Shima Dalirirad1, Daewoo Han1, Andrew J Steckl1. 1. Nanoelectronics Laboratory, Department of Physics, Department of Electrical Engineering and Computer Science, University of Cincinnati, Cincinnati, Ohio 45255-0030, United States.
Abstract
We have developed a disposable point-of-care (POC) aptamer-based biosensor for the detection of salivary cortisol. Nonstressful and noninvasive sampling of saliva compared to that of blood makes saliva an attractive biological matrix in developing POC devices for biomarker monitoring. Aptamers are attractive as recognition elements for multiple reasons, including their specific chemical synthesis, high stability, lack of immunogenicity, and cell-free evolution. A duplex aptamer conjugated to the surface of Au nanoparticles (AuNPs) by Au-S bonds is utilized as the sensor probe in a lateral flow assay (LFA) device. The addition of saliva samples containing cortisol makes the cortisol-aptamer undergo conformational changes and dissociate from the capture probe. Increasing cortisol concentration in the dispensed saliva sample results in increased dissociation and leads to increased binding of AuNP conjugate on the test line. Therefore, the color intensity of the test line on the LFA is a direct function of the concentration of cortisol in saliva. This simple and fast method provides detection in the cortisol range of ∼0.5-15 ng/mL, which is in the clinically accepted range for salivary cortisol. The limit of detection was 0.37 ng/mL, and the accuracy was confirmed by enzyme-linked immunosorbent assay (ELISA) testing results. High selectivity was observed for salivary cortisol against other closely related steroids and stress biomarkers present in saliva.
We have developed a disposable point-of-care (POC) aptamer-based biosensor for the detection of salivary cortisol. Nonstressful and noninvasive sampling of saliva compared to that of blood makes saliva an attractive biological matrix in developing POC devices for biomarker monitoring. Aptamers are attractive as recognition elements for multiple reasons, including their specific chemical synthesis, high stability, lack of immunogenicity, and cell-free evolution. A duplex aptamer conjugated to the surface of Au nanoparticles (AuNPs) by Au-S bonds is utilized as the sensor probe in a lateral flow assay (LFA) device. The addition of saliva samples containing cortisol makes the cortisol-aptamer undergo conformational changes and dissociate from the capture probe. Increasing cortisol concentration in the dispensed saliva sample results in increased dissociation and leads to increased binding of AuNP conjugate on the test line. Therefore, the color intensity of the test line on the LFA is a direct function of the concentration of cortisol in saliva. This simple and fast method provides detection in the cortisol range of ∼0.5-15 ng/mL, which is in the clinically accepted range for salivary cortisol. The limit of detection was 0.37 ng/mL, and the accuracy was confirmed by enzyme-linked immunosorbent assay (ELISA) testing results. High selectivity was observed for salivary cortisol against other closely related steroids and stress biomarkers present in saliva.
Point-of-care
(POC) devices are increasingly being utilized for
monitoring various health-care conditions in a convenient, easy-to-use,
and low-cost format.[1] This ranges from
pregnancy to flu testing to glucose monitoring. These POC platforms
are based on capillary flow through nitrocellulose (paper). This approach
has the advantage that it does not require an external force for fluid
flow, thus further simplifying the operation and reducing the physical
dimensions and the cost of manufacturing. Paper-based lateral flow
assay (LFA) devices use several recognition mechanisms (such as antibody/antigen,
aptamer/aptamer) for detecting the presence of targeted analytes and
displaying the results to the user (primarily as a color change).
Recent reviews of the field of paper-based microfluidics for medical
diagnostics include those by Yamada et al.[2] and Gong and Sinton.[3]The detection
of stress levels using POC devices is an area of
growing importance for both healthy individuals experiencing stressful
events (such as athletes, police, and armed forces personnel) and
individuals with health issues. Multiple stress-related biomarkers
(cortisol, dopamine, serotonin, etc.) present in biological fluids
(blood, sweat, saliva, urine, etc.) at various concentrations provide
an opportunity for developing microfluidic POC diagnostics (see the
review by Steckl and Ray[4]). A variety of
biomarker detection mechanisms have been employed for this purpose,[4] including colorimetric,[5] electrochemical,[6,7] and optical detection.[8]Targeting biomarkers found in blood has
the advantage of being
able to correlate with conventional and very accurate laboratory analysis,
but it suffers from the fact that it requires penetrating the skin
for obtaining samples, which may increase the stress level of patients,
and blood properties change rapidly with time after extraction from
the body because of coagulation. Therefore, the use of noninvasive
biofluids for biomarker detection is quite attractive.Aptamer-based
bioassays[9] are being increasingly
investigated in a wide variety of applications in targeted drug delivery,[10] molecular diagnostics,[11] biosensor systems,[12] and biomarker discovery.[13,14] Aptamers are single-stranded DNA or RNA ligands that are synthesized
toward a wide variety of targets, such as proteins,[15] peptides,[16] small ions,[17] and organic compounds.[18] Aptamers are selected from a DNA library by in vitro selection process
named systematic evolution of ligands by exponential enrichment (SELEX).
