Literature DB >> 16885402

A single ventilator for multiple simulated patients to meet disaster surge.

Greg Neyman1, Charlene Babcock Irvin.   

Abstract

OBJECTIVES: To determine if a ventilator available in an emergency department could quickly be modified to provide ventilation for four adults simultaneously.
METHODS: Using lung simulators, readily available plastic tubing, and ventilators (840 Series Ventilator; Puritan-Bennett), human lung simulators were added in parallel until the ventilator was ventilating the equivalent of four adults. Data collected included peak pressure, positive end-expiratory pressure, total tidal volume, and total minute ventilation. Any obvious asymmetry in the delivery of gas to the lung simulators was also documented. The ventilator was run for almost 12 consecutive hours (5.5 hours of pressure control and more than six hours of volume control).
RESULTS: Using readily available plastic tubing set up to minimize dead space volume, the four lung simulators were easily ventilated for 12 hours using one ventilator. In pressure control (set at 25 mm H2O), the mean tidal volume was 1,884 mL (approximately 471 mL/lung simulator) with an average minute ventilation of 30.2 L/min (or 7.5 L/min/lung simulator). In volume control (set at 2 L), the mean peak pressure was 28 cm H2O and the minute ventilation was 32.5 L/min total (8.1 L/min/lung simulator).
CONCLUSIONS: A single ventilator may be quickly modified to ventilate four simulated adults for a limited time. The volumes delivered in this simulation should be able to sustain four 70-kg individuals. While further study is necessary, this pilot study suggests significant potential for the expanded use of a single ventilator during cases of disaster surge involving multiple casualties with respiratory failure.

Entities:  

Mesh:

Year:  2006        PMID: 16885402      PMCID: PMC7164837          DOI: 10.1197/j.aem.2006.05.009

Source DB:  PubMed          Journal:  Acad Emerg Med        ISSN: 1069-6563            Impact factor:   3.451


  5 in total

1.  Health care system surge capacity recognition, preparedness, and response.

Authors: 
Journal:  Ann Emerg Med       Date:  2005-02       Impact factor: 5.721

2.  Concept of operations for triage of mechanical ventilation in an epidemic.

Authors:  John L Hick; Daniel T O'Laughlin
Journal:  Acad Emerg Med       Date:  2006-01-06       Impact factor: 3.451

3.  Artificial ventilation of the lungs for emergencies.

Authors:  M Darowski; M Englisz
Journal:  Front Med Biol Eng       Date:  2000

Review 4.  Management of post traumatic respiratory failure.

Authors:  Andrew J Michaels
Journal:  Crit Care Clin       Date:  2004-01       Impact factor: 3.598

5.  Health care facility and community strategies for patient care surge capacity.

Authors:  John L Hick; Dan Hanfling; Jonathan L Burstein; Craig DeAtley; Donna Barbisch; Gregory M Bogdan; Stephen Cantrill
Journal:  Ann Emerg Med       Date:  2004-09       Impact factor: 5.721

  5 in total
  54 in total

1.  PEEP/ FIO2 ARDSNet Scale Grouping of a Single Ventilator for Two Patients: Modeling Tidal Volume Response.

Authors:  Vitaly O Kheyfets; Steven R Lammers; Jennifer Wagner; Karsten Bartels; Jerome Piccoli; Bradford J Smith
Journal:  Respir Care       Date:  2020-08       Impact factor: 2.258

2.  Two for one with split- or co-ventilation at the peak of the COVID-19 tsunami: is there any role for communal care when the resources for personalised medicine are exhausted?

Authors:  Steven Dale Pearson; Jesse B Hall; William F Parker
Journal:  Thorax       Date:  2020-04-23       Impact factor: 9.139

3.  A porcine model for initial surge mechanical ventilator assessment and evaluation of two limited-function ventilators.

Authors:  Robert P Dickson; David L Hotchkin; Wayne J E Lamm; Carl Hinkson; David J Pierson; Robb W Glenny; Lewis Rubinson
Journal:  Crit Care Med       Date:  2011-03       Impact factor: 7.598

4.  Indoor fire in a nursing home: evaluation of the medical response to a mass casualty incident based on a standardized protocol.

Authors:  S W Koning; P M Ellerbroek; L P H Leenen
Journal:  Eur J Trauma Emerg Surg       Date:  2014-09-25       Impact factor: 3.693

5.  Innovative Use of High-Fidelity Lung Simulators to Test a Ventilator Splitter Device.

Authors:  Tanna J Boyer; Sally A Mitchell; Johnny F Cartwright; Rami A Ahmed
Journal:  A A Pract       Date:  2020-06

6.  Pandemic influenza: implications for preparation and delivery of critical care services.

Authors:  Mary-Elise Manuell; Mary Dawn T Co; Richard T Ellison
Journal:  J Intensive Care Med       Date:  2011-01-10       Impact factor: 3.510

7.  The role of additive manufacturing and antimicrobial polymers in the COVID-19 pandemic.

Authors:  Jorge M Zuniga; Aaron Cortes
Journal:  Expert Rev Med Devices       Date:  2020-04-30       Impact factor: 3.166

8.  Answering the Challenge of COVID-19 Pandemic Through Innovation and Ingenuity.

Authors:  Kathryn Clare Kelley; Jonathan Kamler; Manish Garg; Stanislaw P Stawicki
Journal:  Adv Exp Med Biol       Date:  2021       Impact factor: 2.622

Review 9.  Preparing Intensive Care Unit in Resource-Constraint Setting Amid COVID-19 Pandemic: Our Experience and Review.

Authors:  Kamal Kajal; B Naveen Naik; Ajay Singh; Shiv Lal Soni; Amarjyoti Hazarika; Kulbhushan Saini; Sanjay Jaswal; Shyam Charan Meena; Naveen Pandey; G D Puri
Journal:  Anesth Essays Res       Date:  2021-03-22

10.  Development of a multi-patient ventilator circuit with validation in an ARDS porcine model.

Authors:  Benjamin P Wankum; Riley E Reynolds; Andrea R McCain; Nathaniel T Zollinger; Keely L Buesing; Russel D Sindelar; Frank M Freihaut; Tariku Fekadu; Benjamin S Terry
Journal:  J Anesth       Date:  2021-06-01       Impact factor: 2.078

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.