Abstract
OBJECTIVES: CARE Dose 4D modulates mAs through several mechanisms according to patient size and shape, whilst maintaining user-defined reference image quality on Siemens Symbia single-photon emission computed tomography (SPECT)-computed tomography (CT) systems. A 20 kg child reference was used in child protocols prior to software version VB10 and a 75 kg adult thereafter. Quality reference mAs conversion factors are estimated for delivering equivalent mAs to children between two comparable SPECT-CT systems using adult and child references for topogram-based patient-size-related dose level adaptations.Entities:
Mesh:
Year: 2021 PMID: 33079892 PMCID: PMC7725139 DOI: 10.1097/MNM.0000000000001312
Source DB: PubMed Journal: Nucl Med Commun ISSN: 0143-3636 Impact factor: 1.698
Computed tomography acquisition and reconstruction parameters for phantom scans on the Symbia T16 and Symbia Intevo Bold
| Parameter type | Parameter | Symbia T16 | Symbia Intevo Bold |
|---|---|---|---|
| Software | Software version | VA60D (syngo CT 2007E) | VB20B (syngo CT VC30) |
| Acquisition | Scan range (body part) and length (cm) | Thorax (20) | Thorax (20) |
| Abdomen (15.5) | Abdomen (15.5) | ||
| Pelvis (18.5) | Pelvis (18.5) | ||
| CARE Dose 4D child protocol reference patient | Child | Adult | |
| Quality reference mAs (mAs) | 10, 20, 30, 40, 50 | 40, 80, 120, 160, 200 | |
| Tube voltage (kV) | 80, 110 | 80, 110 | |
| Tube rotation time (s) | 0.6 | 0.6 | |
| Pitch | 1.5 | 1.5 | |
| Collimation (detector rows × mm) | 16 × 0.6 | 16 × 0.6 | |
| Reconstruction | Reconstruction method | Filtered backprojection | Filtered backprojection |
| Matrix | 512 | 512 | |
| Kernel | B31s (soft tissue) | B31s (soft tissue) | |
| Slice thickness (mm) | 5 | 5 | |
| Increment (mm) | 5 | 5 | |
| Field-of-view (mm) | 650 | 650 |
Fig. 1Quality reference mAs vs. delivered mAs for child reference (Symbia T16) with arms up (a), adult reference (Symbia Intevo Bold) with arms up (b), child reference with arms down (c) and adult reference with arms down (d).
Fig. 2Quality reference mAs conversion factors between adult and child reference patients vs. quality reference mAs with child reference for arms up (a) and down (b).
Fig. 3Slice number vs. delivered mAs ratio (child/adult reference scans) for the thorax (a), abdomen (b) and pelvis (c) with arms down at 40 mAs quality reference.
Proposed conversion factors for child to adult reference patients according to scan range, arm position and tube voltage
| Scan range | Child-to-adult quality reference mAs conversion factor | |||
|---|---|---|---|---|
| Arms up | Arms down | |||
| 80 kV | 110 kV | 80 kV | 110 kV | |
| Thorax | 3.8 | 3.8 | 4.0 | 4.3 |
| Abdomen | 3.8 | 4.1 | 3.7 | 4.0 |
| Pelvis | 4.3 | 4.6 | 3.9 | 4.2 |