A Risk-Adjusted Control Chart to Evaluate Intensity Modulated Radiation Therapy Plan Quality.

Arkajyoti Roy1, Dan Cutright2, Mahesh Gopalakrishnan3, Arthur B Yeh4, Bharat B Mittal3.   

Abstract

PURPOSE: This study aimed to develop a quality control framework for intensity modulated radiation therapy plan evaluations that can account for variations in patient- and treatment-specific risk factors. METHODS AND MATERIALS: Patient-specific risk factors, such as a patient's anatomy and tumor dose requirements, affect organs-at-risk (OARs) dose-volume histograms (DVHs), which in turn affects plan quality and can potentially cause adverse effects. Treatment-specific risk factors, such as the use of chemotherapy and surgery, are clinically relevant when evaluating radiation therapy planning criteria. A risk-adjusted control chart was developed to identify unusual plan quality after accounting for patient- and treatment-specific risk factors. In this proof of concept, 6 OAR DVH points and average monitor units serve as proxies for plan quality. Eighteen risk factors are considered for modeling quality: planning target volume (PTV) and OAR cross-sectional areas; volumes, spreads, and surface areas; minimum and centroid distances between OARs and the PTV; 6 PTV DVH points; use of chemotherapy; and surgery. A total of 69 head and neck cases were used to demonstrate the application of risk-adjusted control charts, and the results were compared with the application of conventional control charts.
RESULTS: The risk-adjusted control chart remains robust to interpatient variations in the studied risk factors, unlike the conventional control chart. For the brainstem, the conventional chart signaled 4 patients with unusual (out-of-control) doses to 2% brainstem volume. However, the adjusted chart did not signal any plans after accounting for their risk factors. For the spinal cord doses to 2% brainstem volume, the conventional chart signaled 2 patients, and the adjusted chart signaled a separate patient after accounting for their risk factors. Similar adjustments were observed for the other DVH points when evaluating brainstem, spinal cord, ipsilateral parotid, and average monitor units. The adjustments can be directly attributed to the patient- and treatment-specific risk factors.
CONCLUSIONS: A risk-adjusted control chart was developed to evaluate plan quality, which is robust to variations in patient- and treatment-specific parameters.
© 2019 The Author(s).

Entities:  

Year:  2019        PMID: 33089020      PMCID: PMC7560572          DOI: 10.1016/j.adro.2019.11.006

Source DB:  PubMed          Journal:  Adv Radiat Oncol        ISSN: 2452-1094


Introduction

Typically, treatment plan evaluation in intensity modulated radiation therapy (IMRT) is performed by comparing the patient’s dose distribution and volume coverage to established thresholds. Often, the thresholds need to be relaxed owing to patient- and treatment-specific risk factors, such as the complexity of a patient’s anatomy, tumor dose requirements, use of chemotherapy, or surgery. This process involves subjective decisions, leading to large variations in IMRT planning criteria across patients and making the planning process irreproducible and plan quality incomparable among patients. Quality control of the treatment planning process can be captured using a control chart framework, as discussed by Chung et al and Pawlicki et al., Most commonly, control charts have been used in quality assurance of dosimetry from the treatment planning system (TPS). Individual, moving-range, standard deviation (SD), and exponentially weighted moving average (EWMA) charts were used to detect unusual trends in dose deviations between the TPS and delivered dose., Alternatively, Nordström et al found control charts to be effective for the continuous verification of monitor units and derived control limits that agree with international guidelines. However, the use of control charts for treatment plan evaluation has been limited due to planning process variability for each patient, including patient- and treatment-specific risk factors. Alternatively, plan evaluations based on patient anatomy have been explored under the knowledge-based planning (KBP) framework. Regression and machine-learning models were used to predict organ-at-risk (OAR) dose based on structure volumes, the planning target volume (PTV) and OAR volume overlap, distance-to-target histograms (DTH), and minimum distance between OAR and PTV.8, 9, 10 In addition, a principal component analysis was used to correlate OAR sparing to anatomic features, such as the median distance between OAR and PTV, portion of OAR volume within an specific distance, OAR and PTV overlap, portion of OAR volume outside the primary treatment field, and DTH., Additionally, overlap volume histograms (OVHs) have been used to capture the spatial configuration of an OAR with respect to a target and suggest possible improvements in OAR dose-volume histograms (DVHs), which were confirmed by replanning. For further examples of plan evaluation under the KBP framework, refer to Ge and Wu. In practice, RapidPlan provides a KBP implementation to predict DVHs with upper and lower bounds at SD. These bounds account for some routine planning process variability, but using SD under normally distributed data substantially increases the probability of a false alarm (type-I error) compared with using SD, as typically used in control chart methodologies. In other words, using SD leads to many more false-positives, which require unnecessary reviews and clinical resources. Furthermore, KBP cannot identify trends in planning criteria because time is not considered. To address these limitations, we propose merging modeling techniques from KBP with control charts using SD limits and plotting plans in chronological order. In this proof of concept, we developed a risk-adjusted control chart that can monitor plan quality after accounting for patient- and treatment-specific risk factors. The proposed control chart employs a risk-adjustment model and a variety of risk factors that are discussed in the following. In addition, we provide details of the patient set used for demonstration. Finally, the risk-adjusted control chart is demonstrated in comparison with conventional control charts.

