Head-to-head comparison of SARS-CoV-2 antigen-detecting rapid test with self-collected nasal swab versus professional-collected nasopharyngeal swab.

Andreas K Lindner1,2, Olga Nikolai1,2, Franka Kausch1, Mia Wintel1, Franziska Hommes1, Maximilian Gertler1, Lisa J Krüger3, Mary Gaeddert3, Frank Tobian3, Federica Lainati3, Lisa Köppel3, Joachim Seybold4, Victor M Corman5,6, Christian Drosten5,6, Jörg Hofmann7, Jilian A Sacks8, Frank P Mockenhaupt1,2, Claudia M Denkinger9,2.   

Abstract

Entities:  

Mesh:

Year:  2021        PMID: 33303544      PMCID: PMC7736752          DOI: 10.1183/13993003.03961-2020

Source DB:  PubMed          Journal:  Eur Respir J        ISSN: 0903-1936            Impact factor:   16.671


× No keyword cloud information.
To the Editor: A number of antigen-detecting rapid diagnostic tests (Ag-RDTs) for severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2) are now commercially available and can result in rapid decisions on patient care, isolation and contact tracing at the point of care [1]. Two Ag-RDTs using nasopharyngeal (NP) swab samples meet World Health Organization (WHO) targets and are now approved through the WHO Emergency Use Listing procedure [2-4]. NP swab samples are frequently perceived as uncomfortable by patients and must be collected by trained healthcare personnel with protective equipment. A more complex sampling technique could also result in an incorrect performance in clinical reality, with a possible consequence for test sensitivity. Evidence supports the use of alternative sampling methods for RT-PCR, including nasal swabs collected by patients, and some tests have received regulatory approval with nasal samples [5, 6]. The term nasal sampling is often not used uniformly but can be differentiated in anterior nasal and nasal mid-turbinate sampling [6]. Considering the ease-of-use of Ag-RDTs, a reliable, simple sampling method would not only allow self-sampling, but may also pave the way for self-testing. The primary objective of this prospective diagnostic accuracy study was a head-to-head comparison (positive and negative percent agreement) of a supervised, self-collected nasal mid-turbinate (NMT) swab sample with a healthcare worker (professional)-collected NP swab sample, using a WHO-listed SARS-CoV-2 Ag-RDT against the reference standard RT-PCR collected from a NP/oropharyngeal (OP) swab. The secondary objective was to assess sensitivity and specificity for different sampling techniques with Ag-RDT. The study was continued until at least 30 positive NP swab samples according to Ag-RDT were obtained, as it was requested by the WHO to approve a novel sample type. This manufacturer-independent study was conducted in partnership with the Foundation of Innovative New Diagnostics (FIND), the WHO collaborating centre for coronavirus disease 2019 (COVID-19) diagnostics. The study protocol was approved by the ethical review committee at Heidelberg University Hospital for the study site in Berlin (registration number S-180/2020). The study took place at the ambulatory SARS-CoV-2 testing facility of Charité University Hospital (Charité-Universitätsmedizin Berlin, Germany) from 23 September to 14 October 2020. The study enrolled adults at high risk for SARS-CoV-2 infection according to clinical suspicion. Participants were excluded if either of the swabs for the Ag-RDT or the RT-PCR reference standard could not be collected. Participants underwent first an instructed, self-collected bilateral NMT swab for the Ag-RDT. Verbal instruction was given to insert the swab horizontally 2–3 cm into the nostril and rotate it for 15 s against the nasal walls on each side. Deviations from the instructed technique were recorded. Subsequently, a combined OP/NP swab (eSwab from Copan with 1mL Amies medium) as per institutional recommendations for RT-PCR, and a separate NP swab for the Ag-RDT were taken from different sides of the nose. The samples for the Ag-RDTs were collected using the swab provided by the manufacturer within the test kit. The Ag-RDT evaluated in this study was the STANDARD Q COVID-19 Ag Test (SD Biosensor, Inc. Gyeonggi-do, Korea; henceforth called STANDARD Q) [7], which is also being distributed by Roche [8]. The test uses the lateral flow assay principle in a cassette-based format with a visual read-out after 15–30 min. The manufacturer's instructions for use were followed. The Ag-RDTs were performed directly after sampling (within 60 min) at point of care by study physicians. The Ag-RDT results were interpreted by two operators, each blinded to the result of the other. The second reader was also blinded to the second Ag-RDT results of individual patients. The visual readout of the Ag-RDT test band was categorised on a semi-quantitative scale as negative, weak positive, positive and strong positive. The Roche Cobas SARS-CoV-2 assay (Pleasanton, CA, USA) or the SARS-CoV-2 E-gene assay from TibMolbiol (Berlin, Germany) were performed for RT-PCR according to routine procedures at the central laboratory. Viral RNA concentrations were calculated using assay specific cycle threshold values, based on external calibration curves [9, 10]. Staff performing the Ag-RDTs were blinded to results of RT-PCR tests and vice versa. Of 303 patients invited, 289 (95.4%) consented to participate. Two patients were excluded as both swabs for the Ag-RDT could not be obtained. The mean±sd age of participants was 34.7±11.0 years; 42.9% were female and 19.0% having comorbidities. On the day of testing, 97.6% of participants had one or more symptoms consistent with COVID-19. Duration of symptoms at the time of presentation on average was 4.4±2.7 days. Among the 289 participants, 39 (13.5%) tested positive for SARS-CoV-2 by RT-PCR (table 1).
TABLE 1

