Shielding effect of a lead apron on the peripheral radiation dose outside the applicator of electron beams from an Elekta linear accelerator.

Huifang He1, Yingdong Zhang1, Jidong Wang1, Xingyu Chen1, Yuyan Yang1, Junjie Wang2.   

Abstract

PURPOSE: To evaluate the shielding effect of lead aprons (LAs) on peripheral radiation doses outside the applicator of electron beams from a linear accelerator.
METHODS: Out-of-field radiation doses of 4-, 6-, 8-, 10-, 12-, and 15-MeV electron beams from an Elekta Synergy linear accelerator (linac) were measured by thermoluminescence dosimeters (TLD) at different depths (0, 0.5, 1.0, and 2.0 cm) and distances from the applicator edge (0-58 cm) in a water-equivalent slab phantom with a different number of layers of LA shielding (0-5 layers). Measurements were performed by 6 × 6, 10 × 10, 14 × 14, and 20 × 20-cm2 applicators at a gantry and collimator angle of 0°. The out-of-field radiation dose profiles were normalized to the maximum dose of every energy and measuring depth.
RESULTS: The out-of-field radiation doses (beyond 3 cm away from the field edge) decreased with an increase in the number of LA layers and distance away from the central beam axis (CAX). After shielding with the LA, the out-of-field doses decreased by up to approximately 99% compared with the no shielding group. For 4-MeV electron beams, there was a peak at 24.5 cm from the CAX, which weakened with an increasing number of LA layers.
CONCLUSION: The shielding effect of the LA varied for a different number of LA layers as well as different depths and distances away from the CAX. Four LA layers were sufficient for shielding out-of-field doses of 4-15-MeV electron beams.
© 2020 The Authors. Journal of Applied Clinical Medical Physics published by Wiley Periodicals, Inc. on behalf of American Association of Physicists in Medicine.

Entities:  

Keywords:  electron beams; out-of-field radiation; shielding effect of LA

Mesh:

Year:  2020        PMID: 33296548      PMCID: PMC7856487          DOI: 10.1002/acm2.13089

Source DB:  PubMed          Journal:  J Appl Clin Med Phys        ISSN: 1526-9914            Impact factor:   2.102


  15 in total

1.  The peripheral dose outside the applicator in electron beams of Oncor linear accelerator.

Authors:  Basak Iktueren; Hatice Bilge; Songul Karacam; Gulyuz Atkovar
Journal:  Radiat Prot Dosimetry       Date:  2011-10-23       Impact factor: 0.972

2.  Levels of leakage radiation from electron collimators of a linear accelerator.

Authors:  K R Das; J A Cramb; R M Millar; M B Kenny; W R Patterson; T L Ackerly; A E Perkins; K Chongkitivitya
Journal:  Med Phys       Date:  1990 Nov-Dec       Impact factor: 4.071

Review 3.  Review of electron beam therapy physics.

Authors:  Kenneth R Hogstrom; Peter R Almond
Journal:  Phys Med Biol       Date:  2006-06-20       Impact factor: 3.609

4.  The peripheral dose outside the applicator in electron beams of an Elekta linear accelerator.

Authors:  Abbas Haghparast; Fatemeh Amiri; Mehran Yarahmadi; Mohammad Rezaei
Journal:  Australas Phys Eng Sci Med       Date:  2018-06-25       Impact factor: 1.430

5.  AAPM TG 158: Measurement and calculation of doses outside the treated volume from external-beam radiation therapy.

Authors:  Stephen F Kry; Bryan Bednarz; Rebecca M Howell; Larry Dauer; David Followill; Eric Klein; Harald Paganetti; Brian Wang; Cheng-Shie Wuu; X George Xu
Journal:  Med Phys       Date:  2017-08-20       Impact factor: 4.071

6.  Evaluating out-of-field doses during radiotherapy of paediatric brain tumours using lead shielding and flattening-filter free beams.

Authors:  Anders Ravnsborg Beierholm; Ditte Eklund Nygaard; Elisabeth Lagoni Juhl; Rune Hansen; Jolanta Hansen
Journal:  Phys Med       Date:  2019-03-20       Impact factor: 2.685

7.  Out-of-field doses in radiotherapy: Input to epidemiological studies and dose-risk models.

Authors:  Roger Harrison
Journal:  Phys Med       Date:  2017-04-06       Impact factor: 2.685

8.  Dose Calculation Accuracy of Radiotherapy Treatment Planning Systems in Out-of-Field Regions.

Authors:  B Farhood; M Ghorbani
Journal:  J Biomed Phys Eng       Date:  2019-04-01

9.  Experimental verification of the application of lateral buildup ratio on the 4-MeV electron beam.

Authors:  James C L Chow; Scott Newman
Journal:  J Appl Clin Med Phys       Date:  2006-02-15       Impact factor: 2.102

10.  Out-of-field doses and neutron dose equivalents for electron beams from modern Varian and Elekta linear accelerators.

Authors:  Carlos E Cardenas; Paige L Nitsch; Rajat J Kudchadker; Rebecca M Howell; Stephen F Kry
Journal:  J Appl Clin Med Phys       Date:  2016-07-08       Impact factor: 2.102

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.