Compared to antibodies, aptamers as potential recognition elements
have several attractive characteristics: easy chemical synthesis,
adaptive modification, small size, high stability, lack of immunogenicity,
and cell-free evolution.[19] Aptamer-based
POC lateral flow assays (LFAs) have been reported for the detection
of stress biomarkers, such as dopamine in urine[20] and cortisol in sweat.[5]In this paper, we report on the detection of cortisol in saliva
using an aptamer recognition mechanism. Saliva is a complex fluid
composed of different peptides/proteins, hormones, nucleic acids,
lipids, electrolytes, ions, and several other components, which reflect
the body’s health and physiological conditions.[21] Saliva has received increasing attention in
POC detection technologies as it contains a wide array of biomarkers
with clinical relevance and liquid samples of sufficient volume can
be readily obtained. Salivary cortisol has a relatively high concentration,[22−24] ranging from 0.6 to 10.4 ng/mL, with an average normal level of
∼6 in the morning and 1 ng/mL in the evening. Cortisol is a
steroid hormone released by the adrenal cortex to regulate blood pressure,
cardiovascular function, and metabolic activities.[25] The body regulates cortisol production based on its level
in the bloodstream, an excess or deficiency of cortisol in the body
leading to Cushing’s syndrome and Addison’s disease,
respectively.[26,27] Salivary cortisol levels for
patients with Cushing’s syndrome are 12 ng/mL in the morning
and 9 ng/mL in the late night,[28] while
for patients with Addison’s disease, it shows 1.5 ng/mL level
in the morning.[29]Normal cortisol
levels in blood are high in the morning (50–230
ng/mL) and low at night (30–160 ng/mL).[4] Related to levels of cortisol in blood, salivary cortisol levels
also change during the day. Yamanaka et al. showed that salivary cortisol
changes during the day with the highest concentration (∼3.7
ng/mL) in the morning and lowest in the evening (0.3 ng/mL).[30] Cortisol levels increase in response to both
physical and psychological stress, such as illness, injury, or depression.[31,32] Monitoring cortisol levels has become an important diagnostic indicator
of stress and physiological state.Cortisol is found primarily
in a conjugated (bound) form with proteins
while circulating in the blood. The major binding proteins are corticosteroid-binding
globulin and albumin, which bind approximately 90 and 7% of cortisol,
respectively.[33] Cortisol is found as a
prominent biomarker molecule in the other main biofluids.[4] However, obtaining accurate cortisol levels in
urine is negatively affected by possible diuretic medication and pregnancy
status. For sweat, the collected sample volume is relatively low and
its biomarker levels are affected by skin conditions.[7] Saliva provides a simple noninvasive and nonstressful source
of sample fluid for analysis. Interestingly, salivary cortisol concentration
has been reported[34] to be independent of
the saliva generation rate and the amount of mucus present. Furthermore,
the transfer of free cortisol from blood to saliva has been reported[34] to reach equilibrium with a coefficient of 0.97
in ∼5 min. The strong correlation between the concentrations
of free cortisol in blood plasma and salivary cortisol indicates the
absence of cortisol-binding proteins in saliva.[35] Therefore, the measurement of salivary cortisol is a promising
approach to assess free cortisol in the blood.A variety of
methods have been reported for the detection of cortisol
in saliva, including chemiluminescent lateral flow immunoassay,[36] radioimmunoassay,[37] enzyme-linked immunosorbent assay,[38] surface
plasmon resonance,[39] liquid chromatography,[40] and antigen–antibody immunoreactions.[41] Lateral flow assays (LFAs) are simple paper-based
devices that meet all requirements expected from POC biosensors: low-cost,
user-friendly, one-step assay performance, high sensitivity, and selectivity
with a low limit of detection (LOD). In LFAs, the recognition antibodies
are immobilized on the nitrocellulose membrane, which provides a substrate
for both reaction and detection. Several groups[7,42−44] have demonstrated salivary biosensors in the clinically
accepted range using various detection elements and methods. Competitive
LFA devices using antibodies have been developed for salivary cortisol
detection, including a commercially available product, as listed in
Supporting Information Table S1. Because
cortisol is a very small molecule (362.46 Da), it does not provide
multiple binding sites and, therefore, the less-sensitive competitive
assay has been utilized rather than the more sensitive sandwich assay.
