Literature DB >> 33075724

Rapid detection of SARS-CoV-2 antibodies using electrochemical impedance-based detector.

Mohamed Z Rashed1, Jonathan A Kopechek2, Mariah C Priddy2, Krystal T Hamorsky3, Kenneth E Palmer3, Nikhil Mittal4, Joseph Valdez4, Joseph Flynn5, Stuart J Williams6.   

Abstract

Coronavirus disease (COVID-19) caused by severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2) was classified as a pandemic by the World Health Organization and has caused over 550,000 deaths worldwide as of July 2020. Accurate and scalable point-of-care devices would increase screening, diagnosis, and monitoring of COVID-19 patients. Here, we demonstrate rapid label-free electrochemical detection of SARS-CoV-2 antibodies using a commercially available impedance sensing platform. A 16-well plate containing sensing electrodes was pre-coated with receptor binding domain (RBD) of SARS-CoV-2 spike protein, and subsequently tested with samples of anti-SARS-CoV-2 monoclonal antibody CR3022 (0.1 μg/ml, 1.0 μg/ml, 10 μg/ml). Subsequent blinded testing was performed on six serum specimens taken from COVID-19 and non-COVID-19 patients (1:100 dilution factor). The platform was able to differentiate spikes in impedance measurements from a negative control (1% milk solution) for all CR3022 samples. Further, successful differentiation and detection of all positive clinical samples from negative control was achieved. Measured impedance values were consistent when compared to standard ELISA test results showing a strong correlation between them (R2=0.9). Detection occurs in less than five minutes and the well-based platform provides a simplified and familiar testing interface that can be readily adaptable for use in clinical settings.
Copyright © 2020 Elsevier B.V. All rights reserved.

Entities:  

Keywords:  Antibodies; Capacitive immunosensors; Impedance spectroscopy; Label-free; SARS-CoV-2

Mesh:

Substances:

Year:  2020        PMID: 33075724      PMCID: PMC7539830          DOI: 10.1016/j.bios.2020.112709

Source DB:  PubMed          Journal:  Biosens Bioelectron        ISSN: 0956-5663            Impact factor:   10.618


