Clinical impact of angiographically insignificant suboptimal poststent findings detected by optical coherence tomography after drug-eluting stent implantation.

Jae Young Cho1, Hyungdon Kook2, Cheol Woong Yu2.   

Abstract

BACKGROUND: Although optical coherence tomography (OCT)-detected suboptimal findings (SF-OCT) such as malapposition, edge dissection, tissue protrusion, thrombus and small minimal stent area (MSA) are frequently observed after the implantation of drug-eluting stents (DES), their clinical implications are controversial. HYPOTHESIS: Clinical outcomes may differ between patients with SF-OCTs and without SF-OCTs after DES implantation.
METHODS: A total of 576 patients undergoing OCT analysis after DES implantation were divided into SF-OCT group (n = 379, 379 lesions) and No SF-OCT group (n = 197, 197 lesions). The study population had no significant abnormal finding in final angiography. Quantification was performed for each SF-OCT. The incidences of major adverse cardiovascular events (MACE: all-cause death, non-fatal myocardial infarction, target vessel revascularization, and stent thrombosis) were compared between the two groups. A median follow-up duration was 21.5 months.
RESULTS: Among 379 patients with SF-OCT, 32.4% had multiple SF-OCTs. Malapposition (32.1%, IQR of maximal depth 315-580 μm) was the most frequent, followed by small MSA (31.6%), edge dissection (12.5%, IQR of maximal flap of opening 0.27-0.52 mm), thrombus (7.6%, IQR of diameter 1.31-1.97mm) and tissue protrusion (6.8%, IQR of diameter 1.05-1.67 mm). The SF-OCT group showed smaller stent diameter and longer stent length, and lower in-stent lumen expansion rate. The incidence of MACE did not differ between the two groups (3.0% for No SF-OCT vs. 5.0% for SF-OCT; HR 1.601; 95% CI 0.639 to 4.011; P = 0.310).
CONCLUSIONS: The presence of angiographically insignificant SF-OCTs were not associated with clinical outcomes in this study.

Entities:  

Mesh:

Year:  2020        PMID: 33075066      PMCID: PMC7571690          DOI: 10.1371/journal.pone.0240860

Source DB:  PubMed          Journal:  PLoS One        ISSN: 1932-6203            Impact factor:   3.240


Introduction

Recent reports of optical coherence tomography (OCT) analysis after drug-eluting stent (DES) implantation revealed that suboptimal findings detected by OCT (SF-OCT) had a high prevalence, while no significant abnormal findings were noted on conventional coronary angiography. The SF-OCTs have been reported to be malapposition, tissue protrusion (TP), edge dissection (ED), thrombus and small minimal stent area (MSA) in previous studies [1-6]. However, their clinical implications are controversial. Clarifying the relationships between SF-OCTs and adverse clinical outcomes could help improve clinical outcomes and avoid unnecessary additional procedures after percutaneous coronary intervention (PCI). Recent studies of SF-OCT focused predominantly on the natural course of these findings, demonstrating that most of them except for malapposition improved or resolved within one year [1-5]. However, data regarding its impact on relevant clinical outcomes is controversial. In addition, there has not been comparative studies with patients without SF-OCT. Therefore, we compared clinical outcomes after DES implantation between patients with or without suboptimal findings differentiated by OCT.

Methods

Study patients

The Optical Coherence Tomography Registry of Korea University Anam Hospital is a single-center registry of patients undergoing OCT imaging of coronary arteries (ClinicalTrials.gov Trial Number: NCT02966262; URL: https://clinicaltrials.gov/ct2/show/NCT02966262?term=NCT02966262&draw=2&rank=1). A total of 576 patients in the OCT registry was retrospectively reviewed for the study at Cardiovascular Center, Korea University Anam Hospital from January 2011 to May 2013. Patients were included into the study based on the following criteria: 1) Patients who had only single lesion to intervene; 2) Patients who underwent sequential OCT immediately after DES implantation or after adjuvant procedures; 3) No significant abnormal finding in final coronary angiography. Patients were allocated to either SF-OCT group or No SF-OCT group, based on whether they exhibited at least one SF-OCT after OCT analysis. After stent implantation, all patients received dual antiplatelet therapy unless contraindicated. The therapy was maintained for at least 12 months. Demographic data, including sex, age, body mass index, comorbidities, prescribed drugs, laboratory data, and clinical presentation, were collected and compared between the two groups. Left ventricular dysfunction was defined as a left ventricular ejection fraction <45%. The primary endpoint was major adverse cardiovascular events (MACE), defined as a composite of cardiac death, non-fatal myocardial infarction (MI), the need for repeated target vessel revascularization (TVR), and stent thrombosis. TVR was ischemia-driven. The secondary endpoint was each component event of MACE. Information of clinical outcomes was collected by the retrospective review of the chart. Collecting data of clinical information and outcomes was performed blindly to angiographic data and OCT findings. This study was approved by the Korea University Hospital Institute Review Board (IRB No. 2016AN0095), and the informed consent was waived due to retrospective study design. This study also complied with the Declaration of Helsinki.