Methods and Materials

Risk-adjusted control charts were developed for 6 OAR DVH points and average monitor units that serve as proxies for plan quality. The 6 DVH points used are D2, D20, D40, D60, D80, and D98, where D represents the amount of the dose delivered to x% of the structure volume. Alternatively, the dose to the absolute volume can be used in this framework. Average monitor units were computed by dividing the total number of monitor units by the number of beams in the plans. To develop the risk-adjusted control charts, we first modeled plan quality using the patient- and treatment-specific risk factors. Then, we used the residuals from the risk adjustment model to develop the risk-adjusted charts.

Risk adjustment model

Patient-specific risk factors, such as complex anatomic geometry or high-dose prescriptions, can have an impact on OAR D or monitor units.7, 8, 9, 10, 11, 12 Treatment-specific risk factors, such as concurrent chemotherapy or surgery, are clinically relevant when evaluating planning criteria. We model for the effect of such risk factors on OAR D for patient i using multiple regression viawhere is the intercept, is the coefficient associated with the jth risk factor z, and residuals are assumed to follow a normal distribution N(0,σ2) with constant variance. To achieve the best fit model, backward elimination was employed for variable selection. For patient i, the residual ε=D− is the difference between the observed D (which ignores the risk factors) and the predicted (which accounts for the risk factors). Therefore, the residual ε represents an adjusted D that effectively is free from the influence of risk factors z Then, the residuals are used within a control chart framework to evaluate and monitor plan quality. Similarly, a risk adjustment model is developed for average monitor units, and the residuals are computed based on the predicted and actual monitor units.

Risk-adjusted control chart

Conventional quality control charts plot the observations, in this case OAR D or average monitor units. Residuals can be used as the plotting statistic if observations change depending on risk factors, and by plotting residuals that are statistically independent of risk factors and plan quality proxy, the effect of the risk factors is removed. Individual control charts can be used to evaluate and monitor individual observations, such as patient i’s residual ε (Eq. 1). The residual chart (hereinafter the risk-adjusted control chart) consists of a center line (CL) estimated as the mean of the residuals, and lower and upper control limits (LCL and UCL). The control limits are SD from the CL, yielding a very low probability of misclassifying a typical plan as unusual. The SD is estimated using the average moving range of all consecutive residual pairs, where the moving range for patient i, MR=|ε− ε| is the absolute difference in residuals between patients i and i+1. is standardized by the constant d2=1.128 to estimate SD (Eqs. 2 and 3). The constant d2 is based on the number of consecutive points used in the MR computation (ie, 2 in this study, ε and ε). Therefore, the LCL and UCL are and The risk-adjusted control chart can identify residuals outside of the control limits (ie, out-of-control [OC]). This indicates an unusual plan quality, signaling the need for further review because the unusual observation likely cannot be attributed to the risk factors.