In the competitive assay, a test line is formed in the absence of
cortisol in the sample. To increase the sensitivity in the competitive
assay LFA, Zangheri et al. and Shirtcliff et al. adopted the chemiluminescence[36] and fluorescence[45] signal probes for the assay, respectively. However, these more complex
detection methods require additional considerations, such as enclosed
adaptors for the signal reader unit and/or the use of a smartphone
camera. For naked-eye detection, Apilux et al. adopted[42] the silver enhancement technique to precipitate
silver ions on gold nanoparticle surfaces, leading to a 3.6×
increase in the colorimetric signal intensity but requiring additional
steps after forming the test line on LFA devices. Oh et al. developed[43] the modified competitive assay with depletion
and detection zones, which gives higher sensitivity and wider detection
range than the conventional competitive assay by evaluating the ratio
of two zones. In this approach, signal interpretation can be less
intuitive, which causes a usability issue without the use of an external
analyzer.Here, we present an aptamer-based POC lateral flow
assay[46,47] that measures cortisol concentration in
saliva by providing a colorimetric
signal in the test strip. We have previously reported[5] the aptamer-based detection of cortisol in sweat using
the mechanism of salt aggregation. In this article, we report on the
aptamer-based detection of cortisol in saliva using the mechanism
of duplex DNA dissociation previously used[20] for dopamine detection in urine. Unlike the urine case, we have
optimized our protocol for the use in salivary detection, which is
more practical for frequent cortisol evaluation. For this work, in
addition to an improved detection mechanism, the sequence of the cortisol-binding
aptamer and the specificity of the DNA capture probes are optimized.
All analytical parameters were investigated to obtain high sensitivity,
selectivity, and reproducibility. This includes the optimization of
the loading density of aptamers on AuNP surfaces, the ratio of cortisol
aptamers (DNA2) to DNA capture probes (DNA1),
AuNP concentration, amount of NaCl needed for aging of the conjugated
AuNP–aptamers, amount of 6-mercapto-1-hexanol (MCH) for blocking
the free space on the AuNP surface, and the concentration of complementary
DNA immobilized on the biosensor.
Experimental
Section
Materials and Solutions
Cortisol-binding
DNA aptamer sequences with complementary biotinylated DNA probes,
shown in Table , and
nuclease-free (NF) water were purchased from Integrated DNA Technologies
(Coralville, Iowa). Au nanoparticles (40 nm diameter) were obtained
from Nanocomposix (San Diego, CA). Cortisol protein (≥98%),
progesterone (>99%), and β-estradiol (≥98%) were purchased
from Fitzgerald (Acton, MA). Sodium phosphate tribasic dodecahydrate
(Na3PO4·12H2O), sodium chloride,
magnesium chloride, Tween 20, Triton X-100, Tris (2-carboxyethyl)
phosphine (TCEP), bovine serum albumin (BSA), neuropeptide Y (NPY;
human, ≥95%), sodium saline citrate (SSC) 20×, 6-mercapto-1-hexanol
(MCH) and phosphate-buffered saline (PBS), Tris buffer (pH 8), nitrocellulose
(NC) membrane (Millipore HF135), cellulose fiber sample/wicking pad
(CFSP001700), Amicon ultrafilters, cysteamine, and sucrose were purchased
from MilliporeSigma (St. Louis, MO). Glass fiber pads (8950) were
purchased from Ahlstrom (Helsinki, Finland). Streptavidin was obtained
from IBA (Göttingen, Germany). Cortisol-competitive human enzyme-linked
immunosorbent assay (ELISA) kit was purchased from Thermo Fisher Scientific
(Waltham, MA).
CTC TCG GGA CGA CGC CCG CAT GTT CCA TGG ATA GTC TTG ACT AGT CGT CCC-
51
T DNA (test line)
biotin- CTC TCG GGA CGA C
13
C DNA (control line)
biotin- AAA AAA AAA AAA AAA AAA AA
20
Italic and underlined bases indicate
complementary
sequences.
Italic and underlined bases indicate
complementary
sequences.Artificial saliva
was obtained from Pickering Laboratories (Mountain
View, CA). The Pure·SAL saliva collection kit was purchased from
Oasis Diagnostics (Vancouver, WA). Human saliva samples were obtained
from healthy volunteers using Pure·SAL, and the samples were
treated prior to usage using Amicon ultrafilters.
Preparation of Duplex Aptamer–AuNP
Conjugates
The 40 nm AuNPs with initial optical density (OD
= 1) were concentrated to OD = 7.5. The consistency of concentrated
AuNPs was validated by UV–vis spectrum measurements at 525
nm (NanoDrop One spectrometer, Thermo Fisher Sci.). Conjugation of
duplex aptamer–AuNP was performed after hybridization between
DNA1 and DNA2 with a ratio of 2:1. DNA1 contains poly T and a short sequence that is a partial complement
to DNA2 (cortisol–aptamer). DNA1 (10
μM) and DNA2 (10 μM) were mixed at a 2:1 ratio
and then heated for 5 min at 95 °C. The mixture was cooled down
at room temperature (RT). After hybridization, 50 μL of thiolated
duplex aptamer (3.3 μM) is activated with 2.5 μL of TCEP
(20 mM) and then incubated for 1 h at RT. TCEP was used to reduce
the disulfide linkage between thiol and sulfhydryl groups in DNA1. Three microliters of the reduced-duplex aptamer was added
to 40 μL of concentrated AuNP solution under mild shaking for
1 h followed by incubation at RT for ∼8 h. Solutions
were aged with NaCl (1 M) to increase the aptamer loading density
without AuNP aggregation. NaCl was slowly added to solutions dropwise
until the final concentration of 70 mM (Figure S1), and then solutions were incubated at 4 °C overnight.