Angiographic analysis

Coronary angiograms were analyzed using a computer-based Telecardiology system, version 2.02 (Medcon Inc., Tel Aviv, Israel) by three radiologic technicians who were blinded to the study purpose. The reference diameter, minimal luminal diameter (MLD), percentage of stenosis, and lesion length were evaluated from diastolic frames using guided catheter magnification calibration in a single, matched view with a computerized quantitative analyzer using a caliper. The average diameter of normal segments proximal and distal to the treated lesion was used as the reference diameter.

OCT acquisition

OCT examination and analysis were performed immediately after stent implantation (LightLab Imaging Inc., Ilumien Offline review workstation, Ver E.4.1, MA, USA). Using a 0.014” guide wire, an OCT imaging catheter (C7 DragonflyTM, LightLab Imaging Inc., MA, USA) was advanced into the distal end of the DES implantation site. The entire length of the stent was imaged with an automatic pullback device moving at 15 mm/s. The whole stent was clearly visualized on each OCT image; in-segment cross-sectional views were also obtained.

OCT analysis

All baseline OCT images were reviewed by an independent observer who was blinded to the clinical presentation, lesion, and procedural characteristics. The analysis encompassed the intra-stent segment, defined by the first and the last cross-sections with a visible strut, and the adjacent vessel segments 5 mm proximal and distal to the stent, defined as edge segments. Quantification was performed for each finding. TP was defined as tissue protruding between adjacent stent struts toward the lumen, with or without disruption of luminal vessel surface continuity (Fig 1A) [7]. TP was distinguished from thrombus by visualizing behind the plaque. TP length was defined as the longest diameter of protruded tissue. The area of TP was also measured (tissue protrusion area).
Fig 1

Representative images of suboptimal findings detected on optical coherence tomography.

(A) Tissue protrusion (white arrows). Maximal length (white two-way arrows) and area (red line) of tissue protrusion. (B) Edge dissection (white arrow). Maximal flap opening (yellow two-way arrow), maximal length (white two-way arrow) and arc (blue line) of edge dissection. (C) Malapposition (white arrows). Lumen area (yellow broken line), stent area (red broken line) and maximal depth (white two-way arrow) of malapposition. Malapposition area was calculated by subtracting stent area from lumen area. (D) Thrombus (white arrow). Diameter (red two-way arrow) of thrombus.

Representative images of suboptimal findings detected on optical coherence tomography.

(A) Tissue protrusion (white arrows). Maximal length (white two-way arrows) and area (red line) of tissue protrusion. (B) Edge dissection (white arrow). Maximal flap opening (yellow two-way arrow), maximal length (white two-way arrow) and arc (blue line) of edge dissection. (C) Malapposition (white arrows). Lumen area (yellow broken line), stent area (red broken line) and maximal depth (white two-way arrow) of malapposition. Malapposition area was calculated by subtracting stent area from lumen area. (D) Thrombus (white arrow). Diameter (red two-way arrow) of thrombus. ED was defined as disruption of the luminal vessel surface in the edge segments (within 5 mm proximal and distal to the stent, with no visible struts) without flow limiting (Fig 1B) [7]. Maximum flap length (from its tip to the joint point with the vessel wall), maximum flap opening (distance from the flap tip to the lumen contour along a line projected through the gravitational center of the lumen), arc of circumferential extension, and longitudinal flap length were measured in the cross-sectional images [8]. Stent malapposition was defined as a clear separation between at least one stent strut and the strut reflection, in addition to a vessel wall +20 μm greater than the actual stent thickness on OCT images (Fig 1C) [7, 9]. The maximum distances from the endoluminal surface of the strut to the vessel wall (maximum depth), the malapposition area (difference between lumen area and stent area), and the longitudinal distance of malapposition were measured. Intracoronary thrombus was defined as a mass protruding beyond the stent strut into the lumen with significant attenuation behind the mass [10, 11]. The presence of a thrombus was assessed quantitatively, and the diameter of the visible thrombus was recorded (Fig 1D). We defined a small MSA as an in-stent minimum area <4.5 mm2 and underexpansion as MSA <80% of the average reference lumen area [6, 12]. For clinical impact, only previously noted significant findings were used for analysis and the following factors were considered significant SF-OCTs [3, 6]: TP: diameter ≥0.5mm and protrusion area/stent area at site of tissue protrusion ≥10% ED: maximum flap opening ≥0.2mm Malapposition: maximum depth ≥200μm Thrombus: diameter ≥0.5mm and thrombus area/stent area at site of thrombus ≥10% Small MSA: in-stent minimum area <4.5 mm2 If adjuvant procedures, such as balloon dilation or stent implantation, were performed based on OCT by physician’s preference, SF-OCTs were analyzed after the adjuvant procedure. OCT-based quantitative measurements were assessed according to well-standardized methods.

