Bacterial colonies of Streptococcus mutans were initially pre‐inoculated from a frozen stock and allowed to grow in 1 ml of Todd Hewitt Broth culture medium (Manufacturer: Becton Dickinson Sparks, MD) overnight.The following day, the bacterial solution was diluted in order to obtain an O.D (optical density) of 480 nm = 0.1 CFU (Colony forming units, i.e., the bacterial quantity). The optimal wavelength for Streptococcus mutans colonies has been validated with previous experiments (Lundgren et al., 2017).Standard curves for OD determinations were generated and compared with direct bacteria count by Acridine orange and subsequent visualization below epifluorescence microscope.The two bacterial counts were consistent with each other; the OD measurement was thus used for subsequent determinations. The bacteria solution, OD 480 Nm = 0.1 CFU, was employed in order to contaminate both microbiological control implants and A, B, and C test groups' implants.The inoculation was carried out under stirring, in the bacterial solution, for 10 hr. Subsequently, the A, B, and C test groups were subjected to the instrumentation treatment as per protocol. Following the instrumental treatment, the implants were inoculated for further 3 hr, under stirring, in a sterile medium. The soil solutions obtained were used for the measurement of the OD. Before being sent to the SEM scanning procedures, the implants were subjected to a bacterial fixing treatment on the surface by immersing 16% Glutaraldehyde for 30 min and then washed with distilled water.
From the analyzes carried out after the instrumentation on the A, B, and C test groups with respect to the microbiological control, it was found that:Group A: there was no reduction of the bacterial count, but an increase of it in the following 3 hours of inoculation with a standard deviation of 3%.Group B: a highly significant reduction in bacterial count occurred with a standard deviation of 8%.Group C: a significant reduction in bacterial count occurred with a standard deviation of 4%.It is recalled, looking at the graph (Figure 1), that for P to have statistical significance, it must be less than 0.01. It is, therefore, deduced that the implants of groups B and C have reached a high statistical significance of bacterial reduction, p < .01.
FIGURE 1
The graph shows for each individual insert the degree of effectiveness in removing the amount of surface bacteria. It can be seen that the insert with the highest bacterial removal capacity is the steel one, in second place the Korean tips, and lastly the peek tip
The graph shows for each individual insert the degree of effectiveness in removing the amount of surface bacteria. It can be seen that the insert with the highest bacterial removal capacity is the steel one, in second place the Korean tips, and lastly the peek tip
Morphological results
Test group (A) vs SEM control at ×500.In the 500x scans, in addition to the scratching and surface smoothing areas, titanium “flakes,” presumably caused by the instrumentation performed with an EMS steel tip, are appreciated. (Figure 2a–d).
FIGURE 2
In the ×500 scans instrumented with EMS steel insert (b–d) we can see, with respect to the virgin control (a), areas of scratching and surface smoothing with the presence of titanium “flakes” presumably caused by the instrumentation made with EMS steel tip. Gruppo test (A) versus Controllo SEM a ×500
In the ×500 scans instrumented with EMS steel insert (b–d) we can see, with respect to the virgin control (a), areas of scratching and surface smoothing with the presence of titanium “flakes” presumably caused by the instrumentation made with EMS steel tip. Gruppo test (A) versus Controllo SEM a ×500Test group (B) Vs SEM control at ×500.In the 500x scans, the marked sanding areas of the surface remain more visible between the normal weaving of the implant with a vertical and horizontal trend, due to the action of the Peek tip mounted on the ultrasound handpiece. (Figure 3a–d).
FIGURE 3
In the ×500 scans instrumented with a peek insert (b–d), compared to the virgin control (a), the areas of marked sanding of the surface with a vertical and horizontal trend, due to the action of the Peek tip mounted on the ultrasonic handpiece, remain more visible between the normal implant texture of the implant. Gruppo test (B) versus Controllo SEM a ×500
In the ×500 scans instrumented with a peek insert (b–d), compared to the virgin control (a), the areas of marked sanding of the surface with a vertical and horizontal trend, due to the action of the Peek tip mounted on the ultrasonic handpiece, remain more visible between the normal implant texture of the implant. Gruppo test (B) versus Controllo SEM a ×500Test group (C) Vs SEM control at ×500.In the 500x scans, it can be seen that the insert has not significantly changed the implant surface, there is a slight smoothing of the surface, due to the action of the tip mounted on the ultrasound handpiece. Even at this magnification, the partial conservation of the surface morphology of the origin of the implant can be observed. (Figure 4a–d).
FIGURE 4
In the ×500 scans, instrumented by means of the IS TiP insert (b–d) with respect to the control (a) it can be seen that the insert has not considerably modified the implant surface, a slight smoothing of the surface is observed, due to the action of the tip mounted on the ultrasound handpiece. Also, at this magnification the partial preservation of the surface morphology of the implant origin can be observed. Scan No. 12 showed the presence of a clear deposit; when analyzed by metallography it was found to be the alloy of which the IS‐TiP‐STS‐3E insert is composed (Figure 6). Gruppo test (C) versus Controllo SEM a ×500
In the ×500 scans, instrumented by means of the IS TiP insert (b–d) with respect to the control (a) it can be seen that the insert has not considerably modified the implant surface, a slight smoothing of the surface is observed, due to the action of the tip mounted on the ultrasound handpiece. Also, at this magnification the partial preservation of the surface morphology of the implant origin can be observed. Scan No. 12 showed the presence of a clear deposit; when analyzed by metallography it was found to be the alloy of which the IS‐TiP‐STS‐3E insert is composed (Figure 6). Gruppo test (C) versus Controllo SEM a ×500
FIGURE 6
(a and b) Graph shows a gold deposit on the surface of the implant found in scan no. 12 due to wear and tear during the instrumentation of the IS‐TiP inserts, which are entirely gold‐plated
Most significant images of the three types of inserts used for mechanical instrumentation (b–d) compared with the control (a). Comparazione a ×500 dei Gruppo test A‐B‐C versus Controllo SEM
Most significant images of the three types of inserts used for mechanical instrumentation (b–d) compared with the control (a). Comparazione a ×500 dei Gruppo test A‐B‐C versus Controllo SEM
Nonresident bacteria in the oral cavity are responsible for many oral diseases; professional oral hygiene in patients who have implant rehabilitation is critical to minimize the risk of peri‐implantitis and avoid implant failure.
Principal findings and practical implications
Although the literature is unfavorable to ultrasonic metal tips, our results tell that TiP STS‐3E metal tip associated with a microbial agent not only minimizes the presence of bacterial colonies but surprisingly is not aggressive to the implant surface.
CONFLICT OF INTEREST
All the authors declare that there are no conflicts of interest regarding the publication of this paper.
AUTHOR CONTRIBUTIONS
Elisabetta Polizzi and Simone Tomasi conceived idea; Bianca D'orto and Giulia Tetè collected and analyzed the data. Elisabetta Polizzi reviewed the paper and was the supervisor.
Authors: J Lincks; B D Boyan; C R Blanchard; C H Lohmann; Y Liu; D L Cochran; D D Dean; Z Schwartz Journal: Biomaterials Date: 1998-12 Impact factor: 12.479
Authors: Bianca D'Orto; Elisabetta Polizzi; Matteo Nagni; Giulia Tetè; Paolo Capparè Journal: Int J Environ Res Public Health Date: 2022-09-17 Impact factor: 4.614