Detection of SARS-CoV-2 within the healthcare environment: a multi-centre study conducted during the first wave of the COVID-19 outbreak in England.

G Moore1, H Rickard2, D Stevenson2, P Aranega-Bou2, J Pitman2, A Crook2, K Davies2, A Spencer2, C Burton2, L Easterbrook2, H E Love2, S Summers2, S R Welch2, N Wand2, K-A Thompson2, T Pottage2, K S Richards2, J Dunning3, A Bennett2.   

Abstract

BACKGROUND: Understanding how severe acute respiratory syndrome coronavirus-2 (SARS-CoV-2) is spread within the hospital setting is essential in order to protect staff, implement effective infection control measures, and prevent nosocomial transmission.
METHODS: The presence of SARS-CoV-2 in the air and on environmental surfaces around hospitalized patients, with and without respiratory symptoms, was investigated. Environmental sampling was undertaken within eight hospitals in England during the first wave of the coronavirus disease 2019 outbreak. Samples were analysed using reverse transcription polymerase chain reaction (PCR) and virus isolation assays.
FINDINGS: SARS-CoV-2 RNA was detected on 30 (8.9%) of 336 environmental surfaces. Cycle threshold values ranged from 28.8 to 39.1, equating to 2.2 x 105 to 59 genomic copies/swab. Concomitant bacterial counts were low, suggesting that the cleaning performed by nursing and domestic staff across all eight hospitals was effective. SARS-CoV-2 RNA was detected in four of 55 air samples taken <1 m from four different patients. In all cases, the concentration of viral RNA was low and ranged from <10 to 460 genomic copies/m3 air. Infectious virus was not recovered from any of the PCR-positive samples analysed.
CONCLUSIONS: Effective cleaning can reduce the risk of fomite (contact) transmission, but some surface types may facilitate the survival, persistence and/or dispersal of SARS-CoV-2. The presence of low or undetectable concentrations of viral RNA in the air supports current guidance on the use of specific personal protective equipment for aerosol-generating and non-aerosol-generating procedures. Crown
Copyright © 2020. Published by Elsevier Ltd. All rights reserved.

Entities:  

Keywords:  Air; Environmental contamination; Environmental surfaces; Hospital; Infection control; SARS-CoV-2

Mesh:

Substances:

Year:  2020        PMID: 33259882      PMCID: PMC7831847          DOI: 10.1016/j.jhin.2020.11.024

Source DB:  PubMed          Journal:  J Hosp Infect        ISSN: 0195-6701            Impact factor:   3.926


