Clinical experience of MRI4D QUASAR motion phantom for latency measurements in 0.35T MR-LINAC.

Taeho Kim1, Benjamin Lewis1, Rajiv Lotey2, Enzo Barberi3, Olga Green1.   

Abstract

PURPOSE: In MRgRT, accuracy of treatment depends on the gating latency, when real-time targeting and gating is enabled. Gating latency is dependent on image acquisition, processing time, accuracy, efficacy of target tracking algorithms, and radiation beam delivery latency. In this report, clinical experience of the MRI4D QUASAR motion phantom for latency measurements on a 0.35-T magnetic resonance-linear accelerator (MR-LINAC) with two imaging speeds and four tracking algorithms was studied. MATERIALS/
METHODS: Beam-control latency was measured on a 0.35-T MR-LINAC system with four target tracking algorithms and two real-time cine imaging sequences [four and eight frames per second (FPS)]. Using an MR-compatible motion phantom, the delays between phantom beam triggering signal and linac radiation beam control signal were evaluated for three motion periods with a rigid target. The gating point was set to be 8 mm above the full exhalation position. The beam-off latency was measured for a total of 24 combinations of tracking algorithm, imaging FPS, and motion periods. The corresponding gating target margins were determined using the target motion speed multiplied by the beam-off latency.
RESULTS: The largest measured beam-off latency was 302 ± 20 ms with the Large Deforming Targets (LDT) algorithm and 4 s motion period imaged with 8-FPS cine MRI. The corresponding gating uncertainty based on target motion speed was 3.0 mm. The range of the average beam-off latency was 128-243 ms in 4-FPS imaging and 47-302 ms in 8-FPS imaging.
CONCLUSIONS: The gating latency was measured using an MRI4D QUASAR motion phantom in a 0.35-T MR-LINAC. The latency measurements include time delay related to MR imaging method, target tracking algorithm and system delay. The gating uncertainty was estimated based on the beam-off latency measurements and the target motion.
© 2020 The Authors. Journal of Applied Clinical Medical Physics published by Wiley Periodicals, Inc. on behalf of American Association of Physicists in Medicine.

Entities:  

Keywords:  MR-LINAC; latency

Mesh:

Year:  2020        PMID: 33336884      PMCID: PMC7856488          DOI: 10.1002/acm2.13118

Source DB:  PubMed          Journal:  J Appl Clin Med Phys        ISSN: 1526-9914            Impact factor:   2.102


  13 in total

1.  First clinical implementation of real-time, real anatomy tracking and radiation beam control.

Authors:  Olga L Green; Leith J Rankine; Bin Cai; Austen Curcuru; Rojano Kashani; Vivian Rodriguez; H Harold Li; Parag J Parikh; Clifford G Robinson; Jeffrey R Olsen; Sasa Mutic; S M Goddu; Lakshmi Santanam
Journal:  Med Phys       Date:  2018-05-28       Impact factor: 4.071

2.  Characterization of imaging latency for real-time MRI-guided radiotherapy.

Authors:  P T S Borman; R H N Tijssen; C Bos; C T W Moonen; B W Raaymakers; M Glitzner
Journal:  Phys Med Biol       Date:  2018-08-06       Impact factor: 3.609

3.  Phase I trial of stereotactic MR-guided online adaptive radiation therapy (SMART) for the treatment of oligometastatic or unresectable primary malignancies of the abdomen.

Authors:  Lauren Henke; Rojano Kashani; Clifford Robinson; Austen Curcuru; Todd DeWees; Jeffrey Bradley; Olga Green; Jeff Michalski; Sasa Mutic; Parag Parikh; Jeffrey Olsen
Journal:  Radiother Oncol       Date:  2017-12-23       Impact factor: 6.280

4.  Technical Note: High temporal resolution characterization of gating response time.

Authors:  Rodney D Wiersma; Bradley P McCabe; Andrew H Belcher; Patrick J Jensen; Brett Smith; Bulent Aydogan
Journal:  Med Phys       Date:  2016-06       Impact factor: 4.071

5.  Technical Note: Evaluation of the latency and the beam characteristics of a respiratory gating system using an Elekta linear accelerator and a respiratory indicator device, Abches.

Authors:  Masahide Saito; Naoki Sano; Koji Ueda; Yuki Shibata; Kengo Kuriyama; Takafumi Komiyama; Kan Marino; Shinichi Aoki; Hiroshi Onishi
Journal:  Med Phys       Date:  2017-12-03       Impact factor: 4.071

6.  Audiovisual Biofeedback Improves Cine-Magnetic Resonance Imaging Measured Lung Tumor Motion Consistency.

Authors:  Danny Lee; Peter B Greer; Joanna Ludbrook; Jameen Arm; Perry Hunter; Sean Pollock; Kuldeep Makhija; Ricky T O'brien; Taeho Kim; Paul Keall
Journal:  Int J Radiat Oncol Biol Phys       Date:  2015-11-18       Impact factor: 7.038

7.  The management of respiratory motion in radiation oncology report of AAPM Task Group 76.

Authors:  Paul J Keall; Gig S Mageras; James M Balter; Richard S Emery; Kenneth M Forster; Steve B Jiang; Jeffrey M Kapatoes; Daniel A Low; Martin J Murphy; Brad R Murray; Chester R Ramsey; Marcel B Van Herk; S Sastry Vedam; John W Wong; Ellen Yorke
Journal:  Med Phys       Date:  2006-10       Impact factor: 4.071

8.  An analysis of thoracic and abdominal tumour motion for stereotactic body radiotherapy patients.

Authors:  Yelin Suh; Sonja Dieterich; Byungchul Cho; Paul J Keall
Journal:  Phys Med Biol       Date:  2008-06-17       Impact factor: 3.609

9.  Using adaptive magnetic resonance image-guided radiation therapy for treatment of inoperable pancreatic cancer.

Authors:  Soumon Rudra; Naomi Jiang; Stephen A Rosenberg; Jeffrey R Olsen; Michael C Roach; Leping Wan; Lorraine Portelance; Eric A Mellon; Anna Bruynzeel; Frank Lagerwaard; Michael F Bassetti; Parag J Parikh; Percy P Lee
Journal:  Cancer Med       Date:  2019-04-01       Impact factor: 4.452

10.  Audiovisual biofeedback guided breath-hold improves lung tumor position reproducibility and volume consistency.

Authors:  Danny Lee; Peter B Greer; Carminia Lapuz; Joanna Ludbrook; Perry Hunter; Jameen Arm; Sean Pollock; Kuldeep Makhija; Ricky T O'Brien; Taeho Kim; Paul Keall
Journal:  Adv Radiat Oncol       Date:  2017-03-10
View more
  1 in total

1.  Stability and Reliability of Enhanced External-Internal Motion Correlation via Dynamic Phase-Shift Corrections Over 30-min Timeframe for Respiratory-Gated Radiotherapy.

Authors:  Andrew Milewski; Guang Li
Journal:  Technol Cancer Res Treat       Date:  2022 Jan-Dec
  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.