Falk Schneider1,2, Pablo Hernandez-Varas3,4, B Christoffer Lagerholm3, Dilip Shrestha1, Erdinc Sezgin1,5, M Julia Roberti6, Giulia Ossato6, Frank Hecht6, Christian Eggeling1,3,7,8, Iztok Urbančič1,9,10. 1. MRC Human Immunology Unit and MRC Weatherall Institute of Molecular Medicine, University of Oxford, Headley Way, Oxford, OX3 9DS, United Kingdom. 2. falk.schneider@rdm.ox.ac.uk. 3. Wolfson Imaging Centre Oxford, MRC Weatherall Institute of Molecular Medicine, University of Oxford, Headley Way, Oxford, OX3 9DS, United Kingdom. 4. Core Facility for Integrated Microscopy, Panum Institute, University of Copenhagen, 2200 Copenhagen N, Denmark. 5. Science for Life Laboratory, Department of Women's and Children's Health, Karolinska Institutet, Stockholm, Sweden. 6. Leica Microsystems CMS GmbH, Am Friedensplatz 3, 68163 Mannheim, Germany. 7. Institute of Applied Optics and Biophysics, Friedrich-Schiller-University Jena, Max-Wien Platz 4, 07743 Jena, Germany. 8. Leibniz Institute of Photonic Technology e.V., Albert-Einstein-Straße 9, 07745 Jena, Germany. 9. Jožef Stefan Institute, Jamova cesta 39, SI-1000 Ljubljana, Slovenia. 10. iztok.urbancic@ijs.si.
Abstract
Probing the diffusion of molecules has become a routine measurement across the life sciences, chemistry and physics. It provides valuable insights into reaction dynamics, oligomerisation, molecular (re-)organisation or cellular heterogeneities. Fluorescence correlation spectroscopy (FCS) is one of the widely applied techniques to determine diffusion dynamics in two and three dimensions. This technique relies on the temporal autocorrelation of intensity fluctuations but recording these fluctuations has thus far been limited by the detection electronics, which could not efficiently and accurately time-tag photons at high count rates. This has until now restricted the range of measurable dye concentrations, as well as the data quality of the FCS recordings, especially in combination with super-resolution stimulated emission depletion (STED) nanoscopy. Here, we investigate the applicability and reliability of (STED-)FCS at high photon count rates (average intensities of more than 1 MHz) using novel detection equipment, namely hybrid detectors and real-time gigahertz sampling of the photon streams implemented on a commercial microscope. By measuring the diffusion of fluorophores in solution and cytoplasm of live cells, as well as in model and cellular membranes, we show that accurate diffusion and concentration measurements are possible in these previously inaccessible high photon count regimes. Specifically, it offers much greater flexibility of experiments with biological samples with highly variable intensity, e.g. due to a wide range of expression levels of fluorescent proteins. In this context, we highlight the independence of diffusion properties of cytosolic GFP in a concentration range of approx. 0.01-1 µm. We further show that higher photon count rates also allow for much shorter acquisition times, and improved data quality. Finally, this approach also pronouncedly increases the robustness of challenging live cell STED-FCS measurements of nanoscale diffusion dynamics, which we testify by confirming a free diffusion pattern for a fluorescent lipid analogue on the apical membrane of adherent cells.
Probing the diffusion of molecules has become a routine measurement across the life sciences, chemistry and physics. It provides valuable insights into reaction dynamics, oligomerisation, molecular (re-)organisation or cellular heterogeneities. Fluorescence correlation spectroscopy (FCS) is one of the widely applied techniques to determine diffusion dynamics in two and three dimensions. This technique relies on the temporal autocorrelation of intensity fluctuations but recording these fluctuations has thus far been limited by the detection electronics, which could not efficiently and accurately time-tag photons at high count rates. This has until now restricted the range of measurable dye concentrations, as well as the data quality of the FCS recordings, especially in combination with super-resolution stimulated emission depletion (STED) nanoscopy. Here, we investigate the applicability and reliability of (STED-)FCS at high photon count rates (average intensities of more than 1 MHz) using novel detection equipment, namely hybrid detectors and real-time gigahertz sampling of the photon streams implemented on a commercial microscope. By measuring the diffusion of fluorophores in solution and cytoplasm of live cells, as well as in model and cellular membranes, we show that accurate diffusion and concentration measurements are possible in these previously inaccessible high photon count regimes. Specifically, it offers much greater flexibility of experiments with biological samples with highly variable intensity, e.g. due to a wide range of expression levels of fluorescent proteins. In this context, we highlight the independence of diffusion properties of cytosolic GFP in a concentration range of approx. 0.01-1 µm. We further show that higher photon count rates also allow for much shorter acquisition times, and improved data quality. Finally, this approach also pronouncedly increases the robustness of challenging live cell STED-FCS measurements of nanoscale diffusion dynamics, which we testify by confirming a free diffusion pattern for a fluorescent lipid analogue on the apical membrane of adherent cells.