Risk factors

When modeling plan quality, various risk factors can be used, including anatomy, treatment history, TPS parameters, physician’s experience, and institutional protocols. Here, we considered tumor dose requirements, use of chemotherapy, surgery, and the patient’s anatomy parametrized through 4 geometry- and 2 proximity-based parameters. As described in Deasy et al, we computed cross-sectional areas, volumes, spreads (maximum distance in x, y, or z dimensions), and surface areas of the PTV and OARs to capture the anatomic geometry. We computed the nonnegative minimum distance between the PTV and OARs, as well as the distance between PTV and OAR centroids to capture the proximity of structures. When the OAR overlaps the PTV, the OAR-PTV overlap volume replaces the null minimum distance. For tumor dose requirements, 6 PTV DVH points D2, D20, D40, D60, D80, and D98 are used. Therefore, a risk adjustment model for a plan quality proxy consists of j=1,…,18 risk factors: 6 PTV DVH points, 4 PTV and 4 OAR geometric measures, 2 proximity measures, use of chemotherapy, and surgery. When modeling the average monitor units, all OAR geometry- and proximity-based risk factors were employed in the original model.

Clinical data

A total of 69 head and neck cancer cases were used for this study. Brainstem and spinal cord were identified as the primary OARs to evaluate. Descriptive statistics about the brainstem and spinal cord D are provided in Table 1. Ipsilateral parotid was also identified as an organ of importance (ipsilateral parotid descriptive statistics are described in the supplemental material (Table E1 and E2) available online at https://doi.org/10.1016/j.adro.2019.11.006, accompanying the statistics describing average monitor units). To parametrize tumor dose, geometry, and proximity consistently, the largest PTV was considered the primary target for each patient. The selection of the largest PTV was deemed reasonable because PTVs were treated sequentially for each patient; therefore, the largest PTV encompassed all other PTVs in the patient’s treatment course.
Table 1

Descriptive statistics of brainstem and cord D

Brainstem
Spinal cord
D2D20D40D60D80D98D2D20D40D60D80D98
Minimum5.502.701.701.300.900.7025.106.904.703.900.300.10
Maximum65.1058.7053.9048.7042.3031.3057.1054.1051.9049.7041.7031.10
Mean42.8435.1826.1616.519.185.6443.3639.3736.4432.3918.813.62
Median44.7037.1026.9014.705.703.9042.9039.5036.7032.1019.501.70
Standard deviation9.2210.4711.3411.168.496.094.485.625.625.9711.605.22

Abbreviations: Dx = doses to x% brainstem volume.

Descriptive statistics of brainstem and cord D Abbreviations: Dx = doses to x% brainstem volume. All patients underwent a computed tomography scan (3-mm slice width) for treatment planning. All target and OAR volumes were delineated and approved by the same physician to reduce variability in contouring and minimize subjective preferences in planning criteria. Outlier detection was performed on the anatomic data to identify inconsistencies in contouring, and resulted in the redelineation of 3 plans. Treatment planning was performed using Pinnacle (Philips) versions 6.2b to 9.10. Treatments were delivered with step-and-shoot intensity modulation. Radiation doses were prescribed for sequential delivery using a standard fractionation scheme of 1.8 to 2.0 Gy or 1.5 Gy twice daily. The low-risk target volume received 45 to 50 Gy while the intermediate-to-high risk target volume received 56 to 66 Gy. The gross tumor volume typically received 70 to 72 Gy. During the planning process, every attempt was made to ensure that 100% of the prescribed dose covered at least 95% of the target volume with a maximum dose <110% of the prescribed dose. For OARs and normal tissue, a combination of maximum dose and maximum DVH objectives were used for optimization. When serial OARs were in close proximity to treatment volumes, priority was given to these OARs while still maintaining adequate coverage for target volumes. A composite DVH was developed from the sequential deliveries for evaluation. The composite plan data were exported to a computational environment for radiation therapy research for data analysis via Digital Imaging and Communications in Medicine (DICOM) file export from Pinnacle.