Introduction

Over the course of 2020, severe acute respiratory syndrome coronavirus-2 (SARS-CoV-2), the causative agent of coronavirus disease 2019 (COVID-19), has spread rapidly across the globe and, as of 15 August 2020, had infected 21 million people and caused over 750,000 deaths [1]. The transmission of respiratory viruses can occur through inhalation of respiratory droplets (particles >5 μm in diameter) and infectious aerosols (<5 μm in diameter), and/or contact with respiratory droplets, either directly or indirectly, via contaminated surfaces. The rapid spread of COVID-19 has led many to conclude that airborne transmission must be involved [2]. However, this is widely debated and, according to current evidence, SARS-CoV-2 is primarily transmitted via droplet and contact routes, although it is acknowledged that airborne transmission could occur in specific circumstances and settings [3]. Healthcare workers (HCWs) and others on the front line are at increased risk of infection [4]. Medical aerosol-generating procedures (AGPs) [e.g. intubation, non-invasive ventilation (NIV), airway suctioning] can produce droplets <5 μm in diameter, and have been associated with increased transmission of SARS-CoV from patients to HCWs [5]. It is argued, however, that there is limited evidence to link AGPs with the transmission of respiratory infections, including COVID-19 [6]. Air samples taken during tracheostomy procedures, high-flow nasal oxygen treatment, NIV and nebulization have not contained SARS-CoV-2 RNA [7], and HCWs exposed to unrecognized cases of COVID-19 undergoing similar high-risk AGPs have not become infected [8]. Nonetheless, occupational exposure has resulted in infection [9], and it has been estimated that patient-to-HCW transmissions could be responsible for 57% of infections among HCWs in England [10]. Nosocomial transmission may also account for 20% of infections in inpatients [10], so understanding how SARS-CoV-2 is spread within the hospital setting is essential to ensure that staff are protected adequately and effective infection control measures are implemented. Several studies, utilizing a range of air and surface sampling methods, have been undertaken to determine the presence and prevalence of SARS-CoV-2 in the healthcare environment [[11], [12], [13], [14], [15], [16], [17], [18], [19], [20], [21]]. The detection of viral RNA in air samples differs between studies, with some reporting widespread airborne contamination [14,18,21] but many reporting low or non-detectable concentrations [13,15,16,19], even in samples collected 10 cm from the face of positive patients [12]. Surfaces touched frequently by HCWs and/or patients are often contaminated with bacterial pathogens. Likewise, SARS-CoV-2 RNA has been detected on high-contact surfaces such as computers, bed rails and door handles. Again, the extent of this surface contamination differs between studies. Reported positivity rates range from 0.8% to >70%, with those studies reporting a higher level of airborne contamination also detecting widespread surface contamination [18,21]. In many cases, sampling was performed before routine cleaning, but the efficacy of cleaning was not assessed [11,13,15]. When comparative samples were taken, SARS-CoV-2 RNA was detected on 61% of surfaces sampled prior to cleaning but was not detected on surfaces after cleaning [17]. The proportion of surfaces contaminated with viral RNA can also differ between ward types. Some studies have detected little to no surface contamination in intensive care units (ICUs) but have detected widespread contamination within general wards [11,21]. In contrast, other studies have reported higher positivity rates within the ICU setting [14,20]. Environmental sampling can provide important information about the spread of healthcare-associated infections. However, this is resource-intensive and time-consuming; as such, many studies investigating SARS-CoV-2 and its contamination of the healthcare environment have focused on a single hospital and, in the context of the COVID-19 pandemic, a single point in time. Sampling frequency is also generally low, meaning that results often represent a snapshot in time and place. In a rapidly evolving outbreak, there is a need to gain rapid understanding of certain trends, and whilst snapshot samples by themselves cannot be considered representative, they can, when taken together, provide useful data relating to type, level and location of environmental contamination. To date, however, differences in study setting, protocol and methodology have led to inconsistency in the results obtained, making it difficult to draw any firm conclusions relating to SARS-CoV-2 and its presence within the healthcare environment. As part of the Public Health England (PHE) national incident response, the presence of SARS-CoV-2 in the air and on environmental surfaces around hospitalized patients, with and without respiratory symptoms, was investigated. Environmental sampling, using standard methods, was undertaken within eight acute hospital trusts in England. Trends, in terms of type and level of surface contamination and the potential for AGPs to disperse SARS-CoV-2, have been identified and these provide evidence to support current infection prevention and control guidance, including the use of personal protective equipment (PPE).

Methods

Between 3rd March 2020 and 12th May 2020, the study team visited eight hospitals (three on more than one occasion; Figure 1 ) and undertook environmental sampling in areas where patients infected with SARS-CoV-2 were receiving care. These included 11 negative pressure isolation rooms, 11 neutral pressure side rooms, six ICU/high-dependency unit (HDU) open cohorts and 12 non-ICU cohort bays. Whilst sampling primarily focused on 44 individual bed spaces (Table I ), samples were also taken from the wider ward environment (e.g. nursing stations, patient toilet areas) and from non-COVID wards. Medical procedures being performed and obvious symptoms such as coughing were observed and recorded. Patient details (hospital number, date of admission, date of diagnosis) were collected for future correlation with clinical virology results. Details regarding routine and terminal (discharge) cleaning were also collected.
Figure 1

Sampling date in relation to the number of laboratory-confirmed cases (red line) and hospital admissions with coronavirus disease 2019 (COVID-19) (blue bars).

Table I

Sampling primarily focused on individual bed spaces (44 different patients)

