| Literature DB >> 33075724 |
Mohamed Z Rashed1, Jonathan A Kopechek2, Mariah C Priddy2, Krystal T Hamorsky3, Kenneth E Palmer3, Nikhil Mittal4, Joseph Valdez4, Joseph Flynn5, Stuart J Williams6.
Abstract
Coronavirus disease (COVID-19) caused by severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2) was classified as a pandemic by the World Health Organization and has caused over 550,000 deaths worldwide as of July 2020. Accurate and scalable point-of-care devices would increase screening, diagnosis, and monitoring of COVID-19 patients. Here, we demonstrate rapid label-free electrochemical detection of SARS-CoV-2 antibodies using a commercially available impedance sensing platform. A 16-well plate containing sensing electrodes was pre-coated with receptor binding domain (RBD) of SARS-CoV-2 spike protein, and subsequently tested with samples of anti-SARS-CoV-2 monoclonal antibody CR3022 (0.1 μg/ml, 1.0 μg/ml, 10 μg/ml). Subsequent blinded testing was performed on six serum specimens taken from COVID-19 and non-COVID-19 patients (1:100 dilution factor). The platform was able to differentiate spikes in impedance measurements from a negative control (1% milk solution) for all CR3022 samples. Further, successful differentiation and detection of all positive clinical samples from negative control was achieved. Measured impedance values were consistent when compared to standard ELISA test results showing a strong correlation between them (R2=0.9). Detection occurs in less than five minutes and the well-based platform provides a simplified and familiar testing interface that can be readily adaptable for use in clinical settings.Entities:
Keywords: Antibodies; Capacitive immunosensors; Impedance spectroscopy; Label-free; SARS-CoV-2
Mesh:
Substances:
Year: 2020 PMID: 33075724 PMCID: PMC7539830 DOI: 10.1016/j.bios.2020.112709
Source DB: PubMed Journal: Biosens Bioelectron ISSN: 0956-5663 Impact factor: 10.618
Fig. 1(A) Image of ACEA Bioscience’s 96-well platform with a (B) schematic of the electrode layout. (C) Actual image of the electrodes themselves within the well and a (D) magnified image of the electrodes. Images A–D are modified from ACEA Bioscience. (E) Schematic of electrical impedance equivalent circuit model of the protein/antibody in solution.
Fig. 2(A) Four arbitrary experimental curves were selected to demonstrate a representative response for each tested concentration and a negative control. (B) Average measured peak impedance magnitude values for each concentration (n = 4); error bars represent +/- one standard deviation.
Fig. S1Measurements with greater time resolution were acquired with a different impedance analyzer (Agilent 4294A, 10 kHz, 0.5 V), demonstrating improved resolution.
Fig. 3(A) A representative impedance spectra curve for a positive and negative clinical sample. (B) Comparison between the measured peak impedance magnitude value (green) when compared to ELISA levels (red). (For interpretation of the references to color in this figure legend, the reader is referred to the web version of this article.)