had to blow once the nose with a tissue. Afterwards, a NMT sample was collected on both sides of the nose, using the specific nasal swab provided in the test kit of the manufacturer, according to the instructions for use, which also correspond to the US Centers for Disease Control and Prevention instructions [3]. Briefly, while tilting the patient's head back 70 degrees, the swab was inserted about 2 cm into each nostril, parallel to the palate until resistance was met at turbinates, then rotated 3–4 times against the nasal walls. Subsequently, a separate NP-swab (provided in the manufacturer test kit) for the Ag-RDT and a combined OP/NP-swab (eSwab from Copan placed in 1 mL Amies medium) as per institutional recommendations for RT-PCR were taken from different sides of the nose. The Ag-RDT evaluated was the STANDARD Q COVID-19 Ag Test (SD Biosensor, Inc., Gyeonggi-do, Korea; henceforth called STANDARD Q) [5]. Study procedures followed the same process as described in the prior study by Lindner et al. [6]. While the test is commercially available as NP sampling kit, the nasal sampling kit is currently available for “research use only” by the manufacturer. The instructions for use of the two test kits showed differences, with a more elaborate extraction process (stirring the swab at least 10 versus five times) and a higher volume of extracted specimen (four versus three drops) used for testing of nasal samples. Of 181 patients invited, 180 (99.4%) consented to participate. One patient was excluded as both swabs for the Ag-RDT could not be obtained. The mean±sd age of participants was 36.2±12.2 years, with 48.0% female and 14.5% having comorbidities. On the day of testing, 96.1% of participants had one or more symptoms consistent with COVID-19. Duration of symptoms at the time of presentation on average was 4.2±2.6 days. Among the 179 participants, 41 (22.9%) tested positive for SARS-CoV-2 by RT-PCR (table 1).
TABLE 1

Antigen-detecting rapid diagnostic test (Ag-RDT) results with a professional-collected nasal mid-turbinate (NMT) swab and nasopharyngeal (NP) swab in RT-PCR positive patients from combined NP/oropharyngeal (OP) swab

NumberNMT swab SD Q Ag-RDTNP swab SD Q Ag-RDTOP/NP swab RT-PCRSymptom duration days
CT valueViral load+
1pos. (+++)pos. (+++)16.419.11
2pos. (++)pos. (+++)19.25#,§7
3pos. (+++)pos. (+++)19.82#8.854
4pos. (++)pos. (+++)17.688.731
5pos. (+++)pos. (+++)17.828.693
6pos. (+++)pos. (+++)18.578.464
7pos. (+++)pos. (+++)22.09#8.174
8pos. (++)pos. (+++)19.968.058
9pos. (++)pos. (+++)22.70#7.995
10pos. (++)pos. (++)20.747.823
11pos. (+++)pos. (+++)23.54#7.753
12pos. (+++)pos. (+++)24.09#7.583
13pos. (+++)pos. (+++)24.37#7.502
14pos. (+++)pos. (++)21.837.505
15pos. (+)pos. (+++)24.41#7.492
16pos. (++)pos. (+++)24.62#7.436
17pos. (+)neg.22.427.3210
18pos. (+++)pos. (+++)25.38#7.204
19pos. (+++)pos. (++)22.857.195
20pos. (++)pos. (+++)23.057.135
21pos. (++)neg.26.48#6.878
22pos. (+++)pos. (++)24.136.813
23pos. (+++)pos. (+++)24.636.662
24pos. (+++)pos. (++)24.716.643
25pos. (+++)pos. (++)27.84#6.472
26neg.neg.25.556.394
27pos. (++)pos. (+++)28.82#6.185
28pos. (+++)pos. (++)26.646.073
29neg.neg.26.796.023
30pos. (+++)pos. (++)29.87#5.874
31neg.neg.30.91#5.567
32pos. (+++)pos. (+++)29.315.275
33neg.neg.29.845.124
34pos. (+)neg.33.05#4.938
35pos. (++)pos. (+)31.564.615
36neg.neg.34.24#4.586
37neg.neg.34.52#4.509
38pos. (+)pos. (++)34.69#4.44
39pos. (+)neg.35.16#4.3110
40neg.neg.34.583.717
41neg.pos. (+)35.533.438
Sensitivity33/41 (80.5%)30/41 (73.2%)
Positive percent agreementƒ93.5% (95% CI 79.3–98.2%)