Antigen-detecting rapid diagnostic test (Ag-RDT) results with a supervised self-collected nasal mid-turbinate (NMT) swab and with a professional-collected nasopharyngeal (NP) swab in RT-PCR positive patients from combined NP/oropharyngeal (OP) swab

NumberNMT swab self-collected SD Q Ag-RDTNP swab professional-collected SD Q Ag-RDTOP/NP swab RT-PCRSymptom duration days
CT valueViral load+
1pos. (+++)pos. (+++)17.33#9.592
2pos. (++)pos. (+++)17.86#9.431
3pos. (+++)pos. (+++)18.01#9.381
4pos. (++)pos. (+++)18.31#9.293
5pos. (+++)pos. (+++)18.40#9.273
6pos. (+++)pos. (+++)18.76#9.164
7pos. (+++)pos. (+++)18.77#9.165
8pos. (+++)pos. (+++)18.78#9.165
9pos. (+++)pos. (+++)19.05#9.083
10pos. (+++)pos. (+++)19.40#8.972
11neg.pos. (+++)19.66#8.901
12pos. (+++)pos. (+++)20.32#8.703
13pos. (+++)pos. (+++)17.818.684
14pos. (++)pos. (+++)20.44#8.672
15pos. (++)pos. (++)20.54#8.635
16pos. (+++)pos. (+++)21.09#8.474
17pos. (+++)pos. (+)18.628.444
18pos. (+)pos. (++)21.87#8.247
19pos. (++)pos. (+++)19.348.235
20pos. (++)pos. (+++)22.05#8.192
21pos. (+++)pos. (+++)19.478.196
22pos. (+++)pos. (+++)22.60#8.036
23pos. (+++)pos. (++)23.66#7.716
24pos. (+)pos. (++)26.42#6.905
25pos. (+++)pos. (+++)26.77#6.795
26neg.neg.24.256.7810
27pos. (++)pos. (+++)24.776.624
28pos. (+++)pos. (++)25.296.462
29pos. (+)pos. (++)29.33#6.035
30neg.neg.29.56#5.973
31neg.neg.29.95#5.853
32pos. (+)pos. (+)30.25#5.764
33neg.neg.27.815.728
34pos. (++)pos. (+)31.20#5.488
35neg.pos. (+)31.61#5.3610
36neg.neg.32.58#5.0710
37neg.neg.32.86#4.992
38neg.neg.34.62#4.477
39neg.neg.35.53#4.2014
Sensitivity29/39 (74.4%)31/39 (79.5%)
Positive percent agreement§90.6% (95% CI 75.8–96.8%)

