Free-breathing R 2 mapping of hepatic iron overload in children using 3D multi-echo UTE cones MRI.

Youngwook Kee1, Christopher M Sandino1, Ali B Syed1, Joseph Y Cheng1, Ann Shimakawa2, Timothy J Colgan3, Diego Hernando3, Shreyas S Vasanawala1.   

Abstract

PURPOSE: To enable motion-robust, ungated, free-breathing R 2 ∗ mapping of hepatic iron overload in children with 3D multi-echo UTE cones MRI.
METHODS: A golden-ratio re-ordered 3D multi-echo UTE cones acquisition was developed with chemical-shift encoding (CSE). Multi-echo complex-valued source images were reconstructed via gridding and coil combination, followed by confounder-corrected R 2 ∗ (=1/ T 2 ∗ ) mapping. A phantom containing 15 different concentrations of gadolinium solution (0-300 mM) was imaged at 3T. 3D multi-echo UTE cones with an initial TE of 0.036 ms and Cartesian CSE-MRI (IDEAL-IQ) sequences were performed. With institutional review board approval, 85 subjects (81 pediatric patients with iron overload + 4 healthy volunteers) were imaged at 3T using 3D multi-echo UTE cones with free breathing (FB cones), IDEAL-IQ with breath holding (BH Cartesian), and free breathing (FB Cartesian). Overall image quality of R 2 ∗ maps was scored by 2 blinded experts and compared by a Wilcoxon rank-sum test. For each pediatric subject, the paired R 2 ∗ maps were assessed to determine if a corresponding artifact-free 15 mm region-of-interest (ROI) could be identified at a mid-liver level on both images. Agreement between resulting R 2 ∗ quantification from FB cones and BH/FB Cartesian was assessed with Bland-Altman and linear correlation analyses.
RESULTS: ROI-based regression analysis showed a linear relationship between gadolinium concentration and R 2 ∗ in IDEAL-IQ (y = 8.83x - 52.10, R2 = 0.995) as well as in cones (y = 9.19x - 64.16, R2 = 0.992). ROI-based Bland-Altman analysis showed that the mean difference (MD) was 0.15% and the SD was 5.78%. However, IDEAL-IQ R 2 ∗ measurements beyond 200 mM substantially deviated from a linear relationship for IDEAL-IQ (y = 5.85x + 127.61, R2 = 0.827), as opposed to cones (y = 10.87x - 166.96, R2 = 0.984). In vivo, FB cones R 2 ∗ had similar image quality with BH and FB Cartesian in 15 and 42 cases, respectively. FB cones R 2 ∗ had better image quality scores than BH and FB Cartesian in 3 and 21 cases, respectively, where BH/FB Cartesian exhibited severe ghosting artifacts. ROI-based Bland-Altman analyses were 2.23% (MD) and 6.59% (SD) between FB cones and BH Cartesian and were 0.21% (MD) and 7.02% (SD) between FB cones and FB Cartesian, suggesting a good agreement between FB cones and BH (FB) Cartesian R 2 ∗ . Strong linear relationships were observed between BH Cartesian and FB cones (y = 1.00x + 1.07, R2 = 0.996) and FB Cartesian and FB cones (y = 0.98x + 1.68, R2 = 0.999).
CONCLUSION: Golden-ratio re-ordered 3D multi-echo UTE Cones MRI enabled motion-robust, ungated, and free-breathing R 2 ∗ mapping of hepatic iron overload, with comparable R 2 ∗ measurements and image quality to BH Cartesian, and better image quality than FB Cartesian.
© 2021 International Society for Magnetic Resonance in Medicine.

