| Literature DB >> 33154622 |
Seth M Pantanelli1, Neal Kansara1, Gerard Smits2.
Abstract
PURPOSE: To compare predictability of postoperative refractive astigmatism (RA) using the Emmetropic Verifying Optical (EVO) Toric Formula v2.0 to one that accounts only for anterior corneal astigmatism.Entities:
Keywords: EVO; astigmatism; calculators; cataract; intraocular lens power; toric
Year: 2020 PMID: 33154622 PMCID: PMC7605955 DOI: 10.2147/OPTH.S276285
Source DB: PubMed Journal: Clin Ophthalmol ISSN: 1177-5467
Demographics
| Parameter | Mean ± SD (Range) |
|---|---|
| Age | 71 ± 8.9 (40–89) |
| Gender | 63 Females (57.8%), 46 Males (42.2%) |
| AL (mm) | 23.66 ± 0.85 (20.95–26.14) |
| Keratometric Astigmatism (D) | 1.74 ± 0.38 (1.24–2.86) |
| IOL Power (D) | 20.62 ± 2.50 (16–27) |
| Postoperative Manifest Refraction |
Abbreviations: AL, axial length; D, diopters; IOL, intraocular lens; SD, standard deviation.
Residual Astigmatism Prediction Errors (Arithmetic and Centroid) by Method of Calculation (n=109)
| enVista | EVO | p-value | |
|---|---|---|---|
| Mean ± SD (D) | 0.63 ± 0.43 | 0.57 ± 0.38 | 0.10 |
| Range (D) | 0.03–2.23 | 0.03–1.84 | |
| Median (D) | 0.56 | 0.51 | |
| Centroid ± SD (D) | 0.37 ± 0.66 @ 179 | 0.17 ± 0.67 @ 090 | x-component: p < 0.0001 |
Abbreviations: D, diopters; EVO, Emmetropia Verifying Optical toric IOL formula v2.0; SD, standard deviation.
Figure 1Residual astigmatism prediction error centroid magnitude by method of calculation.
Figure 2Proportion of eyes with arithmetic residual astigmatism prediction error within 0.5, 0.75, and 1.00 diopters (D).
Residual Astigmatism Prediction Errors by Method of Calculation in Eyes with Against- and with-the-Rule Astigmatism
| enVista | EVO | p-value | |
|---|---|---|---|
| Against-the-rule (n = 67) | |||
| Mean ± SD (D) | 0.58 ± 0.43 | 0.59 ± 0.39 | 0.7673 |
| Centroid ± SD (D) | 0.27 ± 0.67 @ 175 | 0.18 ± 0.69 @ 095 | x-component: p < 0.0001 |
| With-the-rule (n = 38) | |||
| Mean ± SD (D) | 0.73 ± 0.42 | 0.54 ± 0.37 | 0.0055 |
| Centroid ± SD (D) | 0.56 ± 0.64 @ 002 | 0.17 ± 0.65 @ 082 | x-component: p = 0.0055 |
| p-values | |||
| Means | 0.08 | 0.55 | |
| Centroids | 0.0061 | 0.596 |
Abbreviations: D, diopters; EVO, Emmetropia Verifying Optical toric IOL formula v2.0; SD, standard deviation.
Figure 3Double angle plots of the residual astigmatism prediction error by method of calculation and sub-group. (A) enVista against-the-rule (ATR), (B) enVista with-the-rule (WTR), (C) EVO ATR and (D) EVO WTR.
Residual Astigmatism Prediction Errors by Method of Calculation and Intraocular Lens Toricity
| Intraocular Lens | enVista | EVO | p-value |
|---|---|---|---|
| MX60T125 (n = 79) | |||
| Mean ± SD (D) | 0.63 ± 0.44 | 0.57 ± 0.39 | 0.1331 |
| Centroid ± SD (D) | 0.36 ± 0.50 @ 177 | 0.17 ± 0.47 @ 094 | x-component: p < 0.0001 |
| MX60T200 (n = 18) | |||
| Mean ± SD (D) | 0.62 ± 0.46 | 0.57 ± 0.38 | 0.5689 |
| Centroid ± SD (D) | 0.40 ± 0.26 @ 002 | 0.16 ± 0.23 @ 086 | x-component: p < 0.0001 |
| MX60T275 (n = 12) | |||
| Mean ± SD (D) | 0.63 ± 0.38 | 0.61 ± 0.36 | 0.8565 |
| Centroid ± SD (D) | 0.38 ± 0.20 @ 010 | 0.25 ± 0.19 @ 081 | x-component: p < 0.0001 |
| p-values | |||
| Means | 0.9982 | 0.9442 | |
| Centroids | 0.9942 | 0.9942 |
Abbreviations: D, diopters; EVO, Emmetropia Verifying Optical toric IOL formula v2.0.