Cycle threshold (CT) values and viral load (in descending order) of the paired RT-PCR samples are shown, as well as the duration of symptoms per patient. The positive percent agreement between NMT and NP samples on Ag-RDT, as well as the respective sensitivities compared to RT-PCR are shown. SD Q: STANDARD Q COVID-19 Ag Test (SD Biosensor); neg.: negative; pos. (+): weak positive; pos. (++): positive; pos. (+++): strong positive. #: Roche Cobas SARS-CoV-2 assay (E-gene, T2 target); ¶: TibMolbiol assay, E-gene target; +: log10 RNA SARS-CoV-2 per swab; §: including two false-positives on NMT and one on NP.

Antigen-detecting rapid diagnostic test (Ag-RDT) results with a supervised self-collected nasal mid-turbinate (NMT) swab and with a professional-collected nasopharyngeal (NP) swab in RT-PCR positive patients from combined NP/oropharyngeal (OP) swab Cycle threshold (CT) values and viral load (in descending order) of the paired RT-PCR samples are shown, as well as the duration of symptoms per patient. The positive percent agreement between NMT and NP samples on Ag-RDT, as well as the respective sensitivities compared to RT-PCR are shown. SD Q: STANDARD Q COVID-19 Ag Test (SD Biosensor); neg.: negative; pos. (+): weak positive; pos. (++): positive; pos. (+++): strong positive. #: Roche Cobas SARS-CoV-2 assay (E-gene, T2 target); ¶: TibMolbiol assay, E-gene target; +: log10 RNA SARS-CoV-2 per swab; §: including two false-positives on NMT and one on NP. No invalid tests were observed on either NMT or NP samples. Two patients were detected by NP but not by NMT sampling. No patient was detected by NMT sampling only. The positive percent agreement was 90.6% (95% CI 75.8–96.8%; including two false-positive results with NMT and one with NP). The negative percent agreement was 99.2% (95% CI 97.2–99.8%). Inter-rater reliability was near perfect with kappa of 0.98 for NMT and 1.0 for NP samples. The semi-quantitative read-out was more often higher for the NP samples (nine higher on NP, four higher on NMT). Of the two patients detected by NP but not by NMT sampling, one patient collected the NMT swab only with gentle rotation, the second presented 10 days post symptom onset with a low viral load (table 1). The STANDARD Q Ag-RDT with NMT sampling showed a sensitivity of 74.4% (29/39 PCR positives detected; 95% CI 58.9–85.4%) and specificity of 99.2% (95% CI 97.1–99.8%) compared to RT-PCR. The sensitivity with NP sampling was 79.5% (31/39 PCR positives detected; 95% CI 64.5–89.2%) and specificity was 99.6% (95% CI 97.8–100%). In patients with high viral load (>7.0 log10 RNA SARS-CoV-2/swab), the sensitivity of the Ag-RDT with NMT sampling was 95.7% (22/23 PCR positives detected; 95% CI 79.0–99.8%) and 100% (23/23 PCR positives detected; CI 85.7–100) with NP sampling. In contrast, the Ag-RDT frequently did not detect patients with lower viral load or with symptoms >7 days (table 1). For most patients, the application of the flexible swab (meant for NP swab collection) in the nose appeared unpleasant due to a tickling sensation and led to frequent sneezing. The strengths of the study are the rigorous methods, including standardised sampling, two independent blinded readers and an additional semi-quantitative assessment of Ag-RDT results. The cohort was representative, judging from the comparable sensitivity observed in the recent independent validation study of STANDARD Q (sensitivity 76.6%; 95% CI 62.8–86.4%) [4]. The study is limited as it was performed in a single centre. Also, the NP swab was usually rotated against the nasopharyngeal wall for less time than recommended by the manufacturer, which may have a negative impact on the sensitivity of the Ag-RDT with NP sampling, but also reflects the difficulty of collection of this sample type. In conclusion, this study demonstrates that supervised self-sampling from the nose is a reliable alternative to professional nasopharyngeal sampling with STANDARD Q. The data will contribute to WHO recommendations for use of this test. Considering the ease of use of Ag-RDTs, self-sampling and potentially patient self-testing at home may be a future use case. If such testing could be repeated frequently and immediately ahead of situations when transmissions are likely to occur, self-testing with Ag-RDTs may have a significant impact on the pandemic. Further implementation studies on optimised self-sampling techniques and swabs (e.g. less flexible sponge swab) and the correct performance/interpretation of the test by patients themselves, are urgently needed to drive self-testing to scale. This one-page PDF can be shared freely online. Shareable PDF ERJ-03961-2020.Shareable
  3 in total

