Petros Bozidis1, Eleni T Tsaousi1, Charilaos Kostoulas2, Prodromos Sakaloglou1,2, Athanasia Gouni3, Despoina Koumpouli3, Hercules Sakkas1,3, Ioannis Georgiou2, Konstantina Gartzonika1,3.
Abstract
Several SARS-CoV-2 variants have emerged and early detection for monitoring their prevalence is crucial. Many identification strategies have been implemented in cases where sequencing data for confirmation is pending or not available. The presence of B.1.1.318 among prevalent variants was indicated by an unusual amplification pattern in various RT-qPCR commercial assays. Positive samples for SARS-CoV-2, as determined using the Allplex SARS-CoV-2 Assay, the Viasure SARS-CoV-2 Real Time Detection Kit and the GeneFinder COVID-19 Plus RealAmp Kit, presented a delay or failure in the amplification of the N gene, which was further investigated. Whole-genome sequencing was used for variant characterization. The differences between the mean Ct values for amplification of the N gene vs. other genes were calculated for each detection system and found to be at least 14 cycles. Sequencing by WGS revealed that all the N gene dropout samples contained the B.1.1.318 variant. All the isolates harbored three non-synonymous mutations in the N gene, which resulted in four amino acid changes (R203K, G204R, A208G, Met234I). Although caution should be taken when the identification of SARS-CoV-2 variants is based on viral gene amplification failure, such patterns could serve as a basis for rapid and cost-effective screening, functioning as indicators of community circulation of specific variants, requiring subsequent verification via sequencing.Entities:
Keywords: B.1.1.318; SARS-CoV-2; diagnostic quantitative RT-PCR; variant
Year: 2022 PMID: 35454022 PMCID: PMC9029054 DOI: 10.3390/diagnostics12040973
Source DB: PubMed Journal: Diagnostics (Basel) ISSN: 2075-4418
Figure 1(A) All three commercial detection kits show N amplification delay (partial dropout) or failure (complete dropout) in samples infected by the B.1.1.318 strain. (B) 42 samples infected by the B.1.1.7 strain were tested with all three detection kits for comparison. RT-qPCR data as Ct range and Ct mean values for each target of each molecular assay are depicted. (C) Distribution of delta Ct values for cases without dropout infected with B.1.1.7 (black box plots) and with partial dropout infected with B.1.1.318 variant (grey box plots). The visualization of differences in the Ct values of N vs. E or RdRp/S genes in Allplex SARS-CoV-2 assay is presented in left graph and between N vs. Orf1a/b in Viasure SARS-CoV-2 Real Time Detection in right graph both for the groups of the B.1.1.7 and B.1.1.318 variants. The horizontal lines represent mean values of differences of Ct values.
Figure 2(A) The 10 random samples out of 120 that were infected with the B.1.1.318 strain showed similar N gene amplification delays or failures when tested with all three detection kits in parallel. (B) Mutations of the N gene sequence of B.1.1.318 isolates that affect primer–probe annealing.
List of commercial kits used for presumptive identification of specific SARS-CoV-2 variants based on viral gene amplification delays or failures.
| Variant | Gene | Detection Kit | References |
|---|---|---|---|
| B.1.1.7 | S dropout or delay | TaqPath COVID-19 CE-IVD RT-PCR Kit | [ |
| Helix® COVID-19 Test | [ | ||
| RdRp/S double curve or low amplification | Allplex SARS-CoV-2 Assay | [ | |
| N dropout or delay | TaqPath COVID-19 CE-IVD RT-PCR Kit | [ | |
| Allplex SARS-CoV-2/FluA/FluB/RSV Assay | [ | ||
| B.1.258 | S dropout or delay | TaqPath COVID-19 CE-IVD RT-PCR Kit | [ |
| B.1.617.2 | RdRp/S double curve or low amplification | Allplex SARS-CoV-2 Assay | [ |
| B.1.177 (B.1.177.75) | N dropout | TaqPath COVID-19 CE-IVD RT-PCR Kit | [ |
| B.1.1.529 | S dropout | TaqPath COVID-19 Combo Kit | [ |
| B.1.1.318 | N dropout | Amplidiag COVID-19 | [ |
| PerkinElmer SARS-CoV-2 Real-time RT-PCR Assay | [ | ||
| N dropout or delay | Allplex SARS-CoV-2 | This work | |
| VIASURE SARS-CoV-2 Real Time Detection Kit | This work | ||
| GeneFinder COVID-19 Plus RealAmp Kit | This work | ||
| P.1 (or P.1.1) | N dropout | GeneFinder COVID-19 Plus RealAmp Kit | [ |