Fluorescence correlation spectroscopy (FCS) has, since its introduction almost 50
years ago, become a widely applied technique to study diffusion dynamics in
synthetic and biological applications [1-3]. It has greatly
contributed to the understanding of molecular diffusion in model systems and living
cells, both in 2D (in vitro models or cellular membranes) and in 3D
(solution or cellular cytoplasm and nucleus) environments [4-8]. Notably, it has offered fundamental insights into the dynamic
organisation of living systems at the molecular level, e.g. by characterising the
transient, dynamic, yet structured nature of the organisation of fluid membranes
[4, 5, 9].FCS provides a plethora of information about molecular dynamics. The diffusion rates
and local concentrations of fluorescent molecules can be determined directly from
the autocorrelation functions [1,
10, 11]. The spatial variability can be further evaluated
by laser-scanning [12-14] and imaging-based variants of FCS
[15-18]. Further, molecular interactions can be probed
either directly, e.g. binding of molecules detected by cross-correlation (FCCS)
[19], or indirectly via
variations of the apparent diffusion coefficient at different length scales,
measured by spot-variation FCS [20]
providing information on diffusion modes as in single particle tracking [21]. Finally, the combination of FCS
with super-resolution stimulated emission depletion (STED) microscopy allows direct
observation of nanoscale diffusion dynamics, shedding new light on molecular
organisation below the diffraction limit [22].All these invaluable details are extracted from intensity fluctuations due to the
transit of fluorescent molecules through the observation spot of the microscope. As
the fluctuations (i.e. bursts in the fluorescence intensity trace) are most obvious
for sparsely labelled samples, FCS is often considered a single molecule technique,
and has thus been shown multiple times to perform accurately in the range of pico-
and nanomolar concentrations [2].
These concentrations, though, can be far from physiological levels present in living
systems, where molecular abundance can be much higher (e.g. average concentration of
a protein in eukaryotic yeast cells is estimated to be around 1 µm
[23]). Nevertheless, it has
been theoretically predicted and experimentally verified that FCS can perform
similarly and can generate accurate results also for much larger concentrations
(>100 nM) [24]. In this regime,
the main factors for signal quality of FCS, often described by the signal-to-noise
ratio (SNR), are the acquisition time (T), and the number of
detected photons per molecule (i.e. molecular brightness, B, which
depends on the absorption cross section and quantum yield of the dye, the power of
the excitation laser, and the detection efficiency of the measurement setup): SNR
B × T1/2 (see for example [11, 25–28]).For the most efficient and reliable detection of fluorescence fluctuations, sensitive
single-photon-counting detectors are typically used, often coupled to fast
electronics that enable accurate recording of photon arrival times thus also
allowing additional photon filtering in post-processing [29]. One of the main drawbacks of this
instrumentation, however, has been its rather long dead time after each photon
detection (>100 ns) [30],
limiting photon count rates to a few MHz, which is far lower than the typical
repetition rate of excitation lasers running at 20–80 MHz. This has posed a severe
limitation to the accuracy and flexibility in FCS experiments at high fluorophore
concentrations, which are however unavoidable for many applications—for example when
measuring binding dynamics of low affinity, or diffusion dynamics and concentrations
of cellular proteins at different expression levels. Several approaches have been
developed to enable FCS measurements even in such cases: labelling of only a
fraction of the molecules, reduction of the simultaneously visible fluorophores via
fluorescence photoswitching [31,
32], splitting-up of the signal
onto several detectors such as on custom-built detector banks [33], or reduction of the effective observation volume
[34, 35] using for example small sample containers [36], near-field structures [37, 38], plasmonic near-field optics [39-41], or super-resolution STED microscopy [5, 42]. Unfortunately, all of these techniques introduce
more complexity and possible bias, for example due to required controls to check
whether the fraction of labelled or photoswitched molecules truly reflects the
entire population, influence on the sample and fluorescent molecules by surface or
small volume effects, setup complexity, or perplexing photophysics of the
fluorescent label.Here, we demonstrate the straightforward realisation of FCS measurements at high
photon count rates on a commercially available microscope, using novel photon
counting instrumentation. By measuring the diffusion of fluorescent dyes in
solutions, artificial and cell membranes, we explore performance, capabilities,
accuracy, applicability and limitations of confocal FCS and STED-FCS experiments at
photon count rates of up to 20–30 MHz per detection channel, revealing great
potential for FCS experiments at high count rates. As examples, we show the
independence of cytosolic protein diffusion on cellular expression levels and the
application of high count rates to STED-FCS measuring the diffusion behaviour of
lipids in apical cellular plasma membranes.
Materials and methods
Preparation of dyes in solution
Atto655 NHS-ester (AttoTec), Abberior STAR Red NHS-ester also termed KK114
(Abberior), and 20 nm crimson beads (Thermofisher) were stored at
concentration >10 µm and diluted in PBS for
measurements.
Preparation of supported lipid bilayers (SLBs)
SLBs were prepared by spin coating as described previously [43]. Briefly, a solution of 1 mg ml−1
DOPC (1,2-dioleoyl-sn-glycero-3-phosphocholine, Avanti Polar Lipids) dissolved
in 1:2 methanol:chloroform was spin-coated at 3200 rpm for 45 s on a 25 mm
diameter cover slip. The formed lipid film was rehydrated with SLB buffer
(150 mM NaCl, 10 mM HEPES, pH 7.4) and washed several times. All cover slips for
SLB preparation were piranha cleaned (3:1,
H2S04:H2O2) and stored in water.
SLBs were labelled with varying amounts of Abberior STAR Red-DPPE
(1,2-dipalmitoyl-sn-glycero-3-phosphoethanolamine, Abberior).
