Abutments with reduced diameter for both cement and screw retentions: analysis of failure modes and misfit of abutment-crown-connections after cyclic loading.

Izabela Cristina Maurício Moris1, Adriana Cláudia Lapria Faria1, Ricardo Faria Ribeiro1, Renata Cristina Silveira Rodrigues1.   

Abstract

OBJECTIVE: The aim of this study was to analyze failure modes and misfit of abutments with reduced diameter for both cement and screw retentions after cyclic loading.
MATERIAL AND METHODS: Forty morse-taper abutment/implant sets of titanium were divided into four groups (N = 10): G4.8S-4.8 abutment with screw-retained crown; G4.8C-4.8 abutment with cemented crown; G3.8S-3.8 abutment with screw-retained crown; and G3.8C-3.8 abutment with cemented crown. Copings were waxed on castable cylinders and cast by oxygen gas flame and injected by centrifugation. After, esthetic veneering ceramic was pressed on these copings for obtaining metalloceramic crowns of upper canine. Cemented crowns were cemented on abutments with provisional cement (Temp Bond NE), and screw-retained crowns were tightened to their abutments with torque recommended by manufacturer (10 N cm). The misfit was measured using a stereomicroscope in a 10× magnification before and after cyclic loading (300,000 cycles). Tests were visually monitored, and failures (decementation, screw loosening and fractures) were registered. Misfit was analyzed by mixed linear model while failure modes by chi-square test (α = 0.05).
RESULTS: Cyclic loading affected misfit of 3.8C (P ≤ 0.0001), 3.8S (P = 0.0055) and 4.8C (P = 0.0318), but not of 4.8S (P = 0.1243). No differences were noted between 3.8S with 4.8S before (P = 0.1550) and after (P = 0.9861) cyclic loading, but 3.8C was different from 4.8C only after (P = 0.0015) loading. Comparing different types of retentions at the same diameter abutment, significant difference was noted before and after cyclic loading for 3.8 and 4.8 abutments. Analyzing failure modes, retrievable failures were present at 3.8S and 3.8C groups, while irretrievable were only present at 3.8S.
CONCLUSIONS: The cyclic loading decreased misfit of cemented and screw-retained crowns on reduced diameter abutments, and misfit of cemented crowns is greater than screw-retained ones. Abutments of reduced diameter failed more than conventional.
© 2016 John Wiley & Sons A/S. Published by John Wiley & Sons Ltd.

Entities:  

Keywords:  abutments; cyclic loading; dental implants; misfit

Mesh:

Substances:

Year:  2016        PMID: 26923302     DOI: 10.1111/clr.12816

Source DB:  PubMed          Journal:  Clin Oral Implants Res        ISSN: 0905-7161            Impact factor:   5.977


  2 in total

1.  In-vitro fatigue and fracture performance of three different ferrulized implant connections used in fixed prosthesis.

Authors:  Saverio Cosola; Paolo Toti; Enrico Babetto; Ugo Covani; Miguel Peñarrocha-Diago; David Peñarrocha-Oltra
Journal:  J Dent Sci       Date:  2020-08-27       Impact factor: 2.080

2.  Influence of Manufacturing Methods of Implant-Supported Crowns on External and Internal Marginal Fit: A Micro-CT Analysis.

Authors:  Izabela C M Moris; Silas Borges Monteiro; Raíssa Martins; Ricardo Faria Ribeiro; Erica A Gomes
Journal:  Biomed Res Int       Date:  2018-01-23       Impact factor: 3.411