Cycle threshold (CT) values and viral load (in descending order) of the paired RT-PCR samples are shown, as well as the duration of symptoms per patient. The positive percent agreement between NMT and NP samples on Ag-RDT, and the respective sensitivities compared to RT-PCR are shown. SD Q: STANDARD Q COVID-19 Ag Test (SD Biosensor); neg.: negative; pos (+): weak positive; pos. (++): positive; pos. (+++): strong positive. #: Roche Cobas SARS-CoV-2 assay (E-gene, T2 target); ¶: TibMolbiol assay, E-gene target; +: log10 SARS-CoV-2 RNA copies/swab; §: only T1 CT value, T2 invalid due to suspected mutation within oligonucleotide binding region, VL not specified; ƒ: including one false-positive on NMT and one on NP.

Antigen-detecting rapid diagnostic test (Ag-RDT) results with a professional-collected nasal mid-turbinate (NMT) swab and nasopharyngeal (NP) swab in RT-PCR positive patients from combined NP/oropharyngeal (OP) swab Cycle threshold (CT) values and viral load (in descending order) of the paired RT-PCR samples are shown, as well as the duration of symptoms per patient. The positive percent agreement between NMT and NP samples on Ag-RDT, and the respective sensitivities compared to RT-PCR are shown. SD Q: STANDARD Q COVID-19 Ag Test (SD Biosensor); neg.: negative; pos (+): weak positive; pos. (++): positive; pos. (+++): strong positive. #: Roche Cobas SARS-CoV-2 assay (E-gene, T2 target); ¶: TibMolbiol assay, E-gene target; +: log10 SARS-CoV-2 RNA copies/swab; §: only T1 CT value, T2 invalid due to suspected mutation within oligonucleotide binding region, VL not specified; ƒ: including one false-positive on NMT and one on NP. No invalid Ag-RDT results were observed on either NMT or NP samples. Four patients tested positive by NMT but not by NP sampling. One patient was positive by NP sampling only. The positive percent agreement was 93.5% (95% CI 79.3–98.2%), including one false positive result with NMT and one with NP. The negative percent agreement was 95.9% (95% CI 91.4–98.1%). Inter-rater reliability was high (kappa 0.95 for NMT; 0.98 for NP). In the semi-quantitative read-out of the test band intensity in double positive pairs, there was no remarkable difference (eight higher on NMT, nine higher on NP). A third reader was necessary for the agreement on the results of three tests for which the test band was very weak. The STANDARD Q Ag-RDT with NMT sampling showed a sensitivity of 80.5% (33/41 PCR positives detected; 95% CI 66.0–89.8%) and specificity of 98.6% (95% CI 94.9–99.6%) compared to RT-PCR. The sensitivity with NP sampling was 73.2% (30/41 PCR positives detected; 95% CI 58.1–84.3%) and specificity was 99.3% (95% CI 96.0–100%). In patients with high viral load (>7.0 log10 SARS-CoV2 RNA copies per swab), the sensitivity of the Ag-RDT with NMT sampling was 100% (19/19 PCR positives detected; 95% CI 83.9–100%) and 94.7% (18/19 PCR positives detected; 95% CI 76.4–99.7%) with NP sampling. In contrast, the Ag-RDT more frequently did not detect patients with lower viral load or with symptoms >7 days (table 1), as commonly observed in studies on Ag-RDTs [7, 8]. The strengths of the study are the standardised sampling methods, two independent blinded readers and an additional semi-quantitative assessment of Ag-RDT results. The cohort was representative, judging from the comparable sensitivity observed in the recent independent validation study of STANDARD Q (sensitivity 76.6%; 95% CI 62.8–86.4%) [9]. The study is limited as it was performed in a single centre. Theoretically, the previous NMT sample collection could have negatively influenced the test result of the NP sample in patients with a low viral load. In conclusion, this study demonstrates that sensitivity of a WHO-listed SARS-CoV-2 Ag-RDT using professional nasal sampling kit is at least equal to that of NP sampling kit, although confidence intervals overlap. Of note, differences in the instructions for use of the test procedures could have contributed to different sensitivities. NMT sampling can be performed with less training, reduces patient discomfort, and enables scaling of antigen testing strategies. Additional studies of patient self-sampling should be considered to further facilitate scale-up of Ag-RDT testing [6]. This one-page PDF can be shared freely online. Shareable PDF ERJ-04430-2020.Shareable
  2 in total

