Application of digital PCR to determine the reliability of Xpert Xpress SARS-CoV-2 assay with envelope (E) gene negative and nucleocapsid (N2) gene positive results.

River Chun-Wai Wong1, Ann Han Wong2, Yolanda Iok-Ieng Ho2, Gilman Kit-Hang Siu3, Lam-Kwong Lee3, Eddie Chi-Man Leung2, Raymond Wai-Man Lai2.   

Abstract

This study used digital polymerase chain reaction (dPCR) to determine whether envelope (E) gene-negative and nucleocapsid (N2) gene-positive (E-N+) results obtained with the Cepheid Xpert Xpress SARS-CoV-2 assay are reliable. Using droplet digital PCR results as a reference, 18 of 22 E-N+ samples with a low viral load (81.8%) were identified as true positives.
Copyright © 2022 Elsevier Inc. All rights reserved.

Entities:  

Keywords:  Cepheid Xpert Xpress SARS-CoV-2 assay; E gene negative; N2 gene positive; SARS-CoV-2; digital PCR; low viral load

Mesh:

Year:  2022        PMID: 35691105      PMCID: PMC9119709          DOI: 10.1016/j.diagmicrobio.2022.115726

Source DB:  PubMed          Journal:  Diagn Microbiol Infect Dis        ISSN: 0732-8893            Impact factor:   2.983


Introduction

In Cepheid Xpert Xpress SARS-CoV-2 (Xpert Xpress) assay, the dual targets of the envelope (E) and nucleocapsid (N2) genes are amplified. The result is interpreted as positive if either both genes or N2 gene alone is detected [1]. E-N+ results account for 3.2% of the positive SARS-CoV-2 samples tested in our center and similar rate has been reported by Khoshchehreh et al. [2]. Falasca et al. showed that centrifugation to concentrate the Xpert Xpress E-N+ samples by 10 fold, followed by a repeated test, gave a negative result [3]. It is, therefore, recommended that E-N+ results should be interpreted with caution as they may be false positives [3]. Droplet digital PCR (ddPCR) has superior analytical sensitivity, especially for the detection of SARS-CoV-2 in samples with a low viral load [4], [5], [6], [7], [8], [9], [10]. There are 2 types of digital PCR (dPCR) platforms: droplet-based (e.g., Bio-Rad QX200 ddPCR system), which is well-established, and nanoplate-based (e.g., Qiagen QIAcuity One system), which is new to the market. This study aimed to use dPCR to determine whether E-N+ results obtained with the Xpert Xpress assay are reliable. Twenty-two respiratory samples collected from 19 patients with a previous E-N+ result (Ct values ranging from 37.7 to 44.8) from the Xpert Xpress assay were included in the study. The samples were stored at -80°C for 3 months before retrieval for dPCR. 200 μL of samples was used for RNA extraction and eluted in 50 μL as previously described [11]. To assist in the determination of infection status, testing for IgG antibodies against SARS-CoV-2 nucleocapsid protein (anti-NP IgG) was performed on serum samples, if available, using the Alinity i qualitative assay (Abbott, Sligo, Ireland).

Cepheid Xpert Xpress SARS-CoV-2 assay

Samples were analyzed according to the manufacturer's recommendations. The limit of detection (LoD) claimed is 250 copies/mL.

Bio-Rad QX200 ddPCR

The ddPCR test on the QX200 system (Bio-Rad Laboratories, CA, USA) was performed according to the manufacturer's recommendations. All samples were tested in triplicate. The results were analyzed by using QX Manager, standard edition. Samples were interpreted as positive if ≥1 droplets were detected for either one (N1 or N2) or both targets and confirmed by comparing with the positive control. The LoD claimed for the Bio-Rad SARS-CoV-2 ddPCR kit is 150 copies/mL.

Qiagen QIAcuity One

The Qiagen SARS-CoV-2 N1+N2 assay (Qiagen, Hilden, Germany) was performed using the Nanoplate 26k 24-Well protocol according to the manufacturer's recommendations. All samples were tested in triplicate, and results were analyzed by QIAcuity One system. The LoD claimed is 250 copies/mL. Using QX200 ddPCR assay as a reference method, 18 of 22 samples (81.8%), with droplet concentrations ranging from 61 to 1872 copies/mL, were confirmed as positive, while 15 samples (68.2%) were detected with the QIAcuity One assay (Table 1 ). It showed that E-N+ results generated with the Xpert Xpress assay were mostly true positives. The positive percentage agreement between the 2 dPCR platforms was 77.78% (Table 2 ). For the 18 true positive samples confirmed by ddPCR, the Ct values of the N2 gene ranged from 37.7 to 44.4, and 16 of the Ct values were greater than or equal to 39. Of the 4 samples with no positive droplet detected by ddPCR, sample 6 tested positive using the QIAcuity One assay, with a concentration of 50 copies/mL. This was below the LoD of the QX200 ddPCR assay. For the 3 true negative samples (false positives with the Xpert Xpress assay) that were negative by both dPCR assays, the Ct values of the N2 gene ranged from 40.7 to 43.7.
Table 1

Detection of SARS-CoV-2 by digital PCR in 22 samples with a Xpert Xpress SARS-CoV-2 result of E-N+.

