| Literature DB >> 33488800 |
Xiaoxue Wang1, Wei Qin1, Wenqian Chen1, Huifang Liu1, Dan Zhang1, Xianglin Zhang1, Pengmei Li1.
Abstract
Therapeutic drug monitoring is an essential tool when managing the therapeutic use of immunosuppressant cyclosporine A (CsA) in cases with solid organ transplantation. In China, the concentration of CsA is primarily measured using immunoassays. However, existing literature recommends mass spectrometry as the current gold standard for the quantitation of CsA. In the present study, it was attempted to develop a novel application to determine CsA concentrations by using ultra-performance liquid chromatography coupled to high-resolution mass spectrometry (UPLC-HRMS). This technique was then compared with a commercially available chemiluminescent microparticle immunoassay (CMIA) and it was investigated how clinical factors may contribute to quantitation differences between the two methods. An UPLC-Orbitrap-MS method was developed to determine CsA concentrations and this method was validated using guidelines put forward by the Food and Drug Administration from the US. In total, 127 blood samples were acquired from patients undergoing kidney transplantation and analyzed by UPLC-HRMS and CMIA assays. The novel method provided sensitive, accurate and precise results. The mean CsA concentration measured by CMIA was significantly higher than that measured by UPLC-HRMS (85.70±48.99 vs. 67.06±34.56 ng/ml, P<0.0001). Passing Bablok analysis yielded a slope of 1.34 (95% CI: 1.22-1.47) and an intercept of -2.54 (95% CI: -10.29-5.52). A group of samples with a higher metabolic ratio (hydroxylated CsA/CsA>1) exhibited larger discrepancies, while a group of samples taken from patients with a longer post-transplantation time (>10 years) featured narrow 95% CIs from -15.32 to 65.69%, as determined by Bland-Altman analysis. In summary, a reliable, accurate and rapid UPLC-HRMS method for CsA analysis was successfully developed. The measurement of CsA by the CMIA assay in renal transplant patients should be further evaluated with a specific focus on positive bias. Copyright: © Wang et al.Entities:
Keywords: chemiluminescent microparticle immunoassay; cyclosporine; high-resolution mass spectrometry; method comparison
Year: 2021 PMID: 33488800 PMCID: PMC7812591 DOI: 10.3892/etm.2021.9623
Source DB: PubMed Journal: Exp Ther Med ISSN: 1792-0981 Impact factor: 2.447
Figure 1Representative extracted ion chromatograms of cyclosporine. The main metabolites of HCsA and CsA-d4 (internal standard). (A) Extracted ion chromatography of m/z 1202.8485 in blank human whole blood. (B) Blank human whole blood spiked with cyclosporine. (C) Extracted ion chromatography of m/z 1218.8420 in blank human whole blood. (D) Cyclosporine samples from a representative patient in the HCsA/CsA>1 group. (E) Extracted ion chromatography of m/z 1206.8727 in blank human whole blood. (F) Blank human whole blood spiked with an internal standard. m/z, mass to charge ratio; HCsA, hydroxylated cyclosporine A.
Summary of accuracy, precision and stability of cyclosporine A in human blood using ultra-performance liquid chromatography high-resolution mass spectrometry.
| Intraday | Interday | Storage time[ | ||||||||
|---|---|---|---|---|---|---|---|---|---|---|
| Compound (concentration) | Accuracy (%) | Precision (RSD, %) | Accuracy (%) | Precision (RSD, %) | RT/12 h | -80˚C/1 freeze-thaw cycle | -80˚C/2 freeze-thaw cycles | 4˚C/3 days | -80˚C/60 days | 10˚C/24 h (extracted samples) |
| LLOQ (5 ng/ml) | 102.7 | 2.4 | 114.3 | 5.4 | ||||||
| QC1 (15 ng/ml) | 114.9 | 5.6 | 110.4 | 9.2 | 3.5 | -3.8 | -4.2 | -6.4 | -4.0 | 0.1 |
| QC2 (350 ng/ml) | 105.3 | 3.1 | 97.8 | 4.7 | -5.1 | 2.9 | 2.6 | -8.2 | -3.3 | -5.7 |
| QC3 (700 ng/ml) | 98.6 | 2.3 | 95.3 | 6.2 | -2.7 | -4.9 | 3.2 | -7.1 | 5.8 | -7.5 |
aValues are expressed as a deviation (% bias) from the nominal value. RSD, relative standard deviation; RT, room temperature; LLOQ, lower limit of quantification; QC, quality control.