Tissue culture
HeLa cells were cultured at 37 °C at 5% CO2 in high-glucose DMEM
(Thermofisher) supplemented with 10% FBS (Thermofisher), L-Glutamine
(Thermofisher) and penicillin/streptomycin (Thermofisher). Cells were seeded
onto 35 mm IBIDI glass bottom dishes coated with fibronectin (10
µg ml−1 for 5 min and washed with PBS) 24 h
prior to performing the measurements.CHO K1 cells were grown at 37 °C at 5% CO2 in DMEM/F12 (Lonza)
supplemented with 10% FBS (Sigma) and L-Glutamine (Sigma). CHO K1 cells were
labelled in L15 by incubation with a fluorescent lipid analog at room
temperature.The transfections of GFP-SNAP (cytoplasmic GFP) with plasmids obtained from Dr
Katharina Reglinski, were performed with Turbofect (ThermoFisher) according to
the manufacturer’s protocol.
Preparation of giant plasma membrane vesicles (GPMVs)
GPMVs were prepared as described previously [43, 44]. In brief, HeLa cells were cultured as described above but
seeded on 35 mm plastic bottom petri dishes. At a confluency of about 75%, the
cells were washed with GPMV buffer (150 mM NaCl, 2 mM CaCl2, 10 mM
HEPES, pH 7.4) and then incubated with 25 mM PFA and 10 mM DTT in GPMV buffer
for 2 h at 37 °C. The GPMV containing supernatant was collected and labelled
with Abberior STAR Red-PEG(2kDa)-cholesterol (Abberior) at a final concentration
of 0.5 µg ml−1 for 10 min. GPMVs were
non-specifically immobilised on poly-L-lysine (PLL) coated
surfaces as described before [14]. All diffusion measurements in GPMVs were performed on the top
membrane.
Instrumentation and microscopy
All experiments were performed on a Leica SP8 STED FALCON (Leica Microsystems)
equipped with the HC PL APO 100 × /1.40 Oil STED WHITE oil immersion objective
lens (SLB measurements) and the HC PL APO 86 × /1.20 W motCORR STED WHITE
water immersion objective lens with a motorised correction collar (for solution,
cytosolic, and apical cell membrane measurements). The STED WHITE 86 × water
lens has a working distance of 300 µm, and the motorised
correction collar adjusts for refraction index mismatch by optimizing the signal
for every sample (coverslip). It is worth noting that a single initial setting
of the correction collar was sufficient to correct for varying depth over the
investigated range of 100 µm (figure SI 5 (stacks.iop.org/JPhysD/53/164003/mmedia)). We
used the 488 nm and 633 nm lines of a white light laser as the excitation
source. STED-FCS experiments were performed using a 775 nm pulsed laser (80 MHz)
for depletion with laser powers between 0 and 300 mW measured at the objective.
STED delay time was optimised using an SLB sample and minimising the detected
photon count rate under high-power STED illumination. The respective notch
filters (775 nm, 633 nm or 488 nm) were used for emission clean up. For all
measurements with constant excitation power we stayed below saturation intensity
(by checking proportionality of excitation laser power and fluorescence
intensity) as triplet pumping may result in an additional source of deviations
[45]. Fluctuations in laser
intensities, which in some other studies reflected in a pronounced correlation
component with decay times on the order of seconds and had to be corrected for
[33], were not observable
in our case—all FCS curves converged to 0 (see examples in figures 1, SI 1 and SI 7).
Figure 1.
Influence of varying concentration and excitation laser power on FCS
measurements of the dye Atto655 in aqueous solution. (A), (B) Normalized
detected photon count rate data (intensity) over time at different dye
concentrations ((A) 0.1 nM, 50 nM, and 1 µm as
labelled; excitation laser power 15.4 µW) and
excitation laser powers ((B) as labelled; concentration 5 nM),
highlighting the reduction in relative intensity fluctuations for higher
concentrations and excitation laser powers. (C), (D) Detected photon
count rates versus instituted dye concentration ((C) dashed lines
indicate the expected linear increase; excitation laser power 15.4
µW) and instituted excitation laser power for 5 nM
(blue) and 500 nM (green) ((D) dashed lines indicate the expected linear
increase). Values are averages of three repetitions (acquisition time
15 s each), and standard deviations are smaller than the size of the
symbols.
All FCS experiments were performed using the hybrid detectors (HyD-SMDs),
featuring very short dead times, and FALCON electronics allowing acquisition of
TCSPC data at photon count rates of up to 80 Mcps per detection channel without
the necessity for corrections, becoming comparable to or exceeding the
repetition rates of commonly used pulsed excitation lasers. This implementation
is based on sampling the signal from the pulsed laser and detectors using fast
FPGA electronics and applying pattern matching to the resulting bitstreams,
producing as output the photon arrival times with a resolution of 97 ps and dead
time <1.5 ns, at GHz sampling rates (for more technical details, please see
the Leica Falcon Application Note [46]). Though certain other detector types such as avalanche
photodiodes (APDs) offer 2–3-fold higher quantum efficiency in the far-red part
of the spectrum compared to HyD-SMDs, their dead times are typically around 30
ns, i.e. 20-fold longer. The here-employed technology thus offers the highest
currently achievable overall count rates, which are now on pair with the
repetition rate of the excitation laser. Nevertheless, future developments
towards increasing the detectors’ quantum efficiencies will allow further
benefits, e.g. the use of lower excitation laser powers to achieve similarly
high count rates or superior signal-to-noise ratios at high STED powers.Measurement times ranged as indicated from 10 to 60 s. Only for the lifetime
measurements we used 40 MHz pulsing of the white light laser for excitation.
