Characterization and correction of center-frequency effects in X-nuclear eddy current compensations on a clinical MR system.

Mary A McLean1,2, R Scott Hinks3, Joshua D Kaggie1, Ramona Woitek1, Frank Riemer4, Martin J Graves1, Dominick J O McIntyre2, Ferdia A Gallagher1, Rolf F Schulte5.   

Abstract

PURPOSE: The aim of the study was to investigate whether incorrectly compensated eddy currents are the source of persistent X-nuclear spectroscopy and imaging artifacts, as well as methods to correct this.
METHODS: Pulse-acquire spectra were collected for 1 H and X-nuclei (23 Na or 31 P) using the minimum TR permitted on a 3T clinical MRI system. Data were collected in 3 orientations (axial, sagittal, and coronal) with the spoiler gradient at the end of the TR applied along the slice direction for each. Modifications to system calibration files to tailor eddy current compensation for each X-nucleus were developed and applied, and data were compared with and without these corrections for: slice-selective MRS (for 23 Na and 31 P), 2D spiral trajectories (for 13 C), and 3D cones trajectories (for 23 Na).
RESULTS: Line-shape distortions characteristic of eddy currents were demonstrated for X-nuclei, which were not seen for 1 H. The severity of these correlated with the amplitude of the eddy current frequency compensation term applied by the system along the axis of the applied spoiler gradient. A proposed correction to eddy current compensation, taking account of the gyromagnetic ratio, was shown to dramatically reduce these distortions. The same correction was also shown to improve data quality of non-Cartesian imaging (2D spiral and 3D cones trajectories).
CONCLUSION: A simple adaptation of the default compensation for eddy currents was shown to eliminate a range of artifacts detected on X-nuclear spectroscopy and imaging.
© 2020 The Authors. Magnetic Resonance in Medicine published by Wiley Periodicals LLC on behalf of International Society for Magnetic Resonance in Medicine.

Entities:  

Keywords:  MRI; X-nuclei; carbon-13; eddy currents; image artifacts; magnetic resonance spectroscopy; phosphorus-31; sodium-23

Mesh:

Year:  2020        PMID: 33274790      PMCID: PMC7898706          DOI: 10.1002/mrm.28607

Source DB:  PubMed          Journal:  Magn Reson Med        ISSN: 0740-3194            Impact factor:   4.668


  13 in total

1.  Spatiotemporal magnetic field monitoring for MR.

Authors:  Christoph Barmet; Nicola De Zanche; Klaas P Pruessmann
Journal:  Magn Reson Med       Date:  2008-07       Impact factor: 4.668

2.  Temporal and spatial analysis of fields generated by eddy currents in superconducting magnets: optimization of corrections and quantitative characterization of magnet/gradient systems.

Authors:  C Boesch; R Gruetter; E Martin
Journal:  Magn Reson Med       Date:  1991-08       Impact factor: 4.668

3.  A programmable eddy-current compensation system for MRI and localized spectroscopy.

Authors:  M E Fry; S Pittard; I R Summers; W Vennart; F T Goldie
Journal:  J Magn Reson Imaging       Date:  1997 Mar-Apr       Impact factor: 4.813

Review 4.  Quantitative sodium MR imaging: A review of its evolving role in medicine.

Authors:  Keith R Thulborn
Journal:  Neuroimage       Date:  2016-11-24       Impact factor: 6.556

Review 5.  The role of hyperpolarized 129xenon in MR imaging of pulmonary function.

Authors:  Lukas Ebner; Jeff Kammerman; Bastiaan Driehuys; Mark L Schiebler; Robert V Cadman; Sean B Fain
Journal:  Eur J Radiol       Date:  2016-09-16       Impact factor: 3.528

6.  IDEAL spiral CSI for dynamic metabolic MR imaging of hyperpolarized [1-13C]pyruvate.

Authors:  Florian Wiesinger; Eliane Weidl; Marion I Menzel; Martin A Janich; Oleksandr Khegai; Steffen J Glaser; Axel Haase; Markus Schwaiger; Rolf F Schulte
Journal:  Magn Reson Med       Date:  2011-11-29       Impact factor: 4.668

7.  Quantitative sodium imaging with a flexible twisted projection pulse sequence.

Authors:  Aiming Lu; Ian C Atkinson; Theodore C Claiborne; Frederick C Damen; Keith R Thulborn
Journal:  Magn Reson Med       Date:  2010-06       Impact factor: 4.668

8.  Rapid in vivo apparent diffusion coefficient mapping of hyperpolarized (13) C metabolites.

Authors:  Bertram L Koelsch; Galen D Reed; John Kurhanewicz; Peder E Z Larson; Kayvan R Keshari; Myriam M Chaumeil; Robert Bok; Sabrina M Ronen; Daniel B Vigneron
Journal:  Magn Reson Med       Date:  2014-09-11       Impact factor: 4.668

Review 9.  Hyperpolarized 13C MRI: Path to Clinical Translation in Oncology.

Authors:  John Kurhanewicz; Daniel B Vigneron; Jan Henrik Ardenkjaer-Larsen; James A Bankson; Kevin Brindle; Charles H Cunningham; Ferdia A Gallagher; Kayvan R Keshari; Andreas Kjaer; Christoffer Laustsen; David A Mankoff; Matthew E Merritt; Sarah J Nelson; John M Pauly; Philips Lee; Sabrina Ronen; Damian J Tyler; Sunder S Rajan; Daniel M Spielman; Lawrence Wald; Xiaoliang Zhang; Craig R Malloy; Rahim Rizi
Journal:  Neoplasia       Date:  2018-11-23       Impact factor: 5.715

10.  Characterization and correction of center-frequency effects in X-nuclear eddy current compensations on a clinical MR system.

Authors:  Mary A McLean; R Scott Hinks; Joshua D Kaggie; Ramona Woitek; Frank Riemer; Martin J Graves; Dominick J O McIntyre; Ferdia A Gallagher; Rolf F Schulte
Journal:  Magn Reson Med       Date:  2020-12-04       Impact factor: 4.668

View more
  1 in total

1.  Characterization and correction of center-frequency effects in X-nuclear eddy current compensations on a clinical MR system.

Authors:  Mary A McLean; R Scott Hinks; Joshua D Kaggie; Ramona Woitek; Frank Riemer; Martin J Graves; Dominick J O McIntyre; Ferdia A Gallagher; Rolf F Schulte
Journal:  Magn Reson Med       Date:  2020-12-04       Impact factor: 4.668

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.