Fernanda Raphael Escobar Gimenes1, Flávia Fernanda Luchetti Rodrigues Baracioli2, Adriane Pinto de Medeiros1, Patricia Rezende do Prado3, Janine Koepp4, Marta Cristiane Alves Pereira1, Camila Baungartner Travisani1, Soraia Assad Nasbine Rabeh1, Fabiana Bolela de Souza1, Adriana Inocenti Miasso1.
Abstract
AIMS: To identify the types of nasogastric/nasoenteric tube (NGT/NET)-related adverse events and to analyze the degree of harm and the factors associated with mechanical device-related complications.Entities:
Year: 2020 PMID: 33211726 PMCID: PMC7676660 DOI: 10.1371/journal.pone.0241849
Source DB: PubMed Journal: PLoS One ISSN: 1932-6203 Impact factor: 3.240
Type of NGT/NET-related AE and degree of harm (N = 191).
| Variables | % | |
|---|---|---|
| AE—median ± | 1 ± 0.98 (1–6) | - |
| Nasal pressure injury | 1 | 0.5 |
| Epistaxis/nosebleed | 5 | 2.6 |
| Multiple attempts to introduce the tube | 5 | 2.6 |
| Migration/displacement | 6 | 3.1 |
| Bronchoaspiration | 9 | 4.7 |
| Tube obstruction | 25 | 13.1 |
| Inadvertent tube removal | 134 | 70.1 |
| Total | 185 | 96.9 |
| Tubing misconnection | 2 | 1.0 |
| Tube quality | 4 | 2.0 |
| Total | 6 | 3.1 |
| Death | 0 | 0.0 |
| Severe | 3 | 1.6 |
| Moderate | 27 | 14.1 |
| None | 64 | 33.5 |
| Mild | 95 | 49.7 |
| Total | 189 | 98.9 |
AE, Adverse event; SD, Standard deviation.
aCounts do not sum to the total of 191 due to missing values.
Distribution of variables of the patients with NGT/NET-related AE (N = 191).
| Variables | % | |
|---|---|---|
| No. of comorbidities—median ± | 3 ± 2.15 (0–9) | - |
| Intensive | 8 | 4.2 |
| Minimum | 18 | 9.4 |
| Intermediate | 30 | 15.7 |
| Semi-intensive | 36 | 18.8 |
| High dependency | 95 | 49.7 |
| Total | 187 | 97.9 |
| No risk | 16 | 8.4 |
| Low risk | 40 | 20.9 |
| Moderate risk | 55 | 28.8 |
| High risk | 80 | 41.9 |
| Total | 191 | 100.0 |
| Drowsy | 10 | 5.2 |
| Unconscious | 17 | 8.9 |
| Confused | 74 | 38.7 |
| Alert | 88 | 46.1 |
| Total | 189 | 98.9 |
| Yes | 36 | 18.8 |
| No | 154 | 80.6 |
| Total | 190 | 99.5 |
SD, Standard deviation; PCS, Patient Classification System; CCI, Charlson Comorbidity Index; ACDU, Alert, Confused, Drowsy and Unconscious.
aCounts do not sum to the total of 191 due to missing values.
Analysis of the association between the degree of harm and care complexity, disease severity and level of consciousness (N = 191).
| Variables | Degree of harm | ||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|
| None | Low risk | Moderate risk | High risk | Total | |||||||
| n | % | n | % | n | % | n | % | n | % | ||
| Intensive | 1 | 0.5 | 4 | 2.1 | 3 | 1.6 | 0 | 0.0 | 8 | 4.2 | .997 |
| Minimum | 5 | 2.6 | 10 | 5.2 | 12 | 6.3 | 0 | 0.0 | 17 | 8.9 | |
| Intermediate | 8 | 4.2 | 15 | 7.8 | 6 | 3.1 | 0 | 0.0 | 29 | 15.2 | |
| Semi-intensive | 12 | 6.3 | 17 | 8.9 | 5 | 2.6 | 2 | 1.0 | 36 | 18.8 | |
| High dependency | 37 | 19.4 | 45 | 23.6 | 11 | 5.7 | 1 | 0.5 | 94 | 49.2 | |
| Total | 63 | 33.0 | 91 | 47.6 | 27 | 14.1 | 3 | 1.6 | 184 | 96.3 | |
| No risk | 5 | 2.6 | 11 | 5.7 | 0 | 0.0 | 0 | 0.0 | 16 | 8.4 | .794 |
| Low risk | 15 | 7.8 | 16 | 8.8 | 7 | 3.7 | 0 | 0.0 | 38 | 19.9 | |
| Moderate risk | 19 | 9.9 | 28 | 14.6 | 7 | 3.7 | 1 | 0.5 | 55 | 28.8 | |
| High risk | 25 | 13.1 | 39 | 20.4 | 13 | 6.8 | 2 | 1.0 | 79 | 41.4 | |
| Total | 64 | 33.5 | 94 | 49.2 | 27 | 14.1 | 3 | 1.6 | 188 | 98.4 | |
| Drowsy | 1 | 0.5 | 2 | 1.0 | 6 | 3.1 | 1 | 0.5 | 10 | 5.2 | .909 |
| Confused | 16 | 8.4 | 42 | 22.0 | 14 | 7.3 | 1 | 0.5 | 73 | 38.2 | |
| Unconscious | 10 | 5.2 | 5 | 2.6 | 1 | 0.5 | 1 | 0.5 | 17 | 8.9 | |
| Alert | 37 | 19.4 | 43 | 22.5 | 6 | 3.1 | 0 | 0.0 | 86 | 45.0 | |
| Total | 64 | 33.5 | 92 | 48.2 | 27 | 14.1 | 3 | 1.6 | 186 | 97.4 | |
PCS, Patient Classification System; CCI, Charlson Comorbidity Index; ACDU, Alert, Confused, Drowsy and Unconscious.
aCounts do not sum to the total of 191 due to missing values.
bLinear-by-linear chi-square test.