Data analysis
Correlation, time trace cropping, gating and fitting was performed using the
built-in routines in LAS-X (Leica Microsystems). Time gates were applied in
STED-FCS to remove confocal or laser scattering contributions and therefore
improve resolution (see figure SI 7), while not affecting confocal measurements
(figure SI 7; marginal deterioration of signal-to-noise or slightly larger
spread of the fitted parameters were barely noticable). Solution and cytoplasmic
GFP data were fitted with a free 3D diffusion model including offset and a
triplet component [47] as
appropriate (triplet correlation decay time of GFP 40 µs with
relative amplitude fixed to 14%, Atto655 no triplet population [48], Abberior STAR Red triplet
correlation decay time 5 µs with triplet amplitudes within
5%–10%; crimson beads triplet correlation decay time 10–100 µs
with triplet amplitudes 3%–5%). SLB, GPMV and cell membrane data were fitted
with a 2D anomalous diffusion model [49] (including offset and triplet time for Abberior STAR
Red-PEG-Chol or -DPPE: 5 µs with relative amplitude around 10%,
and up to 30% at the highest excitation powers). The parameter optimisation was
performed using a Levenberg–Marquardt non-linear least-squares minimisation
method with data-points weighted by their standard deviations, which were
estimated from variations in ACFs calculated from sub-sections of the intensity
time trace [27]. Some
representative autocorrelation curves together with their fits, non-weighted
fitting residuals, and reduced χ2 values [27] are displayed in figures SI 1
and SI 8.As a measure of data quality and curve smoothness, defining the precision of the
extracted fitted parameters, nRMSD values were calculated by taking the
root-mean-square difference between the measured FCS curve and its fit up to the
transit time, and normalised to the fitted amplitude. Such measure yielded
similar values as the experimental standard deviation of the autocorrelation
curves [27] normalised to the
amplitude (figures SI 1 and SI 8; for easier comparison to nRMSD, the main data
quality assessment tool throughout this work, we there display non-weighted
residuals, but plot also standard deviations to put the residuals in the right
perspective relevant to fitting). Smooth data-points at longer lag times were
excluded, to avoid the influence of model miss-fitting in extreme conditions
applied in this study, e.g. when curves were distorted due to saturation or
photobleaching effects at the highest excitation powers, or by random bright
transits occurring in cells. The constructed measure nRMSD thus correlates well
with the relative standard deviation of the fitted transit times (figure SI
2(D)).For concentration estimation, we assumed a confocal volume of 1 fl. Given that
the correlation amplitude relates to the inverse average number of particles in
the focal volume, concentrations can be estimated [10]. For STED-FCS experiments the diffusion
coefficient was calculated as described before [22] using the following formula: where D is the apparent
diffusion coefficient (in confocal or STED), ω refers to the
full width half max (FWHM) of the observation spot (in confocal or STED) and
τ to the transit time
extracted from the FCS fit. The FWHM was determined by confocal and STED imaging
of fluorescent beads (20 nm crimson beads). Using the assumption that the
fluorescently labelled lipids diffuse freely in a SLB, the FWHM can be
calculated as a function of STED power using the following equation [22]:Note that we used the top membrane of immobilised GPMVs labelled with Abberior
STAR Red-PEG-cholesterol to determine the FWHM far away from the surface using a
water immersion objective.
Results and discussion
Non-saturated photon detection at high dye concentrations and laser
excitation powers
We first tested the advanced photon counting instrumentation, implemented on a
confocal and STED-capable microscope, by recording fluorescence fluctuation data
from a single dye (Atto655, chosen for its low population of triplet states)
diffusing in aqueous solution at different concentrations or excitation laser
powers, resulting in different photon count rates. The photon counting
instrumentation included hybrid detectors with very short dead times and fast
FPGA electronics with real-time GHz sampling and pattern matching, which
together with the 80 MHz pulsed fluorescence excitation allows for detection of
photon count rates of tens of MHz without corrections (as described in detail in
the Materials and methods section). Figures 1(A) and (B) show fluctuations in the normalized photon count rates over time
as recorded for three different dye concentrations and laser excitation powers,
respectively. As expected from theory [10, 25–28], the relative fluctuations
around the average count rate decrease with increasing dye concentration, but
much less so with excitation laser power. Most importantly, we could follow a
linear increase of photon count rate with dye concentration and laser excitation
power (figures 1(C) and (D)), as expected in the absence of
limitations in detection electronics. Note that approximate linearity was
maintained despite employing dye concentrations of up to 1 µm
and registering photon count rates of up to 20–30 MHz. The non-linearity
introduced at excitation laser powers larger than 40 µW (≈30 kW
cm−2, figure 1(D))
were to be expected due to dye photobleaching and saturation of excited state
population and consequently fluorescence emission (i.e. not due to detector
saturation, compare figure 1(C) at
same count rate levels), while slight saturation effects at very high dye
concentrations may result from photon re-absorption and dye self-quenching, as
indicated previously [33, 50]. In due consideration of
acquisition count rate being the limiting factor in conventional equipment, we
present the rest of the data as a function of this parameter.Influence of varying concentration and excitation laser power on FCS
measurements of the dye Atto655 in aqueous solution. (A), (B) Normalized
detected photon count rate data (intensity) over time at different dye
concentrations ((A) 0.1 nM, 50 nM, and 1 µm as
labelled; excitation laser power 15.4 µW) and
excitation laser powers ((B) as labelled; concentration 5 nM),
highlighting the reduction in relative intensity fluctuations for higher
concentrations and excitation laser powers. (C), (D) Detected photon
count rates versus instituted dye concentration ((C) dashed lines
indicate the expected linear increase; excitation laser power 15.4
µW) and instituted excitation laser power for 5 nM
(blue) and 500 nM (green) ((D) dashed lines indicate the expected linear
increase). Values are averages of three repetitions (acquisition time
15 s each), and standard deviations are smaller than the size of the
symbols.
