Detection of SARS-COV N2 Gene: Very low amounts of viral RNA or false positive?

Francesca Falasca1, Ilaria Sciandra1, Daniele Di Carlo2, Massimo Gentile3, Alberto Deales1, Guido Antonelli3, Ombretta Turriziani4.   

Abstract

BACKGROUND: The detection of a low amount of viral RNA is crucial to identify a SARS-CoV-2 positive individual harboring a low level of virus, especially during the convalescent period. However, the detection of one gene at high Cycle threshold (Ct) has to be interpreted with caution. In this study we address this specific issue and report our real-life experience. STUDY
DESIGN: A total of 1639 nasopharyngeal swabs (NPS) were analyzed with Xpert® Xpress SARS-CoV-2. Positive samples showing high Ct values (Ct>35) were concentrated by centrifugation and re-tested with Cepheid or other methods (RealStar SARS-CoV2 RT-PCR, Altona Diagnostics; GeneFinder COVID-19 Plus RealAmp Kit, Elitech).
RESULTS: 1599 (97.5%) negative samples, 36 (2.3%) positive samples and 4 (0.2%) presumptive positive samples were detected. In 17 out of 36 positive patients, very low viral RNA copies were suspected since positivity was detected at high Ct. We confirmed positivity for patients who showed both E and N genes detected and for patients with only N detected but with Ct <39. On the contrary, samples with only gene N detected with Ct values >39 were found negative. NPS taken 24 hours after the first collection confirmed the negativity of the 12 samples. Clinical data sustained these results since only 2 of these 12 patients showed COVID-19-like symptoms.
CONCLUSIONS: These data support our consideration that detection of the N2 gene at high Ct needs to be interpreted with caution, suggesting that collaboration between virologists and clinicians is important for better understanding of results.
Copyright © 2020 Elsevier B.V. All rights reserved.

Entities:  

Keywords:  Ct value; Diagnosis; N2 gene; SARS-CoV-2; false positive

Mesh:

Substances:

Year:  2020        PMID: 33126109      PMCID: PMC7553900          DOI: 10.1016/j.jcv.2020.104660

Source DB:  PubMed          Journal:  J Clin Virol        ISSN: 1386-6532            Impact factor:   3.168


Introduction

The timely diagnosis of COVID-19 cases and subsequent infection control are essential to prevent transmission in healthcare facilities and the community. Rapid SARS-CoV-2 testing, and particularly molecular assays, can have a considerable impact on the ability to make immediate decisions regarding management of the infected patient, including his isolation or the assessment of risk of transmission to healthcare workers performing invasive procedures on critically ill patients. So far, nucleic acid amplification testing is still the gold standard for the diagnosis of SARS-CoV-2 in respiratory samples [1,2]. Despite the good performance of validated nucleic acid amplification assays, a risk of false-negative results still exists. The negativity of the assay may be due to inappropriate sample collection as well as to extraction/Real Time-PCR workflow and to sensitivity of the assays used. Concerning the latter, it is worth noting that the ability of molecular assays to detect SARS-CoV-2 infection can be limited by low amounts of viral RNA (e.g., early or late in COVID-19 disease). Xpert® Xpress SARS-CoV-2 (Cepheid, Sunnyvale, CA) is a rapid molecular diagnostic test utilizing real time RT-PCR technology to detect the nucleocapsid gene (N2 region of the N-gene) and envelope gene (E) in respiratory specimens. The limit of detection (LoD) was reported at 100 copies/ml. Studies have assessed the N2 target’s high specificity and sensitivity for SARS-CoV-2 detection, and cases testing positive with only the N2 region have often been observed in samples containing very low viral RNA copies [3]. The detection of a low amount of viral RNA is crucial to identify a positive individual harboring a low level of virus, especially during the convalescent period. However, the detection of one gene at high Cycle threshold (Ct) introduces problems of interpretation, and the results should be handled with caution. In this study we address this specific issue and report our real-life experience on the above assay

Methods

A total of 1639 nasopharyngeal swabs (NPS) were analyzed with Xpert® Xpress SARS-CoV-2 (Cepheid). The Xpert Xpress SARS-CoV-2 test provides positive results when a signal for the N2 region or signals for both nucleic acid targets (N2 and E) have a Ct within the valid range (<45 Ct) and endpoint above the minimum setting. A presumptive positive result is given when the SARS-CoV-2 signal for only the E nucleic acid target has been detected. Positive samples showing high Ct values (Ct>35) were concentrated by centrifugation (2 hours at 14000 rpm at 4 °C) and re-tested with Cepheid or other methods (RealStar SARS-CoV2 RT-PCR, Altona Diagnostics; GeneFinder COVID-19 Plus RealAmp Kit, Elitech).