To block the unmodified sites and eliminate the nonspecific adsorption
on the surface of AuNPs, 2 μL of MCH (28 μM) was added
to 50 μL of aged AuNP–aptamer solution and incubated
for 1 h. Excess thiolated aptamer and MCH were removed by withdrawing
the supernatant after centrifugation at 8000 rpm for 10 min, and the
AuNP conjugates were resuspended in NF water. This washing process
was repeated three times. The final conjugated AuNP–aptamer
samples were then stored at 4 °C for future experiments.
Preparation of Streptavidin–Biotin–DNA
Conjugates
The test line DNA (T DNA) was designed to capture
the AuNP–DNA1 complex in the test zone. A streptavidin–biotinylated
T DNA conjugate was prepared to be immobilized on the test zone. Streptavidin
was dissolved in 10 mM PBS (pH 7.4) to a final concentration of 1
mg/mL. Streptavidin and T DNA (8μM) with a ratio of 1:3 was
incubated at RT for 1 h to provide stable binding. Sample solutions
were filtered (30 kDa, Millipore Amicon Ultra) for 30 min at 6000
rpm to remove the remaining free T DNA. Conjugated streptavidin–T
DNA, which remained in the filter due to the large molecular size
of streptavidin (55 kDa), was removed from the filter membrane and
resuspended with PBS up to its initial volume. The mixture was stored
at 4 °C for future use. To prepare the capture probe for the
control line, 40 μL of C DNA (10 μM) was mixed with 40
μL of streptavidin (1 mg/mL). After 1 h incubation in ambient
conditions, 60 μL of PBS (10 mM) and 20 μL of SSC 20×
were added to the mixture.
Fabrication of the Lateral
Flow Test Strips
Test strips are composed of four key components:
sample pad, blocking
pad, nitrocellulose membrane, and wicking pad. These are mounted on
an adhesive backing card with 2 mm overlaps to enable smooth fluid
flow between components, as shown in Figure . We used nitrocellulose (NC) membranes (Millipore
HF135) with a 2.5 cm length. Cellulose fiber membranes (CFSP001700,
MilliporeSigma) were used for both the sample pad and the wicking
pad at the end of the test strip. The sample pad was soaked in a buffer
solution (0.15 mM NaCl, 0.05 M Tris, and 0.25% Triton X-100, pH =
8.0) to facilitate the flowing of the sample solution onto the test
strip. The purpose of the “blocking” pad located between
the sample pad and the NC membrane is to release stored blocking agents
that will reduce nonspecific binding of conjugated AuNPs to the NC
membrane as fluid flows through the membrane. Blocking pads were prepared
using a soaked glass fiber membrane (8950, Ahlstrom) with blocking
agents (5% BSA, 20 mM Na3PO4, 10% sucrose, and
0.25% Tween 20). Both sample and blocking pads were soaked for 30
min in their respective buffers and then dried at 45 °C for 90
min before the assembly of the LFA device.
Figure 1
Design of the aptamer-based
lateral flow assay for the detection
of cortisol in saliva: (a) the configuration of the test strip; (b)
binding and colorimetric signal in the presence of cortisol (positive
test); and (c) binding and colorimetric signal in the absence of the
cortisol (negative test).
Design of the aptamer-based
lateral flow assay for the detection
of cortisol in saliva: (a) the configuration of the test strip; (b)
binding and colorimetric signal in the presence of cortisol (positive
test); and (c) binding and colorimetric signal in the absence of the
cortisol (negative test).For immobilizing the test and control DNAs on the NC membrane,
streptavidin was applied as a fixing agent to react with the biotinylated
DNA. T DNA and C DNA were dispensed on the nitrocellulose membrane
by a Biojet AD1500 (BioDot, Irvine, CA) to form the test and control
lines with a separation of 5 mm. Next, the membranes were dried at
65 °C for ∼10 min and stored in a nitrogen-filled box.
C DNA immobilized on the control line can capture AuNP conjugated
with DNA1 to confirm the validity of the test strip. An
automatic strip guillotine cuter (CM4000, BioDot, Irvine, CA) was
used to cut individual 5 mm wide assembled LFA strips.Test line location was evaluated to form consistent test lines
while minimizing any negative interaction between AuNP/aptamer conjugates
and the nitrocellulose membrane. While a location closer to the sample/buffer
pad was preferred to avoid any negative effect on AuNP/aptamer conjugates,
the test line formation was not consistent when it is located too
close to the sample/buffer pad (Figure S2). Also, 80 μL of sample solution was selected because it is
a minimum volume that can reach the wicking pad consistently.
