| Literature DB >> 35582704 |
Mahmoud Al Ahmad1,2, Lillian J A Olule1, Mohammed Meetani3, Farrukh Amin Sheikh4, Rahima Al Blooshi4, Neena G Panicker5, Farah Mustafa6,2, Tahir A Rizvi7,2.
Abstract
This work presents an opto-electrical method that measures the viral nucleocapsid protein and anti-N antibody interactions to differentiate between SARS-CoV-2 negative and positive nasal swab samples. Upon light exposure of the patient nasal swab sample mixed with the anti-N antibody, charge transfer (CT) transitions within the altered protein folds are initiated between the charged amino acids side chain moieties and the peptide backbone that play the role of donor and acceptor groups. A Figure of Merit (FOM) was introduced to correlate the relative variations of the samples with and without antibody at two different voltages. Empirically, SARS-CoV-2 in patient nasal swab samples was detected within two minutes, if an extracted FOM threshold of >1 was achieved; otherwise, the sample wasconsidered negative.Entities:
Keywords: COVID-19; SARS-CoV-2; light intensity; nucleocapsid protein; optical detection
Year: 2021 PMID: 35582704 PMCID: PMC9088791 DOI: 10.1109/JSTSP.2021.3134073
Source DB: PubMed Journal: IEEE J Sel Top Signal Process ISSN: 1932-4553 Impact factor: 7.695