[Effect of diagnostic X-rays on implantable cardiac pacemakers and implantable cardioverter defibrillators, and its management].

Nobuhiro Oda1, Hiroshi Nakajima, Haruhiko Abe, Syuji Koyama, Shingo Kakeda, Yukunori Kourogi.   

Abstract

The effects of diagnostic X-rays on implantable cardiac pacemakers and implantable cardioverter defibrillators (ICDs) were investigated. A total of 33 pacemakers from six manufacturers and nine ICDs from four manufacturers were irradiated using several X-ray units (plain X-ray diagnostic unit, under-table system fluoroscopy unit, over-table system fluoroscopy unit, biplane cardiac digital angiography unit, DSA unit and cone-beam CT unit). No systematic reset phenomena were observed in any pacemakers and ICDs under the X-ray irradiation. Nevertheless, over-sensing associated with the lack of a few pulses was observed in three of 33 pacemakers under radiation exposure. It has been proven that diagnostic X-rays might affect pacemaker function. Since the duration of the over-sensing under the radiation was very short and included transient episodes, this sensing failure, therefore, induced by radiation exposure would not affect the health of pacemaker recipients. ICDs were not affected by radiation exposure at a clinical dose. It is recommended that caution be exercised in direct irradiation to the pacemaker. Using a copper sheet of 2.0 mm thickness or more prevented over-sensing of pacemakers in the present study, especially when serial X-ray exposures were used with an over-table fluoroscopy system or cone-beam CT unit.

Entities:  

Mesh:

Year:  2008        PMID: 18719297     DOI: 10.6009/jjrt.64.805

Source DB:  PubMed          Journal:  Nihon Hoshasen Gijutsu Gakkai Zasshi        ISSN: 0369-4305


  3 in total

Review 1.  [Perioperative management of patients with implanted pacemakers or cardioverter/defibrillators. Recommendations of the Austrian Society for Anaesthesiology, Resuscitation and Intensive Care Medicine, the Austrian Society for Cardiology and the Austrian Society for Surgery].

Authors:  H Gombotz; M Anelli Monti; N Leitgeb; M Nürnberg; B Strohmer
Journal:  Anaesthesist       Date:  2009-05       Impact factor: 1.041

2.  Computed Tomography Scan and ICD Interaction.

Authors:  Jose M Porres; Jose L Cerezuela; Oscar Luque; Pilar Marco
Journal:  Case Rep Med       Date:  2009-11-01

3.  Initial testing of pegfilgrastim (Neulasta Onpro) on-body injector in multiple radiological imaging environments.

Authors:  Zaiyang Long; Anil Nicholas Kurup; Nicole M Jensen; Nicholas J Hangiandreou; Beth A Schueler; Lifeng Yu; Shuai Leng; Christopher P Wood; Joel P Felmlee
Journal:  J Appl Clin Med Phys       Date:  2021-01-04       Impact factor: 2.102