PURPOSE: Although proton therapy has become a well-established radiation modality, continued efforts are needed to improve our understanding of the molecular and cellular mechanisms occurring during treatment. Such studies are challenging, requiring many resources. The purpose of this study was to create a phantom that would allow multiple in vitro experiments to be irradiated simultaneously with a spot-scanning proton beam.Entities:
Year: 2020 PMID: 33274258 PMCID: PMC7707323 DOI: 10.14338/IJPT-20-00007.1
Source DB: PubMed Journal: Int J Part Ther ISSN: 2331-5180
Dosimetric results. Monte Carlo dose and linear energy transfer (LETd) calculations (mean ± 0.5 mm). Markus chamber and radiochromic film (RCF) measurements (mean ± 1 SD).
| BP | Edge | 202.8 (± 0.3) | 204.4 (± 2.8) | 212.3 (± 1.4) | 7.3 (± 0.7) |
| BP | Center | 209.8 (± 0.1) | 202.6 (± 4.0) | 218.7 (± 1.1) | 7.3 (± 0.7) |
| BP | Average | 206.3 (± 3.8) | 203.7 (± 3.4) | 215.5 (± 3.6) | 7.3 (± 0.7) |
| ENT | Edge | 190.1 (± 0.5) | 198.2 (± 5.1) | 195.3 (± 1.8) | 2.2 (± 0.1) |
| ENT | Center | 191.5 (± 0.5) | 196.8 (± 4.7) | 194.7 (± 1.6) | 2.2 (± 0.1) |
| ENT | Average | 190.8 (± 0.8) | 197.0 (± 4.7) | 195.0 (± 1.6) | 2.2 (± 0.1) |
Abbreviations: cGy, centigray; BP, Bragg peak; ENT, entrance.
Biologic verification of the experimental setup using Chinese hamster ovary cells. Fit parameters for the linear quadratic model (± 1 SD) were used to generate cell survival curves. Relative biologic effectiveness (RBE) values at 50% and 10% survival with errors are as reported in Howard et al [27] and Howard et al [28].
| 6-MV x-raysa | — | 0.15 (± 0.03) | 0.04 (± 0.00) | 1.00 | 1.00 |
| 71.3-MeV protonsa | ENT | 0.22 (± 0.04) | 0.03 (± 0.01) | 1.19 (± 0.10) | 1.09 (± 0.04) |
| 71.3-MeV protonsa | BP | 0.34 (± 0.10) | 0.05 (± 0.02) | 1.67 (± 0.23) | 1.46 (± 0.08) |
| 76.8-MeV protons | ENT | 0.22 (± 0.04) | 0.03 (± 0.00) | 1.13 (± 0.09) | 1.02 (± 0.01) |
| 76.8-MeV protons | BP | 0.31 (± 0.05) | 0.04 (± 0.01) | 1.48 (± 0.14) | 1.28 (± 0.04) |
Abbreviations: ENT, entrance; BP, Bragg peak.
Data reproduced from references [27-28].