Visita (hospital)PatientWardLocationDays since symptom onsetDays since admissionDays since first SARS-CoV-2-positive swabNotable treatmentNo. of surfaces ‘positive’ for SARS-CoV-2 RNASARS-CoV-2 RNA detected in the air
1 (A)1IDSingle room (negative pressure)755None0No
2IDSingle room (negative pressure)··45None0No
3IDSingle room (negative pressure)··3··None0No
4ICUSingle room (negative pressure)1045O2 (Venturi)0No
2 (B)5IDSingle room (negative pressure)11··6None0No
6IDSingle room (negative pressure)13··10None0No
7IDSingle room (negative pressure)7··3None4No
3 (C)8IDSingle room (negative pressure)1067None0No
9IDSingle room (negative pressure)533O2 (Venturi)0No
10IDSingle room (negative pressure)811None0No
11IDSingle room (negative pressure)1033None2No
4 (D)12ICUCohort bay444ECMO0No
13ICUCohort bay991ECMO0No
14ICUCohort bay111011ECMO0No
15ICUCohort bay666ECMO0No
16ICUCohort bay121211ECMO0No
17ICUCohort bay171717ECMO0No
5 (D)18GWCohort bay955O2 (Venturi)0No
19GWCohort bay621O2 (Venturi)1Yes
20GWCohort bay866None0Yes
21GWCohort bay911Nebulizer0No
6 (E)22GWSide room (neutral pressure)1033CPAP1No
23GWSide room (neutral pressure)422O2 (Venturi)0No
24ICUSide room (neutral pressure)1777Intubated0No
25ICUCohort bay943Intubated0No
7 (F)26HDUSide room (neutral pressure)1044CPAP1Yes
27HDUSide room (neutral pressure)877CPAP0Yes
8 (G)28HDUCohort bay12109CPAP2No
29HDUCohort bay261616CPAP0No
30HDUCohort bay191212O2 (Venturi)0No
31HDUCohort bay1588CPAP2No
9 (H)32ICUSide room (neutral pressure)944NIV0No
33ICUCohort bay722CPAP0No
34GWSide room (neutral pressure)····6NIV2No
35ICUCohort bay26162Nebulizer1No
10 (H)36ICUCohort bay1061NIV0No
37ICUCohort bay6161Nebulizer0No
11 (F)38HDUSide room (neutral pressure)····2CPAP0No
39HDUSide room (neutral pressure)····4CPAP0No
12 (H)40ICUSide room (neutral pressure)······CPAP0No
13 (D)41GWCohort bay··3434None0No
42GWSide room (neutral pressure)333None3No
43GWCohort bay454544None0No
44GWCohort bay141413None0No

SARS-CoV-2, severe acute respiratory syndrome coronavirus-2; ID, infectious diseases; ICU, intensive care unit; HDU, high-dependency unit; GW, general ward; CPAP, continuous positive airway pressure; NIV, non-invasive ventilation; ECMO, extracorporeal membrane oxygenation; O2 (Venturi), oxygen via a Venturi mask; ··, data not available/collected.

See Figure 1.