Sample no.Digital PCR (copies/mL)
RT-PCRSample type for digital PCR/RT-PCR, time of collection from symptom onsetSARS-CoV-2 IgG anti-NP antibodies, time of collection from symptom onset
Qiagen QIAcuity One dPCRBio-Rad QX200 ddPCRXpert XpressE, N2 gene Ct
147271Not detected, 39.8NPS+TS, 25 daysPositive, 16 days
2435180Not detected, 38.2DTS, 35 daysPositive, 23 days
3Not detected70Not detected, 42.1DTS, asymptomaticPositive, asymptomatic
4Not detected177Not detected, 44.4NPS+TS, 2 daysNot performed
5495744Not detected, 37.7DTS, 23 daysPositive, 13 days
650No positive droplet detectedNot detected, 44.8TA, 32 daysPositive, 6 days
752832Not detected, 40.5DTS, 3 daysNot performed
8Not detected118Not detected, 43.1DTS, 33 daysPositive, 40 days
9Not detected172Not detected, 43.9DTS, 2 daysNot performed
10a199529Not detected, 43.2NPS+TS, asymptomaticPositive, asymptomatic
11a5161Not detected, 43.2DTS, asymptomatic
12a491872Not detected, 42.0NPS+TS, asymptomatic
135085Not detected, 41.4DTS, 2 daysPositive, 14 days
14Not detectedNo positive droplet detectedNot detected, 40.7NPS+TS, 6 daysNot performed
15bNot detectedNo positive droplet detectedNot detected, 43.7DTS, 19 daysPositive, 8 days
16b50141Not detected, 43.3NPA+TS, 19 days
1753137Not detected, 43.2Sputum, 11 daysPositive, 8 days
18102138Not detected, 42.3NPS+TS, 16 daysNot performed
19Not detectedNo positive droplet detectedNot detected, 43.1DTS, 2 daysNot performed
20300650Not detected, 39.0NPS+TS, 210 days, reinfectionPositive, 210 days, reinfection
21236480Not detected, 41.1DTS, asymptomaticPositive, asymptomatic
22336140Not detected, 39.2DTS, 22 daysPositive, 12 days

NPS+TS = Combined swabs collected from nasopharyngeal and throat; DTS = deep throat saliva (also known as posterior oropharyngeal saliva); TA = tracheal aspirate.

Samples 10-12 were collected from the same patient for disease monitoring.

Samples 15 and 16 were collected from the same patient for disease monitoring.

Table 2

Comparison between the Bio-Rad QX200 ddPCR and the Qiagen QIAcuity 1 dPCR results for 22 samples that gave an E-N+ result with the Xpert Xpress SARS-CoV-2 assay.

Bio-Rad QX200 ddPCR
(95% CI)
DetectedNot detectedPPANPA
Qiagen QIAcuityOne dPCRDetected14177.78%75.00%
Not detected43(52.36 to 93.59%)(19.41 to 99.37%)

PPA = positive percent agreement; NPA = negative percent agreement; dPCR = digital polymerase chain reaction; ddPCR = droplet digital polymerase chain reaction; CI = confidence interval.

Detection of SARS-CoV-2 by digital PCR in 22 samples with a Xpert Xpress SARS-CoV-2 result of E-N+. NPS+TS = Combined swabs collected from nasopharyngeal and throat; DTS = deep throat saliva (also known as posterior oropharyngeal saliva); TA = tracheal aspirate. Samples 10-12 were collected from the same patient for disease monitoring. Samples 15 and 16 were collected from the same patient for disease monitoring. Comparison between the Bio-Rad QX200 ddPCR and the Qiagen QIAcuity 1 dPCR results for 22 samples that gave an E-N+ result with the Xpert Xpress SARS-CoV-2 assay. PPA = positive percent agreement; NPA = negative percent agreement; dPCR = digital polymerase chain reaction; ddPCR = droplet digital polymerase chain reaction; CI = confidence interval. Discordant results for the 2 dPCR assays were observed in sample no. 3, 4, 6, 8 and 9. They had viral loads below or close to the LoD. We verified the analytical sensitivity of both dPCR assays using purified intact SARS-CoV-2 viral particles (NATtrol SARS-Related Coronavirus 2; ZeptoMetrix, NY, USA) and determined that the LoDs were 100 copies/mL. Further testing of 5 external quality assurance samples showed 100% concordance with the intended qualitative results, with a similar viral load detected by both platforms (Table 3 ). This provided further evidence that discrepancies occur mainly in samples with viral loads close to the LoD.
Table 3

