Ignacio Torres1, Sandrine Poujois1, Eliseo Albert1, Javier Colomina1, David Navarro2. 1. Microbiology Service, Hospital Clínico Universitario, INCLIVA Research Institute, Valencia, Spain. 2. Microbiology Service, Hospital Clínico Universitario, INCLIVA Research Institute, Valencia, Spain; Department of Microbiology, School of Medicine, University of Valencia, Valencia, Spain. Electronic address: david.navarro@uv.es.
Abstract
OBJECTIVES: There is limited information on the performance of rapid antigen detection (RAD) tests to identify SARS-CoV-2-infected asymptomatic individuals. In this field study, we evaluated the Panbio™ COVID-19 Ag Rapid Test Device (Abbott Diagnostics, Jena, Germany) for this purpose. METHODS: A total of 634 individuals (355 female; median age, 37 years; range, 9-87) were enrolled. Two nasopharyngeal swabs were collected from household (n = 338) and non-household contacts (n = 296) of COVID-19 cases. RAD testing was carried out at the point of care. The RT-PCR test used was the TaqPath COVID-19 Combo Kit (Thermo Fisher Scientific, MA, USA). RESULTS: Household contacts were tested at a median of 2 days (range, 1-7) after diagnosis of the index case, whereas non-household contacts (n = 296) were tested at a median of 6 days (range, 1-7) after exposure. In total, 79 individuals (12.4%) tested positive by RT-PCR, of whom 38 (48.1%) yielded positive RAD results. The overall sensitivity and specificity of the RAD test was 48.1% (95% CI 37.4-58.9) and 100% (95% CI 99.3-100), respectively. Sensitivity was higher in household (50.8%; 95% CI 38.9-62.5) than in non-household (35.7%; 95% CI 16.3-61.2%) contacts. Individuals testing positive by RAD test were more likely (p < 0.001) to become symptomatic than their negative counterparts. DISCUSSION: The Panbio test displays low sensitivity in asymptomatic close contacts of COVID-19 patients, particularly in non-household contacts. Nonetheless, establishing the optimal timing for upper respiratory tract collection in this group seems imperative to pinpoint test sensitivity.
OBJECTIVES: There is limited information on the performance of rapid antigen detection (RAD) tests to identify SARS-CoV-2-infected asymptomatic individuals. In this field study, we evaluated the Panbio™ COVID-19 Ag Rapid Test Device (Abbott Diagnostics, Jena, Germany) for this purpose. METHODS: A total of 634 individuals (355 female; median age, 37 years; range, 9-87) were enrolled. Two nasopharyngeal swabs were collected from household (n = 338) and non-household contacts (n = 296) of COVID-19 cases. RAD testing was carried out at the point of care. The RT-PCR test used was the TaqPath COVID-19 ComboKit (Thermo Fisher Scientific, MA, USA). RESULTS: Household contacts were tested at a median of 2 days (range, 1-7) after diagnosis of the index case, whereas non-household contacts (n = 296) were tested at a median of 6 days (range, 1-7) after exposure. In total, 79 individuals (12.4%) tested positive by RT-PCR, of whom 38 (48.1%) yielded positive RAD results. The overall sensitivity and specificity of the RAD test was 48.1% (95% CI 37.4-58.9) and 100% (95% CI 99.3-100), respectively. Sensitivity was higher in household (50.8%; 95% CI 38.9-62.5) than in non-household (35.7%; 95% CI 16.3-61.2%) contacts. Individuals testing positive by RAD test were more likely (p < 0.001) to become symptomatic than their negative counterparts. DISCUSSION: The Panbio test displays low sensitivity in asymptomatic close contacts of COVID-19patients, particularly in non-household contacts. Nonetheless, establishing the optimal timing for upper respiratory tract collection in this group seems imperative to pinpoint test sensitivity.
