Rodolfo Mastropasqua1, Federica Evangelista2, Francesco Amodei2, Rossella D'Aloisio2, Filomena Pinto2, Emanuele Doronzo2, Pasquale Viggiano2, Annamaria Porreca3, Marta Di Nicola3, Mariacristina Parravano4, Lisa Toto2.
Abstract
Purpose: The purpose of this study is to compare the ability of 3 optical coherence tomography angiography (OCTA) devices to measure lesion area in patients with macular neovascularization (MNV) with type 1, 2 and mixed neovascularization (NV).Entities:
Keywords: macular neovascularization; optical coherence tomography; swept source OCTA
Mesh:
Year: 2020 PMID: 33101783 PMCID: PMC7545059 DOI: 10.1167/tvst.9.11.6
Source DB: PubMed Journal: Transl Vis Sci Technol ISSN: 2164-2591 Impact factor: 3.283
Figure 1.En face SS and SD-OCTA images from the left eye of a 72-year-old man with macular neovascular disease. All images were obtained using the outer retina to choriocapillaris (ORCC) image. (A) PlexElite OCTA 3 × 3-mm2 scan. (B) PlexElite OCTA 6 × 6-mm2. (C) Spectralis OCTA 3 × 3-mm2 scan. (D) Spectralis OCTA 6 × 6-mm2 scan. (E) Optovue OCTA 3 × 3-mm2 scan. (F) Optovue OCTA 6 × 6-mm2.
Optical Coherence Tomography Angiography Instruments
| System | AngioVue | Spectralis OCTA | PLEX Elite 9000 |
|---|---|---|---|
|
| OptoVue, Fremont Inc., CA, USA | Heidelberg Engineering, Heidelberg, Germany | Carl Zeiss Meditec Inc., Dublin, CA, USA |
|
| RTVue XR Avanti SD-OCT | OCT2 SD-OCT | PLEX Elite 9000 SS-OCT |
|
| 2018.1.0.24 | 6.12.4.0 | 1.7.0.27959 |
|
| SSADA | FSADA probabilistic approach | OMAGC |
|
| 840 | 870 | 1050 |
|
| 70,000 | 85,000 | 100,000 |
|
| 304 × 304 | 512 × 512 | 300 × 300 |
|
| 400 × 400 | 512 × 512 | 500 × 500 |
|
| DualTrac | SMARTTrack | FastTrac |
FSADA, full-spectrum amplitude-decorrelation angiography; OMAGC, optical microangiography complex; SSADA, split-spectrum amplitude decorrelation algorithm.
Figure 2.Image processing: in the acquired avascular en face image, the NV lesion was manually delineated by two independent experienced readers. (A) The binarized image (B) was then computed and analyzed.
Figure 3.Detection rate for type 1, type 2 and mixed type 1 and 2 of MNV for the different slabs in each device.
Friedman Test Comparison of Measurements From Three OCTA Images in the Neovascularization Cohort
| AV | ORCC | CC |
| |
|---|---|---|---|---|
|
| 0.160 (0.017–0.441) | 0.268 (0.053–0.631) | 0.236 (0.058–0.745) |
|
|
| 0.039 (0–0.284) | 0.519 (0.083–0.98) | 0.529 (0.058–0.941) |
|
|
| 0.258 (0.143–0.701) | 0.258 (0.143–0.701) | 0.215 (0.018–0.542) |
|
|
|
|
|
|
NV area, mm2, median (IQR).