FCS noise levels at different count rates
The ability to record photon time traces at high count rates consequently allowed
us to acquire FCS data for Atto655 up to 1 µm high dye
concentrations and excitation laser powers up to 40 µW (≈30 kW
cm−2). The autocorrelation curves for these unconventional
conditions show similar decays as for low dye concentrations and laser powers
(figures 2(A) and (B)). From common FCS theory
(assuming large particle concentrations and excluding non-linear photo-physical
effects such as saturation and photobleaching), FCS-derived average transit
times should be independent of dye concentration and laser power, while the
amplitude of the autocorrelation curve should linearly decrease with dye
concentration and stay constant with laser power. We could well recover this
behaviour from fitting our FCS data, even when recorded at count rates of up to
20–30 MHz (figures 2(C) and (D)). As before, deviations at the
highest tested laser powers >40 µW (primarily apparent as a
drop in the values of the transit time and FCS amplitudes, figure 2(D)), can be attributed to dye
photobleaching and fluorescence emission saturation. For dyes other than
Atto655, excessive pumping into their triplet states would result in additional
deterioration at high excitation powers and would thus need to be carefully
considered as well [47, 51].
Figure 2.
Detailed analysis of the FCS data of Atto655 in aqueous solution at
varying instituted dye concentration cdye
(left panels) and excitation laser power
Plaser (right panels). (A), (B)
Representative FCS curves at different concentrations ((A) as labelled,
Plaser = 15.4 µW) and
excitation laser powers ((B) as labelled, cdye = 5 nM).
(C)–(F) Transit time, correlation amplitude ((C) and (D) axes and
colours as labelled, (D) for two cdye as
labelled), and nRMSD (i.e. noise in correlation data, (E) and (F); (F)
for two cdye as labelled) as determined from
FCS for different photon count rates, i.e. different
cdye (C), (E) and
Plaser (D), (F). (Inset (E)) Values of
nRMSD as determined from FCS for different dyes in aqueous solution with
different brightness, i.e. counts-per-molecule (cpm) (average over
cdye ≈ 0.01–1 µm or
1:100–1:2000 dilutions of the stock bead suspension (see figure SI 3),
Plaser = 1.5 µW).
(G), (H) Values of nRMSD as determined from FCS data of Atto655 at
different acquisition times and at varying cdye ((G)
Plaser = 15.4 µW) and
Plaser ((H) cdye = 5 or 500 nM) as labelled.
Data points represent averages and standard deviations (unless smaller
than the symbols) of three repetitions (60 s acquisition time if not
indicated otherwise).
Detailed analysis of the FCS data of Atto655 in aqueous solution at
varying instituted dye concentration cdye
(left panels) and excitation laser power
Plaser (right panels). (A), (B)
Representative FCS curves at different concentrations ((A) as labelled,
Plaser = 15.4 µW) and
excitation laser powers ((B) as labelled, cdye = 5 nM).
(C)–(F) Transit time, correlation amplitude ((C) and (D) axes and
colours as labelled, (D) for two cdye as
labelled), and nRMSD (i.e. noise in correlation data, (E) and (F); (F)
for two cdye as labelled) as determined from
FCS for different photon count rates, i.e. different
cdye (C), (E) and
Plaser (D), (F). (Inset (E)) Values of
nRMSD as determined from FCS for different dyes in aqueous solution with
different brightness, i.e. counts-per-molecule (cpm) (average over
cdye ≈ 0.01–1 µm or
1:100–1:2000 dilutions of the stock bead suspension (see figure SI 3),
Plaser = 1.5 µW).
(G), (H) Values of nRMSD as determined from FCS data of Atto655 at
different acquisition times and at varying cdye ((G)
Plaser = 15.4 µW) and
Plaser ((H) cdye = 5 or 500 nM) as labelled.
Data points represent averages and standard deviations (unless smaller
than the symbols) of three repetitions (60 s acquisition time if not
indicated otherwise).An interesting feature of the FCS data recorded at different dye concentrations
or excitation laser powers, i.e. photon count rates, are the different noise
levels and resulting data quality. From theory [11, 24, 26], the noise
in FCS data should linearly decrease with increasing excitation laser power and
be independent of dye concentration. The latter has for example been
experimentally verified for dye concentrations of up to around 100 nM [24]. Consequently, we set out to
investigate noise levels for FCS data recorded at the large dynamic range of dye
concentrations and laser powers, which became accessible using the new
equipment.Already visual inspection of our FCS data recorded at the different conditions
(figures 2(A) and (B)) indicated notable differences in
noise levels (shown as the spread of the correlation curve), especially for the
different excitation laser powers, which was most pronounced at short lag times
(in this case up to around 40 µs, roughly corresponding to the
transit time of the dye). The non-trivial estimation of noise levels in FCS data
has been the subject of intense investigations [11, 24–29]. In our
experience, the noise levels in the current FCS data were well estimated by the
root-mean-square of the fitting residuals normalized to the amplitude (nRMSD,
calculated for short lag times up to the transit time), with lower values
indicating better data quality (note that nRMSD was not used as the residuals
minimisation metric in our fitting protocol; see Materials and methods section
and figure SI 1 for details). The nRMSD provided us with a single value for the
data quality of each measurement. Conveniently, this measure also roughly
corresponded to the relative standard deviation of the values of the average
transit time as determined from fitting of the data (figure SI 2, the nRMSD
relates closely to the measurement error, i.e. the experimental standard
deviation of the correlation at every data point, figure SI 1). In accordance
with the theoretical predictions [11, 24, 26], the noise in the FCS data
and thus nRMSD were only weakly affected by varying concentration (figure 2(E)), but could be greatly improved
by increased excitation laser power (figure 2(F)). The excitation laser power directly increases
the dyes’ excited state population and thus fluorescence emission rate and the
average detected count rate per single dye (molecular brightness), which is the
reason for the improvements in noise levels. Under comparable measurement
conditions, the nRMSD is therefore also a direct indicator of the molecular
brightness of the investigated dye (inset in figures 2(E) and SI 3). Deteriorated noise levels, i.e.
