| Literature DB >> 33273590 |
T D Do1, S Sawall2, S Heinze3, T Reiner4, C H Ziener5, W Stiller1, H P Schlemmer5, M Kachelrieß2, H U Kauczor1, S Skornitzke6.
Abstract
An evaluation of energy thresholding and acquisition mode for metal artifact reduction in Photon-counting detector CT (PCD-CT) compared to conventional energy-integrating detector CT (EID-CT) was performed. Images of a hip prosthesis phantom placed in a water bath were acquired on a scanner with PCD-CT and EID-CT (tube potentials: 100, 120 and 140 kVp) and energy thresholds (above 55-75 keV) in Macro and Chess mode. Only high-energy threshold images (HTI) were used. Metal artifacts were quantified by a semi-automated segmentation algorithm, calculating artifact volumes, means and standard deviations of CT numbers. Images of a human cadaver with hip prosthesis were acquired on the PCD-CT in Macro mode as proof-of-concept. Images at 140 kVp showed less metal artifacts than 120 kVp or 100 kVp. HTI (70, 75 keV) had fewer artifacts than low energy thresholds (55, 60, 65 keV). Fewer artifacts were observed in the Macro-HTI (8.9-13.3%) for cortical bone compared to Chess-HTI (9.4-19.1%) and EID-CT (10.7-19.0%) whereas in bone marrow Chess-HTI (19.9-45.1%) showed less artifacts compared to Macro-HTI (21.9-38.3%) and EID-CT (36.4-54.9%). Noise for PCD-CT (56-81 HU) was higher than EID-CT (33-36 HU) irrespective of tube potential. High-energy thresholding could be used for metal artifact reduction in PCD-CT, but further investigation of acquisition modes depending on target structure is required.Entities:
Year: 2020 PMID: 33273590 PMCID: PMC7713179 DOI: 10.1038/s41598-020-77904-3
Source DB: PubMed Journal: Sci Rep ISSN: 2045-2322 Impact factor: 4.379
Acquisition protocol and reconstruction parameters for phantom and cadaver scans with energy-integrating detector CT (EID-CT), PCD-CT Macro mode and PCD-CT Chess mode.
| Parameters | Phantom | Cadaver scans | |||
|---|---|---|---|---|---|
| EID-CT | PCD-CT | PCD-CT | EID-CT | PCD-CT Macro-HTI | |
| Tube potential and energy threshold [kVp/keV] | 140 | 140/75 | 140/75 | 140 | 140/75 |
| 140/65 | |||||
| 120 | 120/70 | 120/70 | |||
| 120/60 | |||||
| 100 | 100/65 | 100/65 | |||
| 100/55 | |||||
| CTDIvol | 20 mGy | 20 mGy | 20 mGy | 34.34 mGy | 23.17 mGy |
| Pitch | 0.6 | 0.6 | 0.6 | 0.6 | 0.6 |
| Collimation | 32 × 0.6 mm | 32 × 0.5 mm | 32 × 0.5 mm | 32 × 0.6 mm | 32 × 0.5 mm |
| Reconstruction kernel | B70f. | B70f. | B70f. | B70f. | B70f. |
| Slice thickness/increment | 2/1 mm | 2/1 mm | 2/1 mm | 2/1 mm | 2/1 mm |
Figure 1Femoral phantom with hip prosthesis placed in a water bath. Segmentation of volumes of interest of the bone marrow (brown), cortical bone (turquoise) and water bath (blue) in the axial plane (a) and coronal plane (b). For comparison, images without segmentation in the axial plane (c) and coronal plane (d). Segmentation of reference volumes in the distal femur without prosthesis in the cortical bone (red) and bone marrow (purple) in the axial plane (e) and coronal plane (f). Images were created with the freely available Medical Imaging Interaction Toolkit (version 2018.04.2, available from http://mitk.org/wiki/MITK)[11].
Figure 2Femoral phantom with prosthesis embedded in a water bath and depiction of artifact volumes in the cortical bone, bone marrow and water color-coded in the axial plane (a) and coronal plane (b) on EID-CT at 120 kVp. Image examples of HTIs in Chess mode (c) and in Macro mode (d) highlight the increased artifacts in bone marrow (white arrow) and reduced artifacts in cortical bone (black arrow) for Macro-mode HTIs compared to Chess-mode HTIs. Images were created with the freely available Medical Imaging Interaction Toolkit (version 2018.04.2, available from http://mitk.org/wiki/MITK)[11].
