Abstract
BACKGROUND Fat suppression is an important technique in magnetic resonance imaging (MRI). Comprehensive and quantitative assessment of the influence of fat suppression (FS) on T1-weighted imaging of intracranial vessel wall imaging is needed. In this study, we compared the three-dimensional (3D) variable-flip-angle turbo-spin-echo (CUBE) T1-weighted sequence with and without FS to investigate the differences between the 2 sequences in imaging of the middle cerebral artery (MCA) vessel walls. MATERIAL AND METHODS A 3D CUBE T1-weighted sequence with and without FS by 3.0T MRI was used to obtain intracranial vessel wall images of 105 MCA stenosis patients. The image signal intensity, signal-to-noise ratio, and contrast-to-noise ratio were calculated and compared. Two observers evaluated the image quality of the 2 sequences twice, and interobserver and intraobserver consistency were determined. Differences between the 2 sequences in the area of lumen and plaque were compared. RESULTS The signal intensity, signal-to-noise ratio, and contrast-to-noise ratio of the 3D CUBE T1-weighted sequence without FS were higher, whereas the noise level was lower. In terms of subjective scores, the 3D CUBE T1-weighted sequence without FS performed better. No significant difference was observed in the measurement of the vascular lumen area between the 2 sequences, although there were statistically significant differences in the measurement of plaque area (i.e., the measurement obtained with 3D CUBE T1-weighted sequence without FS was larger). CONCLUSIONS 3D CUBE T1-weighted sequence without FS performed better for MCA vessel walls imaging than 3D CUBE T1-weighted sequence with FS.Entities:
Mesh:
Year: 2020 PMID: 33159730 PMCID: PMC7657061 DOI: 10.12659/MSM.928931
Source DB: PubMed Journal: Med Sci Monit ISSN: 1234-1010
Vessel wall imaging protocol with a GE Discovery 3.0T MRI scanner.
| Parameter | 3D TOF MRA | 3D CUBE T1 with FS | 3D CUBE T1 without FS |
|---|---|---|---|
| TR (ms) | min | 1140.0 | 1140.0 |
| TE (ms) | min | min | min |
| Slice thickness (mm) | 1.4 | 1.0 | 1.0 |
| FOV (cm) | 22.0 | 18.0 | 18.0 |
| NEX | 3 | 1 | 1 |
| Bandwidth | 31.25 | 62.5 | 62.5 |
| Locs per slab | 32 | 60 | 60 |
| Matrix size | 320×256 | 320×288 | 320×288 |
| Scan duration (minutes) | 3: 57 | 3: 50 | 3: 50 |
Locs per slab, total number of locations (slices) generated from a slab; 3D – three dimensional; TR – repetition time; TE – echo time; SNR – signal-to-noise-ratio; FOV – field of view; NEX – number of excitations; TOF – time of flight; MRA – magnetic resonance angiography; FS – fat suppression.
Figure 1Typical images of a patient with MCA stenosis. (A, B) 3D CUBE T1 sequence without fat suppression. (C, D) 3D CUBE T1 sequence with fat suppression. Two ROIs for signal intensity measurements are separately depicted as circles in the lumen (circle 1) and brain white matter (circle 2) on images A and C. An ROI for signal intensity measurement is depicted as a circle (circle 1) in the plaque on images B and D.
Comparison of signal characteristics between the two sequences.
| Group | Signal characteristics | 3D CUBE T1 without FS | 3D CUBE T1 with FS | |
|---|---|---|---|---|
| MCA stenosis patients (N=105) | σN | 34.72±2.82 | 39.66±3.93 | 0.000 |
| SB | 1049.99±16.80 | 803.25±96.95 | 0.000 | |
| SV | 523.09±139.50 | 428.02±113.79 | 0.000 | |
| SL | 251.98±98.52 | 192.57±69.26 | 0.000 | |
| SPQ | 635.52±120.70 | 522.84±103.02 | 0.000 | |
| SNRB | 30.37±3.96 | 20.29±2.56 | 0.000 | |
| SNRV | 14.29±3.02 | 10.89±1.89 | 0.015 | |
| SNRL | 6.88±2.11 | 4.94±2.32 | 0.005 | |
| SNRPQ | 17.91±2.85 | 13.21±1.70 | 0.002 | |
| CNRV | 16.38±5.09 | 9.39±2.39 | 0.002 | |
| CNRL | 23.88±5.21 | 15.23±2.37 | 0.000 | |
| CNRPQ | 12.47±5.37 | 7.04±2.77 | 0.016 |
MCA – middle cerebral artery; FS – fat suppression; S – signal intensity; SNR – signal-to-noise-ratio; CNR – contrast-to-noise-ratio; N – noise; B – brain white matter; V – vascular cross-section; L – lumen; PQ – plaque.
