Nasrullah Undre1, Imran Hussain1, John Meijer2, Johannes Stanta3, Gordon Swan3, Ian Dawson3.
Abstract
BACKGROUND: The calcineurin inhibitor tacrolimus is a narrow therapeutic index medication, which requires therapeutic drug monitoring to optimize dose on the basis of systemic exposure. MITRA microsampling offers a minimally invasive approach for the collection of capillary blood samples from a fingerprick as an alternative to conventional venous blood sampling for quantitation of tacrolimus concentrations.Entities:
Year: 2021 PMID: 33149056 PMCID: PMC8115739 DOI: 10.1097/FTD.0000000000000833
Source DB: PubMed Journal: Ther Drug Monit ISSN: 0163-4356 Impact factor: 3.681
FIGURE 1.The MITRA microsampling device: (A) Example of the clamshell collection kit; (B) collection of capillary blood samples, that is, from a fingerprick. Images provided courtesy of Neoteryx, LCC (Torrance, CA).
Chromatography Gradient Elution Profile
| Time (min) | Mobile Phase A [Ammonium Formate 10 mM (aq):Formic Acid (100:0.2)] | Mobile Phase B [Acetonitrile] |
| Initial | 50 | 50 |
| 0.50 | 50 | 50 |
| 0.51 | 15 | 85 |
| 1.50 | 15 | 85 |
| 1.51 | 5 | 95 |
| 2.00 | 5 | 95 |
| 2.01 | 50 | 50 |
| 2.50 | 50 | 50 |
aq, aqueous.
FIGURE 2.Representative ion chromatograms of (A) a blank sample (B) a blank sample spiked with the internal standard (C) a sample spiked with tacrolimus and internal standard at the LLOQ (1.00 ng/mL), and (D) a matrix blank spiked with tacrolimus at the ULOQ (50.0 ng/mL) and no internal standard. The gray shading shows the integrated signal peak. cps, counts per second. LLOQ, lower limit of quantification; ULOQ, upper limit of quantification.
Assay Accuracy and Precision
| Parameter | LLOQ QC (1.0 ng/mL) | LQC (3.0 ng/mL) | MQC (10.0 ng/mL) | HQC (40.0 ng/mL) |
| n | 24 | 24 | 24 | 24 |
| Mean tacrolimus concentration found (ng/mL) | 1.05 | 3.24 | 10.2 | 40.1 |
| Inter-assay %RSD | 6.8 | 11.8 | 7.0 | 6.1 |
| Inter-assay mean %bias | 5.0 | 8.0 | 2.0 | 0.3 |
%bias, mean percentage bias; %RSD, percentage of relative standard deviation; HQC, high quality control; LLOQ, lower limit of quantification; LQC, low quality control; MQC, medium quality control; QC, quality control.
Effect of Hematocrit on Assay Accuracy and Precision
| HT 20% LQC | HT 20% HQC | HT 30% LQC | HT 30% HQC | HT 40% LQC | HT 40% HQC | HT 50% LQC | HT 50% HQC | |
| n | 6 | 6 | 6 | 6 | 6 | 6 | 6 | 6 |
| Mean tacrolimus concentration (ng/mL) | 2.94 | 40.6 | 3.16 | 42.3 | 3.06 | 40.9 | 2.93 | 45.5 |
| %RSD | 2.8 | 4.4 | 3.7 | 2.6 | 4.1 | 3.5 | 9.0 | 4.7 |
| %bias | −2.0 | 1.5 | 5.3 | 5.8 | 2.0 | 2.3 | −2.3 | 13.8 |
%bias, mean percentage bias; %RSD, percentage of relative standard deviation; HQC, high quality control (40.0 ng/mL); HT, hematocrit; LQC, low quality control (3.0 ng/mL).