Aerosol-scavenging isolation barrier mitigates exposure risk during endonasal procedures in coronavirus-2019.

Alan D Workman1,2, Benjamin S Bleier1,2, Tofigh Sayahi1,2,3, Kerry E Kelly3, Phillip C Song1,2.   

Abstract

Entities:  

Keywords:  COVID-19; aerosol; barrier; endonasal; isolation barrier; mask

Mesh:

Substances:

Year:  2020        PMID: 33258275      PMCID: PMC7753482          DOI: 10.1002/alr.22752

Source DB:  PubMed          Journal:  Int Forum Allergy Rhinol        ISSN: 2042-6976            Impact factor:   5.426


× No keyword cloud information.
The US Centers for Disease Control and Prevention (CDC) and World Health Organization (WHO) advise that severe acute respiratory syndrome‒coronavirus‐2 (SARS‐CoV‐2) may spread from asymptomatic individuals through respiratory droplets and airborne aerosols. , Although masking to limit aerosol spread has been robustly demonstrated, the otolaryngology examination requires unmasking to access the nostrils and mouth. Furthermore, rigid nasal endoscopy (RNE), nasal endoscopy using a nasopharyngolaryngoscopy (NPL) scope, and RNE with sinonasal debridement (RNESD) may produce significant airborne aerosols even without sneezing. , Given these substantial infection control challenges, the purpose of this study was to test the efficacy of a novel isolaer isolation barrier (IIB) (Inquis Medical, Redwood City, CA), which was specifically designed to enable endonasal procedures while capturing both respiratory droplets and airborne aerosols.

Patients and methods

Patient population and study design

This prospective, open‐label study of 30 patients scheduled for RNE, NPL, or RNESD in the rhinology or laryngology clinic was approved by the institutional review board of Mass General Brigham as minimal risk. The IIB consists of a barrier surrounding the nose and mouth with an access window and an elastic strap. The clear flexible plastic apron provides direct facial access while acting as an occlusive physical barrier during aerosolization events. The apron contains a flap valve enabling the direct passage of endoscopes and instruments under the apron, through the valve, or both. The bottom suction port connects to a standard clinic or portable suction source. The continuous suction creates a high‐flow negative‐pressure field within the barrier that turns over exhaled air and scavenges aerosols (Fig. 1A,B). Aerosol generation during positive (without isolation barrier [IB]) and negative (with IB donned, suction at −250 mmHg, flow rate of 50 L/min) control conditions were measured after baseline determination. Control conditions included simulated and intranasal venturi atomization. Clinic procedures were performed with the IB donned and suction applied. Oxygen saturation was monitored throughout the study as a routine precaution.
FIGURE 1

(A) Illustration of the isolaer isolation barrier system consisting of the valve (a), plastic apron (b), and suction port (c). (B) Example of rigid nasal endoscopy with sinonasal debridement (RNESD) using the isolation barrier. In this case, the endoscope is placed directly through the apron valve while a second hand is placed under the apron and through the access window to perform nasal suctioning. (C) Mean aerosol (0.5‐10 µm) production above individual matched ambient baseline levels (in particles/liter ± standard error) among control and clinical procedures (atomizer inside apron, n = 10; sneeze, procedure consisting of rigid nasal endoscopy [RNE], RNESD, and nasopharyngolaryngoscopy [NPL], n = 30; *p < 0.05). (D) Subject body mass index (BMI) vs fold change in aerosol concentration compared with ambient baseline for RNE, NPL, RNESD, and positive control sneeze (without isolation barrier). No significant correlation was observed between BMI and any condition.

(A) Illustration of the isolaer isolation barrier system consisting of the valve (a), plastic apron (b), and suction port (c). (B) Example of rigid nasal endoscopy with sinonasal debridement (RNESD) using the isolation barrier. In this case, the endoscope is placed directly through the apron valve while a second hand is placed under the apron and through the access window to perform nasal suctioning. (C) Mean aerosol (0.5‐10 µm) production above individual matched ambient baseline levels (in particles/liter ± standard error) among control and clinical procedures (atomizer inside apron, n = 10; sneeze, procedure consisting of rigid nasal endoscopy [RNE], RNESD, and nasopharyngolaryngoscopy [NPL], n = 30; *p < 0.05). (D) Subject body mass index (BMI) vs fold change in aerosol concentration compared with ambient baseline for RNE, NPL, RNESD, and positive control sneeze (without isolation barrier). No significant correlation was observed between BMI and any condition.