Rapid antigen detection (RAD) immunoassays have emerged as a valuable alternative to RT-PCR for diagnosis of SARS-CoV-2 infection in patients presenting with clinically compatible COVID-19 [1]. RAD tests are simple to carry out and return results within a short time, thus being well-suited for point-of-care testing (POCT). Moreover, two recently evaluated RAD assays (BD Veritor System and the Panbio™ COVID-19 Ag Rapid Test Device) may be used as a proxy for SARS-CoV-2 cultured from respiratory tract specimens, thus allowing reasonably accurate prediction of contagiousness [2,3]; yet this assumption awaits further validation in larger cohorts. The possibility of using RAD tests to identify SARS-CoV-2-infected asymptomatic contacts of COVID-19patients is appealing, as it could effectively contribute to minimize community SARS-CoV-2 spread through early detection of highly infectious individuals [1], yet little is known about how RAD tests perform in this population group [[4], [5], [6]]. Here, we report on the performance of the Panbio™ COVID-19 Ag Rapid Test Device (Abbott Diagnostic GmbH, Jena, Germany) conducted at POC in this setting.
Materials and methods
Patients
A total of 634 consecutive asymptomatic individuals (female, n = 355; median age, 37 years; range, 9–87 years) attended at the Clínico-Malvarrosa Health Department (Valencia, Spain) were enrolled between 16 October and 20 November 2020. Participants were either household (n = 338) or non-household (n = 296) close contacts of COVID-19patients, as defined by the Spanish Ministry of Health [7]. COVID-19 diagnosis was made upon the presence of compatible signs or symptoms and a positive RT-PCR result in nasopharyngeal (NP) swabs. Timing of sample collection was prescribed at the discretion of either the physician in charge of the index case or local health authorities. The study was approved by the Hospital Clínico de Valencia (HCU) INCLIVA Research Ethics Committee.
SARS-CoV-2 testing
NPs for RAD and RT-PCR testing were collected by experienced nurses at the POC site located at Hospital Malvarrosa, as previously detailed [3]. For each patient, one swab, provided by the manufacturer, was taken from the left nostril for RAD testing, and another one, taken from the right nostril, was immediately placed in 3 mL of Universal Transport Medium (Becton Dickinson, Sparks, MD, USA) and used for RT-PCR testing. RAD testing was carried out at POC immediately after sampling. RT-PCRs were conducted within 24 hr of specimen collection at the Microbiology Service of Hospital Clínico Universitario (Valencia, Spain) with the TaqPath COVID-19 ComboKit (Thermo Fisher Scientific, MA, USA). The AMPLIRUN® TOTAL SARS-COV-2 Control (Vircell S.A., Granada, Spain) was used as the reference material for SARS-CoV-2 RNA load quantitation (in copies/mL, considering RT-PCR CTs for the N gene). An Standard curve was constructed using ten-fold serial dilution (102 to 109 copies/mL) of the reference material; the linear regression equation was: Y = –0.31 × X + 13.77; R
2 = 9.89).
Statistical analyses
Agreement between RAD and RT-PCR tests was assessed using Cohen's Kappa (κ) statistics. Differences between medians were compared using the Mann–Whitney U test. The chi-squared test was used for frequency comparisons. Two-sided p values < 0.05 were considered statistically significant. Statistical analyses were performed using SPSS version 25.0 (SPSS, Chicago, IL, USA).
Results
Overall performance of the RAD test in asymptomatic close contacts
A total of 79 out of 634 individuals (12.4%) tested positive by RT-PCR, of whom 38 (48.1%) returned positive RAD test results. There were no RAD positive/RT-PCR negative cases. Accordingly, concordance between RT-PCR and RAD results was moderate (κ index, 0.61; 95% CI 0.5–0.73). As shown in Fig. 1
, SARS-CoV-2 RNA load in NPs was significantly higher (p < 0.001) in RAD-positive (median, 8.7 log10 copies/ml) than in RAD-negative individuals (4.9 log10 copies/mL).
Fig. 1
RT-PCR cycle thresholds (Ct) (A) and SARS-CoV-2 RNA load (B) in asymptomatic close contacts of COVID-19 patients testing either positive or negative by Panbio™ COVID-19 Ag Rapid Test Device (RAD). The AMPLIRUN® TOTAL SARS-CoV-2 Control (Vircell S.A., Granada, Spain) was used as the reference material for SARS-CoV-2 RNA load quantitation (in copies/mL, considering RT-PCR Cts for the N gene). p Values for comparisons are shown.