Detection Process on Test Strips
Cortisol stock solution
was prepared by dissolving 1 mg in 100 μL
of ethanol and 900 μL of saliva (real or artificial). Sample
solutions containing desired concentrations of cortisol were mixed
with conjugated AuNP–duplex DNA and incubated for 30 min at
37 °C. An 80 μL portion of the mixture was dispensed onto
the sample pad. Distinct red bands on the LFA strip appeared in ∼15
min. Then, the images of the test strips were analyzed using ImageJ
analysis software.[49] The output peak area
was utilized to quantify the cortisol concentration in the sample.
Results and Discussion
Principle
of Cortisol Detection by LFA Biosensor
The principle for
the aptamer-based lateral flow biosensor to detect
cortisol in saliva is based on DNA duplex dissociation in the presence
of the target analyte.[50] In the presence
of cortisol, DNA2 on the conjugated AuNP–duplex
aptamer (Figure a)
binds with cortisol. In the process, DNA2 undergoes conformational
changes and is dehybridized from DNA1, as shown in Figure b. The concentration
of duplex aptamer and its complementary aptamer on the test zone were
optimized to 10 and 8 μM, respectively.
Figure 2
Cortisol detection mechanism
based on duplex aptamer dissociation:
(a) conjugated AuNP–duplex DNA1–DNA2 and (b) DNA2 dissociation from capture strand in the
presence of cortisol molecules.
Cortisol detection mechanism
based on duplex aptamer dissociation:
(a) conjugated AuNP–duplex DNA1–DNA2 and (b) DNA2 dissociation from capture strand in the
presence of cortisol molecules.Two different complementary sequences of DNA1 were immobilized
onto the NC membrane forming the test (T DNA) and control (C DNA)
lines, respectively (Figure ). As illustrated in Figure , when the sample solution is dispensed to the sample
pad, the solution wets the sample pad and moves toward the wicking
pad driven by the capillary action. The dissociated AuNP–DNA1 molecules released by the cortisol–aptamer interaction
are free to be captured by the T DNA on the test line through binding
between complementary sequences (Table ). Therefore, the color intensity on the test line
is directly related to the concentration of cortisol in the solution.
As the liquid continues to flow through the NC membrane, excess free
and/or duplex AuNP–DNA1 aptamers are captured by
C DNA through 20T–20A binding resulting in the formation of
a red band on the control line.
Cortisol
in Artificial Saliva
The
analytical performance for the detection of salivary cortisol was
studied by adding known amounts of cortisol standard solutions (0,
1, 2, 4, 6, 8, 10, and 15 ng/mL) to artificial saliva. Artificial
saliva was diluted 10× in NF water due to its high viscosity,
which did not allow the solution to flow through the NC membrane.
As shown in Figure , it is observed that the test line peak intensity increases monotonically
with increasing cortisol concentration, up to 10 ng/mL. The cortisol
in artificial saliva was detectable as low as 2 ng/mL by the naked
eye.
Figure 3
Test to control (T/C) line ratio obtained by ImageJ analysis to
measure cortisol concentrations (0–15 ng/mL) in artificial
saliva at 15 min after dispensing sample solutions. Error bars indicate
the standard errors of three independent experiments (n = 3). Inset: image of corresponding lateral flow strips tested with
different concentrations of cortisol (from left to right: 0, 1, 2,
4, 6, 8, 10, 15 ng/mL).
Test to control (T/C) line ratio obtained by ImageJ analysis to
measure cortisol concentrations (0–15 ng/mL) in artificial
saliva at 15 min after dispensing sample solutions. Error bars indicate
the standard errors of three independent experiments (n = 3). Inset: image of corresponding lateral flow strips tested with
different concentrations of cortisol (from left to right: 0, 1, 2,
4, 6, 8, 10, 15 ng/mL).Further, using ImageJ,
the corresponding results were analyzed
by measuring the average green channel intensity of the test/control
line (nominally red color) against that of the background nitrocellulose
(white) surrounding it. Color intensities were averaged vertically
and horizontally across both test and background regions to acquire
area averages. The green channel of the RGB color space was found
to provide superior signal resolution compared to that of the red
channel and to grayscale measurements due to the nitrocellulose (white)
having high red values in the RGB color space. Therefore, test/control
line formation (red) correlates to a decrease in the green and blue
channels as opposed to an increase in the red channel. This makes
the green (and blue) channel measurements superior to grayscale measurements
(average of red, green, and blue channels) due to the removal of the
nonchanging red channel. Consequently, a limit of detection <1
ng/mL is achieved for cortisol in artificial saliva using this technique.The selectivity of the aptamer-based LFA (aptasensor) was also
evaluated by testing with several counter-targets and stress biomarkers:
progesterone, β-estradiol, and neuropeptide Y (NPY). The normal
physiological concentration of salivary NPY is ∼0.05 ng/mL.[4] Studies of salivary progesterone show a level
of 47 pg/mL in the follicular phase increasing to 94–235 pg/mL
in the postovulatory period.[51] The mean
salivary β-estradiol concentrations reported by Ham et al.[52] were 9.0 ± 2.6 and 11.7 ± 4.1 pg/mL
in early follicular phase and during ovulation, respectively.The steroid hormone stock solutions were prepared at a concentration
of 1 mg/mL in 10% DMSO and 90% diluted artificial saliva. The concentrations
of all biomarkers were set at 100 ng/mL in diluted artificial saliva.
