Thomas Woolcot1, Evanthia Kousi2, Emma Wells3, Katharine Aitken3, Helen Taylor3, Maria A Schmidt2.
Abstract
We demonstrated a general method to evaluate systematic errors related to Magnetic Resonance (MR) imaging sequences in marker-based co-registration of MR and Computed Tomography (CT) images, and investigated the effect of MR image quality in the co-registration process using clinical MR and CT protocols for stereotactic ablative body radiotherapy (SABR) planning of the liver. Small systematic errors (under 1.6 mm) were detected, unlikely to be a clinical risk to liver SABR. The least favourable marker configuration was found to be a co-planar arrangement parallel to the transaxial image plane.Entities:
Keywords: Fiducial markers; Liver; Registration; Stereotactic radiotherapy
Year: 2018 PMID: 33458402 PMCID: PMC7807725 DOI: 10.1016/j.phro.2018.08.001
Source DB: PubMed Journal: Phys Imaging Radiat Oncol ISSN: 2405-6316
MR sequence parameters.
| MR sequences | T1w | T2w | T2*w |
|---|---|---|---|
| TR/TE (ms) | 5.87/2.47 | 1600/96 | 230/4.92 |
| Voxel size (mm3) | 1.5 × 1.5 × 4 | 1.2 × 1.2 × 4 | 1.5 × 1.5 × 4 |
| Reconstruction/acquisition matrix | 256 × 192/ | 320 × 240/ | 256 × 192/ |
| % Sampling | 75 | 81 | 80 |
| Pixel bandwidth (Hz/px) | 1030 | 710 | 1395 |
| Readout gradient direction | R/L | R/L | R/L |
| Parallel imaging factor | 2 | 2 | 2 |
Fig. 1A. CT Images of the coplanar marker configuration before (left) and after (right) processing to remove the external container shape. The vertical test object orientation makes the arrangement of co-planar markers parallel to the image plane (transaxial) B. A total of N = 12 combinations of test object configuration and MR sequence were registered with CT.
Marker displacements between gold standard registrations and marker-based registrations.
| Mean offset and standard deviation | ||||
|---|---|---|---|---|
| Phase Enc. (A/P) | Slice Sel. (S/I) | Readout (R/L) | ||
| MR sequence | T1w | 0.1 ± 0.9 | 0.0 ± 0.3 | |
| T2*w | 0.0 ± 0.6 | 0.1 ± 0.4 | ||
| T2w | 0.1 ± 0.3 | |||
| p < 0.001 | p < 0.001 | |||
| Marker arrangement | Coplanar | |||
| Non-coplanar | 0.2 ± 0.9 | 0.1 ± 0.4 | ||
| p = 0.13 | p = 0.07 | |||
| Test object orientation | Horizontal | 0.1 ± 0.7 | 0.1 ± 0.6 | 0.0 ± 0.4 |
| Vertical | 0.1 ± 0.5 | |||
| p = 0.94 | p < 0.001 | |||
Results are listed with the p-value from the relevant statistical significance test performed. Here N is the number of measurements. Values shown in bold indicate results which reject the null hypothesis of a population mean equal to zero using a one-sample t-test (p < 0.05).
Fig. 2Sagittal, coronal and axial views of the vector offset for each MR sequence. The phase encoding (PE), frequency encoding (FE) and slice selection (SS) gradient directions are indicated in bold on the axis labels.