Figure 3(a) Artifact percentage depending on tube potential and energy threshold. (b) Metal artifacts depending on location in high-energy threshold images with 140 kVp/75 keV compared to EID-CT.
Mean and SD of CT numbers of high-energy threshold images in cortical bone, bone marrow, water bath and water bath only.
| Scan mode, kVp and keV | Cortical bone | Bone marrow | Water bath with phantom | Water bath only | ||||
|---|---|---|---|---|---|---|---|---|
| Mean [HU] | SD [HU] | Mean [HU] | SD [HU] | Mean [HU] | SD [HU] | Mean [HU] | SD [HU] | |
| Macro-HTI 140/75 | 595.96 | 125.94 | − 647.91 | 112.03 | − 0.03 | 56.87 | 0.18 | 55.23 |
| Chess-HTI 140/75 | 614.68 | 123.64 | − 661.37 | 119.87 | 0.38 | 80.12 | 0.02 | 77.25 |
| Chess-HTI 140/65 | 607.59 | 141.91 | − 632.69 | 117.38 | 1.70 | 65.26 | 1.34 | 63.21 |
| EID-CT 140 | 679.70 | 80.93 | − 674.12 | 81.22 | − 0.34 | 34.62 | − 0.27 | 32.95 |
| Macro-HTI 120/70 | 615.25 | 139.17 | − 641.32 | 111.60 | 1.08 | 57.10 | 1.60 | 55.62 |
| Chess-HTI 120/70 | 640.22 | 124.91 | − 665.98 | 121.25 | 0.73 | 79.89 | 0.26 | 76.92 |
| Chess-HTI 120/60 | 652.25 | 128.28 | − 650.65 | 115.25 | 0.36 | 62.67 | − 0.09 | 60.28 |
| EID-CT 120 | 711.29 | 84.67 | − 676.49 | 81.12 | − 0.36 | 34.60 | − 0.12 | 32.80 |
| Macro-HTI 100/65 | 652.69 | 141.56 | − 645.39 | 104.58 | 0.54 | 57.81 | 0.29 | 55.53 |
| Chess-HTI 100/65 | 671.94 | 142.05 | − 665.54 | 120.07 | 0.09 | 80.84 | 0.17 | 77.37 |
| Chess-HTI 100/55 | 709.94 | 121.90 | − 667.90 | 104.16 | − 0.78 | 61.96 | − 0.66 | 59.24 |
| EID-CT 100 | 761.15 | 86.63 | − 679.67 | 80.87 | − 0.29 | 35.57 | 0.35 | 33.64 |
Figure 4Noise depending on tube potential and energy threshold.
Contrast-to-noise ratio (CNR) when comparing CT numbers of cortical bone and bone marrow to those measured in the water bath with phantom.
| Scan mode, kVp and keV | Cortical bone versus water | Bone marrow versus water |
|---|---|---|
| CNR [HU] | CNR [HU] | |
| Macro-HTI 140/75 | 6.10 | 7.29 |
| Chess-HTI 140/75 | 5.90 | 6.49 |
| Chess-HTI 140/65 | 5.49 | 6.68 |
| EID-CT 140 | 10.93 | 10.79 |
| Macro-HTI 120/70 | 5.77 | 7.25 |
| Chess-HTI 120/70 | 6.10 | 6.49 |
| Chess-HTI 120/60 | 6.46 | 7.02 |
| EID-CT 120 | 11.00 | 10.84 |
| Macro-HTI 100/65 | 6.03 | 7.64 |
| Chess-HTI 100/65 | 5.81 | 6.50 |
| Chess-HTI 100/55 | 7.35 | 7.78 |
| EID-CT 100 | 11.50 | 10.88 |
Figure 5Axial CT of the pelvic region with prosthesis in the left hip. Artifacts in cadaver scans in EID-CT (a) and PCD-CT high-energy threshold Macro at 140kVp/75 keV (b). Note the increased metal artifacts in EID-CT in comparison to PCD-CT (white arrows: streak artifacts at the level of the femoral hip prosthesis). However, PCD-CT showed increased image noise leading to an increased number of “false metal artifacts”, subtracted by the correction factor. Images were created with the freely available Medical Imaging Interaction Toolkit (version 2018.04.2, available from http://mitk.org/wiki/MITK)[11].