Stiffness Analysis of Reinforced Ureteral Stents Against Radial Compression: In vitro Study.

Benoît Vogt1.   

Abstract

INTRODUCTION: Malignant ureteral obstruction caused by cancer diseases may induce renal failure. Indwelling stent is a popular method to release renal obstruction. But adequate stent placement across an obstructed ureter does not necessarily guarantee renal decompression. The aim of the study was to compare, in vitro, the physical characteristics and stiffness of several commercially available reinforced ureteral stents and identify the physical factors that could lead to the obstruction of the stent.
MATERIAL AND METHODS: The test apparatus used for measurements allowed applying a radial compression force on a segment of the stent to stop a water flow through the lumen of the stent. Some reinforced double-pigtail stents Teleflex Medical, Bard, and Coloplast were evaluated.
RESULTS: The best physical-stiffness characteristic was obtained with the Teleflex 8F stent (5.4 N mm-2). The best result against the radial compression was obtained with tandem stents. The radial compressive stresses of the Teleflex stents (4.4 to 5.4 N mm-2) were higher than with the other stents used in the study (1.0 to 2.9 N mm-2). Among the reinforced stents selected in the present study, a wider inner diameter helped increase volumetric flow rate but did not affect the stiffness of the stent. The measurement of inner diameter showed heterogeneity along the tube of some stents.
CONCLUSION: The stiffness of the stent appeared to be an important factor to maintain patency with respect to radial compression forces but the inner diameter of the stent and its preservation may be essential parameters to increase the volumetric flow rate. Some reinforced stents tested in the present study confirmed that it is possible to combine stiffness and wide lumen. The use of tandem stents provided the best stiffness against radial compression and the greatest lumen.
© 2020 Vogt.

Entities:  

Keywords:  compression; diameter; malignant ureteral obstruction; stent; stiffness; ureteral stent

Year:  2020        PMID: 33274188      PMCID: PMC7708781          DOI: 10.2147/RRU.S285031

Source DB:  PubMed          Journal:  Res Rep Urol        ISSN: 2253-2447


  16 in total

1.  The double-J ureteral stent: in vivo and in vitro flow studies.

Authors:  W A Hübner; E G Plas; M L Stoller
Journal:  J Urol       Date:  1992-08       Impact factor: 7.450

Review 2.  The role of biofilm infection in urology.

Authors:  P Tenke; B Kovacs; M Jäckel; E Nagy
Journal:  World J Urol       Date:  2006-01-10       Impact factor: 4.226

3.  Poiseuille and his law.

Authors:  J Pfitzner
Journal:  Anaesthesia       Date:  1976-03       Impact factor: 6.955

4.  Analysis of ureteral stent compression force and its role in malignant obstruction.

Authors:  Matthew S Christman; James O L'esperance; Chong H Choe; Sean P Stroup; Brian K Auge
Journal:  J Urol       Date:  2008-11-17       Impact factor: 7.450

5.  Incidence of bacterial colonisation after indwelling of double-J ureteral stent.

Authors:  Hasan Riza Aydin; Lokman Irkilata; Mustafa Aydin; Selim Gorgun; Hüseyin Cihan Demirel; Senol Adanur; Mevlüt Keles; Aynur Atilla; Mustafa Kemal Atilla
Journal:  Arch Ital Urol Androl       Date:  2016-01-14

6.  A New Customized Ureteral Stent with Nonrefluxing Silicone End-piece to Alleviate Stent-related Symptoms in Malignant Diseases.

Authors:  Benoît Vogt
Journal:  Urology       Date:  2019-12-30       Impact factor: 2.649

7.  Analysis of Urine Flow in Three Different Ureter Models.

Authors:  Kyung-Wuk Kim; Young Ho Choi; Seung Bae Lee; Yasutaka Baba; Hyoung-Ho Kim; Sang-Ho Suh
Journal:  Comput Math Methods Med       Date:  2017-06-04       Impact factor: 2.238

8.  Impact of wall shear stress on initial bacterial adhesion in rotating annular reactor.

Authors:  Thibaut Saur; Emilie Morin; Frédéric Habouzit; Nicolas Bernet; Renaud Escudié
Journal:  PLoS One       Date:  2017-02-16       Impact factor: 3.240

9.  Investigating the flow dynamics in the obstructed and stented ureter by means of a biomimetic artificial model.

Authors:  Francesco Clavica; Xuefeng Zhao; Motaz ElMahdy; Marcus J Drake; Xunli Zhang; Dario Carugo
Journal:  PLoS One       Date:  2014-02-03       Impact factor: 3.240

10.  Association of plasma and urine viscosity with cardiometabolic risk factors and oxidative status. A pilot study in subjects with abdominal obesity.

Authors:  Beatriz Herranz; María Dolores Álvarez; Jara Pérez-Jiménez
Journal:  PLoS One       Date:  2018-10-09       Impact factor: 3.240

View more
  2 in total

1.  10-Year Experience with Reinforced Ureteral Stents for Malignant Ureteral Obstruction.

Authors:  Benoît Vogt; Laure-Hélène Blanchet
Journal:  Res Rep Urol       Date:  2021-08-14

2.  Analysis of Ureteral Tumour Stents for Malignant Ureteral Obstruction: Towards Reshaping an Optimal Stent.

Authors:  Benoît Vogt; Laure-Hélène Blanchet
Journal:  Res Rep Urol       Date:  2021-10-27
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.