Literature DB >> 30270673

Accuracy evaluation of Roche and Siemens tacrolimus and cyclosporine assays in comparison with liquid chromatography-tandem mass spectrometry.

Dae-Hyun Ko1, Eun-Jung Cho2, Woochang Lee2, Sail Chun2, Won-Ki Min2.   

Abstract

Therapeutic drug monitoring of tacrolimus and cyclosporine is crucial to the success of organ transplantation. We evaluated the analytical performances and accuracy of two commercially available tacrolimus and cyclosporine assays (Roche ISD and Siemens) in comparison with liquid chromatography-tandem mass spectrometry (LC-MS/MS). A total of 342 leftover whole blood samples requested for tacrolimus or cyclosporine assays were stored at -20 °C until analysis. Repeatability and between-run imprecision were evaluated using quality control materials provided by the manufacturer. Ring trial samples were used for the assessment of recovery. The results of the Roche ISD assay were compared with those of Siemens tacrolimus and cyclosporine assays and LC-MS/MS. Repeatability and between-run imprecision were 2.1-5.3% and 2.6-7.5%, respectively. Recovery of Roche ISD was 85.7 - 90.6% for cyclosporine and 96.2-98.5% for tacrolimus. The two immunoassays showed slight positive biases relative to LC-MS/MS for cyclosporine. For tacrolimus, Roche ISD produced virtually identical results to those of LC-MS/MS, whereas Siemens showed proportional differences, especially in patients receiving kidney transplantation. The analytical performances of Roche ISD were generally acceptable, especially regarding accuracy. Clinical laboratory staff should be aware of the strengths and weaknesses of commercial immunoassays in order to ensure accurate results.

Entities:  

Keywords:  Therapeutic drug monitoring; accuracy; cyclosporine; liquid chromatography-tandem mass spectrometry; tacrolimus

Mesh:

Substances:

Year:  2018        PMID: 30270673     DOI: 10.1080/00365513.2018.1472801

Source DB:  PubMed          Journal:  Scand J Clin Lab Invest        ISSN: 0036-5513            Impact factor:   1.713


  2 in total

1.  Accuracy evaluation of automated electrochemiluminescence immunoassay for everolimus and sirolimus compared to liquid chromatography-tandem mass spectrometry.

Authors:  Eun Jin Lee; Hyun-Ki Kim; Sunyoung Ahn; Woochang Lee; Hyun Soo Kim; Sail Chun; Won-Ki Min
Journal:  J Clin Lab Anal       Date:  2019-06-14       Impact factor: 2.352

2.  Validation of a novel UPLC-HRMS method for human whole-blood cyclosporine and comparison with a CMIA immunoassay.

Authors:  Xiaoxue Wang; Wei Qin; Wenqian Chen; Huifang Liu; Dan Zhang; Xianglin Zhang; Pengmei Li
Journal:  Exp Ther Med       Date:  2021-01-07       Impact factor: 2.447

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.