Initial testing of pegfilgrastim (Neulasta Onpro) on-body injector in multiple radiological imaging environments.

Zaiyang Long1, Anil Nicholas Kurup1, Nicole M Jensen1, Nicholas J Hangiandreou1, Beth A Schueler1, Lifeng Yu1, Shuai Leng1, Christopher P Wood1, Joel P Felmlee1.   

Abstract

PURPOSE: An increasing number of implantable or external devices can impact whether patients can receive radiological imaging examinations. This study examines and tests the Neulasta (pegfilgrastim) Onpro on-body injector in multiple imaging environments.
METHODS: The injector was analyzed for four imaging modalities with testing protocols and strategies developed for each modality. In x-ray and computed tomography (CT), scans with much higher exposure than clinical protocols were performed with the device attached to an anthropomorphic phantom. The device was monitored until the completion of drug delivery. For magnetic resonance imaging (MRI), the device was assessed using a hand-held magnet and underwent the magnetically induced displacement testing in a 1.5T clinical MRI scanner room. For ultrasound, magnetic field changes were measured around an ultrasound scanner system with three transducers.
RESULTS: For x-ray and CT no sign of device error was identified during or after the high radiation exposure scans. Drug delivery was completed at expected timing with expected volume. For MRI the device showed significant attractive force towards the hand-held magnet and a 50-degree deflection angle at 50 cm from the opening of the scanner bore. No further assessment from the gradient or radiofrequency field was deemed necessary. For ultrasound the maximum magnetic field change from baseline was measured to be +11.7 μT in comparison to +74.2 μT at 4 inches from a working microwave.
CONCLUSIONS: No device performance issue was identified under the extreme test conditions in x-ray or CT. The device was found to be MR Unsafe. Magnetic field changes around an ultrasound system met the limitation set by manufacture. Patient ultrasound scanning is considered safe as long as the transducers do not inadvertently loosen the device.
© 2021 The Authors. Journal of Applied Clinical Medical Physics published by Wiley Periodicals, Inc. on behalf of American Association of Physicists in Medicine.

Entities:  

Keywords:  CT; MRI; neulasta; pegfilgrastim; ultrasound; x-ray

Mesh:

Substances:

Year:  2021        PMID: 33395503      PMCID: PMC7856482          DOI: 10.1002/acm2.13156

Source DB:  PubMed          Journal:  J Appl Clin Med Phys        ISSN: 1526-9914            Impact factor:   2.102


  25 in total

1.  Pre-MRI procedure screening: recommendations and safety considerations for biomedical implants and devices

Authors: 
Journal:  J Magn Reson Imaging       Date:  2000-09       Impact factor: 4.813

2.  Safety of magnetic resonance imaging in patients with permanent pacemakers: a collaborative clinical approach.

Authors:  Barry Anthony Boilson; Anita Wokhlu; Nancy G Acker; Joel P Felmlee; Robert E Watson; Paul R Julsrud; Paul A Friedman; Yong-Mei Cha; Robert F Rea; David L Hayes; Win-Kuang Shen
Journal:  J Interv Card Electrophysiol       Date:  2011-09-21       Impact factor: 1.900

3.  [Effect of diagnostic X-rays on implantable cardiac pacemakers and implantable cardioverter defibrillators, and its management].

Authors:  Nobuhiro Oda; Hiroshi Nakajima; Haruhiko Abe; Syuji Koyama; Shingo Kakeda; Yukunori Kourogi
Journal:  Nihon Hoshasen Gijutsu Gakkai Zasshi       Date:  2008-07-20

4.  Variability in measurement of specific parameters for carotid duplex examination.

Authors:  T R Kohler; Y Langlois; G O Roederer; D J Phillips; K W Beach; J Primozich; R Lawrence; S C Nicholls; D E Strandness
Journal:  Ultrasound Med Biol       Date:  1987-10       Impact factor: 2.998

5.  Safety of magnetic resonance imaging in patients with legacy pacemakers and defibrillators and abandoned leads.

Authors:  Deepak Padmanabhan; Danesh K Kella; Ramila Mehta; Suraj Kapa; Abhishek Deshmukh; Siva Mulpuru; Allan S Jaffe; Joel P Felmlee; Mary L Jondal; Connie M Dalzell; Samuel J Asirvatham; Yong-Mei Cha; Robert E Watson; Paul A Friedman
Journal:  Heart Rhythm       Date:  2017-10-16       Impact factor: 6.343

6.  Febrile neutropenia hospitalization due to pegfilgrastim on-body injector failure compared to single-injection pegfilgrastim and daily injections with reference and biosimilar filgrastim: US cost simulation for lung cancer and non-Hodgkin lymphoma.

Authors:  Ali McBride; Andriy Krendyukov; Nicola Mathieson; Kim Campbell; Sanjeev Balu; Maja Natek; Karen MacDonald; Ivo Abraham
Journal:  J Med Econ       Date:  2019-09-03       Impact factor: 2.448

7.  Safety of magnetic resonance imaging in patients with cardiovascular devices: an American Heart Association scientific statement from the Committee on Diagnostic and Interventional Cardiac Catheterization, Council on Clinical Cardiology, and the Council on Cardiovascular Radiology and Intervention: endorsed by the American College of Cardiology Foundation, the North American Society for Cardiac Imaging, and the Society for Cardiovascular Magnetic Resonance.

Authors:  Glenn N Levine; Antoinette S Gomes; Andrew E Arai; David A Bluemke; Scott D Flamm; Emanuel Kanal; Warren J Manning; Edward T Martin; J Michael Smith; Norbert Wilke; Frank S Shellock
Journal:  Circulation       Date:  2007-11-19       Impact factor: 29.690

Review 8.  Magnetic resonance imaging safety in pacemaker and implantable cardioverter defibrillator patients: how far have we come?

Authors:  Peter Nordbeck; Georg Ertl; Oliver Ritter
Journal:  Eur Heart J       Date:  2015-03-21       Impact factor: 29.983

9.  Do we need to establish guidelines for patients with neuromodulation implantable devices, including spinal cord stimulators undergoing nonspinal surgeries?

Authors:  Ramsis F Ghaly; Tatiana Tverdohleb; Kenneth D Candido; Nebojsa Nick Knezevic
Journal:  Surg Neurol Int       Date:  2016-02-15

10.  Evaluation of projection- and dual-energy-based methods for metal artifact reduction in CT using a phantom study.

Authors:  Zaiyang Long; Michael R Bruesewitz; David R DeLone; Jonathan M Morris; Kimberly K Amrami; Mark C Adkins; Katrina N Glazebrook; James M Kofler; Shuai Leng; Cynthia H McCollough; Joel G Fletcher; Ahmed F Halaweish; Lifeng Yu
Journal:  J Appl Clin Med Phys       Date:  2018-05-10       Impact factor: 2.102

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.