Development of an Interdigitated Electrode-Based Disposable Enzyme Sensor Strip for Glycated Albumin Measurement.

Mika Hatada1, Noya Loew1, Junko Okuda-Shimazaki1, Mukund Khanwalker1, Wakako Tsugawa2, Ashok Mulchandani3, Koji Sode1.   

Abstract

Glycated albumin (GA) is an important glycemic control marker for diabetes mellitus. This study aimed to develop a highly sensitive disposable enzyme sensor strip for GA measurement by using an interdigitated electrode (IDE) as an electrode platform. The superior characteristics of IDE were demonstrated using one microelectrode of the IDE pair as the working electrode (WE) and the other as the counter electrode, and by measuring ferrocyanide/ferricyanide redox couple. The oxidation current was immediately reached at the steady state when the oxidation potential was applied to the WE. Then, an IDE enzyme sensor strip for GA measurement was prepared. The measurement of fructosyl lysine, the protease digestion product of GA, exhibited a high, steady current immediately after potential application, revealing the highly reproducible measurement. The sensitivity (2.8 nA µM-1) and the limit of detection (1.2 µM) obtained with IDE enzyme sensor strip were superior compared with our previously reported sensor using screen printed electrode. Two GA samples, 15 or 30% GA, corresponding to healthy and diabetic levels, respectively, were measured after protease digestion with high resolution. This study demonstrated that the application of an IDE will realize the development of highly sensitive disposable-type amperometric enzyme sensors with high reproducibility.

Entities:  

Keywords:  diabetes mellitus; disposable enzyme sensor; fructosyl amino acid oxidase; glycated albumin; interdigitated electrode; point of care testing

Mesh:

Substances:

Year:  2021        PMID: 33572552      PMCID: PMC7866809          DOI: 10.3390/molecules26030734

Source DB:  PubMed          Journal:  Molecules        ISSN: 1420-3049            Impact factor:   4.411


  29 in total

1.  Modification of reduced graphene/Au-aptamer to develop an electrochemical based aptasensor for measurement of glycated albumin.

Authors:  Azad Farzadfard; Javad Shabani Shayeh; Mehran Habibi-Rezaei; Meysam Omidi
Journal:  Talanta       Date:  2020-01-07       Impact factor: 6.057

2.  Sensitive Detection of Glycated Albumin in Human Serum Albumin Using Electrochemiluminescence.

Authors:  Yuki Inoue; Mikako Inoue; Masato Saito; Hiroyuki Yoshikawa; Eiichi Tamiya
Journal:  Anal Chem       Date:  2017-05-16       Impact factor: 6.986

3.  Interfacial capacitance immunosensing using interdigitated electrodes: the effect of insulation/immobilization chemistry.

Authors:  F Rafael Castiello; James Porter; Paresa Modarres; Maryam Tabrizian
Journal:  Phys Chem Chem Phys       Date:  2019-07-17       Impact factor: 3.676

4.  A facile route to glycated albumin detection.

Authors:  Nadra Bohli; Olivier Meilhac; Philippe Rondeau; Syrine Gueffrache; Laurence Mora; Adnane Abdelghani
Journal:  Talanta       Date:  2018-03-12       Impact factor: 6.057

5.  Lucica GA-L glycated albumin assay kit: a new diagnostic test for diabetes mellitus.

Authors:  Takuji Kohzuma; Masafumi Koga
Journal:  Mol Diagn Ther       Date:  2010-02-01       Impact factor: 4.074

6.  Specific detection of oxytetracycline using DNA aptamer-immobilized interdigitated array electrode chip.

Authors:  Yeon Seok Kim; Javed H Niazi; Man Bock Gu
Journal:  Anal Chim Acta       Date:  2008-12-25       Impact factor: 6.558

7.  Microbead-based electrochemical immunoassay with interdigitated array electrodes.

Authors:  Jennifer H Thomas; Sang Kyung Kim; Peter J Hesketh; H Brian Halsall; William R Heineman
Journal:  Anal Biochem       Date:  2004-05-15       Impact factor: 3.365

Review 8.  Glycated albumin and diabetes mellitus.

Authors:  Norihiro Furusyo; Jun Hayashi
Journal:  Biochim Biophys Acta       Date:  2013-05-12

9.  Electrochemical impedance spectroscopy biosensor with interdigitated electrode for detection of human immunoglobulin A.

Authors:  Ryuzo Ohno; Hitoshi Ohnuki; Huihui Wang; Takuya Yokoyama; Hideaki Endo; Daiju Tsuya; Mitsuru Izumi
Journal:  Biosens Bioelectron       Date:  2012-08-10       Impact factor: 10.618

10.  Electrochemical Detection of C-Reactive Protein in Human Serum Based on Self-Assembled Monolayer-Modified Interdigitated Wave-Shaped Electrode.

Authors:  Somasekhar R Chinnadayyala; Jinsoo Park; Young Hyo Kim; Seong Hye Choi; Sang-Myung Lee; Won Woo Cho; Ga-Yeon Lee; Jae-Chul Pyun; Sungbo Cho
Journal:  Sensors (Basel)       Date:  2019-12-16       Impact factor: 3.576

View more
  2 in total

Review 1.  Recent Developments in Biomarkers for Diagnosis and Screening of Type 2 Diabetes Mellitus.

Authors:  Margarita Ortiz-Martínez; Mirna González-González; Alexandro J Martagón; Victoria Hlavinka; Richard C Willson; Marco Rito-Palomares
Journal:  Curr Diab Rep       Date:  2022-03-10       Impact factor: 5.430

Review 2.  Current Challenges and Future Trends of Enzymatic Paper-Based Point-of-Care Testing for Diabetes Mellitus Type 2.

Authors:  Margarita Ortiz-Martínez; Raquel Flores-DelaToba; Mirna González-González; Marco Rito-Palomares
Journal:  Biosensors (Basel)       Date:  2021-11-27
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.