Aerosol sampling

Aerosol quantification was performed in a 10.3‐m2 room with 6 air changes per hour using a particle sizer (OPS 3330; TSI, Inc, Shoreview, MN)3 25 cm from the subject's nares. Total particle counts were quantified and patient comfort was assessed using a satisfaction form. A sample size of 30 was determined to have a 2‐sided significance of 0.001 and a power of 0.95 to detect a difference between the IIB and sneeze control. Particles were quantified as total per 10‐second interval and averaged in procedures lasting up to 70 seconds. Individualized background particle concentration obtained for each patient were subtracted from each condition, as described elsewhere.

Results

Patients’ demographics

The mean age was 54.3 ± 19.6 years and 33% of subjects had facial hair. Mean donning and doffing times were 19.5 ± 7.1 seconds and 5.2 ± 0.9 seconds, respectively (Table 1).
TABLE 1

Patient demographics, procedural data, and satisfaction rating

Number of subjects30
Age (years)54.3 ± 19.6
Sex 
 Male13 (43%)
 Female17 (57%)
Body mass index 26.4 ± 4.6
Facial hair10 (33%)
Donning time (seconds)19.5 ± 7.1
Doffing time (seconds)5.2 ± 0.9
Baseline particulate (0.5‐10 µm/L)338.4 ± 15.1
Particles (0.5‐10 µm/L) above baseline
 Atomizer (positive control)15,507 ± 4541.0
 Sneeze (positive control)268.6 ± 91.8
 Atomizer with barrier31.6 ± 15.9
 Sneeze with barrier−32.2 ± 20.8
 Procedure with barrier47.1 ± 13.1
Mean particle size (µm) 
 Baseline1.11 ± 0.07
 Atomizer positive control0.98 ± 0.01
 Sneeze positive control1.26 ± 0.07
 Procedure with barrier1.04 ± 0.06
Barrier satisfaction (0‐10 scale)9.5 ± 0.8
 Face soreness1 (3%)
 Skin irritation1 (3%)
 Barrier moved2 (7%)
 Difficulty fitting barrier3 (10%)
 Claustrophobia0 (0%)
 Dry mouth0 (0%)
 Dry nose0 (0%)
 Dry eyes0 (0%)
 Difficulty breathing out0 (0%)
 Difficulty breathing in1 (3%)
 Barrier too noisy1 (3%)
Patient demographics, procedural data, and satisfaction rating

Aerosol mitigation

The mean ambient baseline was 338.4 ± 15.1 particles/L. Mean particles over baseline during atomization and sneeze was 15,507 ± 4541.0 particles/L (fold change [FC] from background, 46.8; n = 10 subjects; p < 0.05) and 268.6 ± 91.8 particles/L (FC from background, 1.79; n = 30 subjects; p < 0.05), respectively. Among both negative control and procedure conditions (n = 30 subjects), aerosol production was not significantly different from ambient baseline, regardless of facial hair presence/absence (Fig. 1C). No significant correlation was observed between body mass index (BMI) and FC in particulates over baseline during any clinical procedure or sneeze control (Fig. 1D).

Patient satisfaction

Using a visual analog scale with 1 being “least satisfied” and 10 being “most satisfied,” patients rated their mean satisfaction with wearing the IIB at 9.5 ± 0.8 (Table 1).

Discussion

SARS‐CoV‐2 can spread by respiratory droplets and airborne aerosols and remains viable in air for 3 to 16 hours. Surgical mask efficacy is limited and may still allow significant airborne aerosol exposure. The otolaryngology exam requires direct access to the nose and mouth, thereby obviating even the partial protection afforded by surgical masks. The performance of endonasal procedures carries all of the hallmarks of a potential infectious encounter as nasal endoscopy and RNESD can produce significant aerosols, , may induce cough and/or sneeze, and occur within a small clinic room in close proximity to an unmasked patient. In parallel, epidemiologic evidence of asymptomatic spread even in the absence of overt coughing or sneezing has come to light. Therefore, regardless of their generative source, unmasked patient‐derived aerosols produced during the totality of the endonasal exam pose a potential infectious risk to both the provider team and the clinic air supply. The IIB tested within this study comprises a physical barrier that provides a window enabling procedural access to nose and mouth while effectively scavenging airborne aerosols through high flow and negative pressure, as recommended by the US Food and Drug Administration. The results demonstrate that the IIB was well tolerated and no small droplet or airborne aerosol escape were detected during procedural conditions. In addition, the aerosolizing control conditions were successfully mitigated using the barrier. Finally, IIB efficacy did not appear to be dependent on BMI, presence of facial hair, or tested particle size. The degree to which otolaryngology providers have been exposed to unmitigated patient aerosol plumes is now evident. The future of the coronavirus‐2019 (COVID‐19) pandemic remains uncertain in light of ongoing questions regarding vaccine efficacy, transmission blocking activity, hesitance, and durability. Even in a post‒COVID‐19 world, seasonal exposure to “common cold” viruses and the potential for more frequent epi/pandemics will all remain persistent threats to infectious transmission within the otolaryngology clinic. The integration of routine control strategies will help to mitigate these risks and enable us to continue to provide safe, high‐level care to our patients.
  8 in total