Sampling date in relation to the number of laboratory-confirmed cases (red line) and hospital admissions with coronavirus disease 2019 (COVID-19) (blue bars). Sampling primarily focused on individual bed spaces (44 different patients) SARS-CoV-2, severe acute respiratory syndrome coronavirus-2; ID, infectious diseases; ICU, intensive care unit; HDU, high-dependency unit; GW, general ward; CPAP, continuous positive airway pressure; NIV, non-invasive ventilation; ECMO, extracorporeal membrane oxygenation; O2 (Venturi), oxygen via a Venturi mask; ··, data not available/collected. See Figure 1. Surfaces deemed to be high-contact sites were sampled using nylon flocked swabs (Copan, Bresica, Italy) wetted with universal transport medium. Tryptone soya agar contact plates (Oxoid Ltd, Basingstoke, UK) were also used (from 27th March 2020) in order to provide an indication of general surface cleanliness. Air samples were taken using two types of active air sampler: a Coriolis μ air sampler (Bertin Instruments, Montigny-le-Bretonneux, France), operating at 300 L/min and collecting into 15 mL RNase-free phosphate buffered saline (PBS); and an MD8 air sampler (Sartorius, Göttingen, Germany), operating at 50 L/min and collecting on to a gelatine membrane filter. Both samplers were positioned close to patients (<1 m) with and without respiratory symptoms and operated for 10 min. The type and duration of AGP, if any, was noted. Ambient temperature and relative humidity were monitored. All samples were returned to PHE Porton Down. Agar contact plates were incubated at 37oC for 48 h whilst the air and swab samples (for virus detection) were frozen at -80oC prior to processing. Laboratory-based validation experiments confirmed that neither the transport nor storage conditions had an adverse effect on subsequent reverse transcription polymerase chain reaction (RT-PCR) analysis. RNA was extracted from aliquots (140 μL) of each swab and Coriolis air sample using the QIAamp Viral RNA Mini Kit (Qiagen Ltd, Manchester, UK). The remaining Coriolis sample was concentrated to <1 mL using a Vivaspin 20 centrifugal concentrator. Each gelatine membrane was dissolved in 10 mL Minimum Essential Medium (MEM). Aliquots (140 μL) of both were extracted. In total, 425 samples [surface swabs (N=336) and air (N=89)] were analysed for SARS-CoV-2 using RT-PCR. All samples were screened in duplicate using one of the following targets: RNA-dependent RNA polymerase (RdRp) with probe 2, envelope (E) or nucleocapsid (N) and ORF1ab (Viasure; CerTest Biotec, Zaragoza, Spain). A sample was considered positive when amplification was detected in both replicates, or ‘suspect’ when it was detected in a single replicate. ‘Suspect’ samples were re-analysed and considered positive if amplification was detected in both replicates. All positive samples were quantified using the N target on the Viasure platform. Amplification in a single replicate was considered sufficient for quantification. Samples that could not be quantified were re-extracted and quantification was re-attempted. Virus isolation was performed on all positive samples with a cycle threshold (Ct) value <34. Vero E6 cells (Vero C1008; ATCC CRL-1586) in culture medium [MEM supplemented with GlutaMAX-I, 10% (v/v) fetal bovine serum (FBS), 1X (v/v) non-essential amino acids and 25 mM HEPES] were incubated at 37oC. Cells (1 x 106 cells/25 cm2 flask) were washed with 1X PBS and inoculated with ≤1 mL environmental sample and incubated at 37°C for 1 h. Cells were washed with 1X PBS and maintained in 5 mL culture medium (4% FBS) with added antibiotic–antimycotic (4X), incubated at 37°C for 7 days and monitored for cytopathic effects (CPE). Cell monolayers that did not display CPE were subcultured up to three times, providing continuous cultures of ∼30 days.

Results

Environmental sampling was undertaken in and around the bed space of 44 different patients, 35 (80%) of whom were male (Table I). Twenty-three patients had been admitted to an ICU (N=15) or a respiratory HDU (N=8), whilst 21 patients occupied beds in a non-ICU setting. These included 10 patients who, after being diagnosed early in the outbreak, were admitted to infectious diseases units. At the time of sampling, 21 patients were receiving mechanical ventilation either invasively (N=8) or non-invasively (N=13), six patients were receiving oxygen via a Venturi mask, and three patients required drugs or saline to be administered by nebulization. All patients had tested positive for SARS-CoV-2 and the median time since diagnosis was 5 days (range 1–44 days). Time since symptom onset ranged from 3 to 45 days. In total, 336 surfaces were sampled for bacteria and/or SARS-CoV-2. The mean aerobic colony count was 1 colony-forming unit (cfu)/cm2. Of those surfaces with more extensive bacterial contamination (>2.5 cfu/cm2), 18 (70%) were associated with a patient's bed (bed rail, bed control, nurse call button) or mobile phone. SARS-CoV-2 RNA was detected on 30 (8.9%) of the 336 surfaces sampled (Table II ). Of the 44 individual bed spaces, 10 were contaminated with viral RNA and accounted for 19 (63%) of all positive sites. In addition to nurse call buttons (N=4), bed control panels (N=3) and mobile phones (N=3), viral RNA was also detected on bedside equipment (e.g. monitor screens, syringe drivers, computer keyboards), particularly in the ICU/HDU setting. However, in the non-ICU setting, 27% of surfaces contaminated with SARS-CoV-2 RNA were located outside the patient bed area. These included toilet door handles and portable vital signs monitors, which together accounted for 26% of all positive sites.
Table II

Severe acute respiratory syndrome coronavirus-2 (SARS-CoV-2) RNA was detected on 30 of 336 surfaces sampled across eight acute hospital trusts. All positive samples were quantified using the N target on the Viasure platform (CerTest Biotec, Zaragoza, Spain)

