Danyao Nie1, Pisong Yan2, Zonghui Yan1, Jing Zhang1, Wenling He1, Ming Li1, Liangnan Sun1, Mingquan Huang1, Xinhua Liu1, Haotian Lin3.
Abstract
BACKGROUND: To evaluate the clinical results and rotational stability of V4c toric implantable collamer lens (TICL, STAAR Surgical Company, Monrovia, CA, USA) in patients with moderate to high myopic astigmatism. Retrospective, interventional case series was performed at Shenzhen Eye Hospital, Shenzhen, Guangdong, China.Entities:
Keywords: Astigmatism; polar value analysis; posterior chamber intraocular lenses; refractive surgery; vector analysis
Year: 2021 PMID: 33569441 PMCID: PMC7867885 DOI: 10.21037/atm-20-7835
Source DB: PubMed Journal: Ann Transl Med ISSN: 2305-5839
Preoperative patient demographic data
| Characteristic | Valuea |
|---|---|
| Age (y) | 25.57±4.51 (18 to 42) |
| Sex, female (n) | 30 (69.8%) |
| Operated eyes (n) | |
| Right | 36 |
| Left | 36 |
| Both | 29 |
| Preoperative spherical refraction (D) | −8.33±2.39 (−14.25 to −4.00) |
| Preoperative cylindrical refraction (D) | −1.91±0.73 (−5.50 to −1.00) |
| Preoperative MRSE (D) | −9.29±2.41 (−15.25 to −4.50) |
| V4c TICL spherical power (D) | −9.64±2.48 (−15.00 to −5.00) |
| V4c TICL cylinder power (D) | −1.80±0.80 (−6.00 to −1.00) |
| V4c TICL SE (D) | −10.51±2.42 (−16.50 to −5.50) |
| V4c TICL size (mm) | 12.85±0.39 (12.10 to 13.70) |
| WTW (mm) | 11.61±0.38 (10.76 to 12.47) |
| ACD (mm) | 3.29±0.28 (2.87 to 4.04) |
| Axial length (mm) | 26.80±1.32 (25.41 to 29.27) |
a, results are expressed as mean ± standard deviation (range) unless otherwise noted. D, diopters; MRSE, manifest refraction spherical equivalent; TICL, toric implantable collamer lens; WTW, white-to-white distance; ACD, anterior chamber depth.
Visual and clinical outcomes in eyes undergoing V4c TICL implantation for myopic astigmatism (n=72 eyes)
| Parameter | Preop, mean ± SD (range) | Postop 1 m, mean ± SD (range) | P value |
|---|---|---|---|
| UCVA (logMAR) | 1.19±0.30 (0.70 to 1.70) | −0.01±0.05 (−0.18 to 0.10) | <0.001 |
| BSCVA (logMAR) | 0.04±0.06 (0.00 to 0.30) | −0.03±0.04 (−0.18 to 0.05) | <0.001 |
| SER (D) | −9.29±2.41 (−15.25 to −4.50) | −0.37±0.55 (−1.68 to 0.88) | <0.001 |
| Endothelial cell density (cells/mm2) | 2,950.07±285.57 (2,231 to 3,614) | 2,819.90±418.87 (1,773 to 3,759.20) | 0.002 |
| Corneal astigmatism (D) | 1.85±0.88 (0.20 to 4.40) | 1.81±1.00 (−0.50 to 3.75) | 0.770 |
Preop, preoperative visit; Postop 1 m, 1 month follow-up; SER, spherical equivalent refraction; SER, sphere + 1/2*cylinder; SD, standard deviation; D, diopter; UCVA, uncorrected visual acuity; BSCVA, best-spectacle corrected visual acuity.
Figure 1Achieved versus attempted spherical equivalent refraction (SER) at the 1-month follow-up. The area between the two dotted lines represents the postoperative SER within ±0.50 D, and that between the two solid grey lines represents the postoperative SER within ±1.00 D. The solid red line depicts linear regression.