Comparison of 5 QCMD EQAP SARS-CoV-2 samples analyzed with the Bio-Rad QX200 ddPCR and the Qiagen QIAcuity 1 dPCR assays.

Qiagen QIAcuity One dPCRBio-Rad QX200 ddPCRQCMD EQAP (for reference only)
Sample no.log10 copies/mLlog10 copies/mLlog10 copies/mL
SCV2_21C1B-014.774.964.13
SCV2_21C1B-023.483.462.51
SCV2_21C1B-032.822.952.00
SCV2_21C1B-043.703.842.94
SCV2_21C1B-053.863.993.15

QCMD = Quality Control for Molecular Diagnostics; EQAP = external quality assurance program.

Comparison of 5 QCMD EQAP SARS-CoV-2 samples analyzed with the Bio-Rad QX200 ddPCR and the Qiagen QIAcuity 1 dPCR assays. QCMD = Quality Control for Molecular Diagnostics; EQAP = external quality assurance program. There was no positive droplet being detected by QX200 ddPCR on 2 samples (no. 6 & 15) and 3 samples (no. 3, 8 &15) were not detected by QIAcuity One assay, while these patients were seropositive that indicates a prior COVID-19 infection. Sample no. 15 & 16 were collected from the same patient on the same day and discrepancies in dPCR results was probably associated with sample quality as no. 15 was a saliva sample that self-collected by patient while no. 16 was a nasopharyngeal sample that collected by healthcare workers. These patients were likely have resolving infections with prolonged viral RNA shedding in their respiratory samples. As Hong Kong has adopted a Zero-COVID strategy, clinically stable patients will be monitored for the presence of SARS-CoV-2 before discharge. Our findings suggested majority of E-N+ results obtained with the Xpert Xpress assay were true positives, which is discrepant to the results reported by Falasca et al. [3]. It may due to different approaches used for the confirmation of such weakly positive results. In our laboratory, any sample with a Ct value ≥35 is confirmed using another RT-PCR platform, and a positive result is only reported if SARS-CoV-2 RNA is detected by both platforms. Otherwise, the result is reported as indeterminate. Clinicians are strongly advised to collect another sample (preferably a lower respiratory tract specimen) from patients with indeterminate results and manage them as presumptive positive while recommending repeat sampling. Testing for the presence of Anti-NP IgG could also be considered to investigate the possible false positive RT-PCR result. Stoichiometric variability may also have contributed to the discrepancies between the 2 dPCR assays. The input volume of RNA was different, with 5.5 μL used for each of the 22 μL reactions in the QX200 ddPCR assay and 10 μL used for each of the 40 μL reactions in the QIAcuity One assay. Such a difference in the volume of template added may affect the amount of viral RNA available for the RT-PCR reaction. In conclusion, our study showed that most of the E-N+ Xpert Xpress results were true positives. There are several limitations of this study. First, the sample size was small. Second, dPCR was performed retrospectively and not a “head-to-head” comparison with the Xpert Xpress assay, which may have introduced bias into the analysis. Third, RNA may be degraded after prolonged storage or multiple freeze and/or thaw processes. Further prospective studies are therefore recommended.

Authors' contribution

River Chun-Wai Wong: Conceptualization, Data curation, Formal analysis, Validation, Investigation, Writing-original draft, Writing-review & editing. Ann Han Wong: Writing-review & editing. Yolanda Iok-Ieng Ho: Writing-review & editing. Gilman Kit-Hang Siu: Supervision, Writing-review & editing. Lam-Kwong Lee: Writing-review & editing. Eddie Chi-Man Leung: Writing-review & editing. Raymond Wai-Man Lai: Project administration, Supervision.

Declaration of competing interest

The authors declare that we have no conflicts of interest.
  10 in total

1.  Rapid adaptation and continuous performance evaluation of SARS-CoV-2 envelope gene (E-gene) real-time RT-PCR assays to support the hospital surge in test demand.