As shown in Figure , a significant signal response was observed only toward cortisol,
while other biomarkers presented minor signals comparable to the control
even at these concentrations extremely higher than their normal physiological
levels, indicating the excellent selectivity of this aptamer-based
lateral flow biosensor for cortisol in saliva.
Figure 4
Selectivity of the aptamer-based
LFAs for detecting cortisol vs
other biomarkers in artificial saliva (100 ng/mL, 80 μL) at
15 min after dispensing sample solutions, average, and standard error
for n = 3 replicates.
Selectivity of the aptamer-based
LFAs for detecting cortisol vs
other biomarkers in artificial saliva (100 ng/mL, 80 μL) at
15 min after dispensing sample solutions, average, and standard error
for n = 3 replicates.
Salivary
Cortisol Measurements
For
the present study, whole saliva samples were collected from four healthy
adult volunteers in our laboratory using the Pure·Sal saliva
kit[53] (Oasis Diagnostics, Vancouver). The
Pure·Sal kits were used according to the manufacturer’s
instructions by resting the absorbent swab inside the mouth until
it was saturated (5 min). Then, the swab was placed into a syringe
that contains a filter for immediate compression and filtration. Volunteers
were asked to donate morning saliva samples before eating and brushing/rinsing
their mouths. For the collection of afternoon saliva samples, volunteers
were asked to rinse their mouth and wait for at least 10 min before
using Pure·Sal. The saliva samples were initially kept at −20
°C and then transferred into a −80 °C freezer for
future study. Saliva is a complex biological fluid that contains large
proteins like mucins that could interfere with the assay interactions
and contribute to what is known as the matrix effect of saliva. UV–vis
optical absorption measurements were performed on the whole saliva
(WS) samples taken with/without Pure·Sal and compared to the
use of a 50 K Amicon ultrafilter. As shown in Figure , the large peak at λ = 285 nm indicating
protein absorbance is significantly reduced using Pure·Sal, reaching
a value closer to that obtained using the 50 K Amicon ultrafilters.
Figure 5
UV absorbance
of whole human saliva samples without filtering and
after using either a Pure·Sal collector or a 50 K Amicon ultrafilter.
UV absorbance
of whole human saliva samples without filtering and
after using either a Pure·Sal collector or a 50 K Amicon ultrafilter.Although the Pure·Sal collector removes a
significant proportion
of large molecules in human saliva samples, the presence of some remaining
proteins still interferes with the assay performance on test strips.
High concentrations of ions or proteins in biological samples like
saliva increase van der Waals attractive forces between particles,
which induce aggregations of AuNPs in our devices.[54] We have investigated Amicon ultrafilters with various mass
cutoff values (50, 30, 10, and 3 K) for saliva samples to determine
the optimum molecular cutoff needed to allow assay interactions. Morning
(am) and afternoon (pm) saliva samples were centrifuged at 6000 rpm
for 30 min. One hundred microliters of filtrated saliva with different
molecular cutoff levels were introduced into the 3 μL conjugated
AuNP–aptamer. As shown in Figure S3, filtering the saliva samples with a 10 K mass filter prevented
AuNP aggregation (maintaining the red color), while filtering with
the 50 K filter mostly resulted in aggregation (indicated by a purple/dark
blue color). To overcome the protein and ion variations in saliva
matrix, which exist among individuals and also in the same subjects
at various times, Amicon ultrafilters with the 10 K molecular cutoff
were used for pretreatment of all human saliva samples.To obtain
a benchmark for the concentration of cortisol in human
saliva samples, commercial enzyme-linked immunosorbent assay (ELISA)
kits were utilized (cortisol-competitive human ELISA kit, Thermo Fisher
Sc, Waltham, MA). Essentially zero salivary cortisol concentration
was measured by ELISA in all afternoon samples, while morning saliva
samples showed measurable detection results (Table S2).To evaluate the response of the prepared cortisol
aptasensor, cortisol-free
human saliva samples as determined using ELISA cortisol kits were
spiked with varying cortisol concentrations (0, 0.5, 1, 2, 3, 4, 5,
10, and 15 ng/mL) and then added to conjugated AuNP–aptamer
solutions. After 5 min of incubation at 37 °C, 80 μL of
the final solution was dispensed on the test strips. The optimized
assay allows the detection of salivary cortisol at very low concentrations,
with a visual LOD of 0.5 ng/mL, as shown in Figure .