Sample locationSurface sampledMean Ct valueMean genomic copies/swab
General wardWider wardToilet door handle28.802.19 x 105
General wardWider wardToilet door handle38.941.90 x 102
General wardCohort bayToilet door handle38.161.56 x 102
Infectious diseasesIsolation roomToilet door handle38.453.72 x 102
General wardSide roomDoor handle37.959.94 x 101
General wardSide roomNurse call button30.712.89 x 104
Infectious diseasesIsolation roomNurse call button33.309.80 x 103
General wardSide roomNurse call button36.211.27 x 103
HDUSide roomNurse call button36.261.26 x 103
HDUWider wardPortable vital signs monitor35.891.58 x 103
General wardCohort bayPortable vital signs monitor36.709.03 x 102
General wardCohort bayPortable vital signs monitor37.824.17 x 102
General wardCohort bayPortable vital signs monitor38.971.87 x 102
Infectious diseasesIsolation roomMobile phone30.347.49 x 104
General wardCohort bayMobile phone36.984.15 x 102
General wardCohort bayMobile phone37.263.08 x 102
General wardSide roomBed rail35.561.01 x 103
Infectious diseasesIsolation roomBed control35.122.76 x 103
General wardCohort bayBed control38.103.43 x 102
HDUCohort bayBed control38.92Unable to quantifya
HDUCohort bayMonitor35.728.97 x 102
HDUCohort bayMonitor36.117.41 x 102
HDUCohort baySyringe driver37.023.64 x 102
ICUCohortBedside computer39.115.91 x 101
General wardSide roomBedside computer38.71Unable to quantifya
Infectious diseasesIsolation roomChair arm37.844.23 x 102
General wardCohort bayCurtain37.983.72 x 102
General wardSide roomWindowsill38.057.63 x 101
Infectious diseasesIsolation roomAir vent37.522.75 x 102
A&EResuscitation bayTrolley drawer37.898.66 x 101

Ct, cycle threshold; ICU, intensive care unit; HDU, high-dependency unit; A&E, accident and emergency.

SARS-CoV-2 detected on initial screening but quantification was unsuccessful.

Severe acute respiratory syndrome coronavirus-2 (SARS-CoV-2) RNA was detected on 30 of 336 surfaces sampled across eight acute hospital trusts. All positive samples were quantified using the N target on the Viasure platform (CerTest Biotec, Zaragoza, Spain) Ct, cycle threshold; ICU, intensive care unit; HDU, high-dependency unit; A&E, accident and emergency. SARS-CoV-2 detected on initial screening but quantification was unsuccessful. RT-PCR Ct values ranged from 28.8 to 39.1 which, when quantified, equated to 2.2 x 105 to 59 genomic copies/swab. Samples with a Ct value <34 were incubated on Vero E6 cells. No CPE or decrease in Ct values across the course of three serial passages were observed, suggesting that the samples did not contain infectious virus. Ambient temperature and relative humidity differed between wards and ranged from 21oC to 25oC and from 21% to 41%, respectively. Air samples were collected using two types of high-volume air sampler, but SARS-CoV-2 RNA was only detected in four (7.3%) of the 55 samples taken using the Coriolis μ sampler. Two of these samples were taken in two different single rooms (neutral pressure). In both cases, the sampler was positioned close (<1 m) to a patient being treated with continuous positive airway pressure (CPAP) via a mask that covered the nose and mouth. Time since diagnosis was 4 and 7 days, with both patients reporting symptoms at least 8 days prior to sampling. Viral RNA was also detected in two air samples taken in two four-bed cohort bays. On one of these occasions, the air sampler was positioned close to a patient who was receiving oxygen via a Venturi mask. This patient had tested positive for SARS-CoV-2 the previous day with a Ct value of 21.35. The second patient, diagnosed 6 days earlier (Ct value of 17.68), was not receiving any notable treatment. However, approximately 30–40 min before sampling was undertaken, there was a ‘crash call’ elsewhere within the bay. There was no intubation or cardiopulmonary resuscitation, but a significant increase in staff activity was observed and may have facilitated the dispersal of airborne particles. The total volume of each air sample was 3 m3, and the associated Ct values ranged from 37 to 39 which, when quantified, equating to 460 to <10 genomic copies/m3 air.