Introduction

The severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2) was classified as a world pandemic by the World Health Organization (WHO) due to its rapid human to human transmission nature with an estimated basic reproduction number of 2.2 (Li et al., 2020). Until reliable management of this disease is implemented there is a considerable need to develop rapid diagnostic tests that are readily available and can be integrated with current healthcare protocols to in order to monitor the spread of the virus. Immunoassays detect binding of antibodies to a specific target molecule and can be used for detecting the presence and concentration of specific antibodies to determine if a patient has been previously infected and evaluate the prevalence of the disease. The standard diagnostic method for precise and quantitative antibody detection for most common pathogen including SARS-CoV-2, is the enzyme-linked immunosorbent assay (ELISA) (Law et al., 2015); however, this detection method can take hours to perform due to sample processing and incubation steps. Over the past few decades, biosensors have emerged as a powerful tool to complement ELISA methods. Biosensors are the integration of a biorecognition element which remains in contact with the target molecule and a transduction element that confirms target binding and detects the event with a measurable electrical signal. A variety of biosensing platforms have been developed for respiratory viral infections (Saylan et al., 2019) and COVID-19 (Choi, 2020). There are many types of biosensors for antibodies (Liu and Jiang, 2017), many rely on optical, chemical, and/or electrical mechanisms to produce a measureable signal. The method discussed herein is electrochemical impedance-based sensing (EIS). EIS detection methods are cost-effective, typically label-free, and enable high precision for detection of antibodies or pathogens (Cesewski and Johnson, 2020, Leva-Bueno et al., 2020). Here, an equivalent circuit (containing combination of resistors, capacitors, and capacitive phase elements) is modeled after the sensing region and the measured impedance and phase angle data is used to extract values of the fit circuit elements to monitor changes in system behavior. Non-faradic capacitive elements typically represent the behavior of the ionic double layer at the electrode–electrolyte interface, which can provide essential information about the nature and rate of probe-target binding (Berggren et al., 2001, Daniels and Pourmand, 2007). Even though measuring the impedance across wide range of frequencies is useful for comprehensive biosensing applications, fixed or single frequency measurements can still provide rapid and improved transient information about rapid changes in double-layer capacitance due to increasing sampling (Daniels and Pourmand, 2007, Gebbert et al., 1992, Quoc et al., 2017, Peters et al., 2015). Also, operating at fixed frequencies simplifies EIS platform hardware, eliminating the need for complex frequency generators and enabling portable analysis. Capacitive immunosensing is an emerging EIS technique that focuses on sensing changes in the electric double layer (EDL) through sensing binding events between the probe immobilized on the electrode surface and suspended target (Berggren et al., 2001). The performance of capacitive immunosensors can be adjusted by modifying the electrode surface area, insulation (coating) layer, and the electrode and coating material. Highly sensitive sensors can be achieved by making thin conformal coating layer which ensure stable capacitance changes in response to probe/target binding (Castiello et al., 2019). Several studies reported the use of capacitance immunosensors for detection of pathogens (Cesewski and Johnson, 2020, Hanna et al., 2006, Luka et al., 2019, Teeparuksapun et al., 2012, Wang et al., 2019, Wang et al., 2017), thus demonstrating that it can be used as a reliable label-free sensing method. In this study, we report a non-faradic capacitive immunosensing assay using a commercially-available impedance detection system that uses specialized well-plates that have integrated sensing electrodes from ACEA Biosciences (Fig. 1). ACEA Biosciences’ 16-well plate xCELLigence system (RTCA S16) design was intended for non-invasive EIS detection of cell proliferation, morphology change, and attachment quality. Each well contains an array of specially-designed interdigitated electrodes fused to polyethylene terephthalate. The plate interface independently addresses each well to acquire single frequency measurements (10 kHz) every few seconds. Here, we demonstrate that it can be used to successfully detect SARS-CoV-2 antibodies. We hypothesize that the detected mechanism corresponds to binding kinetics between the SARS-CoV-2 spike protein receptor binding domain (RBD) and the anti-SARS-CoV-2 antibody. This approach provides quick and reliable results, is relatively compact and portable, and has the potential to be scaled up (simultaneous scanning of up to four 384 well plates is possible); thus, this approach is a good candidate for rapid screening of serum samples for SARS-CoV-2 antibodies.
Fig. 1

(A) Image of ACEA Bioscience’s 96-well platform with a (B) schematic of the electrode layout. (C) Actual image of the electrodes themselves within the well and a (D) magnified image of the electrodes. Images A–D are modified from ACEA Bioscience. (E) Schematic of electrical impedance equivalent circuit model of the protein/antibody in solution.

(A) Image of ACEA Bioscience’s 96-well platform with a (B) schematic of the electrode layout. (C) Actual image of the electrodes themselves within the well and a (D) magnified image of the electrodes. Images A–D are modified from ACEA Bioscience. (E) Schematic of electrical impedance equivalent circuit model of the protein/antibody in solution.

