Suchitra Kumari1, Jyotirmayee Bahinipati2, Tapaswini Pradhan2, Durgesh P Sahoo3.
Abstract
BACKGROUND: The primary health-care center (PHC) and community health center (CHC) are not well equipped with laboratory services. Semiauto analyzer-based reporting could be an effective modality, provided that the performance standard is comparable to that of the fully automatic analyzer. So, the objective of this study was to analyze the test results of biochemical parameters in semiauto and fully automatic analyzer and to compare the quality performance.Entities:
Keywords: Auto analyzer; biochemical parameters; performance standard; semiauto analyzer
Year: 2020 PMID: 33110800 PMCID: PMC7586617 DOI: 10.4103/jfmpc.jfmpc_94_20
Source DB: PubMed Journal: J Family Med Prim Care ISSN: 2249-4863
Descriptive statistics, skewness, kurtosis, and correlation between two measurement methods
| Mean | Std. deviation | Skewness | Kurtosis | Spearman correlation | |||
|---|---|---|---|---|---|---|---|
| Semiauto urea | 31.74 | 26.118 | 149 | 3.432 | 15.801 | 0.691 | 0.0001 |
| Auto urea | 41.58 | 41.073 | 149 | 2.724 | 8.032 | ||
| Semiauto cholesterol | 142.79 | 61.498 | 149 | 1.522 | 2.373 | 0.798 | 0.0001 |
| Auto cholesterol | 161.30 | 67.555 | 149 | 1.037 | 0.693 | ||
| Semiauto TG | 136.17 | 67.470 | 149 | 2.258 | 5.989 | 0.821 | 0.0001 |
| Auto TG | 160.51 | 92.825 | 149 | 2.596 | 7.402 | ||
| Semiauto OT | 42.42 | 76.576 | 149 | 4.533 | 20.42 | 0.653 | 0.0001 |
| Auto OT | 56.95 | 102.384 | 149 | 4.187 | 17.455 | ||
| Semiauto PT | 37.52 | 41.046 | 149 | 5.427 | 35.923 | 0.551 | 0.0001 |
| Auto PT | 59.99 | 104.540 | 149 | 7.995 | 77.454 |
Figure 1Correlation between two measurement methods. (a) Correlation between manual and automated urea. (b) Correlation between manual and automated cholesterol. (c) Correlation between manual and automated triglyceride. (d) Correlation between manual OT and automated OT. (e) Correlation between manual and automated PT
Bland-Altman analysis of agreement between two measurement methods
| Mean | Std. deviation | Limits of agreement (LOA) | Cronbach α | 95% CI | |||
|---|---|---|---|---|---|---|---|
| Upper | Lower | Upper | Lower | ||||
| Difference urea=Semiauto-auto | -9.85 | 23.997 | 37.189 | -56.880 | 0.862 | 0.809 | 0.90 |
| Difference Cholesterol=Semiauto-auto | -18.52 | 39.261 | 58.434 | -95.468 | 0.898 | 0.859 | 0.926 |
| Difference TG=Semiauto-auto | -24.34 | 38.513 | 51.144 | -99.829 | 0.94 | 0.918 | 0.957 |
| Difference OT=Semiauto-auto | -14.53 | 49.354 | 82.204 | -111.264 | 0.919 | 0.889 | 0.942 |
| Difference PT=Semiauto-auto | -22.47 | 73.150 | 120.905 | -165.845 | 0.731 | 0.628 | 0.805 |
Correlation between the means and the differences to confirm the uniformity of variance
| Kendall’s tau_b | Difference urea=manual-auto | |
|---|---|---|
| Mean urea=(manual + auto) /2 | Correlation coefficient | -0.386 |
| 0.0001 | ||
| Difference cholesterol=manual-auto | ||
| Mean cholesterol=(manual + auto) /2 | Correlation coefficient | -0.181 |
| 0.0001 | ||
| Difference TG=manual-auto | ||
| Mean TG=(manual + auto) /2 | Correlation coefficient | -0.224 |
| 0.0001 | ||
| Difference OT=manual-auto | ||
| Mean OT=(manual + auto) /2 | Correlation coefficient | -0.222 |
| 0.0001 | ||
| Difference PT=manual-auto | ||
| Mean PT=(manual + auto) /2 | Correlation coefficient | -0.231 |
| 0.0001 |
Figure 2Agreement between manual and auto methods (a) Urea (b) Cholesterol (c) TG (d) OT (e) PT