Discussion

When sampling the healthcare environment, many variables can impact the results obtained. This can make interpretation of the data difficult, particularly if a frame of reference is lacking. In this study and to provide context, agar contact plates were used to provide an aerobic bacterial colony count and an indication of surface cleanliness. Whilst no microbiological standards exist for healthcare surfaces, a benchmark of <2.5 cfu/cm2 has been suggested [22]. ‘Universal’ disinfectant/detergent wipes were used for damp dusting in all but one of the wards visited, where a chlorine-dioxide-based solution with disposable cloths was used instead. Terminal cleaning was chlorine-based. SARS-CoV-2 RNA was detected on 30 (8.9%) of the 336 surfaces sampled (Table II). The proportion of surfaces positive for viral RNA differed between hospitals and ranged from 0% to 27%. This likely reflects the fact that sampling was undertaken on different types of ward occupied by different types of patient requiring different types of care and/or treatment (Table I), rather than differences in cleaning product or protocol. Overall, however, the results are similar to those of other studies [13,15,20] and suggest that, whilst SARS-CoV-2 can contaminate healthcare surfaces, widespread contamination is unlikely [17]. The bacterial load on the majority (89%) of surfaces sampled was <2.5 cfu/cm2 suggesting that, in general and despite increased pressure on beds and workload, the routine cleaning performed by the nursing and domestic staff across all eight hospitals was effective. Nonetheless, contamination of the healthcare environment can occur, and SARS-CoV-2 RNA was detected on the same type of surface in multiple hospitals (Table II), implying that, despite the effectiveness of the cleaning protocols employed, some types of surface could facilitate the survival, persistence and/or dispersal of SARS-CoV-2. Patients consider the nurse call button a direct conduit to care, and many patients were observed to hold the button close even whilst dozing. Intensity and frequency of contact can increase microbial transfer from hands to surface [23], and SARS-CoV-2 RNA was detected on four (17%) of the nurse call buttons sampled. Ct values ranged from 30.8 to 36.2, equating to 2.9 x 104 to 1.2 x 103 genomic copies/swab (Table II). To reduce the risk of transmission of SARS-CoV-2 in the hospital setting, it is recommended that surfaces such as over-bed tables, bed rails and nurse call buttons should be cleaned at least twice daily [24]. The median number of bacteria recovered from nurse call buttons was 50 cfu/25cm2 (2 cfu/cm2). Fewer bacteria (<1 cfu/cm2) were recovered from tables and bed rails, suggesting that these surfaces are (and can be) cleaned effectively. Heavy contamination of the nurse call button has been described previously [25], and staff should be reminded that routine cleaning should include all aspects of the patient bed. Future consideration should be given to design modification and/or improving the ability to clean nurse call buttons. Patient mobility can contribute greatly to the spread of bacteria within a ward [25]. Similarly, SARS-CoV-2 RNA has been detected on patient contact sites outside the immediate bed space [13,17] and, in the current study, outside of cohort bays – specifically, toilet door handles. The presence of SARS-CoV-2 RNA on door handles has been reported previously [13,14,20], and the contact area between the hand and handle and the grip pressure likely facilitates transfer to and from the hands. In this study, the amount of SARS-CoV-2 RNA detected on one door handle was 2.2 x 105 genomic copies/swab, implying significant transfer from a contaminated hand. Despite this, the authors were unable to culture viable virus. The lowest genomic copy number (N gene) required to isolate virus from clinical samples is reportedly 5 x 105 genomic copies/mL [26], which is higher than the copy number in any of the environmental samples collected during this study. Subjecting the samples to multiple freeze–thaw cycles may also have impacted infectivity by disrupting virion and genome integrity [26]. Regardless, there is potential for viable virus to contaminate a single door handle and to be transferred to the hands of numerous successive contacts and, as a consequence, to other inanimate surfaces [27]. SARS-CoV-2 RNA was detected on 4.9% (7/143) and 13.8% (22/159) of surfaces sampled in the ICU/HDU and non-ICU wards, respectively. In contrast to patients admitted to cohort wards, patients in ICUs/HDUs are more likely to be bed bound and be receiving mechanical ventilation. Reduced patient mobility likely contributed to the less-frequent detection of SARS-CoV-2 in ICUs/HDUs. However, viral RNA was still detected on staff contact sites (e.g. monitor screens, syringe drivers; Table II). Disposable gloves are an important element of PPE and can prevent the hands of HCWs from acquiring pathogens. However, during routine patient care, the glove surface itself can become contaminated. If gloves are not changed regularly and appropriately, contamination of surfaces via gloved hands can occur [19]. When caring for patients with COVID-19, particularly in the ICU/HDU setting, the requirement to don full PPE presents additional challenges in terms of preserving PPE and ensuring that staff know how to implement appropriate hand hygiene within an outbreak setting [14,19]. Non-critical medical devices (e.g. blood