Saverio Cosola1,2, Paolo Toti1,3, Enrico Babetto4, Ugo Covani3, Miguel Peñarrocha-Diago2, David Peñarrocha-Oltra2.
Abstract
BACKGROUND/Entities:
Keywords: Fatigue test; Ferrulized implant neck; Prosthesis fixation
Year: 2020 PMID: 33384826 PMCID: PMC7770321 DOI: 10.1016/j.jds.2020.08.002
Source DB: PubMed Journal: J Dent Sci ISSN: 1991-7902 Impact factor: 2.080
Figure 1Three configurations: (A) AL: abutment Level without ferrulized neck; (B) ILS: implant Level with ferrulized neck and screw-retained crown; (C) ILC: Implant Level with ferrulized neck and cement-retained crown. (B) Schematic of test set-up for systems (ISO14801:2016). Ltot: distance from the center of the hemisphere to the clamping plane, Lp: nominal bone level, Li: exposed screw threads. (C) Loading apparatus Instron ElectroPuls® E10000.
Implants and components used in the study with the ultimate failure load and maximum bending moment, and failure modes in a static load test. (A) AL: Abutment Level without ferrulized neck; (B) ILS: Implant Level with ferrulized neck and screw-retained crown; (C) ILC: Implant Level with ferrulized neck and cement-retained crown.
| Configuration | Fixture | Abutment | Material | Ultimate failure load (N) | Bending moment, (N mm) | Failure mode |
|---|---|---|---|---|---|---|
| AL | LA-ZT-380-130 | A-ABU-330-1 | CP-Ti grade 4 | 421.6 ± 12.5 | 2318.8 ± 68.8 | Deformation |
| ILS | LA-ZT-380-130 | A-ABU-330-1 | CP-Ti grade 4 | 362.8 ± 23.8 | 1995.4 ± 130.9 | Deformation |
| ILC | LA-ZT-380-130 | A-ABU-330-1 | CP-Ti grade 4 | 445.7 ± 23.6 | 2451.4 ± 129.8 | Deformation |
Figure 2View of load–displacement curves for (A) AL: Abutment Level without ferrulized neck; (B) ILS: Implant Level with ferrulized neck and screw-retained crown; (C) ILC: Implant Level with ferrulized neck and cement-retained crown. Permanent deformations occurred after maximum load/displacement of the loading device.
Figure 3Results of cyclic load test. The significant pair-wise difference between 2 groups was marked by asterisk (∗). The red point was the maximum value of the force for which the prosthetic system guarantees 5 millions of cycles. (A) AL: Abutment Level without ferrulized neck; (B) ILS: Implant Level with ferrulized neck and screw-retained crown; (C) ILC: Implant Level with ferrulized neck and cement-retained crown.
Figure 4Failure modes in a static load test: (A) AL: Abutment Level without ferrulized neck; (B) ILS: Implant Level with ferrulized neck and screw-retained crown; (C) ILC: Implant Level with ferrulized neck and cement-retained crown.