higher nRMSD values, were again observed at dye concentrations around 1
µm, but were much less pronounced at the highest excitation
laser powers above 20–40 µW, despite saturation of photon count
rates due to photophysical limitations of the dye (figure 1(D)) and deviations of values of the transit times
and correlation amplitudes (figure 2(D)) as highlighted above.
FCS noise levels at different acquisition times
Predicted from theory and to a certain extent verified experimentally [11, 24, 26], the noise in FCS data should decrease with the square root of
the acquisition time. We could well reproduce this dependence for different dye
concentrations and excitation laser powers (figures 2(G) and (H); again, the same issues as outlined above caused deviations at
high dye concentrations and excitation laser powers). This data establishes
unique possibilities of adapting to experimental conditions. Due to the absence
of saturation effects in photon count rates in the 5–500 nM concentration range,
the possibility of increasing the laser power and detecting correspondingly
higher photon count rates does not only increase the data quality (i.e. lower
nRMSD values), but alternatively allows for a significant reduction (up to two
orders of magnitude) in the acquisition time required for generating similar
data quality (figures 2(G), (H) and SI 2).
FCS of cytosolic GFP in live cells at various expression levels
The possibility of acquiring accurate FCS data in a wide range of dye
concentrations allows simplification or realization of experiments under
challenging conditions. For example, FCS-based measurements of diffusion or
concentration of fluorescent proteins in cells are usually challenged by the
naturally varying expression levels of the fluorescent proteins, as shown in the
representative confocal image in figure 3(A) for HeLa cells expressing cytoplasmic GFP (green fluorescent
protein). Using standard FCS instrumentation, only carefully chosen dim cells
would be measurable, which may represent only a small, and not necessarily
representative fraction of all cells, introducing a potential source of
bias.
Figure 3.
Acquiring FCS data in live cells at different expression levels of the
cytosolic fluorescent protein GFP-SNAP. (A) Representative confocal
image of HeLa cells transiently transfected with cytoplasmic GFP-SNAP,
highlighting different brightness and thus expression levels. Scale bar:
10 µm. (B)–(D) Values of transit time (B), correlation
amplitude ((C) with the grey guideline indicating the predicted inverse
relationship), and nRMSD (i.e. noise in correlation data (D)) as
determined from FCS for different photon count rates originating from
different expression levels. Averages and standard deviations of at
least three measurements per cell (acquisition time 15 s, excitation
laser power 0.6 µW).
Acquiring FCS data in live cells at different expression levels of the
cytosolic fluorescent protein GFP-SNAP. (A) Representative confocal
image of HeLa cells transiently transfected with cytoplasmic GFP-SNAP,
highlighting different brightness and thus expression levels. Scale bar:
10 µm. (B)–(D) Values of transit time (B), correlation
amplitude ((C) with the grey guideline indicating the predicted inverse
relationship), and nRMSD (i.e. noise in correlation data (D)) as
determined from FCS for different photon count rates originating from
different expression levels. Averages and standard deviations of at
least three measurements per cell (acquisition time 15 s, excitation
laser power 0.6 µW).Using our current setup, we could now record FCS data for all HeLa cells
irrespective of their fluorescence intensity, revealing average transit times of
GFP over a wide range of photon count rates and thus concentrations resulting
from different expression levels (figure 3(B)). The photon count rates are correlated with the correlation
amplitude (figure 3(C)), which is
inversely proportional to the average number of fluorescent molecules in the
observation volume (see Materials and methods) and thus concentration and
expression level of GFP can be inferred. Taking our observation volume of about
1 fl, we can estimate concentrations of GFP of approx. 0.01–5
µm between the differently expressing cells (see Materials and
methods section). These data indicate that within the tested range the mobility
of GFP is independent of expression level. In addition, the quality of the FCS
data as quantified by the nRMSD values was maintained over the range of tested
expression levels (figure 3(D)), as
predicted from the behaviour of the organic dye in solution (see figure 2). Only in the regime beyond 10 MHz
(in our case corresponding to concentrations around 1 µm), we
observed a slight signal deterioration due to, for example, possible
out-of-focus and self-absorption contributions, reflected in an increase of the
nRMSD (figure 3(D)) and a decrease
in the amplitude beyond the predicted inverse relationship with the count rate
(grey dashed guide line in figure 3(C)).
STED-FCS of lipid dyes in model membranes
The main strength of FCS on a super-resolution STED microscope, STED-FCS, is the
ability to directly report on nanoscale molecular mobility and thus determine
apparent values of diffusion coefficients from the average transit times (see
Materials and methods section) for different observation spot sizes—from
conventional confocal spot sizes with lateral diameters of around 200 nm, down
to STED microscopy recordings with observation spot diameters of 30–40 nm. From
the dependency of the apparent values of the diffusion coefficient on the
observation spot diameter, STED-FCS has provided insights into the molecular
diffusion modes, similar to spot-variation FCS [20], but now at the relevant molecular scale,
which is particularly valuable for the elucidation of the nanoscale architecture
of biological membranes [22].