Polar analysis of refractive astigmatism at 1 month after V4c TICL implantation. All astigmatic polar values with the steeper meridian of the preoperative refractive astigmatism as the reference plane
| Parameter | AKP | AKP+45 | Net astigmatism |
|---|---|---|---|
| Preop | 1.91±0.73 | 0 | 1.91±0.73 @ 0° |
| Postop | 0.54±0.41 | 0.25±0.31 | 0.60±0.36 @ 12.42° |
| Paired | <0.001 | <0.001 | – |
| SIA | −1.37±0.84 | 0.25±0.31 | 1.39±0.51 @ 84.83° |
| TIA | −1.80±0.80 | 0 | 1.80±0.80 @ 90° |
| Paired | <0.001 | <0.001 | – |
| Intended refraction | 0.11±0.28 | 0 | 0.11±0.28 @ 90° |
| Error of astigmatism | 0.43±0.54 | 0.25±0.31 | 0.50±0.41 @ 15.08° |
Preop, preoperative; Postop, postoperative; SIA, surgically induced astigmatism; TIA, target induced astigmatism.
Figure 2Target-induced astigmatism (TIA) against surgically induced astigmatism (SIA) in patients with anterior corneal astigmatism at 1 month postoperatively, expressed as AKP. The two dotted lines represent values within ±0.50 D. The two solid grey lines represent values within ±1.00 D. The solid red line represents linear regression.
Polar value analysis of changes in error of treatment stratified by the magnitude of V4c TICL power at 1 month postoperatively
| Parameter | ≤2.0 D (n=54) | >2.0 D (n=18) | F value | P value |
|---|---|---|---|---|
| Preop | ||||
| AKP | 1.46±041 | 2.82±0.83 | 84.199 | <0.001 |
| AKP + 45 | 0 | 0 | – | – |
| Postop 1 m | ||||
| AKPerror | 0.33±0.52 | 0.72±0.53 | 7.584 | 0.007 |
| AKP + 45error | 0.24±0.28 | 0.26±0.42 | 0.043 | 0.837 |
| │AKPerror│ | 0.47±0.39 | 0.75±0.49 | 6.245 | 0.015 |
TICL, toric implantable collamer lens; Preop, preoperative; Postop, postoperative.
Figure 3Bivariate analysis of the error of treatment. The means are shown as follows: preoperative astigmatic polar value = triangle on the right; Target-induced astigmatism (TIA, expressed as AKP) = star on the left; Surgically induced astigmatism (SIA, expressed as AKP) = circle on the left; the combined mean of the error of treatment = square in the middle.
Figure 4V4c TICL rotation at 1 month after surgery. Ax, axis; TICL, toric implantable collamer lens.
Comparison of ocular and V4c TICL parameters between eyes with postoperative rotation <10 degrees and eyes with rotation of ≥10 degrees
| Parameter | V4c TICL rotation | F value | P value | |
|---|---|---|---|---|
| ≥10° (n=20) | <10° (n=52) | |||
| Age (y) | 24.80±5.66 | 25.87±4.01 | 0.803 | 0.373 |
| V4c TICL spherical power (D) | −10.10±2.09 | −11.91±2.42 | 8.672 | 0.004 |
| V4c TICL cylinder power (D) | 1.55±0.67 | 1.90±0.83 | 2.817 | 0.098 |
| K power (D) | 44.13±1.58 | 43.77±1.58 | 0.729 | 0.396 |
| V4c TICL size (mm) | 12.83±0.48 | 12.86±0.35 | 0.107 | 0.745 |
| WTW (mm) | 11.65±0.48 | 11.60±0.34 | 0.250 | 0.619 |
| ACD (mm) | 3.34±0.24 | 3.27±0.29 | 0.764 | 0.385 |
| Axial length (mm) | 26.83±1.42 | 26.79±1.35 | 0.003 | 0.957 |
| V4c TICL central vault (μm) | 681.11±206.26 | 625.82±221.79 | 0.932 | 0.338 |
| Absolute value of V4c TICL rotation (degrees) | 19.59±12.87 | 3.96±2.87 | 69.180 | <0.001 |
| Absolute value of AKPerror | 0.85±0.34 | 0.42±0.41 | 17.132 | <0.001 |
TICL, toric implantable collamer lens; D, diopter; WTW, white-to-white distance; ACD, anterior chamber depth.
Absolute degree of rotation based on the V4c TICL fixation angle at 1 month postoperatively
| Parameter | Group 1 (n=16) | Group 2 (n=43) | Group 3 (n=13) | F value | P value |
|---|---|---|---|---|---|
| Fixation angle | 0 | 2.67±1.49 | 8.00±1.87 | 108.294 | <0.001 |
| Absolute degree of rotation | 5.60±4.36 | 7.26±7.55 | 15.59±17.21 | 4.434 | 0.016 |
TICL, toric implantable collamer lens.