Abstract

We investigated the microstructures, tensile properties, fatigue strengths, and durability limits of hot-forged Ti-15Zr-4Nb (Ti-15-4) alloy artificial hip stems. These properties were compared with those of Ti-15Zr-4Nb-4Ta (Ti-15-4-4) and Ti-6Al-4V (Ti-6-4) alloy stems annealed after selective laser melting. The tensile and fatigue properties of test specimens cut from Ti-15-4 stems annealed after hot forging were excellent compared with those of the Alloclassic Zweymüller Stepless (SL) stem, which is used globally. The 0.2% proof stress (σ0.2%PS), ultimate tensile strength (σUTS), total elongation (TE) at breaking, and fatigue strength (σFS) after 107 cycles were 919 ± 10, 983 ± 9 MPa, 21 ± 1%, and 855 ± 14 MPa, respectively. The durability limit (PD) after 5 × 106 cycles of Ti-15-4 stems was excellent compared with that of the SL stem. The σUTS values of 90°- and 0°-direction-built Ti-15-4-4 rods were 1032 ± 1 and 1022 ± 2 MPa, and their TE values were 14 ± 1% and 16 ± 1%, respectively. The σFS values of annealed 90°-direction-built Ti-15-4-4 and Ti-6-4 rods were 640 ± 11 and 680 ± 37 MPa, respectively, which were close to that of the wrought Ti-15-4 rod (785 ± 17 MPa). These findings indicate that the hot forging and selective laser melting (SLM) techniques can also be applied to the manufacture of artificial hip prostheses. In particular, it was clarified that selective laser melting using Ti-15-4-4 and Ti-6-4 powders is useful for the low-cost manufacturing of custom-made artificial joint prostheses and other prosthetic implants.

Entities:  

Keywords:  artificial hip joints; biocompatible titanium alloy; custom-made implant; durability; fatigue property; hot forging; microstructure; selective laser melting; tensile property

Year:  2021        PMID: 33557357      PMCID: PMC7915577          DOI: 10.3390/ma14040732

Source DB:  PubMed          Journal:  Materials (Basel)        ISSN: 1996-1944            Impact factor:   3.623


  20 in total

1.  Modelling evaluation of the testing condition influence on the maximum stress induced in a hip prosthesis during ISO 7206 fatigue testing.

Authors:  M T Raimondi; R Pietrabissa
Journal:  Med Eng Phys       Date:  1999-06       Impact factor: 2.242

2.  Computed tomography-based custom-made stem for dysplastic hips in Japanese patients.

Authors:  Kenji Kawate; Yutaka Ohneda; Tetsuji Ohmura; Hiroshi Yajima; Kazuya Sugimoto; Yoshinori Takakura
Journal:  J Arthroplasty       Date:  2008-07-09       Impact factor: 4.757

3.  Is there a need of custom-made prostheses for total hip arthroplasty? Gait analysis, clinical and radiographic analysis of customized femoral components.

Authors:  Christian Götze; Dieter Rosenbaum; Jan Hoedemaker; Friedrich Bottner; Wolfram Steens
Journal:  Arch Orthop Trauma Surg       Date:  2008-10-31       Impact factor: 3.067

4.  Fretting wear in a modular neck hip prosthesis.

Authors:  M Viceconti; M Baleani; S Squarzoni; A Toni
Journal:  J Biomed Mater Res       Date:  1997-05

5.  Effect of proximal femoral bone support on the fixation of a press-fit noncemented total hip replacement femoral component.

Authors:  Sophia N Sangiorgio; Edward Ebramzadeh; Sean L Borkowski; Daniel A Oakes; Jeremy J Reid; Benjamin C Bengs
Journal:  J Appl Biomater Funct Mater       Date:  2013-06-24       Impact factor: 2.604

6.  Periprosthetic DXA after total hip arthroplasty with short vs. ultra-short custom-made femoral stems: 37 patients followed for 3 years.

Authors:  Carlina V Albanese; Francesco S Santori; Laura Pavan; Ian D Learmonth; Roberto Passariello
Journal:  Acta Orthop       Date:  2009-06       Impact factor: 3.717

7.  Custom cementless stem improves hip function in young patients at 15-year followup.

Authors:  Xavier Flecher; Oliver Pearce; Sebastien Parratte; Jean-Manuel Aubaniac; Jean-Noel Argenson
Journal:  Clin Orthop Relat Res       Date:  2009-08-19       Impact factor: 4.176

8.  Custom-design implants for severe distorted proximal anatomy of the femur in young adults followed for 4-8 years.

Authors:  Panagiotis Koulouvaris; Kosmas Stafylas; Thomas Sculco; Theodore Xenakis
Journal:  Acta Orthop       Date:  2008-04       Impact factor: 3.717

9.  Microstructures and Mechanical Properties of Laser-Sintered Commercially Pure Ti and Ti-6Al-4V Alloy for Dental Applications.

Authors:  Yoshimitsu Okazaki; Akira Ishino
Journal:  Materials (Basel)       Date:  2020-01-29       Impact factor: 3.623

10.  Biological Safety Evaluation and Surface Modification of Biocompatible Ti-15Zr-4Nb Alloy.

Authors:  Yoshimitsu Okazaki; Shin-Ichi Katsuda
Journal:  Materials (Basel)       Date:  2021-02-04       Impact factor: 3.623

View more
  1 in total

1.  Design of Low-Cost and High-Strength Titanium Alloys Using Pseudo-Spinodal Mechanism through Diffusion Couple Technology and CALPHAD.

Authors:  Chaoyi Ding; Chun Liu; Ligang Zhang; Di Wu; Libin Liu
Journal:  Materials (Basel)       Date:  2021-05-28       Impact factor: 3.623

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.