Experimental and methods

Methods

Fig. 1 shows ACEA Bioscience’s well platform with its integrated electrodes. The specialized well-plate is attached to an interface and is connected via USB to a laptop. Software controls the timing of measurements for each independently addressable well. Impedance measurements with respect to time, , were acquired continuously at 10 kHz and 22 mV. Results herein are reported as a change in impedance magnitude between the real-time measured impedance () with respect to a measured background impedance (), . The impedance model in Fig. 1E depicts the mechanism of the spike RBD antigen proteins, medium solution, and anti-SARS-CoV-2 antibodies. The total impedance includes the surface impedance which combines both EDL capacitance, , in parallel with double layer resistance, . The surface impedance is in series with the ionic medium resistance, , which arises from the conductivity of the ions in the solution and is not altered by the binding between probe protein and target antibodies. For simplicity, we are neglecting any insulative effects that may be present on the electrode surface and their effect on the total impedance. The total measured dynamic impedance magnitude, , is a function of in the EDL resistance and capacitance described with (Quoc et al., 2017): where . From Eq. (1), the series capacitances term, , can be extracted from the imaginary part of the impedance as the following: As can be seen from Eq. (2), is frequency dependent. At high frequencies, it will converge to , however, at low frequencies, will be dependent on , , and . Hence, protein/antibody binding events will influence the EDL resistance and capacitance at low frequencies (Gebbert et al., 1992) and thus, 10 kHz is an appropriate sensing frequency for this study.

Experimental

To initially validate our approach, a monoclonal anti-SARS-CoV-2 antibody (CR3022, Abcam, Cambridge, MA, USA) was tested. The well plates were coated with 2.5 g/mL SARS-CoV-2 spike RBD protein (obtained from BEI Resources, NR-52306) diluted in PBS, 50 L/well was added, and plates were sealed and refrigerated at 4 °C for 1–3 days. Next, wells were decanted and washed with 200 L of 0.1% PBS/Tween-20 (PBS-T) solution three times using manual pipetting. To perform the test, the wells were filled with 100 L of blocking solution (3% milk) and incubated for 1 h. The measurement wells were decanted and 100 L of 1% milk solution buffer (background buffer) was added to each well, then a background impedance measurement () was acquired. Next, half of the background buffer was removed (50 L) and 50 L of the sample solution containing CR3022 antibody (0.1 g/ml, 1.0 g/ml, or 10 g/ml) diluted in 1% milk solution was applied to the well while simultaneous impedance measurements () were continuously acquired. Tests with six human serological samples were provided (Serum samples were collected following protocols approved by the University of Louisville Institutional Review Board) and the impedance team was blinded to clinical information about the samples (i.e., whether they were positive or negative for anti-SARS-CoV-2 antibodies as validated by ELISA). Wells were prepared and background impedance measurements were acquired as previously described for the CR3022 tests. Next, 50 L was aspirated from each well and replaced with a 50 L of human serological samples (1:100 dilution factor:following recommended values in literature (Freeman et al., 2020) while continuous impedance measurements were acquired. Impedance measurements were compared to ELISA measurements on the same serological samples. ELISA was performed by coating immunoplates with of SARS-CoV-2 spike RBD protein overnight, washing three times in PBS-T, blocking for 1 h, incubating for 30 min with human serological samples diluted 1:100, washing 3 times in PBS-T, and incubating 1 h with HRP-labeled secondary anti-human IgG antibody. To perform the peroxidase reaction, wells were washed three times, TMB solution was added for 3 min, the reaction was stopped by addition of 1M HCl, and optical density of each well was measured on a platereader. Impedance measurements were exported and analyzed further in MATLAB. Impedance magnitude peaks were identified as measurements that increased at least ten times greater than background noise oscillations (¿ 78 m) when compared to its baseline value. The action of adding 50 L buffer, as will be shown later, generated an increase in impedance of approximately 50 m, which is less than our specified detection criteria. The reported peak value is the mean of the first recorded peak value and the subsequent four impedance measurements. The time between impedance measurements for the same well was approximately two seconds.