1.  Swabs Collected by Patients or Health Care Workers for SARS-CoV-2 Testing.

Authors:  Yuan-Po Tu; Rachel Jennings; Brian Hart; Gerard A Cangelosi; Rachel C Wood; Kevin Wehber; Prateek Verma; Deneen Vojta; Ethan M Berke
Journal:  N Engl J Med       Date:  2020-06-03       Impact factor: 91.245

2.  Detection of 2019 novel coronavirus (2019-nCoV) by real-time RT-PCR.

Authors:  Victor M Corman; Olfert Landt; Marco Kaiser; Richard Molenkamp; Adam Meijer; Daniel Kw Chu; Tobias Bleicker; Sebastian Brünink; Julia Schneider; Marie Luisa Schmidt; Daphne Gjc Mulders; Bart L Haagmans; Bas van der Veer; Sharon van den Brink; Lisa Wijsman; Gabriel Goderski; Jean-Louis Romette; Joanna Ellis; Maria Zambon; Malik Peiris; Herman Goossens; Chantal Reusken; Marion Pg Koopmans; Christian Drosten
Journal:  Euro Surveill       Date:  2020-01

3.  International external quality assessment for SARS-CoV-2 molecular detection and survey on clinical laboratory preparedness during the COVID-19 pandemic, April/May 2020.

Authors:  Veerle Matheeussen; Victor M Corman; Oliver Donoso Mantke; Elaine McCulloch; Christine Lammens; Herman Goossens; Daniela Niemeyer; Paul S Wallace; Paul Klapper; Hubert Gm Niesters; Christian Drosten; Margareta Ieven
Journal:  Euro Surveill       Date:  2020-07
  3 in total
  53 in total

1.  SARS-CoV-2 testing for public health use: core principles and considerations for defined use settings.

Authors:  Catharina Boehme; Emma Hannay; Rangarajan Sampath
Journal:  Lancet Glob Health       Date:  2021-01-28       Impact factor: 26.763

2.  Evaluation of the Roche antigen rapid test and a cell culture-based assay compared to rRT- PCR for the detection of SARS-CoV-2: A contribution to the discussion about SARS-CoV-2 diagnostic tests and contagiousness.

Authors:  Jacqueline Steinlin-Schopfer; Maria Teresa Barbani; Richard Kamgang; Martina Zwahlen; Franziska Suter-Riniker; Ronald Dijkman
Journal:  J Clin Virol Plus       Date:  2021-05-09

Review 3.  Rapid, point-of-care antigen tests for diagnosis of SARS-CoV-2 infection.

Authors:  Jacqueline Dinnes; Pawana Sharma; Sarah Berhane; Susanna S van Wyk; Nicholas Nyaaba; Julie Domen; Melissa Taylor; Jane Cunningham; Clare Davenport; Sabine Dittrich; Devy Emperador; Lotty Hooft; Mariska Mg Leeflang; Matthew Df McInnes; René Spijker; Jan Y Verbakel; Yemisi Takwoingi; Sian Taylor-Phillips; Ann Van den Bruel; Jonathan J Deeks
Journal:  Cochrane Database Syst Rev       Date:  2022-07-22

Review 4.  Performance of Antigen Detection Tests for SARS-CoV-2: A Systematic Review and Meta-Analysis.