However, measurements at various sizes of the effective observation spot
inherently impose a large variation in the average number of fluorescent
molecules in the observation spot (N) and thus detected photon
count levels (figure 4(A),
schematics). Large observation spots at the confocal recordings entail already
high count rates and high values of N at rather low dye
concentrations, while the smaller observation spots at the STED microscopy
recordings require relatively large dye concentrations to reach photon count
rates and values of N that are high enough for allowing
reasonably low acquisition times (it has also been shown theoretically that too
low count rates or concentrations lead to noisy and inaccurate FCS data [11, 24, 26]). This has limited the range of useful dye concentrations in
STED-FCS measurements using conventional detection electronics.
Figure 4.
STED-FCS in supported lipid bilayers (SLBs) at various dye
concentrations. (A) (Right) Sketch illustrating the shrinking of the
observation spot from the confocal (green) to the STED (purple) modus
for a high or low concentration of dye molecules (yellow dots). (Left)
Average fluorescence intensity experimentally recorded for different
concentrations of a fluorescent lipid analogue (Abberior STAR Red-DPPE;
different symbols) in DOPC SLBs for increasing powers of the STED laser.
From these data, the noise level of FCS data (nRMSD, (B), the average
fitted transit time (C), and calculated apparent STED resolution (D) for
confocal (green) and STED-FCS recordings (purple, 150 mW depletion
power) are plotted against the respective confocal acquisition count
rates, which are the limiting factor in each STED-FCS experiment.
Excitation power 2.3 µW, acquisition time 15 s. Each
data-point represents the average and standard deviation of at least
three measurements per SLB preparation.
STED-FCS in supported lipid bilayers (SLBs) at various dye
concentrations. (A) (Right) Sketch illustrating the shrinking of the
observation spot from the confocal (green) to the STED (purple) modus
for a high or low concentration of dye molecules (yellow dots). (Left)
Average fluorescence intensity experimentally recorded for different
concentrations of a fluorescent lipid analogue (Abberior STAR Red-DPPE;
different symbols) in DOPC SLBs for increasing powers of the STED laser.
From these data, the noise level of FCS data (nRMSD, (B), the average
fitted transit time (C), and calculated apparent STED resolution (D) for
confocal (green) and STED-FCS recordings (purple, 150 mW depletion
power) are plotted against the respective confocal acquisition count
rates, which are the limiting factor in each STED-FCS experiment.
Excitation power 2.3 µW, acquisition time 15 s. Each
data-point represents the average and standard deviation of at least
three measurements per SLB preparation.To evaluate the performance of the new detection electronics in STED-FCS, we
recorded data at varying concentrations of the fluorescent lipid analogue
Abberior STAR Red-DPPE (DPPE) diffusing in a fluid supported lipid membrane
bilayer (SLB, composed of lipids DOPC), which is a convenient and well
characterised model membrane system, often used as a control sample in STED-FCS
experiments [22]. While
increasing the lipid analogue concentration and pushing the confocal count rates
beyond the conventional FCS range (figure 4(A)) did not influence the noise in confocal measurements (figure
4(B), green data points), it—as
expected—significantly improved the data quality of the STED-FCS recordings (low
nRMSD values, figure 4(B), purple
data points; note that for plotting of these, the respective confocal count
rates, which limit the experimental conditions for STED recordings, were used as
the x-value). Also as expected from theory, the values of
transit times were smaller in the STED compared to the confocal recordings (due
to the reduced observation spot size in the STED mode, here roughly 40 nm in
diameter) and hardly changed with instituted concentrations of the fluorescent
probe (i.e. at increased count rate, figure 4(C)). Similarly, the observation spot diameters as
determined from the recorded FCS data remained constant (figure 4(D), see Materials and methods for
details about its calculation), all in all highlighting the great flexibility
and improvement in STED-FCS experiments when employing non-saturating detection
electronics.For simplicity, we demonstrated the above effects at a single STED laser power
(150 mW, observation spot diameter of 40 nm), but the conclusions held also true
for other STED laser powers (and thus observation spot sizes). Similar or even
larger improvements of STED-FCS data quality as by increasing the dye
concentration were achieved by increasing the excitation laser power (figure SI
4; note that we, in contrast to the previous data of Atto655, now included an
additional decay due to triplet state population in FCS data model, see
materials and methods and figure SI 4E). Effects due to saturation of the
excited states (such as the triplet state) of the fluorescent label started to
deteriorate the signal quality and bias the extracted parameter values only at
laser powers beyond 25 µW, supporting the benefits of high
count rates for STED-FCS experiments with the excellent dyes available
nowadays.
STED-FCS in live cell membranes
Finally, we verified the reliability of STED-FCS measurements at high photon
count rates for measurements in living cells. We labelled the plasma membrane of
live HeLa cells using the fluorescent lipid analogue cholesterol-PEG-Abberior
STAR Red (Chol-PEG-KK114, figure 5(A)), for which previous studies have consistently indicated free
diffusion in cellular plasma membranes [52]. The resulting STED-FCS data show low nRMSD
values, i.e. low noise levels in the correlation data, which can in the STED
microscopy mode be significantly improved to almost confocal quality by
increasing either the excitation laser power or concentration of Chol-PEG-KK114
(i.e. total count rate, figure 5(B)) without biasing the resulting values of transit times (figure
5(C)).
Figure 5.
