The peripheral dose outside the applicator in electron beams of Oncor linear accelerator.

Basak Iktueren1, Hatice Bilge, Songul Karacam, Gulyuz Atkovar.   

Abstract

In this study, the peripheral dose outside the applicator was measured using electron beams produced by an Oncor linear accelerator and compared with the data of the treatment planning system (TPS). The dose profiles have been measured, by using a water-equivalent slab phantom and a parallel plate ionisation chamber, at 6, 9 and 15 MeV energy levels in 5×5, 10×10, 15×15, 20×20 and 25×25 cm(2) applicators and at 0, 10 and 20° gantry angles; and at the surface, 0.2, 0.5, 1 cm and d(max) depth for each electron energy level. The peripheral dose has been determined with these profiles by normalisation at the field central beam axis (CAX). It has been noticed that, using a 10×10 cm(2) applicator, there is a 1.4 % dose peak on the surface 6 cm away from the field edge where the field CAX is at 100 %, at a gantry angle of 0° with 6 and 9 MeV electron beams; also for the 15 MeV electron beam there is a 2.3 % dose peak. It has been discovered that the peak dose approaches a minimum depending on the increase in depth and reaches 2.5-4 % depending on the growth of the field dimension. At gantry angles of 10 and 20°, 6 and 9 MeV electron beams created small peaks and a maximum dose could be reached at 0.2 and 1 cm depth. Electron beam of 15 MeV did not peak at depths of 0.2 and 1 cm at gantry angles of 10 and 20°. The measured peripheral dose outside the applicators has been compared with the data from a TPS's computer using the Pencil Beam algorithm; it has been stated that dose calculations can be made as far as 3 cm outside the field. In conclusion, the TPS is not sufficient to measure the peripheral dose outside the applicators, and this dose can only be determined by direct measurement.

Entities:  

Mesh:

Year:  2011        PMID: 22025738     DOI: 10.1093/rpd/ncr392

Source DB:  PubMed          Journal:  Radiat Prot Dosimetry        ISSN: 0144-8420            Impact factor:   0.972


  3 in total

1.  Experimental assessment of out-of-field dose components in high energy electron beams used in external beam radiotherapy.

Authors:  Mohamad M Alabdoaburas; Jean-Pierre Mege; Jean Chavaudra; Jérémi Vũ Bezin; Atilla Veres; Florent de Vathaire; Dimitri Lefkopoulos; Ibrahima Diallo
Journal:  J Appl Clin Med Phys       Date:  2015-11-08       Impact factor: 2.102

2.  Shielding effect of a lead apron on the peripheral radiation dose outside the applicator of electron beams from an Elekta linear accelerator.

Authors:  Huifang He; Yingdong Zhang; Jidong Wang; Xingyu Chen; Yuyan Yang; Junjie Wang
Journal:  J Appl Clin Med Phys       Date:  2020-12-09       Impact factor: 2.102

3.  Verification of electron beam parameters in an intraoperative linear accelerator using dosimetric and radiobiological response methods.

Authors:  Agnieszka Dróżdż; Martyna Waluś; Marcin Zieliński; Bożena Malesa; Marta Kruszyna-Mochalska; Katarzyna Kulcenty; Beata Adamczyk; Piotr Nowaczyk; Julian Malicki; Jacek Pracz
Journal:  Rep Pract Oncol Radiother       Date:  2021-12-30
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.