Salla J Kiiskinen1, Oskari Luomala2, Teija Häkkinen3, Susanna Lukinmaa-Åberg4, Anja Siitonen1.
Abstract
Timely and reliable laboratory diagnostics is a necessity for patient safety and good patient management. Success in external quality assessment (EQA) reflects on the everyday work in a clinical laboratory. This study evaluated the reliability of serological point-of-care (POC) testing for the Epstein-Barr virus (EBV) that causes infectious mononucleosis (IM). Data from the results of 95 external quality control (EQC) samples, altogether 18 885 results during an eight-year period (2010-2017) were collected from 273 Finnish testing sites. Diagnosing acute infectious mononucleosis (EBV IM) is based on clinical, haematological and serological findings. Heterophile antibody tests are used for this purpose because they can be carried out at POC and are cheap and robust to perform. In this study, the data showed that the testing sites used 3 test methods and 17 different test kits; of the kits, 4 were used during the whole study period. The most commonly used test methods were immunochromatographic assays (12 test kits, 17 959 EQC results). Latex agglutination (4 test kits, 504 results) and immunofiltration test methods (one kit, 422 results) were also used. The overall success rate was 99.3% (for positive samples 99.6%, for negative samples 99.1%). The success rates of the different test methods varied from 94.3% for the immunofiltration method to 99.6% for the latex agglutination method. The lowest success rates were found for negative samples: 82.0% (QuickVue, Quidel [immunochromatographic method]), 91.3% (RDT EBV IgM Assay, Bio-Rad [immunofiltration method]). The results of the negative samples that represented old EBV immunity were the most difficult to interpret with a success rate of 98.9% compared to success rates of clearly positive (99.6%) and negative (99.5%) samples (P < .001). Especially the immunofiltration method (RDT EBV IgM Assay) produced 13.7% false positive results for samples of old immunity. The data showed that 42 of the studied 95 EBV IM EQA rounds were reported as expected (true positive or true negative) by all testing sites.Entities:
Keywords: Epstein-Barr virus; clinical laboratory; external quality assessment; external quality control; infectious mononucleosis; proficiency testing; serological point-of-care test
Year: 2020 PMID: 33328737 PMCID: PMC7720297 DOI: 10.1177/1178636120977481
Source DB: PubMed Journal: Microbiol Insights ISSN: 1178-6361
The external quality control results, test methods and used test kits during 2010–2017.
| Test methodTest kit (manufacturer) | Expected result | Reported results: true/all | Success % |
|---|---|---|---|
| Latex agglutination tests for heterophile antibodies | 502/504 | 99.6 | |
| Monogen[ | pos. | 132/133 | 99.2 |
| neg. | 177/178 | 99.4 | |
| Monospot[ | pos. | 64/64 | 100 |
| neg. | 102/102 | 100 | |
| Latex agglutination, other[ | pos. | 12/12 | 100 |
| neg. | 15/15 | 100 | |
| Immunochromatographic assays for heterophile antibodies | 17 855/17 959 | 99.4 | |
| Clearview[ | pos. | 4609/4618 | 99.8 |
| neg. | 7099/7148 | 99.3 | |
| InstAlert[ | pos. | 1800/1811 | 99.4 |
| neg. | 2964/2981 | 99.4 | |
| Mononucleosis[ | pos. | 235/238 | 98.7 |
| neg. | 454/456 | 99.6 | |
| Mnitop (All. Diag) | pos. | 87/87 | 100 |
| neg. | 162/162 | 100 | |
| Nadal Mononucleosis[ | pos. | 36/38 | 97.7 |
| neg. | 99/100 | 99.0 | |
| OSOM Mono Test (Sekisui Diagnostics) | pos. | 28/28 | 100 |
| neg. | 67/67 | 100 | |
| QuickVue (Quidel) | pos. | 26/26 | 100 |
| neg. | 41/50 | 82.0 | |
| Immunochromatographic, other[ | pos. | 57/57 | 100 |
| neg. | 91/92 | 98.9 | |
| Immunofiltration test for IgM antibodies | 398/422 | 94.3 | |
| RDT EBV IgM Assay[ | pos. | 146/146 | 100 |
| neg. | 252/276 | 91.3 | |
| All test methods | 18 755/18 885 | 99.3 | |
| 17 test kits | pos. | 7232/7258 | 99.6 |
| neg. | 11 523/11 627 | 99.1 |
The test kit was in use for the whole eight-year study period.
The test kit was only used in large laboratories.
Contains the results from 2 tests: Avitex-IM (Omega Diagnostics) and MNI Test (Fumouze).
Contains one possible sample mix-up.
Contains two possible sample mix-ups.
The test kit was only used in small laboratories.
Contains the results from 5 tests: MNITOP optima im (Biosynex), Diaquick Mononucleosis Cassette (Dialab), Mono Rapid Test Cassette (Hangzhou Alltest biotech), Immunocard Stat Mono (Meridian) and Mononucleosis Test Card (ulti med).
The test kit only had manufacturer validation for serum samples.
External quality control results according to the used test methods in different external quality control sample groups (I, II, III). Success was compared between sample groups.
| Test method | GROUP I | GROUP II | GROUP III | Total | ||||
|---|---|---|---|---|---|---|---|---|
| Positive, recent EBV infection | Success % | Negative, no EBV antibodies | Success % | Negative, old EBV immunity | Success % | Total | Success % | |
| Latex agglutination test | 208/209 | 99.5 | 105/106 | 99.1 | 189/189 | 100 | 502/504 | 99.6 |
| Immunochromatographic assay | 6 878/6 903 | 99.6 | 4203/4222 | 99.5 | 6774/6834 | 99.1 | 17 855/17 959 | 99.4 |
| Immunofiltration test[ | 146/146 | 100 | 126/128 | 98.4 | 126/148 | 86.3 | 398/422 | 94.3 |
| All test methods | 7232/7258 | 99.6 | 4434/4456 | 99.5 | 7089/7171 | 98.9 | 18 755/18 885 | 99.3 |
| .729 | .106 | <.001[ | ||||||
The test kit only had the manufacturer’s validation for serum samples.
Between Groups I and II.
The correct results from the negative Epstein-Barr virus infectious mononucleosis external quality control sample representing old EBV immunity (Sample 6/2010).
| Test methodTest kit (manufacturer) | Reported results | |
|---|---|---|
| True/all | Success % | |
| Latex agglutination tests for heterophile antibodies | 8/8 | 100 |
| Monogen (Biokit) | 6/6 | 100 |
| Monospot (Meridian) | 2/2 | 100 |
| Immunochromatographic assays for heterophile antibodies | 194/217 | 89.4 |
| Clearview (Unipath/Alere) | 150/172 | 87.2 |
| InstAlert (Innovacon) | 41/42 | 97.6 |
| Diaquick Mononucleosis Cassette (Dialab) | 3/3 | 100 |
| All test methods | 202/225 | 89.8 |