| Literature DB >> 33392548 |
Walter A Schrading1, Ben McCafferty2, Jordan Grove2, David B Page3.
Abstract
OBJECTIVE: To determine the correlation between 3 lightweight portable pulse oximeter devices compared to a standard wall mount pulse oximetry device.Entities:
Keywords: COVID‐19/novel coronavirus; hypoxia; oxygen; oxygenation status; pulse oximetry; respiratory physiological phenomena
Year: 2020 PMID: 33392548 PMCID: PMC7771801 DOI: 10.1002/emp2.12292
Source DB: PubMed Journal: J Am Coll Emerg Physicians Open ISSN: 2688-1152
FIGURE 1The 3 portable pulse oximeters used in this study. From left to right: Santa Medical SM‐165 (SM), Walgreens’ OxyWatch C20 (OxyWatch), and Nonin Onyx II 9550 (Onyx)
Baseline population characteristics (N = 200)
|
|
|
|---|---|
|
| |
| Median (IQR; Range) | 58 (51–69; 21–97) |
|
| |
| White | 96 (48.5%) |
| African American | 101 (50.5%) |
| Other | 2 (1%) |
| Missing | 1 (0.5%) |
|
| |
| Male | 108 (54%) |
| Female | 92 (46%) |
|
| |
| Yes | 68 (34%) |
| No | 129 (64.5%) |
| Missing | 3 (1.5%) |
|
| |
| Yes | 4 (2%) |
| No | 196 (98%) |
|
| |
| Yes | 0 (0%) |
| No | 200 (100%) |
|
| |
| Median (IQR) | 0 (0–2.25) |
|
| |
| Yes | 60 (30%) |
| No | 140 (70%) |
|
| |
| Shortness of breath | 131(66%) |
| Chest pain | 78 (39%) |
| Exac COPD | 27(14%) |
| SpO2 ≤ 94% | 17(9%) |
CPAP, continuous positive airway pressure; Exac COPD, exacerbated chronic obstructive pulmonary disease; IQR, interquartile range.
FIGURE 2SpO2 measurements obtained by each respective device are shown here. Each data point on each device represents the oxygen reading for one patient. The oxygen percentage data shown includes outliers as well as mean lines and markers. SM, Santa Medical SM‐165; OxyWatch, Walgreens’ OxyWatch C20; Onxy, Nonin Onyx II 9550
Distribution of oxygen saturation and measured difference by instrument
| Measurement Instrument | ||||
|---|---|---|---|---|
| Measure | Control | OxyWatch | SM | Onyx |
| Total patients (N) | 200 | 198 | 200 | 200 |
| Oxygen saturation (%) | ||||
| Mean (SD)* | 96.08% (3.84) | 94.23% (4.70) | 95.02% (4.49) | 95.08% (3.90) |
| Min, Max SpO2 | 59%, 100% | 59%, 100% | 62%, 100% | 68%, 100% |
| Pearson Correlation Coefficient ( | Ref | 0.79 (< 0.001) | 0.76 (< 0.001) | 0.79 (0.001) |
| Difference in SpO2 (Control‐Measured; %) | ||||
| Mean (95% CI) | Ref | 1.87% (1.46–2.28) | 1.07% (0.65–1.48) | 1.00% (0.65–1.35) |
| Min, Max | Ref | −5%, 19% | −6%, 23% | −9%, 15% |
CI, confidence interval; SM, Santa Medical.
*Analysis of variance: P = 0 < 0.001; F value = 15.85.
Distribution of oxygen saturation and measured differences by instrument in patients with control SpO2 ≤ 94% (N = 60)
| Measurement Instrument | ||||
|---|---|---|---|---|
| Measure | Control | OxyWatch | SM | Onyx |
| Measured SpO2 | ||||
| Mean (SD) | 92.15% (4.57) | 90.25% (5.76) | 91.37% (5.10) | 91.60% (4.68) |
| Min, Max SpO2 | 59%, 94% | 59%, 99% | 62%, 98% | 68%, 100% |
| Pearson's Correlation Coefficient ( | Ref | 0.75 (0.001) | 0.79 (0.001) | 0.70 (0.001) |
| Difference in measured SpO2 (Control‐Measured) | ||||
| Mean (95% CI) | Ref | 1.95% (0.96–2.94) | 0.78% (−0.04–1.60) | 0.55% (−0.37–1.47) |
| Min, Max | Ref | −5%, 13% | −6%, 10% | −9%, 15% |
CI, confidence interval; SM, Santa Medical.
Analysis of variance: P < 0.0001, F value = 14.49.
FIGURE 3Bland‐Altman plots for Walgreens’ OxyWatch C20
FIGURE 5Bland‐Altman plots for Nonin Onyx II 9550
Comparison of test devices to standard bedside oximeter for detecting hypoxia (SpO2 ≤94%)
| Measurement Instrument | |||
|---|---|---|---|
| OxyWatch | SM | Onyx | |
| Hypoxia (≤94%) | |||
| Sensitivity (95% CI) | 92.17% (87.27–97.08) | 90.70% (85.69–95.71) | 92.06% (87.34–96.78) |
| Specificity (95% CI) | 60.00% (49.59–70.41) | 67.61% (56.72–78.49) | 67.57% (56.90–78.23) |
| Positive predictive value (95% CI) | 75.71% (68.61–82.82) | 83.57% (77.43–89.71) | 82.86% (76.61–89.10) |
| Negative predictive value (95% CI) | 85.00% (75.97–94.03) | 80.00% (69.88–90.12) | 83.33% (73.90–92.76) |
| McNemar's Test (Sensitivity)‐ | 0.001 (14.53–1) | 0.063 (3.46–1) | 0.0165 (5.76–1) |
Chi2, chi square; CI, confidence interval; DOF, degree of freedom; SM, Santa Medical.