Comparison of image quality between two sequences datasets as scored by two readers.
| Characteristics | Reader 1, 1st | Reader 2, 1st | Reader 1, 2nd | Reader 2, 2nd | ||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|
| MCA stenosis patients (n=105) | 3D CUBE T1 without FS | 3D CUBE T1 with FS | 3D CUBE T1 without FS | 3D CUBE T1 with FS | 3D CUBE T1 without FS | 3D CUBE T1 with FS | 3D CUBE T1 without FS | 3D CUBE T1 with FS | ||||
| Noise | 4.73± 0.54 | 4.04± 0.61 | 0.000 | 4.68± 0.37 | 4.05± 0.58 | 0.001 | 4.74± 0.38 | 4.06± 0.57 | 0.000 | 4.72± 0.37 | 4.04± 0.59 | 0.001 |
| Vessel wall visualization | 4.70± 0.57 | 4.37± 0.71 | 0.000 | 4.68± 0.36 | 4.26± 0.66 | 0.000 | 4.71± 0.37 | 4.39± 0.67 | 0.000 | 4.67± 0.39 | 4.26± 0.68 | 0.000 |
| Lumen visualization | 4.84± 0.41 | 4.81± 0.45 | 0.280 | 4.88± 0.19 | 4.83± 0.46 | 0.109 | 4.86± 0.33 | 4.80± 0.42 | 0.170 | 4.83± 0.17 | 4.82± 0.46 | 0.401 |
| Vessel wall and surrounding tissue resolution | 4.68± 0.56 | 4.15± 0.63 | 0.000 | 4.67± 0.41 | 4.20± 0.71 | 0.000 | 4.70± 0.56 | 4.16± 0.42 | 0.000 | 4.68± 0.40 | 4.23± 0.75 | 0.000 |
| Plaque visualization | 4.65± 0.60 | 4.33± 0.71 | 0.000 | 4.58± 0.56 | 4.40± 0.71 | 0.025 | 4.66± 0.54 | 4.38± 0.75 | 0.000 | 4.59± 0.52 | 4.40± 0.71 | 0.025 |
| Plaque and Vessel wall resolution | 4.58± 0.64 | 4.25± 0.70 | 0.000 | 4.56± 0.54 | 4.35± 0.75 | 0.033 | 4.59± 0.77 | 4.27± 0.63 | 0.000 | 4.57± 0.50 | 4.37± 0.57 | 0.033 |
MCA – middle cerebral artery; FS – fat suppression.
ICC values of intraobserver, and interobserver reproducibility for two sequences.
| Characteristic | 3D CUBE T1 without FS-ICC (95% CI) | 3D CUBE T1 with FS-ICC (95% CI) | 3D CUBE T1 without FS-ICC (95% CI) | 3D CUBE T1 with FS-ICC (95%CI) | ||||
|---|---|---|---|---|---|---|---|---|
| MCA stenosis patients (n=105) | Reader 1 1st
| Reader 2 1st
| Reader 1 1st
| Reader 2 1st
| 1st Reader 1 | 2nd Reader 1 | 1st Reader 1 | 2nd Reader 1 |
| Noise | 0.97 (0.98–0.99) | 0.97 (0.95–0.98) | 0.98 (0.98–0.99) | 0.94 (0.91–0.95) | 0.76 (0.57–0.80) | 0.78 (0.61–0.81) | 0.79 (0.72–0.85) | 0.78 (0.71–0.83) |
| Vessel wall visualization | 0.96 (0.94–0.98) | 0.98 (0.97–0.98) | 0.96 (0.94–0.98) | 0.99 (0.98–0.99) | 0.74 (0.56–0.80) | 0.75 (0.57–0.82 | 0.78 (0.59–0.82) | 0.76 (0.51–0.79) |
| Lumen visualization | 0.98 (0.97–0.98) | 0.94 (0.92–0.96) | 0.99 (0.99–0.99 | 0.98 (0.97–0.98) | 0.77 (0.65–0.82) | 0.78 (0.68–0.84) | 0.79 (0.66–0.81) | 0.78 (0.63–0.80) |
| Vessel wall and surrounding tissue resolution | 0.99 (0.98–0.99) | 0.99 (0.99–0.99) | 0.99 (0.99–0.99 | 0.99 (0.99–0.99) | 0.75 (0.59–0.81) | 0.75 (0.62–0.83) | 0.75 (0.60–0.73) | 0.76 (0.69–0.77) |
| Plaque visualization | 0.98 (0.97–0.98) | 0.98 (0.97–0.98) | 0.98 (0.97–0.98) | 0.98 (0.97–0.98) | 0.82 (0.73–0.87) | 0.81 (0.71–0.85) | 0.76 (0.65–0.84) | 0.78 (0.68–0.87) |
| plaque and Vessel wall resolution | 0.98 (0.98–0.99) | 0.97 (0.97–0.98) | 0.98 (0.97–0.98) | 0.94 (0.90–0.95) | 0.81 (0.71–0.87) | 0.74 (0.57–0.77) | 0.76 (0.61–0.82) | 0.73 (0.57–0.77) |
MCA – middle cerebral artery; ICC – intraclass correlation coefficient; CI – confidence interval; FS – fat suppression.