Results and discussion

Fig. 2A shows a representative impedance curve for each tested CR3022 dilution. Fig. 2B shows the reported average magnitude peak value for each dilution (n = 4 wells each). The addition of CR3022 antibody to the RBD-coated wells caused a rapid increase in measured impedance followed by a decay during the next few minutes. There is inherent error associated with the measured maximum peak value due to the timing of the recorded impedance measurements; greater sensitivity and precision is possible if data acquisition for each well occurred with a period faster than the tested two-second intervals. To demonstrate the need for faster transient measurements, we measured the addition of CR3022 (10 g/ml) and a control (i.e., decanting and adding 50 L of buffer) with continuous impedance acquisition using a different impedance analyzer system (Agilent 4294A at 10 kHz). Results (Supplemental Fig. S1) demonstrated that the measured peak rapidly decreases in less than one second before subsequently decaying more slowly; thus, increased sensitivity and accuracy is possible at higher temporal resolution.
Fig. 2

(A) Four arbitrary experimental curves were selected to demonstrate a representative response for each tested concentration and a negative control. (B) Average measured peak impedance magnitude values for each concentration (n = 4); error bars represent +/- one standard deviation.

Fig. S1

Measurements with greater time resolution were acquired with a different impedance analyzer (Agilent 4294A, 10 kHz, 0.5 V), demonstrating improved resolution.

Fig. 3A shows a representative impedance curve for two positive and one negative serological sample. The positive impedance signature resembles closely to the CR3022 measurements in that there was a sharp peak followed by a slower decay. The two positive serological samples were from two patients with different antibody levels; similar to the CR3022 levels, impedance spectroscopy can be used to quantify antibody levels in serological samples. Fig. 3B compares the measured impedance peak value to ELISA levels. These results show a clear correlation between impedance values and the concentration of antibodies within the sample.
Fig. 3

(A) A representative impedance spectra curve for a positive and negative clinical sample. (B) Comparison between the measured peak impedance magnitude value (green) when compared to ELISA levels (red). (For interpretation of the references to color in this figure legend, the reader is referred to the web version of this article.)

(A) Four arbitrary experimental curves were selected to demonstrate a representative response for each tested concentration and a negative control. (B) Average measured peak impedance magnitude values for each concentration (n = 4); error bars represent +/- one standard deviation. The rapid changes in total impedance in the presented experiments is due to the association and disassociation of the protein/antibody which is happening within few seconds after the introduction of the sample (Kulin et al., 2002). The binding events can alter the double layer charge distribution, shape and dimension, which in turns influence the conductivity and dielectric constant of the adsorbed layer above the electrode, thus altering the overall impedance (Berggren et al., 2001, Bergveld, 1991). The bound layer manifest itself as an extra insulator layer above the electrode which decreases the measured capacitance (Bergveld, 1991). Moreover, dipoles in the bound antigen headgroup can contribute to the measured capacitance because they affect the dielectric constant () of the surface impedance. During the binding event, the dielectric constant of the surface layer drops from (for PBS) to  (Daniels and Pourmand, 2007, Sondag-Huethorst and Fokkink, 1995, Świetlow et al., 1992) which subsequently decreases the capacitance as well, and therefore, increases the overall measured system impedance. It is noted that these dipoles would not react or contribute to the total impedance if the applied signal frequency exceeds 1 MHz (Bordi et al., 2004, Feldman et al., 2003), which rationalize the operation at a low signal frequency (10 kHz). (A) A representative impedance spectra curve for a positive and negative clinical sample. (B) Comparison between the measured peak impedance magnitude value (green) when compared to ELISA levels (red). (For interpretation of the references to color in this figure legend, the reader is referred to the web version of this article.) The limit of detection using this approach is limited by hardware noise and variations in sample handling. Although the lower limit of detection was not demonstrated here, these experiments demonstrate that this implementation of EIS can detect clinically relevant antibody concentrations. The ability to automize the system can offer high throughput operation. Impedance systems also lend themselves to be portable, including potential smartphone-based analytical systems. Further improvements can be made to enhance sensing and throughput of the platform. For example, decreasing the spacing between interdigitated electrodes will reduce the influence of . Also, electrodes with a thin conformal coating of insulative material would enhance the stability and sensitivity of the capacitance measurement (Castiello et al., 2019). Future tests will determine if measurements acquired at other frequencies and/or voltages (i.e., other than 10 kHz, 22 mV) will further improve platform performance. There are alternative electrochemical detection schemes that could be used for antibody detection. Field-effect transistor-based biosensors (Gaurav and Shukla, 2020, Seo et al., 2020) offer a good approach for detection of the antibodies through the induced variations in source–drain channel conductivity that arise from the electric field of the sample environment due to the binding of the target with the bioreceptors immobilized on the metal/polymer gate. However, such platforms are either expensive, require complicated fabrication methods, or need sophisticated equipment to operate. Further, a limitation in applying FET-based sensors is that their performance may be critically obstructed by the presence of multiple ligands and proteins in serum samples. Moreover, in terms of electrical detection, the detection of biomolecules is severely hindered by ionic screening effect caused by high ionic strength of physiological environment (Vu and Chen, 2019). Antibodies against SARS-CoV-2 are generally detected using either the recombinant spike protein or the smaller RBD portion of the spike protein. This study only tested detected binding of antibodies to RBD but similar results would be expected with the recombinant spike protein as well (Premkumar et al., 2020). Coating the wells with anti-SARS-CoV-2 antibodies instead of spike RBD antigen may enable rapid EIS detection of viral particles in patient samples, although further testing is needed to determine the limit of detection for that approach. There is a need for rapid patient testing due to the highly contagious nature of SARS-CoV-2 particularly when compared to viruses from the same family including, Middle East respiratory syndrome coronavirus (MERS-CoV) (fatality rate: 34%) and severe acute respiratory syndrome coronavirus (SARS-CoV) (mortality rate: 11%).