Entities:  

Keywords:  3D multi-echo UTE cones k-space sampling trajectory; chemical-shift-encoded MRI; confounder-corrected zzm321990 zzm321990 Rzzm321990 2zzm321990 zzm321990 zzm321990 zzm321990 ; free-breathing liver zzm321990 zzm321990 Rzzm321990 2zzm321990 zzm321990 zzm321990 zzm321990 mapping; hepatic iron overload

Mesh:

Year:  2021        PMID: 33432613      PMCID: PMC8886621          DOI: 10.1002/mrm.28610

Source DB:  PubMed          Journal:  Magn Reson Med        ISSN: 0740-3194            Impact factor:   4.668


  24 in total

1.  Noninvasive measurement and imaging of liver iron concentrations using proton magnetic resonance.

Authors:  Timothy G St Pierre; Paul R Clark; Wanida Chua-anusorn; Adam J Fleming; Gary P Jeffrey; John K Olynyk; Pensri Pootrakul; Erin Robins; Robert Lindeman
Journal:  Blood       Date:  2004-07-15       Impact factor: 22.113

2.  Design and analysis of a practical 3D cones trajectory.

Authors:  Paul T Gurney; Brian A Hargreaves; Dwight G Nishimura
Journal:  Magn Reson Med       Date:  2006-03       Impact factor: 4.668

3.  Deep residual network for off-resonance artifact correction with application to pediatric body MRA with 3D cones.

Authors:  David Y Zeng; Jamil Shaikh; Signy Holmes; Ryan L Brunsing; John M Pauly; Dwight G Nishimura; Shreyas S Vasanawala; Joseph Y Cheng
Journal:  Magn Reson Med       Date:  2019-05-22       Impact factor: 4.668

4.  Quantitative ultrashort echo time imaging for assessment of massive iron overload at 1.5 and 3 Tesla.

Authors:  Axel J Krafft; Ralf B Loeffler; Ruitian Song; Aaryani Tipirneni-Sajja; M Beth McCarville; Matthew D Robson; Jane S Hankins; Claudia M Hillenbrand
Journal:  Magn Reson Med       Date:  2017-01-16       Impact factor: 4.668

5.  Clinical Implementation of a Focused MRI Protocol for Hepatic Fat and Iron Quantification.

Authors:  B Dustin Pooler; Diego Hernando; Scott B Reeder
Journal:  AJR Am J Roentgenol       Date:  2019-03-27       Impact factor: 3.959

6.  Ultra-short echo time images quantify high liver iron.

Authors:  Eamon K Doyle; Kristin Toy; Bertin Valdez; Jonathan M Chia; Thomas Coates; John C Wood
Journal:  Magn Reson Med       Date:  2017-06-22       Impact factor: 4.668

Review 7.  Quantification of liver iron with MRI: state of the art and remaining challenges.

Authors:  Diego Hernando; Yakir S Levin; Claude B Sirlin; Scott B Reeder
Journal:  J Magn Reson Imaging       Date:  2014-03-03       Impact factor: 4.813

8.  3D R 2 * mapping of the placenta during early gestation using free-breathing multiecho stack-of-radial MRI at 3T.

Authors:  Tess Armstrong; Dapeng Liu; Thomas Martin; Rinat Masamed; Carla Janzen; Cass Wong; Teresa Chanlaw; Sherin U Devaskar; Kyunghyun Sung; Holden H Wu
Journal:  J Magn Reson Imaging       Date:  2018-08-24       Impact factor: 4.813

9.  Multipeak fat-corrected complex R2* relaxometry: theory, optimization, and clinical validation.

Authors:  Diego Hernando; J Harald Kramer; Scott B Reeder
Journal:  Magn Reson Med       Date:  2013-01-28       Impact factor: 4.668

10.  Complex confounder-corrected R2* mapping for liver iron quantification with MRI.

Authors:  Diego Hernando; Rachel J Cook; Naila Qazi; Colin A Longhurst; Carol A Diamond; Scott B Reeder
Journal:  Eur Radiol       Date:  2020-08-12       Impact factor: 5.315

View more
  1 in total

1.  Radiomics nomograms based on R2* mapping and clinical biomarkers for staging of liver fibrosis in patients with chronic hepatitis B: a single-center retrospective study.

Authors:  Die Zhang; Yi Cao; Yi Sun; Xia Zhao; Cheng Peng; Jing Zhao; Xiaohui Bao; Lifei Wang; Chen Zhang
Journal:  Eur Radiol       Date:  2022-09-23       Impact factor: 7.034

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.