Results

First, we provide the descriptive statistics of all risk factors included in the modeling. Then, we focus on the risk adjustment model and control charts for brainstem and spinal cord D2 because this reflects the maximum dose criteria commonly used to evaluate brainstem and spinal cord doses for head and neck cases. The regression outputs, diagnostic plots, and control charts for D20, D40, D60, D80, and D98 on the brainstem and spinal cord accompanied the 6 DVH points for the ipsilateral parotid and average monitor units in the supplemental material (available online at https://doi.org/10.1016/j.adro.2019.11.006). The descriptive statistics on the PTV DVH points for D2, D20, D40, D60, D80, and D98 for the 69 head and neck cases are summarized in Table 2. One patient received a low dose to the PTV, leading to a minimum D40=8.2, D60=3.6, D80=2.6, and D98=0.1. This was unusual compared with the remaining plans, with the second lowest D40=58.2, D60=51.1, D80=42.6, and D98=1.6.
Table 2

Descriptive statistics of PTV D

D2D20D40D60D80D98
Minimum61.537.08.23.62.60.1
Maximum83.178.676.772.867.860.2
Mean76.072.268.562.353.945.6
Median76.473.370.162.854.547.2
Standard deviation3.65.78.58.68.28.9

Abbreviations: Dx = doses to x% brainstem volume.

Descriptive statistics of PTV D Abbreviations: Dx = doses to x% brainstem volume. Statistics on the PTV, brainstem, and spinal cord geometries are summarized in Table 3. The mean cross-sectional area, volume, and surface area for the PTV were larger than the brainstem and spinal cord. However, the spinal cord spread was observed to be larger than the brainstem and the PTV.
Table 3

Minimum, mean, median, maximum, and standard deviation of the PTV, brainstem, and spinal cord geometries

StructureCross-sectional area (cm2)
Volume (cm3)
MinMeanMedianMaxSDMinMeanMedianMaxSD
PTV27.068.767.0145.120.5289.91317.71254.52451.5448.0
Brainstem1.74.24.110.51.59.427.426.447.36.7
Spinal cord0.51.01.01.80.310.821.321.034.44.8

Abbreviations: max = maximum; min = minimum; PTV = planning target volume; SD = standard deviation.

Minimum, mean, median, maximum, and standard deviation of the PTV, brainstem, and spinal cord geometries Abbreviations: max = maximum; min = minimum; PTV = planning target volume; SD = standard deviation. Statistics on proximity measures, such as minimum and centroid distances between the OARs and PTV are summarized in Table 4. One patient had a PTV that overlapped with the brainstem, leading to a no-minimum distance between the structures. All remaining patients had positive minimum distances between the PTV and OARs. Furthermore, the average distance between the brainstem and PTV centroids was larger than the average distance between the spinal cord and PTV centroids. Treatment history was incorporated by accounting for concurrent chemotherapy (66 patients) and surgery (39 patients). These geometry-, proximity-, PTV dose-, and treatment-based risk factors are incorporated into the risk adjustment model.
Table 4

Descriptive statistics of proximity between PTV and organs at risk

Brainstem – PTV
Spinal cord – PTV
Minimum distance (cm)Distance between centroids (cm)Minimum distance (cm)Distance between centroids (cm)
Minimum0.06.50.53.9
Maximum2.612.51.714.0
Mean0.810.10.96.0
Median0.710.20.95.8
Standard deviation0.41.20.21.4

Abbreviation: PTV = planning target volume.

Descriptive statistics of proximity between PTV and organs at risk Abbreviation: PTV = planning target volume. The regression output from the risk adjustment model for the brainstem and spinal cord D2 are reported in Tables 5 and 6, respectively, with accompanying diagnostic plots shown in Figure 1. Although all 18 risk factors were included in every OAR model, the final model with the best fit risk factors using the Akaike information criterion is reported in the regression output in Tables 5 and 6. The Akaike information criterion is most commonly used for model selection, and measures the relative quality of the regression model when risk factors are removed iteratively to improve model quality.
Table 5

