Literature DB >> 33465155

Simultaneous ventilation in the Covid-19 pandemic. A bench study.

Claude Guérin1,2,3, Martin Cour1,2, Neven Stevic1, Florian Degivry1, Erwan L'Her4,5, Bruno Louis3, Laurent Argaud1,2.   

Abstract

COVID-19 pandemic sets the healthcare system to a shortage of ventilators. We aimed at assessing tidal volume (VT) delivery and air recirculation during expiration when one ventilator is divided into 2 test-lungs. The study was performed in a research laboratory in a medical ICU of a University hospital. An ICU (V500) and a lower-level ventilator (Elisée 350) were attached to two test-lungs (QuickLung) through a dedicated flow-splitter. A 50 mL/cmH2O Compliance (C) and 5 cmH2O/L/s Resistance (R) were set in both A and B test-lungs (A C50R5 / B C50R5, step1), A C50-R20 / B C20-R20 (step 2), A C20-R20 / B C10-R20 (step 3), and A C50-R20 / B C20-R5 (step 4). Each ventilator was set in volume and pressure control mode to deliver 800mL VT. We assessed VT from a pneumotachograph placed immediately before each lung, pendelluft air, and expiratory resistance (circuit and valve). Values are median (1st-3rd quartiles) and compared between ventilators by non-parametric tests. Between Elisée 350 and V500 in volume control VT in A/B test- lungs were 381/387 vs. 412/433 mL in step 1, 501/270 vs. 492/370 mL in step 2, 509/237 vs. 496/332 mL in step 3, and 496/281 vs. 480/329 mL in step 4. In pressure control the corresponding values were 373/336 vs. 430/414 mL, 416/185 vs. 322/234 mL, 193/108 vs. 176/ 92 mL and 422/201 vs. 481/329mL, respectively (P<0.001 between ventilators at each step for each volume). Pendelluft air volume ranged between 0.7 to 37.8 ml and negatively correlated with expiratory resistance in steps 2 and 3. The lower-level ventilator performed closely to the ICU ventilator. In the clinical setting, these findings suggest that, due to dependence of VT to C, pressure control should be preferred to maintain adequate VT at least in one patient when C and/or R changes abruptly and monitoring of VT should be done carefully. Increasing expiratory resistance should reduce pendelluft volume.

Entities:  

Mesh:

Year:  2021        PMID: 33465155      PMCID: PMC7815120          DOI: 10.1371/journal.pone.0245578

Source DB:  PubMed          Journal:  PLoS One        ISSN: 1932-6203            Impact factor:   3.240


  32 in total

1.  Use of a single ventilator to support 4 patients: laboratory evaluation of a limited concept.

Authors:  Richard D Branson; Thomas C Blakeman; Bryce Rh Robinson; Jay A Johannigman
Journal:  Respir Care       Date:  2011-10-12       Impact factor: 2.258

2.  Positive-pressure ventilation equipment for mass casualty respiratory failure.

Authors:  Lewis Rubinson; Richard D Branson; Nicki Pesik; Daniel Talmor
Journal:  Biosecur Bioterror       Date:  2006

3.  Fast and frugal innovations in response to the COVID-19 pandemic.

Authors:  Matthew Harris; Yasser Bhatti; Jim Buckley; Dhananjaya Sharma
Journal:  Nat Med       Date:  2020-06       Impact factor: 53.440

4.  Personalized Ventilation to Multiple Patients Using a Single Ventilator: Description and Proof of Concept.

Authors:  Jay S Han; Azad Mashari; Devin Singh; Jose Dianti; Ewan Goligher; Michael Long; William Ng; Marcin Wasowicz; David Preiss; Alex Vesely; Robert Kacmarek; Shaf Keshavjee; Laurent Brochard; Joseph A Fisher; Arthur S Slutsky
Journal:  Crit Care Explor       Date:  2020-05-06

5.  Shared Ventilation in the Era of COVID-19: A Theoretical Consideration of the Dangers and Potential Solutions.

Authors:  Jacob Herrmann; Andrea Fonseca da Cruz; Monica L Hawley; Richard D Branson; David W Kaczka
Journal:  Respir Care       Date:  2020-05-06       Impact factor: 2.258

6.  Differential Ventilation Using Flow Control Valves as a Potential Bridge to Full Ventilatory Support during the COVID-19 Crisis.

Authors:  Matthew A Levin; Anjan Shah; Ronak Shah; Erica Kane; George Zhou; James B Eisenkraft; Martin D Chen
Journal:  Anesthesiology       Date:  2020-10-01       Impact factor: 7.892

7.  A single ventilator for multiple simulated patients to meet disaster surge.

Authors:  Greg Neyman; Charlene Babcock Irvin
Journal:  Acad Emerg Med       Date:  2006-08-02       Impact factor: 3.451

8.  COVID-19 Does Not Lead to a "Typical" Acute Respiratory Distress Syndrome.

Authors:  Luciano Gattinoni; Silvia Coppola; Massimo Cressoni; Mattia Busana; Sandra Rossi; Davide Chiumello
Journal:  Am J Respir Crit Care Med       Date:  2020-05-15       Impact factor: 21.405

9.  COVID-19 pneumonia: different respiratory treatments for different phenotypes?

Authors:  Luciano Gattinoni; Davide Chiumello; Pietro Caironi; Mattia Busana; Federica Romitti; Luca Brazzi; Luigi Camporota
Journal:  Intensive Care Med       Date:  2020-04-14       Impact factor: 17.440

10.  Shared Ventilation: Toward Safer Ventilator Splitting in Resource Emergencies.

Authors:  Anne D Cherry; Jhaymie Cappiello; Muath Bishawi; Melanie G Hollidge; David B MacLeod
Journal:  Anesthesiology       Date:  2020-09       Impact factor: 7.892

View more
  1 in total

1.  The eSpiro Ventilator: An Open-Source Response to a Worldwide Pandemic.

Authors:  Nicolas Terzi; Fabrice Rastello; Christophe Déhan; Marion Roux; Florian Sigaud; Guillaume Rigault; Cyril Fromentin; Adrien Farrugia; Claude Guérin
Journal:  J Clin Med       Date:  2021-05-27       Impact factor: 4.241