pressure cuffs, temperature probes) have been implicated in nosocomial infection [28]. SARS-CoV-2 was detected on four (31%) of 13 portable vital signs monitors (Table II). The highest level of viral RNA (1.6 x 103 genomic copies/swab) was detected on a fingertip pulse oximeter associated with a machine that had been removed from a single room occupied by a patient with COVID-19. The other three machines were located in cohort bays. When (or on whom) these machines were last used or when they were last cleaned was not known, and the results demonstrate the presence and/or persistence of viral RNA and not infectious virus. Nonetheless, contact pressure has been shown to significantly affect viral transfer to and from fingerpads [29]. In the absence of cleaning, fingertip pulse oximeters could facilitate transmission of SARS-CoV-2, particularly between asymptomatic and non-infected patients. SARS-CoV-2 RNA was not detected in any air sample taken using the MD8 sampler. This may have been due to the relatively low volume of air sampled (0.5 m3) and/or the inability to concentrate the viscous dissolved gelatine membrane. In contrast, viral RNA was detected in four (7.3%) of the high volume (3 m3) air samples taken using the Coriolis μ sampler. It is not known what may have contributed to this airborne contamination, but two of these samples were taken <1 m from two patients receiving CPAP therapy (Table I). CPAP is considered to be an AGP. However, air samples were taken close to 11 other patients receiving NIV, seven of whom had also tested positive for COVID-19 <7 days earlier. No viral RNA was detected. The make/model of CPAP machine used to treat these two patients was not used elsewhere, and it is possible that the equipment used to deliver NIV to patients may promote the generation and/or release of aerosols [30]. How the apparatus is used or tolerated may also have an effect. During sampling, one of the two patients was observed to turn over in bed multiple times, and on one occasion, disconnected the CPAP machine to aid movement. The dispersal distance of exhaled air from a jet nebulizer and Venturi-type oxygen mask is estimated to be 0.8 m and 0.4 m, respectively [30]. In this study, SARS-CoV-2 was not detected in any air samples collected during drug nebulization. Viral RNA was detected <1 m from one of six patients receiving oxygen. Time since diagnosis and time since symptom onset were 1 and 6 days, respectively, which were comparatively earlier than many of the other patients (Table I). Others hypothesize that the concentration of SARS-CoV-2 in the air and/or on high-touch surfaces is highest during the first week of illness [11], suggesting that new admissions to hospital may have greater potential to transmit the virus to others. It has been suggested that placing suspected cases of COVID-19 in single rooms or bays that are fully disinfected between admissions could reduce nosocomial infection rates by 80% [10]. No formal assessment of air exchange and ventilation efficiency was performed as part of this study. It was assumed that, in accordance with UK guidelines [31], the pressure differential between isolation rooms and corridors was monitored continually (and that negative pressure was maintained), and that the air change rate in critical care areas was higher than in neutral pressure single rooms and general wards, including those areas in which SARS-CoV-2 RNA was detected in the air. Whilst this could infer the benefit of increased air exchange, in this study, air samples were taken <1 m from patients so the impact of ventilation on aerosol levels would have been minimal. In all four cases where SARS-CoV-2 was detected in air samples, the concentration of viral RNA was low and ranged from 460 to <10 genomic copies/m3 air. As discussed, samples containing this level of viral nucleic acid are unlikely to contain viable (infectious) virus [27], and this finding, together with the inability to detect SARS-CoV-2 RNA in all other air samples, supports current guidance on the use of specific PPE for AGPs and non-AGPs. It is acknowledged, however, that many of the procedures believed to generate aerosols and droplets were not captured during this study, and that samples were only collected over a 10-min period. Unprotected, prolonged exposure to an infected patient has been linked to transmission [9]. In a rapidly evolving outbreak situation, there is a need to gain rapid understanding of certain trends; in this case, contamination of the healthcare environment. Despite its limitations, this multi-centre study supports the findings of other studies [13,15,19,20] and should provide assurance to HCWs. SARS-CoV-2 may be present on frequently touched surfaces, but effective cleaning should reduce the risk of fomite transmission [21] and limit the concentration of SARS-CoV-2 in aerosols [17]. Recommendations to clean frequently touched surfaces regularly are warranted, and the need to clean items such as door handles, nurse call buttons and multi-use patient monitoring equipment should be emphasized. In wards caring for patients with COVID-19, viral RNA in the air was either not detected or was present at a very low concentration. These results suggest that, if worn and used correctly, the PPE recommended in the UK, including components to protect against aerosol exposures when indicated, should provide adequate protection against the potential virus exposure risks identified in this study.
  23 in total