Conclusion

In this study demonstrated the feasibility of using a quantitative EIS technique with commercially available equipment for rapid and accurate detection of SARS-CoV-2 antibodies at clinically relevant concentrations. This approach could enable rapid screening of patient samples, expanded serological surveys to assess anti-SARS-CoV-2 antibody levels in the community, and potentially enhance assessment of vaccine activity.

CRediT authorship contribution statement

Mohamed Z. Rashed: Methodology, Investigation, Data curation, Writing - original draft. Jonathan A. Kopechek: Conceptualization, Methodology, Validation, Resources, Formal analysis, Funding acquisition, Writing - review & editing. Mariah C. Priddy: Data curation, Writing - review & editing. Krystal T. Hamorsky: Conceptualization, Resources, Methodology, Writing - review & editing. Kenneth E. Palmer: Resources, Conceptualization, Methodology, Writing - review & editing. Nikhil Mittal: Resources. Joseph Valdez: Resources. Joseph Flynn: Resources, Writing - review & editing. Stuart J. Williams: Conceptualization, Methodology, Validation, Resources, Formal analysis, Supervision, Project administration, Visualization, Writing - review & editing.

Declaration of Competing Interest

The authors declare that they have no known competing financial interests or personal relationships that could have appeared to influence the work reported in this paper.
  20 in total

1.  Real-time measurement of spontaneous antigen-antibody dissociation.

Authors:  Simone Kulin; Rani Kishore; Joseph B Hubbard; Kristian Helmerson
Journal:  Biophys J       Date:  2002-10       Impact factor: 4.033

Review 2.  Electrochemical biosensors for pathogen detection.

Authors:  Ellen Cesewski; Blake N Johnson
Journal:  Biosens Bioelectron       Date:  2020-04-12       Impact factor: 10.618

3.  Capacitive immunosensor for the detection of host cell proteins.

Authors:  Kosin Teeparuksapun; Martin Hedström; Proespichaya Kanatharana; Panote Thavarungkul; Bo Mattiasson
Journal:  J Biotechnol       Date:  2011-11-09       Impact factor: 3.307

Review 4.  Human stem cell-derived cardiomyocytes in cellular impedance assays: bringing cardiotoxicity screening to the front line.