Review 1.  Mucosal immunity and vaccines.

Authors:  Jan Holmgren; Cecil Czerkinsky
Journal:  Nat Med       Date:  2005-04       Impact factor: 53.440

2.  Effectiveness of facemasks to reduce exposure hazards for airborne infections among general populations.

Authors:  A C K Lai; C K M Poon; A C T Cheung
Journal:  J R Soc Interface       Date:  2011-09-21       Impact factor: 4.118

3.  Airborne Aerosol Generation During Endonasal Procedures in the Era of COVID-19: Risks and Recommendations.

Authors:  Alan D Workman; Aria Jafari; D Bradley Welling; Mark A Varvares; Stacey T Gray; Eric H Holbrook; George A Scangas; Roy Xiao; Bob S Carter; William T Curry; Benjamin S Bleier
Journal:  Otolaryngol Head Neck Surg       Date:  2020-05-26       Impact factor: 3.497

4.  Persistence of Severe Acute Respiratory Syndrome Coronavirus 2 in Aerosol Suspensions.

Authors:  Alyssa C Fears; William B Klimstra; Paul Duprex; Amy Hartman; Scott C Weaver; Kenneth S Plante; Divya Mirchandani; Jessica Ann Plante; Patricia V Aguilar; Diana Fernández; Aysegul Nalca; Aysegul Totura; David Dyer; Brian Kearney; Matthew Lackemeyer; J Kyle Bohannon; Reed Johnson; Robert F Garry; Doug S Reed; Chad J Roy
Journal:  Emerg Infect Dis       Date:  2020-06-22       Impact factor: 6.883

5.  Characterization of Aerosols Generated During Patient Care Activities.

Authors:  Caroline A O'Neil; Jiayu Li; Anna Leavey; Yang Wang; Matthew Hink; Meghan Wallace; Pratim Biswas; Carey-Ann D Burnham; Hilary M Babcock
Journal:  Clin Infect Dis       Date:  2017-10-15       Impact factor: 9.079

6.  Airborne transmission of SARS-CoV-2: The world should face the reality.

Authors:  Lidia Morawska; Junji Cao
Journal:  Environ Int       Date:  2020-04-10       Impact factor: 9.621

7.  Aerosol-scavenging isolation barrier mitigates exposure risk during endonasal procedures in coronavirus-2019.

Authors:  Alan D Workman; Benjamin S Bleier; Tofigh Sayahi; Kerry E Kelly; Phillip C Song
Journal:  Int Forum Allergy Rhinol       Date:  2020-12-10       Impact factor: 5.426

8.  Quantification of Aerosol Particle Concentrations During Endoscopic Sinonasal Surgery in the Operating Room.

Authors:  Alex Murr; Nicholas R Lenze; William Colby Brown; Mark W Gelpi; Charles S Ebert; Brent A Senior; Brian D Thorp; Adam M Zanation; Adam J Kimple
Journal:  Am J Rhinol Allergy       Date:  2020-10-04       Impact factor: 2.467

  8 in total
  2 in total

1.  Stop the leak!: Mitigating potential exposure of aerosolized COVID-19 during laparoscopic surgery.

Authors:  John M Uecker; Farshid Alembeigi; Christopher R Idelson; Austin Fagerberg; Naser Ahmad; Alexander Cohen; Mitchell Gilkey
Journal:  Surg Endosc       Date:  2020-09-24       Impact factor: 4.584

2.  Aerosol-scavenging isolation barrier mitigates exposure risk during endonasal procedures in coronavirus-2019.

Authors:  Alan D Workman; Benjamin S Bleier; Tofigh Sayahi; Kerry E Kelly; Phillip C Song
Journal:  Int Forum Allergy Rhinol       Date:  2020-12-10       Impact factor: 5.426