RT-PCR cycle thresholds (Ct) (A) and SARS-CoV-2 RNA load (B) in asymptomatic close contacts of COVID-19patients testing either positive or negative by Panbio™ COVID-19 Ag Rapid Test Device (RAD). The AMPLIRUN® TOTAL SARS-CoV-2 Control (Vircell S.A., Granada, Spain) was used as the reference material for SARS-CoV-2 RNA load quantitation (in copies/mL, considering RT-PCR Cts for the N gene). p Values for comparisons are shown.Overall sensitivity and specificity of RAD was 48.1% and 100% (Table 1
). For the above-mentioned prevalence (12.4%), the negative predictive value (NPV) of the RAD test was 93.1.5%. As expected, overall RAD sensitivity was directly related to SARS-CoV-2 load in NP specimens (Table S1), reaching 96.8% when specimens with viral load ≥7.4 log10 copies/mL (Ct ≤ 20) were analysed separately.
Table 1
Performance of the Panbio™ COVID-19 Ag Rapid Test Device for SARS-CoV-2 detection in asymptomatic household and non-household close contacts
Parameter
Population group
All individuals
Household contacts
Non-household contacts
Sensitivity % (95% CI)
48.1 (37.4–58.9)
50.8 (38.9–62.5)
35.7 (16.3–61.2)
Specificity% (95% CI)
100 (99.3–100)
100 (98.6–100)
100 (98.7–100)
Negative predictive valuea
93.1 (90.8–94.9)
89.5 (85.9–92.5)
96.9 (94.2-98.4)
Positive predictive valuea
100 (90.8–100)
100 (89.6–100)
100 (56.6–100)
Adjusted to actual prevalence in the respective population group.
Performance of the Panbio™ COVID-19 Ag Rapid Test Device for SARS-CoV-2 detection in asymptomatic household and non-household close contactsAdjusted to actual prevalence in the respective population group.
Performance of RAD test in household and non-household asymptomatic close contacts
Household contacts (n = 338; median age, 36.5; range, 10–86 years; 175 female) were tested at a median of 2 days (range, 1–7) after diagnosis of the presumed index case. Sixty-five (19.2%) tested positive by RT-PCR, of whom 33 (50.7%) were positive by RAD test. The likelihood of obtaining either a positive or a negative RAD result was unrelated to the time elapsed since diagnosis of the index case (p 0.33).Non-household contacts (n = 296; median age, 38.5 years; range, 9–87 years; 180 female) were tested at a median of 6 days (range, 1–7) after self-reported exposure. Five individuals yielded RT-PCR-positive/RAD-positive results (1.6%) and 9 had RT-PCR-positive/RAD-negative results (3.0%). Overall, median time from exposure to testing was similar among individuals displaying either positive or negative RAD results (p 0.89).The agreement level between RT-PCR and RAD results was significantly higher (p < 0.001) for household (κ, 0.61; 95% CI 0.50–0.75) than for non-household (κ, 0.51; 95% CI 0.20–0.83) contacts. RAD sensitivity was significantly higher (p < 0.001) in household contacts, while the opposite was true for NPV (Table 1).SARS-CoV-2 RNA load was comparable (p 0.21) across household (median, 6.8 log10 copies/mL; range, 3.4–10.9) and non-household (median, 5.9 log10 copies/mL; range, 3.5–10.6) contacts, and was significantly higher (p < 0.001) in RAD-positive than in RAD-negative individuals, irrespective of the subcohort considered.
Clinical outcomes
Thirty-nine out of the 79 individuals testing positive by RT-PCR eventually became mildly symptomatic (49.3%), without requiring hospitalization. Individuals testing positive by RAD, either household or non-household contacts, were more likely (p < 0.001) to develop clinically compatible COVID-19 (30 out of 38) than those who did not (9 out of 41).