Regression output for modeling brainstem D2

CoefficientsEstimateStandard errort statisticP-value
Intercept74.73138.13579.1866.56E-13
PTV cross-sectional area−0.16710.0412−4.0570.00015
PTV0.01190.00244.9536.66E-06
PTV surface area−0.00400.0026−1.5560.125181
Minimum distance brainstem – PTV−13.46981.5823−8.5138.52E-12
Centroid distance brainstem – PTV−2.95350.7141−4.1360.000115
Brainstem volume−0.21130.1070−1.9750.052981
Brainstem spread2.48170.83622.9680.004352
PTV D40−0.31210.2250−1.3870.170833
PTV D600.82020.32712.5070.014983
PTV D80−0.55010.1853−2.9690.004341
Multiple R281.58%f statistic25.69
Adjusted R278.40%P-value2.2e-16
Residual standard error4.284

Abbreviations: Dx = doses to x% brainstem volume; PTV = planning target volume.

Table 6

Regression output for modeling spinal cord D2

CoefficientsEstimateStandard errort statisticP-value
Intercept52.937311.04904.7911.19E-05
PTV cross-sectional area−0.06970.0311−2.2410.0289
PTV0.00730.00233.2380.00199
PTV spread−0.41880.2779−1.5070.13728
PTV surface area−0.00340.0021−1.6380.10689
Minimum distance cord – PTV−14.28422.3202−6.1577.51E-08
Cord volume−0.19300.0959−2.0130.04881
PTV D20.30030.14452.0780.04213
PTV D40−0.44750.1873−2.3890.02019
PTV D600.80870.25853.1280.00275
PTV D80−0.51250.1526−3.3580.00139
Multiple R249.38%f statistic5.657
Adjusted R240.65%P-value7.47e-06
Residual standard error3.452

Abbreviations: Dx = doses to x% brainstem volume; PTV = planning target volume.

Figure 1

Residual diagnostic plots for (a, b) brainstem, and (c, d) spinal cord doses to 2% brainstem volume. Residual versus fitted plot shows linear relationship in variables and homoscedasticity in residuals and normal Q–Q plot shows normality of residuals.

Regression output for modeling brainstem D2 Abbreviations: Dx = doses to x% brainstem volume; PTV = planning target volume. Regression output for modeling spinal cord D2 Abbreviations: Dx = doses to x% brainstem volume; PTV = planning target volume. Residual diagnostic plots for (a, b) brainstem, and (c, d) spinal cord doses to 2% brainstem volume. Residual versus fitted plot shows linear relationship in variables and homoscedasticity in residuals and normal Q–Q plot shows normality of residuals. When modeling brainstem D2, model fit and significance are reported by R = 81.6% and P = 2.2e-16, respectively, as shown in Table 5. For spinal cord D2, R = 49.38% and P = 7.47e-06 (Table 6). Both best fit models produced significant results; however, the fit of the brainstem model was substantially higher than the spinal cord model. For both risk adjustment models, the assumptions of linearity, normality, and homoscedasticity appear valid. The residual versus fitted values plot in Figure 1A shows a random scattering of residuals around a mean of 0 and within a horizontal band, demonstrating linearity in the relationships between x and y, and homoscedasticity in the residuals. The normal Q–Q plot shows the residuals lying approximately on the y=x line in Figure 1B, demonstrating the normality of the residuals. Similar observations are made in Figures 1C and D to reflect valid assumptions for spinal cord D2 The residuals from the regression models are used as the charting variable for the risk-adjusted control charts as described in the following section. Conventional individual control charts have been used for the quality control of planning parameters in radiation therapy. Therefore, we compared the results from the proposed risk-adjusted control charts with those of the conventional individual control charts for the brainstem and spinal cord D2 Figure 2A demonstrates the conventional chart plotting brainstem D2, with an UCL=63.5Gy and LCL=22.2 Gy. Patient 4 was identified as out OC based on a high D2=65.1 Gy compared with the remaining plans. Conversely, patients 42, 51, and 52 were identified as OC based on their low brainstem D2=20.3Gy, 6.1Gy, and 5.5Gy, respectively.
Figure 2

(a) Conventional and (b) risk-adjusted control charts for brainstem doses to 2% brainstem volume and (c) conventional and (d) risk-adjusted control charts for spinal cord doses to 2% brainstem volume.