Results

A total of 1639 NPS were analyzed with Xpert® Xpress SARS-CoV-2 from 01 April 2020 to 31 July 2020. One thousand five-hundred ninety-nine (97.5%) negative samples, 36 (2.3%) positive samples and 4 (0.2%) presumptive positive samples were detected. In 17 out of 36 positive patients, very low viral RNA copies were suspected since positivity was detected at high Ct. Specifically, in 14 patients only the N2 gene was detected with Ct between 38.0 and 43.4 (Table 1 ), while in 3 samples the E and N genome regions were both amplified with high Ct (Table 1: Pt 2, gene E 45 Ct and gene N 38.9; Pt 9 gene E 37.6 and gene N 42 Ct; Pt 16 gene E 41.7 and gene N 40.1 Ct). To confirm positivity, samples were re-tested with Cepheid and other methods (Altona or Elitech), after 10-fold virus concentration.
Table 1

Interpretation of Ct values and results after nasopharyngeal swab concentration.

SamplesCt Gene ECt Gene NInterpretationResultafter virus concentration
Pt 142.1positivenegative
Pt 245.038.9positivepositive
Pt 342.1positivenegative
Pt 442.3positivenegative
Pt 542.2positivenegative
Pt 641.9positivenegative
Pt 738.3positivepositive
Pt 838.0positivepositive
Pt 937.642.0positivepositive
Pt 1042.2positivenegative
Pt 1141.6positivenegative
Pt 1242.9positivenegative
Pt 1341.7positivenegative
Pt 1442.7positivenegative
Pt 1543.4positivenegative
Pt 1641.740.1positivepositive
Pt 1742positivenegative
Interpretation of Ct values and results after nasopharyngeal swab concentration. Surprisingly, we confirmed positivity for Pt 2, Pt 9 and Pt 16, which showed both E and N genes detected and for patients with only N detected but with Ct <39 (Pt 7 and Pt 8). On the contrary, samples with only gene N detected with Ct values >39 were found negative. NPS taken 24 hours after the first collection confirmed the negativity of the 12 samples. Notably, only 2 of these 12 patients were admitted to hospital with COVID-19-like symptoms; the rest were screened preoperatively or prior to admission to hospital wings, as recommended in order to identify possible asymptomatic infections.

Discussion

The qualitative real time PCR (qRT-PCR) assay is being widely used for detection of SARS-CoV-2 infection. However, the problem of qRT-PCR with inaccurate results is increasingly reported. Studies have shown that a significant number of SARS-CoV-2 false negative samples is inevitably due to low viral load in patients’ throats and to the limited sensitivity of PCR technology. PCR is easily affected by sample inhibitors, poor amplification efficiency, and less precision in low-concentration samples [4]. False negative results may compromise the timely diagnosis, early treatment, prevention of transmission, and assessment of discharge criteria [5]. For all these reasons, several publications have focused their attention on false negative results [6,7]. However, false positive results have been observed, even if less frequently, due to contaminations of commercial primers/probe sets or poor test specificity [8,9]. Katz et al. reported false-positive reverse transcriptase polymerase chain reaction screening for SARS-CoV-2 in the setting of urgent head and neck surgery and otolaryngologic emergencies [10]. The implications for the patients were significant, including delay of urgent surgery and transfer to COVID-19-designated units. A false positive result may have several effects: delay of correct diagnosis in patients with breathing difficulties or other symptoms; administration of inappropriate treatment; wasteful consumption of personal protective equipment; reduction in healthcare-workers leading to uncareful management of patients; unnecessary stress in isolated individuals including family members. Finally, epidemiological analysis could be distorted due to false results. In this analysis we found 17 SARS-CoV-2 positive samples with high Ct value. Surprisingly, when we re-tested the samples after viral concentration with Cepheid or other methods, only 5 samples were confirmed positive: 3 samples with both the E and N genes detected and 2 samples with only the N gene detected with Ct<39. In this context, a critical analysis and a careful evaluation of the test results is fundamental to minimize the false positive rate in order to achieve better management of patients. Based on our experience, we suggest suspecting false positive results when only the N gene is detected with Ct values >39. For all 12 patients, NPS collected 24 hours after the first negative NPS were still negative, supporting the idea that the low positivity detected in the first samples was not due to a recent infection. Alternatively, the results may be due to a late stage of infection. In this situation, serology testing could aid in discriminating true positivity, especially in a final phase of infection. IgM and IgG antibodies to SARS-CoV-2 were tested in only two individuals and no antibodies were found in either. However, since these individuals were asymptomatic and their anamnesis did not report any previous contact with SARS-CoV-2 positive patients, the detection of only N2 gene at very high Ct (>39) suggests a false positive result. These data support our consideration that detection of the N2 gene at high Ct needs to be interpreted with caution and highlight the importance of virologist-clinician collaboration for better understanding of results.