Figure 2Method comparison using Passing-Bablok regression (left: A, C, E and G) and relative (%) Bland-Altman analysis (right: B, D, F and H) of the new Ultra-Performance LC High-Resolution MS method vs. the CMIA assay. Sample points for each subgroup include (A and B) metabolic ratio, (C and D) age, (E and F) gender and (G and H) the length of time after transplantation. The lines of correlation are represented by solid blue lines, and the 95% limits of agreement are represented by dashed lines. NA, information not available; HCsA, hydroxylated cyclosporine A; SD, standard deviation; LC-MS, liquid chromatography tandem mass spectrometry; CMIA, chemiluminescent microparticle immunoassay.
Figure 3Method comparison using Passing-Bablok regression (left) and relative (%) Bland-Altman analysis (right) of the new UPLC-HRMS method vs. the CMIA assay. The liner correlation graphs are represented by solid blue lines and the 95% limits of agreement are represented by dashed lines. PT, post-transplantation; HCsA, hydroxylated cyclosporine A; SD, standard deviation; LC-MS, liquid chromatography tandem mass spectrometry; CMIA, chemiluminescent microparticle immunoassay; UPLC-HRMS, ultra-performance LC high-resolution MS.
Comparison between UPLC-HRMS and CMIA for the determinations of CsA in human blood.
| Item | No. of samples | Concentration according to UPLC-HRMS, ng/ml | Concentration according to CMIA, ng/ml | Passing-Bablok regression equation | 95% CI of the intercept | 95% CI of the slope | Spearman correlation coefficient (P-value) | P-value for comparison of correlation coefficients | Bland-Altman Mean bias (lower-upper limit) |
|---|---|---|---|---|---|---|---|---|---|
| HCsA/CsA | 0.059 | ||||||||
| ≤1 | 65 | 67.9[ | 91.9 (69-375.2) | y=2.13+1.31x | -9.68-13.34 | 1.14-1.48 | 0.860 (<0.0001) | 25.41 (-22.52-73.34) | |
| >1 | 62 | 66.9[ | 77.0 (9.1-200.8) | y=-8.81+1.41x | -22.95-2.51 | 1.20-1.65 | 0.740 (<0.0001) | 25.52 (-28.36-79.39) | |
| Age, years | 0.539 | ||||||||
| ≤50 | 59 | 68.3[ | 88.8 (20.0-217.5) | y=3.67+1.20x | -7.56-16.32 | 1.04-1.40 | 0.841 (<0.0001) | 27.10 (-20.01-74.23) | |
| >50 | 68 | 66.8[ | 84.7 (9.1-375.2) | y=-7.29+1.42x | -17.68-2.65 | 1.28-1.60 | 0.805 (<0.0001) | 24.03 (-29.80-77.86) | |
| Sex | 0.840 | ||||||||
| Female | 55 | 68.7[ | 91.9 (9.1-228.2) | y=8.25+1.17x | -4.77-17.87 | 1.02-1.39 | 0.817 (<0.0001) | 25.34 (-27.78-78.47) | |
| Male | 72 | 66.9[ | 82.4 (20.0-375.2) | y=-12.45+1.49x | -24.68-1.25 | 1.32-1.70 | 0.829 (<0.0001) | 25.55 (-23.62-74.72) | |
| Timepost-transplantation (years) | 0.155 | ||||||||
| ≤10 | 45 | 60.3[ | 87.9 (20.0-206.1) | y=-6.34+1.39x | -21.62-4.13 | 1.18-1.66 | 0.768 (<0.0001) | 28.38 (-24.93-81.69) | |
| >10 | 60 | 66.9[ | 83.4 (25.6-195.9) | y=4.76+1.23x | -7.34-14.91 | 1.03-1.42 | 0.863 (<0.0001) | 25.18 (-15.32-65.69) |
aSignificantly different from the concentration measured by CMIA. Values are expressed as the median (range). UPLC-HRMS, ultra-performance liquid chromatography high-resolution mass spectrometry; CMIA, chemiluminescent microparticle immunoassay; HCsA, hydroxylated cyclosporine A.