1.  Research please.

Authors:  R Vasant; A Haigh
Journal:  Br Dent J       Date:  2021-02       Impact factor: 1.626

Review 2.  A Bibliometric Analysis of Personal Protective Equipment and COVID-19 Researches.

Authors:  Yu Zhang; Man Hu; Junwu Wang; Pingchuan Wang; Pengzhi Shi; Wenjie Zhao; Xin Liu; Qing Peng; Bo Meng; Xinmin Feng; Liang Zhang
Journal:  Front Public Health       Date:  2022-04-29

Review 3.  What is the risk of acquiring SARS-CoV-2 from the use of public toilets?

Authors:  Stephanie J Dancer; Yuguo Li; Alwyn Hart; Julian W Tang; Davey L Jones
Journal:  Sci Total Environ       Date:  2021-06-11       Impact factor: 7.963

Review 4.  Contact transmission of SARS-CoV-2 on fomite surfaces: surface survival and risk reduction.

Authors:  Abhimanyu Tharayil; R Rajakumari; Miran Mozetic; Gregor Primc; Sabu Thomas
Journal:  Interface Focus       Date:  2021-12-10       Impact factor: 3.906

5.  Protocol for the COG-UK hospital-onset COVID-19 infection (HOCI) multicentre interventional clinical study: evaluating the efficacy of rapid genome sequencing of SARS-CoV-2 in limiting the spread of COVID-19 in UK NHS hospitals.

Authors:  James Blackstone; Oliver Stirrup; Fiona Mapp; Monica Panca; Andrew Copas; Paul Flowers; Leanne Hockey; James Price; David Partridge; Christine Peters; Thushan de Silva; Gaia Nebbia; Luke B Snell; Rachel McComish; Judith Breuer
Journal:  BMJ Open       Date:  2022-04-19       Impact factor: 3.006

6.  Adsorption of SARS CoV-2 spike proteins on various functionalized surfaces correlates with the high transmissibility of Delta and Omicron variants.

Authors:  Daniela Dobrynin; Iryna Polishchuk; Lotan Portal; Ivan Zlotver; Alejandro Sosnik; Boaz Pokroy
Journal:  Mater Today Bio       Date:  2022-04-19

7.  Inactivation of Human Coronavirus by FATHHOME's Dry Sanitizer Device: Rapid and Eco-Friendly Ozone-Based Disinfection of SARS-CoV-2.

Authors:  Timsy Uppal; Amir Khazaieli; Antoine M Snijders; Subhash C Verma
Journal:  Pathogens       Date:  2021-03-14

8.  Controlling risk of SARS-CoV-2 infection in essential workers of enclosed food manufacturing facilities.

Authors:  Julia S Sobolik; Elizabeth T Sajewski; Lee-Ann Jaykus; D Kane Cooper; Ben A Lopman; Alicia Nm Kraay; P Barry Ryan; Juan S Leon
Journal:  medRxiv       Date:  2021-05-18

9.  Persistence of Severe Acute Respiratory Syndrome Coronavirus 2 (SARS-CoV-2) Virus and Viral RNA in Relation to Surface Type and Contamination Concentration.

Authors:  Susan Paton; Antony Spencer; Isobel Garratt; Katy-Anne Thompson; Ikshitaa Dinesh; Paz Aranega-Bou; David Stevenson; Simon Clark; Jake Dunning; Allan Bennett; Thomas Pottage
Journal:  Appl Environ Microbiol       Date:  2021-06-25       Impact factor: 4.792

10.  Evaluating splatter and settled aerosol during orthodontic debonding: implications for the COVID-19 pandemic.

Authors:  Hayley Llandro; James R Allison; Charlotte C Currie; David C Edwards; Charlotte Bowes; Justin Durham; Nicholas Jakubovics; Nadia Rostami; Richard Holliday
Journal:  Br Dent J       Date:  2021-01-08       Impact factor: 2.727

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.