Authors:  Matthew F Peters; Sarah D Lamore; Liang Guo; Clay W Scott; Kyle L Kolaja
Journal:  Cardiovasc Toxicol       Date:  2015-04       Impact factor: 3.231

Review 5.  Rapid methods for the detection of foodborne bacterial pathogens: principles, applications, advantages and limitations.

Authors:  Jodi Woan-Fei Law; Nurul-Syakima Ab Mutalib; Kok-Gan Chan; Learn-Han Lee
Journal:  Front Microbiol       Date:  2015-01-12       Impact factor: 5.640

Review 6.  Development of Point-of-Care Biosensors for COVID-19.

Authors:  Jane Ru Choi
Journal:  Front Chem       Date:  2020-05-27       Impact factor: 5.221

7.  Early Transmission Dynamics in Wuhan, China, of Novel Coronavirus-Infected Pneumonia.

Authors:  Qun Li; Xuhua Guan; Peng Wu; Xiaoye Wang; Lei Zhou; Yeqing Tong; Ruiqi Ren; Kathy S M Leung; Eric H Y Lau; Jessica Y Wong; Xuesen Xing; Nijuan Xiang; Yang Wu; Chao Li; Qi Chen; Dan Li; Tian Liu; Jing Zhao; Man Liu; Wenxiao Tu; Chuding Chen; Lianmei Jin; Rui Yang; Qi Wang; Suhua Zhou; Rui Wang; Hui Liu; Yinbo Luo; Yuan Liu; Ge Shao; Huan Li; Zhongfa Tao; Yang Yang; Zhiqiang Deng; Boxi Liu; Zhitao Ma; Yanping Zhang; Guoqing Shi; Tommy T Y Lam; Joseph T Wu; George F Gao; Benjamin J Cowling; Bo Yang; Gabriel M Leung; Zijian Feng
Journal:  N Engl J Med       Date:  2020-01-29       Impact factor: 176.079

8.  A sensitive DNA capacitive biosensor using interdigitated electrodes.

Authors:  Lei Wang; Milena Veselinovic; Lang Yang; Brian J Geiss; David S Dandy; Tom Chen
Journal:  Biosens Bioelectron       Date:  2016-09-02       Impact factor: 10.618

Review 9.  Field-Effect Transistor Biosensors for Biomedical Applications: Recent Advances and Future Prospects.

Authors:  Cao-An Vu; Wen-Yih Chen
Journal:  Sensors (Basel)       Date:  2019-09-28       Impact factor: 3.576

Review 10.  A review on impedimetric immunosensors for pathogen and biomarker detection.

Authors:  J Leva-Bueno; Sally A Peyman; P A Millner
Journal:  Med Microbiol Immunol       Date:  2020-04-03       Impact factor: 3.402

View more
  53 in total

1.  Diagnosis of COVID-19, vitality of emerging technologies and preventive measures.

Authors:  Muhammad Asif; Yun Xu; Fei Xiao; Yimin Sun
Journal:  Chem Eng J       Date:  2021-05-07       Impact factor: 13.273

2.  Portable Tools for COVID-19 Point-of-Care Detection: A Review.

Authors:  Elga F Saki; Samuel A Setiawan; Dedy H B Wicaksono
Journal:  IEEE Sens J       Date:  2021-09-07       Impact factor: 3.301

3.  Rapid SARS-CoV-2 diagnosis using disposable strips and a metal-oxide-semiconductor field-effect transistor platform.

Authors:  Chan-Wen Chiu; Minghan Xian; Jenna L Stephany; Xinyi Xia; Chao-Ching Chiang; Fan Ren; Cheng-Tse Tsai; Siang-Sin Shan; Yu-Te Liao; Josephine F Esquivel-Upshaw; Santosh R Rananaware; Long T Nguyen; Nicolas C Macaluso; Piyush K Jain; Melanie N Cash; Carla N Mavian; Marco Salemi; Marino E Leon; Chin-Wei Chang; Jenshan Lin; Stephen J Pearton
Journal:  J Vac Sci Technol B Nanotechnol Microelectron       Date:  2022-02-09