Funding

This work was supported by a grant to GA from the Italian Ministry of Health: COVID-19 A.F. 2020-2021 N. “COVID-2020-12371817.

Declaration of Competing Interest

The authors report no declarations of interest.
  9 in total

1.  SARS-CoV-2 viral load assessment in lung transplantation.

Authors:  R Novysedlak; J Vachtenheim; I Stříž; O Viklický; R Lischke; Z Strizova
Journal:  Physiol Res       Date:  2021-12-16       Impact factor: 1.881

2.  Unusual N Gene Dropout and Ct Value Shift in Commercial Multiplex PCR Assays Caused by Mutated SARS-CoV-2 Strain.

Authors:  Petros Bozidis; Eleni T Tsaousi; Charilaos Kostoulas; Prodromos Sakaloglou; Athanasia Gouni; Despoina Koumpouli; Hercules Sakkas; Ioannis Georgiou; Konstantina Gartzonika
Journal:  Diagnostics (Basel)       Date:  2022-04-13

3.  Application of digital PCR to determine the reliability of Xpert Xpress SARS-CoV-2 assay with envelope (E) gene negative and nucleocapsid (N2) gene positive results.

Authors:  River Chun-Wai Wong; Ann Han Wong; Yolanda Iok-Ieng Ho; Gilman Kit-Hang Siu; Lam-Kwong Lee; Eddie Chi-Man Leung; Raymond Wai-Man Lai
Journal:  Diagn Microbiol Infect Dis       Date:  2022-05-20       Impact factor: 2.983

4.  Molecular diagnosis of SARS-CoV-2 in seminal fluid.

Authors:  D Paoli; F Pallotti; G Nigro; L Mazzuti; M N Hirsch; M B Valli; S Colangelo; C M Mastroianni; G Antonelli; A Lenzi; O Turriziani; F Lombardo
Journal:  J Endocrinol Invest       Date:  2021-04-30       Impact factor: 4.256

5.  Weak positive SARS-CoV-2 N2 gene results using the Xpress Xpert assay: the need for an alternate interpretative criteria in a low prevalence setting.

Authors:  Hemalatha Varadhan; Vishal Ahuja; Catherine Pitman; Dominic E Dwyer
Journal:  Pathology       Date:  2021-11-23       Impact factor: 5.306

Review 6.  Performance Evaluation of Lateral Flow Assays for Coronavirus Disease-19 Serology.

Authors:  Lucy Ochola; Paul Ogongo; Samuel Mungai; Jesse Gitaka; Sara Suliman
Journal:  Clin Lab Med       Date:  2021-11-03       Impact factor: 1.935

7.  SARS-CoV2 RT-PCR assays: In vitro comparison of 4 WHO approved protocols on clinical specimens and its implications for real laboratory practice through variant emergence.

Authors:  Mariem Gdoura; Imen Abouda; Mehdi Mrad; Imen Ben Dhifallah; Zeineb Belaiba; Wasfi Fares; Anissa Chouikha; Maroua Khedhiri; Kaouther Layouni; Henda Touzi; Amel Sadraoui; Walid Hammemi; Zina Meddeb; Nahed Hogga; Sihem Ben Fadhel; Sondes Haddad-Boubaker; Henda Triki
Journal:  Virol J       Date:  2022-03-28       Impact factor: 4.099

8.  Detection of SARS-CoV-2 at the point of care.

Authors:  Michael J Loeffelholz; Yi-Wei Tang
Journal:  Bioanalysis       Date:  2021-07-22       Impact factor: 2.681

9.  Combination of rRT-PCR and Anti-Nucleocapsid/Anti-Spike Antibodies to Characterize Specimens with Very Low Viral SARs-CoV-2 Load: A Real-Life Experience.

Authors:  Zoe Florou; Meropi Zigra; Philippos Kartalidis; Katerina Tsilipounidaki; Georgia Papadamou; Aikaterini Belia; George C Fthenakis; Efthymia Petinaki
Journal:  Microorganisms       Date:  2021-06-10
  9 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.