Authors:  Yolanda I I Ho; Ann H Wong; Eddie C M Leung; River C W Wong; Raymond W M Lai
Journal:  J Med Virol       Date:  2020-11-22       Impact factor: 2.327

2.  Sensitivity assessment of droplet digital PCR for SARS-CoV-2 detection.

Authors:  Luca Falzone; Nicolò Musso; Giuseppe Gattuso; Dafne Bongiorno; Concetta Ilenia Palermo; Guido Scalia; Massimo Libra; Stefania Stefani
Journal:  Int J Mol Med       Date:  2020-07-13       Impact factor: 4.101

3.  Evaluation on testing of deep throat saliva and lower respiratory tract specimens with Xpert Xpress SARS-CoV-2 assay.

Authors:  River Chun-Wai Wong; Ann Han Wong; Yolanda Iok-Ieng Ho; Eddie Chi-Man Leung; Raymond Wai-Man Lai
Journal:  J Clin Virol       Date:  2020-08-16       Impact factor: 3.168

4.  Highly accurate and sensitive diagnostic detection of SARS-CoV-2 by digital PCR.

Authors:  Lianhua Dong; Junbo Zhou; Chunyan Niu; Quanyi Wang; Yang Pan; Sitong Sheng; Xia Wang; Yongzhuo Zhang; Jiayi Yang; Manqing Liu; Yang Zhao; Xiaoying Zhang; Tao Zhu; Tao Peng; Jie Xie; Yunhua Gao; Di Wang; Xinhua Dai; Xiang Fang
Journal:  Talanta       Date:  2020-10-27       Impact factor: 6.057

5.  ddPCR: a more accurate tool for SARS-CoV-2 detection in low viral load specimens.

Authors:  Tao Suo; Xinjin Liu; Jiangpeng Feng; Ming Guo; Wenjia Hu; Dong Guo; Hafiz Ullah; Yang Yang; Qiuhan Zhang; Xin Wang; Muhanmmad Sajid; Zhixiang Huang; Liping Deng; Tielong Chen; Fang Liu; Ke Xu; Yuan Liu; Qi Zhang; Yingle Liu; Yong Xiong; Guozhong Chen; Ke Lan; Yu Chen
Journal:  Emerg Microbes Infect       Date:  2020-12       Impact factor: 7.163

6.  Digital droplet PCR accurately quantifies SARS-CoV-2 viral load from crude lysate without nucleic acid purification.

Authors:  Harish N Vasudevan; Peng Xu; Venice Servellita; Steve Miller; Leqian Liu; Allan Gopez; Charles Y Chiu; Adam R Abate
Journal:  Sci Rep       Date:  2021-01-12       Impact factor: 4.379

7.  Analytical and Clinical Performance of Droplet Digital PCR in the Detection and Quantification of SARS-CoV-2.

Authors:  Kyoung Bo Kim; Hayoung Choi; Gun Dong Lee; Jaewoong Lee; Seungok Lee; Yonggoo Kim; Sung-Yeon Cho; Dong-Gun Lee; Myungshin Kim
Journal:  Mol Diagn Ther       Date:  2021-07-28       Impact factor: 4.074

8.  Detection of SARS-COV N2 Gene: Very low amounts of viral RNA or false positive?

Authors:  Francesca Falasca; Ilaria Sciandra; Daniele Di Carlo; Massimo Gentile; Alberto Deales; Guido Antonelli; Ombretta Turriziani
Journal:  J Clin Virol       Date:  2020-10-14       Impact factor: 3.168

9.  Detection and quantification of SARS-CoV-2 by droplet digital PCR in real-time PCR negative nasopharyngeal swabs from suspected COVID-19 patients.

Authors:  Claudia Alteri; Valeria Cento; Maria Antonello; Luna Colagrossi; Marco Merli; Nicola Ughi; Silvia Renica; Elisa Matarazzo; Federica Di Ruscio; Livia Tartaglione; Jacopo Colombo; Chiara Grimaldi; Stefania Carta; Alice Nava; Valentino Costabile; Chiara Baiguera; Daniela Campisi; Diana Fanti; Chiara Vismara; Roberto Fumagalli; Francesco Scaglione; Oscar Massimiliano Epis; Massimo Puoti; Carlo Federico Perno
Journal:  PLoS One       Date:  2020-09-08       Impact factor: 3.240

10.  A needle in the haystack? Assessing the significance of envelope (E) gene-negative, nucleocapsid (N2) gene-positive SARS-CoV-2 detection by the Cepheid Xpert Xpress SARS-COV-2 assay.

Authors:  Mahdi Khoshchehreh; Noah Wald-Dickler; Paul Holtom; Susan M Butler-Wu
Journal:  J Clin Virol       Date:  2020-10-29       Impact factor: 3.168

  10 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.