Authors:  Anastasia Tapari; Georgia G Braliou; Maria Papaefthimiou; Helen Mavriki; Panagiota I Kontou; Georgios K Nikolopoulos; Pantelis G Bagos
Journal:  Diagnostics (Basel)       Date:  2022-06-04

5.  Automated amplification-free digital RNA detection platform for rapid and sensitive SARS-CoV-2 diagnosis.

Authors:  Hajime Shinoda; Tatsuya Iida; Asami Makino; Mami Yoshimura; Junichiro Ishikawa; Jun Ando; Kazue Murai; Katsumi Sugiyama; Yukiko Muramoto; Masahiro Nakano; Kotaro Kiga; Longzhu Cui; Osamu Nureki; Hiroaki Takeuchi; Takeshi Noda; Hiroshi Nishimasu; Rikiya Watanabe
Journal:  Commun Biol       Date:  2022-05-26

6.  Antigen Detection Tests for SARS-CoV-2: a systematic review and meta-analysis on real world data.

Authors:  Matteo Riccò; Silvia Ranzieri; Simona Peruzzi; Marina Valente; Federico Marchesi; Nicola Luigi Bragazzi; Davide Donelli; Federica Balzarini; Pietro Ferraro; Vincenza Gianfredi; Carlo Signorelli
Journal:  Acta Biomed       Date:  2022-05-11

7.  High-Frequency Self-Testing by Schoolteachers for Sars-Cov-2 Using a Rapid Antigen Test–Results of the Safe School Hesse study.

Authors:  Sebastian Hoehl; Barbara Schenk; Olga Rudych; Stephan Göttig; Ivo Foppa; Niko Kohmer; Onur Karaca; Tuna Toptan; Sandra Ciesek
Journal:  Dtsch Arztebl Int       Date:  2021-04-09       Impact factor: 5.594

8.  Diagnosing SARS-CoV-2 with Antigen Testing, Transcription-Mediated Amplification and Real-Time PCR.

Authors:  Sascha Dierks; Oliver Bader; Julian Schwanbeck; Uwe Groß; Michael S Weig; Kemal Mese; Raimond Lugert; Wolfgang Bohne; Andreas Hahn; Nicolas Feltgen; Setare Torkieh; Fenja R Denker; Peer Lauermann; Marcus W Storch; Hagen Frickmann; Andreas Erich Zautner
Journal:  J Clin Med       Date:  2021-05-29       Impact factor: 4.241

9.  Accuracy of a Novel SARS-CoV-2 Antigen-Detecting Rapid Diagnostic Test from Standardized Self-Collected Anterior Nasal Swabs.

Authors:  Bilgin Osmanodja; Klemens Budde; Daniel Zickler; Marcel G Naik; Jörg Hofmann; Maximilian Gertler; Claudia Hülso; Heike Rössig; Philipp Horn; Joachim Seybold; Stephanie Lunow; Melanie Bothmann; Astrid Barrera-Pesek; Manuel Mayrdorfer
Journal:  J Clin Med       Date:  2021-05-13       Impact factor: 4.241

10.  Impact of repeated nasal sampling on detection and quantification of SARS-CoV-2.

Authors:  Joshua M Levy; Jennifer K Frediani; Erika A Tyburski; Anna Wood; Janet Figueroa; Russell R Kempker; Paulina A Rebolledo; Mark D Gonzalez; Julie Sullivan; Miriam B Vos; Jared O'Neal; Greg S Martin; Wilbur A Lam; Jesse J Waggoner
Journal:  Sci Rep       Date:  2021-07-21       Impact factor: 4.996

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.