4.  Detection of SARS-CoV-2 in COVID-19 Patient Nasal Swab Samples Using Signal Processing.

Authors:  Mahmoud Al Ahmad; Lillian J A Olule; Mohammed Meetani; Farrukh Amin Sheikh; Rahima Al Blooshi; Neena G Panicker; Farah Mustafa; Tahir A Rizvi
Journal:  IEEE J Sel Top Signal Process       Date:  2021-12-09       Impact factor: 7.695

5.  Debulking different Corona (SARS-CoV-2 delta, omicron, OC43) and Influenza (H1N1, H3N2) virus strains by plant viral trap proteins in chewing gums to decrease infection and transmission.

Authors:  Henry Daniell; Smruti K Nair; Hancheng Guan; Yuwei Guo; Rachel J Kulchar; Marcelo D T Torres; Md Shahed-Al-Mahmud; Geetanjali Wakade; Yo-Min Liu; Andrew D Marques; Jevon Graham-Wooten; Wan Zhou; Ping Wang; Sudheer K Molugu; William R de Araujo; Cesar de la Fuente-Nunez; Che Ma; William R Short; Pablo Tebas; Kenneth B Margulies; Frederic D Bushman; Francis K Mante; Robert P Ricciardi; Ronald G Collman; Mark S Wolff
Journal:  Biomaterials       Date:  2022-07-18       Impact factor: 15.304

6.  Ultra-precise label-free nanosensor based on integrated graphene with Au nanostars toward direct detection of IgG antibodies of SARS-CoV-2 in blood.

Authors:  Seyyed Alireza Hashemi; Sonia Bahrani; Seyyed Mojtaba Mousavi; Navid Omidifar; Nader Ghaleh Golab Behbahan; Mohammad Arjmand; Seeram Ramakrishna; Kamran Bagheri Lankarani; Mohsen Moghadami; Mansoureh Shokripour; Mohammad Firoozsani; Wei-Hung Chiang
Journal:  J Electroanal Chem (Lausanne)       Date:  2021-05-08       Impact factor: 4.464

7.  Fast SARS-CoV-2 virus detection using disposable cartridge strips and a semiconductor-based biosensor platform.

Authors:  Minghan Xian; Hao Luo; Xinyi Xia; Chaker Fares; Patrick H Carey; Chan-Wen Chiu; Fan Ren; Siang-Sin Shan; Yu-Te Liao; Shu-Min Hsu; Josephine F Esquivel-Upshaw; Chin-Wei Chang; Jenshan Lin; Steven C Ghivizzani; Stephen J Pearton
Journal:  J Vac Sci Technol B Nanotechnol Microelectron       Date:  2021-05-18

Review 8.  Biosensing strategies for the electrochemical detection of viruses and viral diseases - A review.

Authors:  Laís Canniatti Brazaca; Pãmyla Layene Dos Santos; Paulo Roberto de Oliveira; Diego Pessoa Rocha; Jéssica Santos Stefano; Cristiane Kalinke; Rodrigo Alejandro Abarza Muñoz; Juliano Alves Bonacin; Bruno Campos Janegitz; Emanuel Carrilho
Journal:  Anal Chim Acta       Date:  2021-03-12       Impact factor: 6.911

Review 9.  Emerging Electrochemical Biosensing Trends for Rapid Diagnosis of COVID-19 Biomarkers as Point-of-Care Platforms: A Critical Review.

Authors:  Sasya Madhurantakam; Sriram Muthukumar; Shalini Prasad
Journal:  ACS Omega       Date:  2022-04-05

Review 10.  Nanomaterials to tackle the COVID-19 pandemic.

Authors:  Parsa Pishva; Meral Yüce
Journal:  Emergent Mater       Date:  2021-02-12
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.