Seth Kwadjo Angmorterh1, Andrew England2, Sonia Aboagye3, Eric Kwasi Ofori1, Peter Hogg4.
Abstract
INTRODUCTION: The incidence of pressure ulcers (PUs) presents a substantial threat to patients, especially geriatric patients, those with restricted mobility, and patients suffering from chronic diseases such as cancer. PUs creates a huge financial burden on healthcare authorities and patients, costing billions to treat and manage. Radiography and radiotherapy patients may experience medical device related (MDR) PUs and studies have shown that high interface pressure (IP) values exist for the head when placed on an X-ray table without a mattress. These high IP values pose a PU risk to patients, especially those accessing prolonged radiography/radiology and radiotherapy procedures. The current study assessed the impact on IP values for the head from using a thin silicone gel surface overlay during radiographic procedures and identified whether this reduced the risk of PUs.Entities:
Year: 2020 PMID: 33299607 PMCID: PMC7710438 DOI: 10.1155/2020/3246531
Source DB: PubMed Journal: Radiol Res Pract ISSN: 2090-195X
Physical characteristics of the five static surface overlays.
| Name | Image | Material | Dimension ( | Weight (kg) |
|---|---|---|---|---|
| Gel table/hip pad |
| Gel | 45 × 45 × 1.4 | 3.4 |
| Silicone gel flat pad |
| Silicone gel | 45 × 45 × 0.7 | 1.4 |
| Blue hollow overlay |
| Elastic | 6 × 6 × 1.6 | <0.09 |
| Sundance SUN Z3-S |
| Fluidised | 18 × 18 × 1.7 | 0.2 |
| Small round gel |
| Gel | 8.5 × 8.5 × 0.6 | <0.09 |
∗L = length, W = width, and T = thickness.
Mean recorded dose at low and high tube potentials with corresponding percentage changes.
| Control (no overlay) | Mean dose for high kV (75 kV, 2 mAs) | 37.3 mGy | ||
|---|---|---|---|---|
| Mean dose for low kV (50 kV, 2 mAs) | 10.7 mGy | |||
| Surface overlay | Mean dose at 75 kV, 2 mAs | Percentage decrease (%) | Mean dose at 50 kV, 2 mAs | Percentage decrease (%) |
|
| ||||
| Gel table/hip pad | 32.1 mGy | 14.0 | 8.2 mGy | 23.4 |
| Silicone gel flat pad | 34.2 mGy | 8.3 | 9.6 mGy | 10.3 |
| Blue hollow overlay | 35.1 mGy | 5.9 | 10.3 mGy | 3.7 |
| Sundance SUN Z3-S | 30.0 mGy | 19.6 | 7.8 mGy | 27.0 |
| Small round gel | 34.3 mGy | 8.0 | 9.6 mGy | 10.3 |
Exposure and deviation indices for the surface overlays at high and low kVs.
| High kV (120, 1.2 mAs) | Low kV (60, 1.2 mAs) | |
|---|---|---|
| Control | EI–2940.10 | EI224.47 |
| TI–250 | TI–250 | |
| DI–10.70 | DI–−0.46 | |
|
| ||
| Gel table/hip pad | EI–2705.38 | EI–195.26 |
| TI 250 | TI–250 | |
| DI–10.34 | DI–−1.07 | |
|
| ||
| Silicone gel flat pad | EI–2926.91 | EI–213.16 |
| TI–250 | TI–250 | |
| DI–10.68 | DI–−0.69 | |
|
| ||
| Blue hollow surface | EI–2842.58 | EI–207.48 |
| TI–250 | TI–250 | |
| DI–10.55 | DI–−0.80 | |
|
| ||
| Sundance SUN Z3-S | EI–2773.14 | EI–197.91 |
| TI–250 | TI–250 | |
| DI–10.45 | DI–−1.01 | |
|
| ||
| Small round gel | EI–3020.51 | EI–222.46 |
| TI–250 | TI–250 | |
| DI–10.82 | DI–−0.50 | |
EI, TI, and DI all in units of microgray (μGy).
Radiographic images of the five surfaces at high and low kVs.
| Surface overlay | High kV | Low kV | High kV with hand | Low kV with hand | Artefact present |
|---|---|---|---|---|---|
| Gel table pad |
|
|
|
| Yes |
| Silicone gel flat pad |
|
|
|
| No |
| Blue hollow gel overlay |
|
|
|
| Yes |
| Sundance SUN Z3-S |
|
|
|
| Yes |
| Small round gel |
|
|
|
| Yes |
The mean Hounsfield unit (HU) with standard deviation (SD) for the various surface overlays.
| Surface no | Surface overlay | Mean HU ± SD |
|---|---|---|
| 1 | Gel table/hip pad | −0.67 ± 22.93 |
| 2 | Silicone gel flat pad | −12.54 ± 26.80 |
| 3 | Blue hollow surface | −508.96 ± 37.93 |
| 4 | Sundance SUN Z3-S | −0.7 ± 15.3 |
| 5 | Small round gel | −18.17 ± 7.06 |
Figure 1Comparison of mean IP and PPI with standard deviation on the X-ray table with and without the intervention.