Analysis of the association between mechanical device-related complication and care complexity, disease severity, level of consciousness and use of an invasive breathing device (N = 717).
| Variables | Mechanical device-related complication | ||||||
|---|---|---|---|---|---|---|---|
| Yes | No | Total | |||||
| % | % | % | |||||
| .862 | |||||||
| Intensive | 7 | 1.0 | 17 | 2.4 | 24 | 3.3 | |
| Minimum | 14 | 2.0 | 78 | 10.9 | 92 | 12.8 | |
| Semi-intensive | 24 | 3.3 | 108 | 15.1 | 132 | 18.4 | |
| Intermediate | 44 | 6.1 | 91 | 12.7 | 135 | 18.8 | |
| High dependency | 89 | 12.4 | 215 | 30.0 | 304 | 42.4 | |
| Total | 178 | 24.8 | 509 | 71.0 | 687 | 95.8 | |
| .041 | |||||||
| No risk | 13 | 1.8 | 29 | 4.0 | 42 | 5.8 | |
| Low risk | 38 | 5.3 | 83 | 11.6 | 121 | 16.9 | |
| Moderate risk | 52 | 7.3 | 123 | 17.1 | 175 | 24.4 | |
| High risk | 87 | 12.1 | 289 | 40.3 | 376 | 52.4 | |
| Total | 190 | 26.5 | 529 | 73.8 | 714 | 99.6 | |
| .041 | |||||||
| Drowsy | 6 | 0.8 | 50 | 7.0 | 56 | 7.8 | |
| Unconscious | 17 | 2.4 | 64 | 8.9 | 81 | 11.3 | |
| Confused | 48 | 6.7 | 111 | 15.5 | 159 | 22.2 | |
| Alert | 117 | 16.3 | 300 | 41.8 | 417 | 58.1 | |
| Total | 188 | 26.2 | 525 | 73.2 | 713 | 99.4 | |
| .967 | |||||||
| Yes | 28 | 3.9 | 91 | 12.7 | 119 | 16.6 | |
| No | 162 | 22.6 | 435 | 60.7 | 597 | 83.3 | |
| Total | 190 | 26.5 | 526 | 73.4 | 716 | 99.9 | |
PCS, Patient Classification System; CCI, Charlson Comorbidity Index; ACDU, Alert, Confused, Drowsy and Unconscious.
aCounts do not sum to the total of 717 due to missing values.
bCochran-Armitage Chi-Square test.
cFisher’s exact test.
Independent variables included in the final logistic regression model for the analysis of predictors of mechanical device-related complications in tube fed patients.
| Variables | Estimate | |||
|---|---|---|---|---|
| (Intercept) | -1.1215 | 0.421 | -2.661 | 0.007 |
| Age | -0.0104 | 0.005 | -1.837 | 0.066 |
| High dependency care | 1.3974 | 0.348 | 4.007 | <0.001 |
| Intensive care | 1.6369 | 0.613 | 2.667 | 0.007 |
| Intermediate care | 1.2647 | 0.364 | 3.469 | <0.001 |
| Semi-intensive care | 0.5240 | 0.404 | 1.295 | 0.195 |
| Confused | -0.2039 | 0.233 | -0.873 | 0.382 |
| Unconscious | -0.2497 | 0.340 | -0.734 | 0.462 |
| Drowsy | -1.3978 | 0.471 | -2.965 | 0.003 |
| Death | -0.6221 | 0.249 | -2.489 | 0.012 |
SD, Standard Deviation; PCS, Patient Classification System; ACDU, Alert, Confused, Drowsy and Unconscious.
Model of logistic regression analysis with the response variable mechanical device-related complication (N = 197).
| Variables | Gross OR (95%CI) | Adjusted OR (95%CI) | ||
|---|---|---|---|---|
| Minimum care (reference) | ||||
| High dependency care | 2.64 (1.39–5.02) | .003 | 3.56 (1.83–6.93) | .001 |
| Intensive care | 1.98 (0.66–5.97) | .224 | 4.72 (1.43–15.52) | .011 |
| Intermediate care | 2.74 (1.37–5.49) | .004 | 3.23 (1.60–6.54) | .001 |
| Semi-intensive care | 1.26 (0.60–2.67) | .545 | 1.60 (0.72–3.52) | .247 |
| Alert (reference) | ||||
| Confused | 0.98 (0.63–1.51) | .918 | 0.81 (0.52–1.28) | .371 |
| Unconscious | 0.61 (0.34–1.1) | .103 | 0.76 (0.39–1.48) | .423 |
| Drowsy | 0.32 (0.13–0.78) | .012 | 0.24 (0.10–0.61) | .003 |
| Death vs. Non-Death | 0.59 (0.37–0.93) | .023 | 0.51 (0.31–0.83) | .007 |
OR, Odds Ratio; 95%CI, 95% Confidence Interval; PCS, Patient Classification System; ACDU, Alert, Confused, Drowsy and Unconscious.
ap values were calculated using the Likelihood Ratio test.