Figure 6
Assay results for the detection of cortisol
in human saliva: (a)
ImageJ peak area intensity for images taken at 15 min after assay
start for saliva samples with cortisol concentrations from 0 to 15
ng/mL; error bars represent the standard error from four different
saliva samples, each performed in triplicate (n =
12); and (b) photographs of lateral flow strips tested with increasing
cortisol concentrations from top: 0, 0.5, 1, 2, 3, 4, 5, 10, and 15
ng/mL.
Assay results for the detection of cortisol
in human saliva: (a)
ImageJ peak area intensity for images taken at 15 min after assay
start for saliva samples with cortisol concentrations from 0 to 15
ng/mL; error bars represent the standard error from four different
saliva samples, each performed in triplicate (n =
12); and (b) photographs of lateral flow strips tested with increasing
cortisol concentrations from top: 0, 0.5, 1, 2, 3, 4, 5, 10, and 15
ng/mL.The numerical signal values from
ImageJ analysis were obtained
by subtracting the background color from the peak intensity of color
lines. The resulting graph in Figure a shows the values at different cortisol concentration
and the standard error calculated for each concentration from 12 replicated
experiments.
Calibration Curve
Since the PM saliva
samples were determined by ELISA to have no measurable cortisol amount,
they were used for generating the calibration curve for human saliva.
To quantify cortisol concentrations of the saliva samples, the color
intensity of both T and C lines was analyzed by ImageJ. Quantitatively,
the color intensity of the test line is directly related to the concentration
of cortisol in the sample. The T/C ratio for each cortisol concentration
(I) is then divided by the T/C ratio of the blank
case (I0, no added cortisol). Finally,
the normalized intensity (I/I0) is plotted vs cortisol concentration to generate a calibration
plot (Figure ). All
LFA tests have been implemented on LFA strips produced in the same
batch.
Figure 7
Calibration curves for cortisol obtained in artificial (LOD = 0.99
ng/mL) and human saliva (LOD = 0.37 ng/mL). Images were analyzed with
ImageJ software to quantify the signal intensity of the T and C lines
and plotted as described in Section .
Calibration curves for cortisol obtained in artificial (LOD = 0.99
ng/mL) and human saliva (LOD = 0.37 ng/mL). Images were analyzed with
ImageJ software to quantify the signal intensity of the T and C lines
and plotted as described in Section .Using the obtained calibration curve, the aptasensor showed an
LOD of 0.37 and 0.99 ng/mL for cortisol in human and artificial saliva,
respectively. Interestingly, this is similar to the reported[5] LOD for aptamer-based detection of cortisol in
sweat of ∼1 ng/mL.It is not totally clear at this point
what causes the different
signal intensities in human and artificial saliva. It is possible
that some critical components present (and not filtered) in human
saliva samples enhance the binding between cortisol and the aptamers.The limit of detection (LOD) was calculated based[55] on the value of the limit of blank (LOB), which is the
highest apparent analyte concentration for blank samples (no analyte).
The LOD was calculated using the following equationsThe current working range of the LFA assay
is from ∼0.5 to ∼15 ng/mL, which covers the normal range
of cortisol in saliva of ∼1–10 ng/mL. Signal saturation
of the test line intensity is present for concentrations > 10 ng/mL,
possibly due to the absence of additional available binding sites
on the test line. In the current work, the experimental conditions
were optimized to increase the test strip sensitivity under normal
concentrations of salivary cortisol. For highly stressful conditions,
cortisol levels can be higher than the normal range. To quantify the
cortisol concentration for individuals with persistently high levels,
one can either lower the sensitivity (e.g., lower AuNP concentration,
increasing the aptamer to AuNP ratio, increasing the complementary
aptamer length, etc.) or simply dilute the collected saliva samples
in the appropriate ratio.We assume that the presence of proteins
and ions in real saliva
provided positive matrix interference that led to lower LOD in real
saliva. Proteins and ions have provided higher intensities at both
T and C lines by inducing small aggregation on conjugated AuNP–DNA1. In addition, human saliva samples in the nitrocellulose
membrane presented the higher rate of interactions with the immobilized
reagents at the T and C lines.To investigate the accuracy of
the proposed aptasensor, morning
saliva samples from four volunteers were analyzed. The saliva samples
were pretreated with 10 K Amicon ultrafilters and then added to the
conjugated AuNP–aptamer solution. Following incubation at 37
°C for 5 min, samples were dispensed on the test strips. The
cortisol concentration of the saliva samples was calculated from the
intensity of the I/I0 ratio of test strips for each subject
and the real saliva calibration curve. The experimental results of
four independent measurements for each subject were compared to ELISA
measurements (n = 2) from the same saliva samples,
as shown in Figure . The data points and error bars represent the mean and relative
standard errors, respectively. A good correlation between the aptasensor
values and the ELISA analysis was achieved for all samples (98.6%),
with recovery values in the range from 89 to 114%.