(a) Conventional and (b) risk-adjusted control charts for brainstem doses to 2% brainstem volume and (c) conventional and (d) risk-adjusted control charts for spinal cord doses to 2% brainstem volume. Figure 2B displays the conventional chart for spinal cord D2, with UCL=54.2Gy and LCL=32.5Gy. Patient 1 was identified as OC based on their spinal cord D2=57.1Gy that exceeds the UCL. Patient 43 was identified as OC based on their spinal cord D2=25.1Gy that is unusually low compared with the remaining patients. For our proposed method, Figure 2C depicts the risk-adjusted individual chart plotting the residuals from the brainstem D2 model, with UCL=11.6 Gy and LCL=−11.6Gy. Unlike the conventional chart, no patients had an OC brainstem D2 after accounting for the best fit risk factors. Figure 2D illustrates the risk-adjusted control chart for spinal cord D2 with UCL=9.0 Gy and LCL=−9.0 Gy, where only patient 19 was identified as an unusual plan with a low spinal cord D2=36.9 Gy. For the remaining brainstem D, the risk-adjusted control chart identified patients 4 and 58 with unusually high D80 and D98, and patient 61 was identified with unusually low D20 (Fig E4; available online at https://doi.org/10.1016/j.adro.2019.11.006). For the remaining spinal cord D, the plans for patients 1 and 3 produced an unusually high spinal cord D98 (Fig E5; available online at https://doi.org/10.1016/j.adro.2019.11.006). For the ipsilateral parotid, patient 1 was OC owing to a low D2 and a low D20, patient 51 was OC due to a low D40, and patients 52 and 54 were OC due to a low D60 (Fig E6; available online at https://doi.org/10.1016/j.adro.2019.11.006). When evaluating average monitor units, 2 plans were identified as OC in the conventional chart, with patients 31 and 61 having an unusually high number of average monitor units. Patient 31 remained OC in the risk-adjusted chart, and patient 61 was adjusted to be in control (Fig E7; available online at https://doi.org/10.1016/j.adro.2019.11.006).

