Literature DB >> 33384288

Postinterventional Assessment after Stent and Flow-Diverter Implantation Using CT: Influence of Spectral Image Reconstructions and Different Device Types.

C Zaeske1, T Hickethier2, J Borggrefe2, L Goertz2,3, R Dettmeyer4, M Schlamann2, N Abdullayev2, C Kabbasch2.   

Abstract

BACKGROUND AND
PURPOSE: CTA provides a noninvasive alternative technique to DSA in the follow-up after endovascular aneurysm treatment to evaluate aneurysm occlusion and exclude intraluminal narrowing after stent or flow-diverter implantation; however, assessability may be impeded by stent material artifacts. The objective of this in vitro study was to compare the visual assessability of different conventional stents and flow diverters as well as different reconstructions of dual-layer CT images.
MATERIALS AND METHODS: Four conventional intracranial stents and 4 flow diverters were implanted in identical aneurysm phantoms. Conventional and monoenergetic images (40, 50, 60, 90, 120, 180 keV) were acquired to evaluate attenuation alteration, visible lumen diameter, and SNR. Image quality was rated subjectively by 2 independent radiologists using a 4-point Likert scale.
RESULTS: Low kiloelectron volt (40-60 keV) monoenergetic reconstructions showed an improved SNR and an improved lumen density ratio compared with high kiloelectron volt reconstructions (90-180 keV) and conventional reconstructions, however without reaching significance compared with the latter. Assessment of the adjacent aneurysm and subjective evaluation was not affected by the imaging technique and stent type. Artifact susceptibility varied with the device used and increased among flow diverters.
CONCLUSIONS: Low kiloelectron volt reconstructions improved the assessment of the stent lumen in comparison with high kiloelectron volt reconstructions. No significant improvement in image quality could be shown compared with conventional images. For some devices, iodine-specific reconstructions led to severe artifacts and are therefore not recommended. There was no relevant improvement in the assessability of the adjacent aneurysm.
© 2021 by American Journal of Neuroradiology.

Entities:  

Mesh:

Year:  2020        PMID: 33384288      PMCID: PMC7959427          DOI: 10.3174/ajnr.A6952

Source DB:  PubMed          Journal:  AJNR Am J Neuroradiol        ISSN: 0195-6108            Impact factor:   3.825


  17 in total

1.  Fourth update on CT angiography of coronary stents: in vitro evaluation of 24 novel stent types.

Authors:  Tilman Hickethier; Justus Wenning; Jonas Doerner; David Maintz; Guido Michels; Alexander C Bunck
Journal:  Acta Radiol       Date:  2017-11-28       Impact factor: 1.990

2.  Comparison of Recanalization and In-Stent Stenosis Between the Low-Profile Visualized Intraluminal Support Stent and Enterprise Stent-Assisted Coiling for 254 Intracranial Aneurysms.

Authors:  Xin Feng; Zenghui Qian; Peng Liu; Baorui Zhang; Luyao Wang; Erkang Guo; Xiaolong Wen; Wenjuan Xu; Chuhan Jiang; Zhongxue Wu; Youxiang Li; Aihua Liu
Journal:  World Neurosurg       Date:  2017-09-27       Impact factor: 2.104

3.  Paediatric cerebrovascular CT angiography-towards better image quality.

Authors:  Stefanie C Thust; Wui Khean Kling Chong; Roxana Gunny; Asif Mazumder; Marius Poitelea; Anna Welsh; Ash Ederies; Kshitij Mankad
Journal:  Quant Imaging Med Surg       Date:  2014-12

4.  Stent assisted coiling versus non-stent assisted coiling for the management of ruptured intracranial aneurysms: a meta-analysis and systematic review.

Authors:  Xiaoxi Zhang; Qiao Zuo; Haishuang Tang; Gaici Xue; Pengfei Yang; Rui Zhao; Qiang Li; Yibin Fang; Yi Xu; Bo Hong; Qinghai Huang; Jianmin Liu
Journal:  J Neurointerv Surg       Date:  2019-03-06       Impact factor: 5.836

Review 5.  Computed tomography imaging and angiography - principles.

Authors:  Shervin Kamalian; Michael H Lev; Rajiv Gupta
Journal:  Handb Clin Neurol       Date:  2016

6.  Dual energy computed tomography of lung nodules: differentiation of iodine and calcium in artificial pulmonary nodules in vitro.

Authors:  Naomi Knöss; Beata Hoffmann; Bernhard Krauss; Martin Heller; Jürgen Biederer
Journal:  Eur J Radiol       Date:  2010-11-27       Impact factor: 3.528

7.  The p64 Flow Diverter-Mid-term and Long-term Results from a Single Center.

Authors:  S Sirakov; A Sirakov; P Bhogal; M Penkov; K Minkin; K Ninov; H Hristov; V Karakostov; R Raychev
Journal:  Clin Neuroradiol       Date:  2019-08-09       Impact factor: 3.649

8.  Subtraction CTA: An Alternative Imaging Option for the Follow-Up of Flow-Diverter-Treated Aneurysms?

Authors:  M P Duarte Conde; A M de Korte; F J A Meijer; R Aquarius; H D Boogaarts; R H M A Bartels; J de Vries
Journal:  AJNR Am J Neuroradiol       Date:  2018-10-04       Impact factor: 3.825

9.  Overview of Different Flow Diverters and Flow Dynamics.

Authors:  Georgios A Maragkos; Adam A Dmytriw; Mohamed M Salem; Vincent M Tutino; Hui Meng; Christophe Cognard; Paolo Machi; Timo Krings; Vitor Mendes Pereira
Journal:  Neurosurgery       Date:  2020-01-01       Impact factor: 4.654

10.  Evaluation of soft-plaque stenoses in coronary artery stents using conventional and monoenergetic images: first in-vitro experience and comparison of two different dual-energy techniques.

Authors:  Tilman Hickethier; Justus Wenning; Grischa Bratke; David Maintz; Guido Michels; Alexander C Bunck
Journal:  Quant Imaging Med Surg       Date:  2020-03
View more