Figure 8
Cortisol levels obtained
from aptasensor assay and standard ELISA
method. Cortisol concentrations were obtained from the calibration
curve in human saliva.
Cortisol levels obtained
from aptasensor assay and standard ELISA
method. Cortisol concentrations were obtained from the calibration
curve in human saliva.
Biosensor
Stability Analysis
To investigate
the sensor stability both in solution and during lateral flow operation,
we tested strips at different storage times. A set of 27 strips was
fabricated under identical conditions and stored in a nitrogen-filled
box at RT for 2 months. Cortisol at a concentration of 1 ng/mL in
artificial saliva was tested on three test strips at storage days
0, 5, 10, 15, 20, 25, 30, 40, and 60 to determine the immobilized
biochemical component stability. As shown in Figure a, the signal intensity of the test line
(peak area) on the test trip was essentially constant and without
loss of activity, with the average of signal intensity and SD of 11.1
and 0.18, respectively.
Figure 9
Stability test of the biosensor for a period
of 60 days: (a) ImageJ
signal from the test strips at 1 ng/mL cortisol and (b) conjugated
AuNP–aptamer solutions. Error bars indicate the standard errors
of three independent experiments.
Stability test of the biosensor for a period
of 60 days: (a) ImageJ
signal from the test strips at 1 ng/mLcortisol and (b) conjugated
AuNP–aptamer solutions. Error bars indicate the standard errors
of three independent experiments.UV–visible absorption spectra (Figure b) were obtained from samples (n = 3) of conjugated AuNP–aptamer solutions maintained at 4
°C for 2 months. The spectra exhibit nearly identical absorbance
peaks at λ = 260 and 525 nm, representing the DNA and the dispersed
AuNPs, respectively. This demonstrates the high stability of the prepared
solutions without any aggregation or degradation in a 2 month period.
Summary and Conclusions
We have presented
the first report of an aptamer-based lateral
flow biosensor to quantify cortisol levels in human saliva. The low-cost
and high-sensitivity LFA assay is intended as single-use, disposable
point-of-care test. The assay was designed based on the dissociation
of a duplex cortisol–aptamer upon the introduction of the target
cortisol in the clinically accepted concentration range. Increasing
concentration of the target in the sample leads to increased dissociation
of the duplex aptamer and higher intensity at the test line of the
LFA. Cortisol at various concentrations was added to cortisol-free
human saliva samples to obtain the limit of detection and unknown
concentration of human salivary cortisol based on the calibration
curve. Under optimal conditions, a detection range of the cortisol
concentration within 15 min was from 0.5 to 15 ng/mL and the LOD was
0.37 ng/mL in human saliva. This indicates that the aptamer-based
biosensor has similar LOD to antibody-based assays while being able
to benefit from the advantages of aptamers (such as lower cost, higher
stability, potential improvement with future SELEX procedures). The
fact that the calibration of the assays was performed against human
saliva (not buffer solution or artificial saliva) makes the results
more accurate. To further improve the LFA performance, more sensitive
and selective aptamers can be developed by repeating SELEX processes
more times with additional modifications. In this study, to eliminate
the matrix effect on AuNP aggregation, human saliva collection using
Pure·Sal collectors followed by centrifugation was implemented.
To develop an onsite detection biosensor, further studies need to
be done on the surface chemistry of AuNPs to increase their stability
in the presence of ionic solutions. One possible approach is optimizing
salt-aging conditions to further increase the aptamer density on AuNP
surfaces. Moreover, assay sensitivity could be enhanced by incorporating
a different nanoparticle, such as upconverting phosphor (UCP), that
has a strong sample tolerance in various matrices. Studies have shown
that the sensitivity of the test trips incorporating UCP is increased
by 10–100× compared with using AuNPs.[56] The drawback of using UCPs is the need for an instrument
to read the result, which could be an obstacle to performing POC testing
in certain situations. While here we report on the use of aptamer-based
detection of salivary cortisol, the approach could be applied for
rapid quantitative screening of cortisol and other biomarkers in various
biofluids.
Authors: Bo A G Jönsson; Birgitta Malmberg; Asa Amilon; Anne Helene Garde; Palle Orbaek Journal: J Chromatogr B Analyt Technol Biomed Life Sci Date: 2003-01-25 Impact factor: 3.205
Authors: Nako Nakatsuka; Kyung-Ae Yang; John M Abendroth; Kevin M Cheung; Xiaobin Xu; Hongyan Yang; Chuanzhen Zhao; Bowen Zhu; You Seung Rim; Yang Yang; Paul S Weiss; Milan N Stojanović; Anne M Andrews Journal: Science Date: 2018-09-06 Impact factor: 47.728