Discussion

Many patients were identified as OC in the conventional charts across the DVH points for all 3 OARs and average monitor units. This was expected owing to the large variability in the DVHs and monitor units across patients. The risk-adjusted control chart typically reduced the number of OC plans by attributing some of the variability between patients to their respective risk factors. For example, compared with the entire patient set, the minimum distances between the PTVs and brainstems for patients 51 and 52 were >3.4 SD above the mean. This contributes to the unusually low brainstem D2. Similarly, for patient 42, the minimum distance was >2.8 SD above the mean. For patient 4, the minimum distance was 0 owing to volume overlap between the brainstem and PTV. Because this minimum distance is almost 2 SD below the mean, the close proximity to the PTV contributed to the high brainstem D2. In the case of patient 1, the minimum distance was >1.7 SD below the mean. The close proximity to the PTV contributed to a high spinal cord D2. Conversely, for patient 43, the minimum distance was >3.7 SD above the mean, which contributed to the low spinal cord D2. Patient 19’s spinal cord spinal cord D2 remained in control in the conventional chart, but the risk-adjusted chart signals OC after accounting for the risk factors. For this patient, the larger PTV (1.7 SD above the mean) and higher PTV dose (above average PTV D60 and D80) was used to adjust the spinal cord D2 to be unusually low in the risk-adjusted chart. Similar observations were made from other best fit risk factor variables that directly led to the adjustments between the conventional and risk-adjusted control charts for the remaining Dx on the brainstem, spinal cord, ipsilateral parotid, and average monitor units. At clinics, the risk-adjusted control chart can potentially improve the workflow and quality of radiation treatments. For example, the proposed chart can evaluate an optimized plan to be of high quality (in control) after accounting for risk factors, even if the OAR planning criteria are not met. In other words, the control chart can identify plans where improving the OAR planning criteria is only possible at the cost of worsening other risk factors, such as tumor dose distribution. Thus, the proposed chart can help reduce planning time and improve quality for such cases by identifying truly high-quality plans without having to iteratively modify planning criteria. Furthermore, the typical treatment planning approval process in a clinical setting involves comparing maximum dose and DVH criteria for the OARs to established guidelines, such as Quantitative Analyses of Normal Tissue Effects in the Clinic. These guidelines are directly related to risk of toxicity. Of note, a proxy for plan quality, as discussed herein, is not necessarily correlated to toxicity probability. The proposed quality control methodology aims to identify unusual plans based on historical patient data and is independent of meeting toxicity constraints. For patients who need to be treated for secondary cancers or re-treated for a primary cancer, the reduction of low doses that already meet the toxicity guidelines will be of value. An unusually low OAR dose is of less concern, but an OAR dose that exceeds the UCL does increase the potential for adverse effects and can be a cause for an additional review of the plan. The proposed chart can also be used to monitor future patient’s plans rather seamlessly. Given an existing database of already treated patient plans, the risk adjustment regression Equation 1 can be used to model any plan quality proxy, such as an OAR D. Next, residuals obtained from the adjustment model can be used to establish control limits using Equations (2), (3). Similarly, for a new patient, Equation 1 can be used to model for the same quality proxy, and Equations 2 and 3 can be used to compute the residual. Finally, the residual is plotted on the risk-adjusted control chart, reflecting the quality of the new patient’s treatment plan. If the residual is within the limits, the plan is acceptable. However, if the residual is OC, further review is needed. Another appealing feature of a control chart is the ability to capture the trend of the planning parameter over time because patients are plotted in chronological order. Not only can trends in planning criteria be visually observed, there are additional order-based quality evaluation criteria that accompany the control limits, which can be employed in the risk-adjusted control chart setting (out of the scope of this work). These include 8 consecutive points above or below the centerline, 4 of 5 consecutive points beyond SD, and 2 of 3 consecutive points beyond SD that signal a plan to be unusual. These additional measures of quality over time do not exist and cannot be accomplished within the current KPB framework. In addition to the anatomic parameters discussed herein, other parameterizations, such as portions of OAR outside the primary target fields and tumor site, can be easily considered., Beam information, such as beam angles and field arrangements, can also be included. However, given that there are many such planning parameters, external factors, and anatomic parametrizations that could be included in this framework, an exhaustive model considering all risk factors is outside of the scope of this proof of concept. Unlike control charts that monitor mean, range, or SD that use large subgroup sizes and rely on the central limit theorem, individual charts include a subgroup size of one and thus are sensitive to nonnormal distributions. Nonparametric control charts have been studied in the past, and change-point methods have been proposed to detect shifts in observations over a period in time., Although effective to detect prolonged shifts, these approaches are limited in identifying sporadic OC points and do not account for risk factors that may affect the process quality. As a result, developing nonparametric risk-adjusted control charts serve as an extension of this work. In the same context, extreme outliers can affect the width of the control limits when SD is used to capture the variability in the data. In such a case, a more robust estimate of variability, such as median absolute deviation, can be used to compute the control limits. Furthermore, standardizing the contouring of OARs can reduce outliers, leading to more acceptable control limits. Moreover, the quality control of OAR DVH can also be extended to a multivariate setting or profile monitoring, where multiple OAR D or the entire OAR DVH profile can be parametrized for risk adjustment. Capability indices have been used by Breen et al and Sanghangthum et al in radiation therapy to demonstrate that their dosimetry process was within acceptable user specification limits. Extending the indices to the risk-adjusted setting will be considered in future work. Ideally, clinics should have the ability to readily apply the control chart framework to their patient data without tediously curating and formatting the data, as was required for this proof of concept. For example, DVH Analytics is an open-source DICOM radiation therapy database software that can read large batches of DICOM files and compute various anatomic and dose values, store them in an SQL database, and provide a graphical user interface to explore the data. Furthermore, as the treatment planning process continues to evolve toward fully automated planning, the quality assurance of planning criteria becomes even more important because optimization algorithms often produce locally optimum solutions in radiation therapy. The proposed control chart methodology can employ large databases of existing plans to help identify locally optimal plans that could be improved through reoptimization of planning parameters.