STED-FCS of Abberior STAR-Red-PEG-cholesterol in membranes of live HeLa
cells. (A) Confocal image of the apical plane (roughly 10
µm above the cover slip) and mid plane of HeLa
cells, membrane labelled with the fluorescent cholesterol analogue
(image sizes 60 × 27 µm2). (B) The noise
level of FCS data (nRMSD values) and (C) average fitted transit time for
confocal (green) and STED-FCS recordings (125 mW depletion laser power,
purple), in the top membrane of HeLa cells, plotted against their
respective count rates. (D) Diffusion coefficient determined from FCS
data in confocal and STED mode in the top membrane of the HeLa cells.
Every dot represents a single FCS measurement. Excitation laser 2.3
µW, acquisition time 15 s.
STED-FCS of Abberior STAR-Red-PEG-cholesterol in membranes of live HeLa
cells. (A) Confocal image of the apical plane (roughly 10
µm above the cover slip) and mid plane of HeLa
cells, membrane labelled with the fluorescent cholesterol analogue
(image sizes 60 × 27 µm2). (B) The noise
level of FCS data (nRMSD values) and (C) average fitted transit time for
confocal (green) and STED-FCS recordings (125 mW depletion laser power,
purple), in the top membrane of HeLa cells, plotted against their
respective count rates. (D) Diffusion coefficient determined from FCS
data in confocal and STED mode in the top membrane of the HeLa cells.
Every dot represents a single FCS measurement. Excitation laser 2.3
µW, acquisition time 15 s.Note that we here measured the diffusion in the apical membrane of HeLa cells,
i.e. 5–10 µm above the microscope coverslip (figure 5(A)), rather than in the basal
membrane as before [52]. This
avoids potential biasing effects by the coverslip surface. Yet, penetration
through the aqueous cellular environment over such a distance causes spherical
aberrations when employing a traditional oil-immersion STED microscope objective
(due to the refractive index mismatch between water and oil) [53], having detrimental effects
on STED-FCS experiments (figure SI 5). Such aberrations can either be corrected
for using adaptive optics [54,
55] or employing a water
immersion objective [56], which
shows constant signal levels and performance of for example STED and STED-FCS
experiments in a wide range of focal depths (5–100 µm above the
coverslip) without the need for depth-dependent readjustments of the correction
collar (figure SI 5). Taking the observation spot size as determined from
calibration data (figure SI 6, diameter of 100 nm), we could calculate apparent
values of the diffusion coefficient D for both the confocal and
STED mode (spot diameters of 280 and 100 nm, respectively; see Materials and
methods), which were both in the same range (D ≈ 0.45
µm2 s−1, figure 5(D)) as observed before for the basal membrane
[52], highlighting free
diffusion and indicating similar diffusion characteristics of the probe in the
apical and basal plasma membrane [52], i.e. negligible bias by the coverslip surface.
Conclusions
We systematically evaluated the reduction in error and bias of FCS measurements
recorded at high photon count rates, as enabled by novel detection electronics
integrated into a turn-key microscope. We were able to record highly accurate FCS
data with detected photon count rates of up to about 10 MHz, i.e. dye concentrations
up to 1 µm. This improved performance introduces huge flexibility
for performing FCS experiments to measure diffusion or concentration, previously
impossible due to limitations in the detection electronics (e.g. allowing only
recordings of photon count rates of up to 1 MHz). This now enables: (1) FCS
measurements at high dye concentrations for e.g. low-affinity binding assays, (2)
the recording of fluctuation data with reduced acquisition times by increasing the
excitation power to higher count rates (dye photophysics permitting), (3) performing
live-cell experiments in a wide range of expression levels of fluorescently tagged
proteins, and (4) optimization of the data quality of STED-FCS recordings over a
wide range of observation spot sizes by increasing dye concentration and/or
excitation laser power. Using these features we could for example show that
cytosolic diffusion of GFP was independent of expression level in live HeLa cells,
and that the fluorescent lipid analogue was diffusing freely in the apical membrane
similarly as reported before for the basal membrane [52].Improved detection instrumentation as the one presented here are becoming
increasingly available and will be further optimized, pushing the ease of use of FCS
or related measurements, such as fluorescence cross correlation spectroscopy (FCCS)
[57], fluorescence lifetime
correlation spectroscopy (FLCS) [58], number and brightness (N&B) analysis [59], or line- and raster-scanning correlation
spectroscopy (RICS) [12, 60]. In combination with
high-throughput methods this could enable the systematic evaluation of
overexpression of fluorescent proteins [61], tracking of dynamically changing diffusion properties, or other
previously unattainable applications.
Authors: Jörn Heine; Christian A Wurm; Jan Keller-Findeisen; Andreas Schönle; Benjamin Harke; Matthias Reuss; Franziska R Winter; Gerald Donnert Journal: Rev Sci Instrum Date: 2018-05 Impact factor: 1.523
Authors: Luca Lanzanò; Lorenzo Scipioni; Melody Di Bona; Paolo Bianchini; Ranieri Bizzarri; Francesco Cardarelli; Alberto Diaspro; Giuseppe Vicidomini Journal: Nat Commun Date: 2017-07-06 Impact factor: 14.919
Authors: Melody Di Bona; Michael A Mancini; Davide Mazza; Giuseppe Vicidomini; Alberto Diaspro; Luca Lanzanò Journal: Biophys J Date: 2019-02-14 Impact factor: 4.033
Authors: Matthew J Laurence; Timothy S Carpenter; Ted A Laurence; Matthew A Coleman; Megan Shelby; Chao Liu Journal: Membranes (Basel) Date: 2022-03-31