1.  Head-to-head comparison of SARS-CoV-2 antigen-detecting rapid test with self-collected nasal swab versus professional-collected nasopharyngeal swab.

Authors:  Andreas K Lindner; Olga Nikolai; Franka Kausch; Mia Wintel; Franziska Hommes; Maximilian Gertler; Lisa J Krüger; Mary Gaeddert; Frank Tobian; Federica Lainati; Lisa Köppel; Joachim Seybold; Victor M Corman; Christian Drosten; Jörg Hofmann; Jilian A Sacks; Frank P Mockenhaupt; Claudia M Denkinger
Journal:  Eur Respir J       Date:  2021-04-15       Impact factor: 16.671

2.  Diagnostic accuracy of two commercial SARS-CoV-2 antigen-detecting rapid tests at the point of care in community-based testing centers.

Authors:  Alice Berger; Marie Therese Ngo Nsoga; Francisco Javier Perez-Rodriguez; Yasmine Abi Aad; Pascale Sattonnet-Roche; Angèle Gayet-Ageron; Cyril Jaksic; Giulia Torriani; Erik Boehm; Ilona Kronig; Jilian A Sacks; Margaretha de Vos; Frédérique Jacquerioz Bausch; François Chappuis; Adriana Renzoni; Laurent Kaiser; Manuel Schibler; Isabella Eckerle
Journal:  PLoS One       Date:  2021-03-31       Impact factor: 3.752

  2 in total
  13 in total

1.  Evaluation of the Roche antigen rapid test and a cell culture-based assay compared to rRT- PCR for the detection of SARS-CoV-2: A contribution to the discussion about SARS-CoV-2 diagnostic tests and contagiousness.

Authors:  Jacqueline Steinlin-Schopfer; Maria Teresa Barbani; Richard Kamgang; Martina Zwahlen; Franziska Suter-Riniker; Ronald Dijkman
Journal:  J Clin Virol Plus       Date:  2021-05-09

Review 2.  Rapid, point-of-care antigen tests for diagnosis of SARS-CoV-2 infection.

Authors:  Jacqueline Dinnes; Pawana Sharma; Sarah Berhane; Susanna S van Wyk; Nicholas Nyaaba; Julie Domen; Melissa Taylor; Jane Cunningham; Clare Davenport; Sabine Dittrich; Devy Emperador; Lotty Hooft; Mariska Mg Leeflang; Matthew Df McInnes; René Spijker; Jan Y Verbakel; Yemisi Takwoingi; Sian Taylor-Phillips; Ann Van den Bruel; Jonathan J Deeks
Journal:  Cochrane Database Syst Rev       Date:  2022-07-22

Review 3.  Performance of Antigen Detection Tests for SARS-CoV-2: A Systematic Review and Meta-Analysis.

Authors:  Anastasia Tapari; Georgia G Braliou; Maria Papaefthimiou; Helen Mavriki; Panagiota I Kontou; Georgios K Nikolopoulos; Pantelis G Bagos
Journal:  Diagnostics (Basel)       Date:  2022-06-04

4.  Accuracy of rapid point-of-care antigen-based diagnostics for SARS-CoV-2: An updated systematic review and meta-analysis with meta-regression analyzing influencing factors.