Conclusions

A risk-adjusted control chart was developed to detect unusual planning criteria after accounting for patient- and treatment-specific risk factors, allowing for multiple plan comparisons on a large scale. When evaluating 69 head and neck cases, the risk-adjusted control charts were robust to the interpatient variations in anatomy, tumor dose prescription, use of chemotherapy, and surgery. Furthermore, the charts reduced the number of unusual plans compared with conventional control charts. Detecting unusually high OAR doses in risk-adjusted charts provides the clinician with a chance to review the treatment plan within the context of historical patient data and probe for possible improvements to reduce the potential for adverse effects.
  23 in total

1.  CERR: a computational environment for radiotherapy research.

Authors:  Joseph O Deasy; Angel I Blanco; Vanessa H Clark
Journal:  Med Phys       Date:  2003-05       Impact factor: 4.071

2.  Control chart analysis of data from a multicenter monitor unit verification study.

Authors:  Fredrik Nordström; Sacha af Wetterstedt; Stefan Johnsson; Crister Ceberg; Sven J Bäck
Journal:  Radiother Oncol       Date:  2012-01-10       Impact factor: 6.280

3.  Process control analysis of IMRT QA: implications for clinical trials.

Authors:  Todd Pawlicki; Sua Yoo; Laurence E Court; Sharon K McMillan; Roger K Rice; J Donald Russell; John M Pacyniak; Milton K Woo; Parminder S Basran; Arthur L Boyer; Claribel Bonilla
Journal:  Phys Med Biol       Date:  2008-08-26       Impact factor: 3.609

4.  Statistical process control for IMRT dosimetric verification.

Authors:  Stephen L Breen; Douglas J Moseley; Beibei Zhang; Michael B Sharpe
Journal:  Med Phys       Date:  2008-10       Impact factor: 4.071

5.  Multiple local minima in radiotherapy optimization problems with dose-volume constraints.

Authors:  J O Deasy
Journal:  Med Phys       Date:  1997-07       Impact factor: 4.071

6.  Predicting dose-volume histograms for organs-at-risk in IMRT planning.

Authors:  Lindsey M Appenzoller; Jeff M Michalski; Wade L Thorstad; Sasa Mutic; Kevin L Moore
Journal:  Med Phys       Date:  2012-12       Impact factor: 4.071

7.  Quantitative analysis of the factors which affect the interpatient organ-at-risk dose sparing variation in IMRT plans.

Authors:  Lulin Yuan; Yaorong Ge; W Robert Lee; Fang Fang Yin; John P Kirkpatrick; Q Jackie Wu
Journal:  Med Phys       Date:  2012-11       Impact factor: 4.071

8.  State of dose prescription and compliance to international standard (ICRU-83) in intensity modulated radiation therapy among academic institutions.

Authors:  Indra J Das; Aaron Andersen; Zhe Jay Chen; Andrea Dimofte; Eli Glatstein; Jeremy Hoisak; Long Huang; Mark P Langer; Choonik Lee; Matthew Pacella; Richard A Popple; Roger Rice; Jennifer Smilowitz; Patricia Sponseller; Timothy Zhu
Journal:  Pract Radiat Oncol       Date:  2016-11-13

9.  Modeling the dosimetry of organ-at-risk in head and neck IMRT planning: an intertechnique and interinstitutional study.

Authors:  Jun Lian; Lulin Yuan; Yaorong Ge; Bhishamjit S Chera; David P Yoo; Sha Chang; FangFang Yin; Q Jackie Wu
Journal:  Med Phys       Date:  2013-12       Impact factor: 4.071

10.  Significant improvement in normal tissue sparing and target coverage for head and neck cancer by means of helical tomotherapy.

Authors:  Claudio Fiorino; Italo Dell'Oca; Alessio Pierelli; Sara Broggi; Elena De Martin; Nadia Di Muzio; Barbara Longobardi; Ferruccio Fazio; Ricardo Calandrino
Journal:  Radiother Oncol       Date:  2006-03-20       Impact factor: 6.280

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.