Discussion
In this field study, overall sensitivity of the Panbio™ COVID-19 Ag Rapid Test Device for identification of SARS-CoV-2-infected individuals among asymptomatic close contacts of confirmed COVID-19 cases was 48.1%, close to the figures reported by Linares et al. (54.5%) [4], Fenollar et al. (45.4%) [5] and Bulilete et al. (59.0%) [6], in apparently comparable cohorts. However, in two of these studies [4,5], the RAD test was carried out at a central laboratory, and timing of sample collection was not disclosed [4,5]. In the study by Bulilete et al. [5] most participants (70.6%) were tested within 5 days of exposure. Sensitivity of the Panbio™ test was lower than was previously found [[3], [4], [5], [6]] in symptomatic patients (around 80%), yet as reported for the latter patients, RAD sensitivity was directly related to the magnitude of SARS-CoV-2 RNA load in NP specimens. Such a striking difference might reflect dissimilarities across symptomatic and asymptomatic individuals in the kinetics of SARS-CoV-2 load in the upper respiratory tract [8,9]. While it is well known that SARS-CoV-2 load peaks around the time of symptoms onset in the former group [10,11], the timing is uncertain in asymptomatic cases.Interestingly, individuals testing positive by RAD were more likely to become (mildly) symptomatic than their negative counterparts, pointing to a pathogenetic link between SARS-CoV-2 RNA load and development of overt COVID-19.The strength of the current study is that it reflects the real-life performance of the RAD test at POC. Among its limitations are the relative low number of cases, and the possibility that samples were collected too early after exposure, particularly in non-household contacts, in whom RAD sensitivity was strikingly low. In this sense, Linares et al. [4] reported the sensitivity of the Panbio™ test as very low in close contacts at less than 7 days from exposure.In summary, we found the Panbio™ test to display low overall sensitivity in asymptomatic contacts of COVID-19patients, although it may identify those displaying higher SARS-CoV-2 RNA loads, who are presumed to be more contagious. Nevertheless, establishing the optimal timeframe for NP collection in household and non-household contacts seems crucial to accurately determine the sensitivity of the test.
Transparency declaration
This work received no public or private funds. The authors declare no conflicts of interest.
Author contributions
I.T., S.P. and E.A.: Methodology and data validation. E.A., I.T. and J.C.: Formal analysis. D.N.: Conceptualization, supervision, writing the original draft. All authors reviewed the original draft.
Authors: Jacqueline Dinnes; Pawana Sharma; Sarah Berhane; Susanna S van Wyk; Nicholas Nyaaba; Julie Domen; Melissa Taylor; Jane Cunningham; Clare Davenport; Sabine Dittrich; Devy Emperador; Lotty Hooft; Mariska Mg Leeflang; Matthew Df McInnes; René Spijker; Jan Y Verbakel; Yemisi Takwoingi; Sian Taylor-Phillips; Ann Van den Bruel; Jonathan J Deeks Journal: Cochrane Database Syst Rev Date: 2022-07-22
Authors: Anastasia Tapari; Georgia G Braliou; Maria Papaefthimiou; Helen Mavriki; Panagiota I Kontou; Georgios K Nikolopoulos; Pantelis G Bagos Journal: Diagnostics (Basel) Date: 2022-06-04
Authors: Christina C Bartenschlager; Selin Temizel; Alanna Ebigbo; Vivian Gruenherz; Petra Gastmeier; Helmut Messmann; Jens O Brunner; Christoph Römmele Journal: Value Health Date: 2022-06-01 Impact factor: 5.101
Authors: Lukas E Brümmer; Stephan Katzenschlager; Sean McGrath; Stephani Schmitz; Mary Gaeddert; Christian Erdmann; Marc Bota; Maurizio Grilli; Jan Larmann; Markus A Weigand; Nira R Pollock; Aurélien Macé; Berra Erkosar; Sergio Carmona; Jilian A Sacks; Stefano Ongarello; Claudia M Denkinger Journal: PLoS Med Date: 2022-05-26 Impact factor: 11.613
Authors: Marina Di Domenico; Alfredo De Rosa; Francesca Di Gaudio; Pietro Internicola; Cinzia Bettini; Nicola Salzano; Davide Castrianni; Andrea Marotta; Mariarosaria Boccellino Journal: Int J Environ Res Public Health Date: 2021-06-10 Impact factor: 3.390