| Literature DB >> 33487834 |
Shreyasi Ray1, Jyotirmay Kirtania2.
Abstract
BACKGROUND: Passive movements of head and neck are sometimes unavoidable during surgery under general anesthesia due to patient positioning according to the needs of the surgery or transmitted movements from surgical manipulations. AIMS: This prospective crossover randomized study evaluates the effects of passive movements of the head and neck on the performance of i-gel® supraglottic airway device in spontaneously breathing patients under general anesthesia.Entities:
Keywords: General anesthesia; passive head-and-neck movement; pressure support ventilation with positive end-expiratory pressure; spontaneous breathing; supraglottic airway
Year: 2020 PMID: 33487834 PMCID: PMC7819417 DOI: 10.4103/aer.AER_73_20
Source DB: PubMed Journal: Anesth Essays Res ISSN: 2229-7685
Figure 1CONSORT diagram representing the enrollment process of the study participants
Demographic characteristics of the study population
| Parameter | Mean (SD) | Median (IQR) |
|---|---|---|
| Age (years) | 37.8 (15.8) | 35 (26.5) |
| Height (cm) | 154.1 (4.1) | 154.0 (6.0) |
| Weight (kg) | 49.2 (8.6) | 48 (9.3) |
| BMI (kg.m−2) | 20.7 (3.6) | 20.9 (4.2) |
| Sex | ||
| Male | 44 | |
| Female | 16 | |
| ASA physical status | ||
| I | 50 | |
| II | 10 | |
| i gel® size | ||
| Size 3 | 36 | |
| Size 4 | 24 | |
Data presented as mean (SD); median (IQR) for continuous data, and actual numbers (counts) for nominal data. Total participants, n=60. SD=Standard deviation, IQR=Interquartile range, ASA=American Society of Anesthesiologists, BMI=Body mass index
Performance of i-gel® in different positions of head and neck under inhalation general anesthesia with spontaneous breathing on pressure support ventilation plus positive end-expiratory pressure
| Parameter | Median (IQR) | ||||||
|---|---|---|---|---|---|---|---|
| Neutral | Flexion | Extension | Left flexion | Right flexion | Left rotation | Right rotation | |
| Exhaled minute volume (L) | 5.0 (1.5) | 5.0 (1.5) | 4.6 (1.4) | 4.9 (1.6) | 4.9 (1.6) | 4.9 (1.6) | 4.8 (1.6) |
| ETCO2 | |||||||
| mmHg | 36 (2.0) | 36 (2.0) | 39 (3.3) | 38 (2.3) | 38 (2.3) | 37 (2.0) | 38 (2.3) |
| kPa | 4.8 (0.3) | 4.8 (0.3) | 5.2 (0.4) | 5.1 (0.3) | 5.1 (0.3) | 4.9 (0.3) | 5.1 (0.3) |
| Peak airway pressure | |||||||
| cmH2O | 16 (4) | 15 (5) | 14 (4) | 14 (4) | 14 (4) | 15 (5) | 14 (4) |
| kPa | 1.6 (0.4) | 1.5 (0.5) | 1.4 (0.4) | 1.4 (0.4) | 1.4 (0.4) | 1.5 (0.5) | 1.4 (0.4) |
| Pressure support | |||||||
| cmH2O | 11 (5) | 11 (5) | 11 (5) | 11 (5) | 11 (5) | 11 (5) | 11 (5) |
| kPa | 1.1 (0.5) | 1.1 (0.5) | 1.1 (0.5) | 1.1 (0.5) | 1.1 (0.5) | 1.1 (0.5) | 1.1 (0.5) |
| PEEP | |||||||
| cmH2O | 5 (2) | 5 (2) | 5 (2) | 5 (2) | 5 (2) | 5 (2) | 5 (2) |
| kPa | 0.5 (0.2) | 0.5 (0.2) | 0.5 (0.2) | 0.5 (0.2) | 0.5 (0.2) | 0.5 (0.2) | 0.5 (0.2) |
| SpO2 (%) | 100 (0) | 100 (0) | 99 (0) | 99 (0) | 99 (0) | 99 (0) | 99 (0) |
| Audible leak of airway gases (no/yes) | |||||||
| No | 58 | 57 | 16 | 22 | 23 | 55 | 53 |
| Yes | 2 | 3 | 44 | 38 | 37 | 5 | 7 |
| Visible displacement of i-gel® (no/yes) | |||||||
| No | 60 | 58 | 17 | 38 | 40 | 55 | 54 |
| Yes | 0 | 2 | 43 | 22 | 20 | 5 | 6 |
Continuous data were subjected to Shapiro-Wilk test of normality (P<0.05 for all), and found to be nonnormally distributed. Continuous data presented as median and IQR. Nominal data presented as actual numbers (counts). IQR=Interquartile range, PEEP=Positive end-expiratory pressure, ETCO2=End-tidal carbon dioxide
Comparison of medians of exhaled minute volume in neutral, flexion, extension, left flexion, right flexion, left rotation, and right rotation positions
| Neutral versus flexion | Neutral versus extension | Neutral versus left flexion | Neutral versus right flexion | Neutral versus left rotation | Neutral versus right rotation | |
|---|---|---|---|---|---|---|
| Exhaled minute volume | 0.642 | <0.001* | <0.001* | <0.001* | <0.001* | <0.001* |
| ETCO2 | 0.784 | <0.001* | <0.001* | <0.001* | <0.001* | <0.001* |
*P<0.05 was considered statistically significant for paired-sample, two-sided, Wilcoxon signed-rank test with continuity correction. ETCO2=End-tidal carbon dioxide
Audible air leak and visible displacement of i-gel® in neutral, flexion, extension, left flexion, right flexion, left rotation, and right rotation positions
| Neutral versus flexion | Neutral versus extension | Neutral versus left flexion | Neutral versus right flexion | Neutral versus left rotation | Neutral versus right rotation | |
|---|---|---|---|---|---|---|
| Audible air leak | 0.564 | <0.001* | <0.001* | <0.001* | 0.317 | 0.095 |
| Visible displacement of i-gel® | 0.174 | <0.001* | <0.001* | <0.001* | 0.033* | 0.020* |
*P<0.05 was considered statistically significant for post hoc pair-wise analysis with McNemar test