Authors:  Lukas E Brümmer; Stephan Katzenschlager; Sean McGrath; Stephani Schmitz; Mary Gaeddert; Christian Erdmann; Marc Bota; Maurizio Grilli; Jan Larmann; Markus A Weigand; Nira R Pollock; Aurélien Macé; Berra Erkosar; Sergio Carmona; Jilian A Sacks; Stefano Ongarello; Claudia M Denkinger
Journal:  PLoS Med       Date:  2022-05-26       Impact factor: 11.613

5.  A prospective diagnostic evaluation of accuracy of self-taken and healthcare worker-taken swabs for rapid COVID-19 testing.

Authors:  Helen R Savage; Lorna Finch; Richard Body; Rachel L Watkins; Gail Hayward; Eloïse Cook; Ana I Cubas-Atienzar; Luis E Cuevas; Peter MacPherson; Emily R Adams
Journal:  PLoS One       Date:  2022-06-30       Impact factor: 3.752

6.  Head-to-head performance comparison of self-collected nasal versus professional-collected nasopharyngeal swab for a WHO-listed SARS-CoV-2 antigen-detecting rapid diagnostic test.

Authors:  Julian A F Klein; Lisa J Krüger; Frank Tobian; Mary Gaeddert; Federica Lainati; Paul Schnitzler; Andreas K Lindner; Olga Nikolai; B Knorr; A Welker; Margaretha de Vos; Jilian A Sacks; Camille Escadafal; Claudia M Denkinger
Journal:  Med Microbiol Immunol       Date:  2021-05-24       Impact factor: 3.402

7.  ESCMID COVID-19 guidelines: diagnostic testing for SARS-CoV-2.

Authors:  Paraskevi C Fragkou; Giulia De Angelis; Giulia Menchinelli; Fusun Can; Federico Garcia; Florence Morfin-Sherpa; Dimitra Dimopoulou; Elisabeth Mack; Adolfo de Salazar; Adriano Grossi; Theodore Lytras; Chrysanthi Skevaki
Journal:  Clin Microbiol Infect       Date:  2022-02-23       Impact factor: 13.310

8.  Diagnosing SARS-CoV-2 with Antigen Testing, Transcription-Mediated Amplification and Real-Time PCR.

Authors:  Sascha Dierks; Oliver Bader; Julian Schwanbeck; Uwe Groß; Michael S Weig; Kemal Mese; Raimond Lugert; Wolfgang Bohne; Andreas Hahn; Nicolas Feltgen; Setare Torkieh; Fenja R Denker; Peer Lauermann; Marcus W Storch; Hagen Frickmann; Andreas Erich Zautner
Journal:  J Clin Med       Date:  2021-05-29       Impact factor: 4.241

9.  Accuracy of novel antigen rapid diagnostics for SARS-CoV-2: A living systematic review and meta-analysis.

Authors:  Lukas E Brümmer; Stephan Katzenschlager; Mary Gaeddert; Christian Erdmann; Stephani Schmitz; Marc Bota; Maurizio Grilli; Jan Larmann; Markus A Weigand; Nira R Pollock; Aurélien Macé; Sergio Carmona; Stefano Ongarello; Jilian A Sacks; Claudia M Denkinger
Journal:  PLoS Med       Date:  2021-08-12       Impact factor: 11.069

10.  Self-testing for the detection of SARS-CoV-2 infection with rapid antigen tests for people with suspected COVID-19 in the community.

Authors:  Joep J J M Stohr; Vivian F Zwart; Gabriel Goderski; Adam Meijer; Carla R S Nagel-Imming; Marjolein F Q Kluytmans-van den Bergh; Suzan D Pas; Femke van den Oetelaar; Marloes Hellwich; Kim H Gan; Ariene Rietveld; Jaco J Verweij; Jean-Luc Murk; Wouter van den Bijllaardt; Jan A J W Kluytmans
Journal:  Clin Microbiol Infect       Date:  2021-08-04       Impact factor: 13.310

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.