Retrospective Distortion and Motion Correction for Free-Breathing DW-MRI of the Kidneys Using Dual-Echo EPI and Slice-to-Volume Registration.

Jaume Coll-Font1,2, Onur Afacan2,3, Scott Hoge2,4, Harsha Garg3, Kumar Shashi3, Bahram Marami5, Ali Gholipour2,3, Jeanne Chow2,3, Simon Warfield2,3, Sila Kurugol2,3.   

Abstract

BACKGROUND: Diffusion-weighted MRI (DW-MRI) of the kidneys is a technique that provides information about the microstructure of renal tissue without requiring exogenous contrasts such as gadolinium, and it can be used for diagnosis in cases of renal disease and assessing response-to-therapy. However, physiological motion and large geometric distortions due to main B0 field inhomogeneities degrade the image quality, reduce the accuracy of quantitative imaging markers, and impede their subsequent clinical applicability.
PURPOSE: To retrospectively correct for geometric distortion for free-breathing DW-MRI of the kidneys at 3T, in the presence of a nonstatic distortion field due to breathing and bulk motion. STUDY TYPE: Prospective.
SUBJECTS: Ten healthy volunteers (ages 29-38, four females). FIELD STRENGTH/SEQUENCE: 3T; DW-MR dual-echo echo-planar imaging (EPI) sequence (10 b-values and 17 directions) and a T2 volume. ASSESSMENT: The distortion correction was evaluated subjectively (Likert scale 0-5) and numerically with cross-correlation between the DW images at b = 0 s/mm2 and a T2 volume. The intravoxel incoherent motion (IVIM) and diffusion tensor (DTI) model-fitting performance was evaluated using the root-mean-squared error (nRMSE) and the coefficient of variation (CV%) of their parameters. STATISTICAL TESTS: Statistical comparisons were done using Wilcoxon tests.
RESULTS: The proposed method improved the Likert scores by 1.1 ± 0.8 (P < 0.05), the cross-correlation with the T2 reference image by 0.13 ± 0.05 (P < 0.05), and reduced the nRMSE by 0.13 ± 0.03 (P < 0.05) and 0.23 ± 0.06 (P < 0.05) for IVIM and DTI, respectively. The CV% of the IVIM parameters (slow and fast diffusion, and diffusion fraction for IVIM and mean diffusivity, and fractional anisotropy for DTI) was reduced by 2.26 ± 3.98% (P = 6.971 × 10-2 ), 11.24 ± 26.26% (P = 6.971 × 10-2 ), 4.12 ± 12.91% (P = 0.101), 3.22 ± 0.55% (P < 0.05), and 2.42 ± 1.15% (P < 0.05). DATA
CONCLUSION: The results indicate that the proposed Di + MoCo method can effectively correct for time-varying geometric distortions and for misalignments due to breathing motion. Consequently, the image quality and precision of the DW-MRI model parameters improved. LEVEL OF EVIDENCE: 2 TECHNICAL EFFICACY STAGE: 1.
© 2020 International Society for Magnetic Resonance in Medicine.

Entities:  

Keywords:  diffusion weighted imaging (DWI); distortion compensation; free-breathing; kidney imaging; motion compensation

Mesh:

Year:  2020        PMID: 33382173      PMCID: PMC8046730          DOI: 10.1002/jmri.27473

Source DB:  PubMed          Journal:  J Magn Reson Imaging        ISSN: 1053-1807            Impact factor:   4.813


  36 in total

1.  Reducing distortions in diffusion-weighted echo planar imaging with a dual-echo blip-reversed sequence.

Authors:  Daniel Gallichan; Jesper L R Andersson; Mark Jenkinson; Matthew D Robson; Karla L Miller
Journal:  Magn Reson Med       Date:  2010-08       Impact factor: 4.668

2.  Prospective pediatric study comparing glomerular filtration rate estimates based on motion-robust dynamic contrast-enhanced magnetic resonance imaging and serum creatinine (eGFR) to 99mTc DTPA.

Authors:  Sila Kurugol; Onur Afacan; Richard S Lee; Catherine M Seager; Michael A Ferguson; Deborah R Stein; Reid C Nichols; Monet Dugan; Alto Stemmer; Simon K Warfield; Jeanne S Chow
Journal:  Pediatr Radiol       Date:  2020-01-27

3.  Motion-robust parameter estimation in abdominal diffusion-weighted MRI by simultaneous image registration and model estimation.

Authors:  Sila Kurugol; Moti Freiman; Onur Afacan; Liran Domachevsky; Jeannette M Perez-Rossello; Michael J Callahan; Simon K Warfield
Journal:  Med Image Anal       Date:  2017-05-03       Impact factor: 8.545

4.  Reliable estimation of incoherent motion parametric maps from diffusion-weighted MRI using fusion bootstrap moves.

Authors:  Moti Freiman; Jeannette M Perez-Rossello; Michael J Callahan; Stephan D Voss; Kirsten Ecklund; Robert V Mulkern; Simon K Warfield
Journal:  Med Image Anal       Date:  2013-01-03       Impact factor: 8.545

5.  Estimation of the effective self-diffusion tensor from the NMR spin echo.

Authors:  P J Basser; J Mattiello; D LeBihan
Journal:  J Magn Reson B       Date:  1994-03

6.  Motion-Robust Spatially Constrained Parameter Estimation in Renal Diffusion-Weighted MRI by 3D Motion Tracking and Correction of Sequential Slices.

Authors:  Sila Kurugol; Bahram Marami; Onur Afacan; Simon K Warfield; Ali Gholipour
Journal:  Mol Imaging Reconstr Anal Mov Body Organs Stroke Imaging Treat (2017)       Date:  2017-09-09

7.  Spatially-constrained probability distribution model of incoherent motion (SPIM) for abdominal diffusion-weighted MRI.

Authors:  Sila Kurugol; Moti Freiman; Onur Afacan; Jeannette M Perez-Rossello; Michael J Callahan; Simon K Warfield
Journal:  Med Image Anal       Date:  2016-04-01       Impact factor: 8.545

8.  Simultaneous Motion and Distortion Correction Using Dual-Echo Diffusion-Weighted MRI.

Authors:  Onur Afacan; W Scott Hoge; Tess E Wallace; Ali Gholipour; Sila Kurugol; Simon K Warfield
Journal:  J Neuroimaging       Date:  2020-05-06       Impact factor: 2.486

9.  Dipy, a library for the analysis of diffusion MRI data.

Authors:  Eleftherios Garyfallidis; Matthew Brett; Bagrat Amirbekian; Ariel Rokem; Stefan van der Walt; Maxime Descoteaux; Ian Nimmo-Smith
Journal:  Front Neuroinform       Date:  2014-02-21       Impact factor: 4.081

10.  Diffusion Tensor Imaging of the Kidney: Design and Evaluation of a Reliable Processing Pipeline.

Authors:  Pasquale Borrelli; Carlo Cavaliere; Luca Basso; Andrea Soricelli; Marco Salvatore; Marco Aiello
Journal:  Sci Rep       Date:  2019-09-04       Impact factor: 4.379

View more
  1 in total

1.  Comparing the clinical utility of single-shot, readout-segmented and zoomit echo-planar imaging in diffusion-weighted imaging of the kidney at 3 T.

Authors:  Wenguang Liu; Hui Liu; Simin Xie; Ismail Bilal Masokano; Yu Bai; Xiao Wang; Linhui Zhong; Yi Wu; Jilin Nie; Gaofeng Zhou; Yigang Pei; Wenzheng Li
Journal:  Sci Rep       Date:  2022-07-20       Impact factor: 4.996

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.