Statistical analysis

Data are expressed as mean± standard deviation for continuous variables, whereas data for categorical variables are expressed as number and percentage of patients. The chi-square test was used to compare categorical variables. Event rates were estimated using Kaplan-Meier survival analysis at 5 years, and hazard ratios (HR) were generated using Cox regression analysis. Because patients may have experienced more than 1 MACE, each patient was assessed until the occurrence of his or her first event and only once during analysis. To determine the association between clinical characteristics and outcomes, univariate and multivariate Cox regression analyses were performed for the entire population. The Cox regression model included the following variables, which were considered to be related with clinical outcomes: age, body mass index, hypertension, diabetes mellitus, clinical diagnosis, smoking history, lipid profile, peak CK-MB, use of renin-angiotensin system blocker and beta blocker, stent diameter, stent length, MSA, proximal and distal reference area, underexpansion, SF-OCT, components and quantitative measurements of SF-OCT. Additionally, selected variables were tested for logistic univariate regression associated with SF-OCTs; if P-value <0.05, they were simultaneously entered into a logistic multivariate regression model to identify independent predictors of SF-OCTs and to calculate their adjusted odds ratios (OR) with associated 95% confidence intervals (CI). The logistic regression model included the following variables, which were considered to be related with SF-OCTs: age, male sex, hypertension, diabetes mellitus, clinical diagnosis, lipid profile, peak CK-MB, stent diameter, stent length, MSA, proximal reference area, distal reference area, underexpansion and adjuvant procedure. SPSS version 20.0 (IBM SPSS Statistics, IBM Corporation, Armonk, New York) was used for all analyses. A P-value <0.05 was considered statistically significant.

Results

The study protocol is diagrammed in Fig 2. Overall, a total of 576 patients with 576 lesions who underwent OCT immediately after DES implantation was analyzed. Adjuvant procedures after OCT examination were performed in 132 patients (132 lesions). Among overall study population, 379 patients (379 lesions) showed one or more SF-OCT (SF-OCT group) and 197 patients (197 lesions) showed no SF-OCT in the final OCT examination (No SF-OCT group). The baseline characteristics were not different between the No SF-OCT group versus the SF-OCT group (Table 1). Table 2 shows the quantitative coronary angiography results. Smaller stent diameter (2.98±0.42 vs. 3.27±0.43 mm, P <0.001) and longer stent length was used (23.24±7.56 vs. 20.53±7.20 mm, P <0.001) in the SF-OCT group. MSA (5.27 ± 2.02 vs. 6.85 ± 2.08 mm2, P<0.001) and percentage of in-stent lumen expansion (72.53 ± 11.06 vs. 78.95 ± 9.24%, P<0.001) were also smaller in the SF-OCT group. The incidence of underexpansion was more frequent in the SF-OCT group (77.3% vs. 51.3%, P<0.001).
Fig 2

Study flow chart.

PCI: percutaneous coronary intervention DES: drug-eluting stent; OCT: optical coherence tomography; CAG: coronary angiography; SF-OCT: optical coherence tomography detected suboptimal findings.

Table 1

Baseline patient characteristics.

VariableNo SF-OCTSF-OCTP-value
(197 patients)(379 patients)
Age (year)61.2 ± 11.862.0 ± 10.90.415
Male sex (%)142 (72.1)289 (76.3)0.312
Body mass index (kg/m2)24.7 ± 3.124.9 ± 3.10.775
Smoking, n (%)
    Previous22 (11.2)66 (17.4)0.057
    Current57 (28.9)121 (31.9)
Comorbidity
    Hypertension, n (%)109 (55.3)240 (63.3)0.072
    Diabetes mellitus, n (%)59 (29.9)145 (38.3)0.054
    CVA history, n (%)6 (3.0)15 (4.0)0.647
    Left ventricular dysfunction, n (%)12 (6.1)26 (6.9)0.860
    Previous PCI history, n (%)23 (11.7)44 (11.6)1.000
Laboratory Data
    White blood cell count (x103/uL)8.2 ± 3.57.8 ± 3.30.274
    Creatinine (mg/dL)1.12 ± 1.311.21 ± 1.480.466
    Total cholesterol (mg/dL)168.6 ± 66163.7 ± 67.40.405
    Triglyceride (mg/dL)140.6 ± 130.1138.3 ± 103.80.816
    HDL-cholesterol (mg/dL)41.1 ± 13.939.0 ± 15.60.117
    LDL-cholesterol(mg/dL)110.8 ± 48.5102.4 ± 53.40.066
    Peak CK-MB (ng/mL)47.7 ± 98.542.9 ± 92.70.580
    hs-CRP (mg/dL)3.4 ± 13.84.8 ± 18.80.363
    ESR (mm/hr)7.3 ± 8.58.6 ± 13.70.174
Drug
    Aspirin, n (%)195 (99.0)377 (99.5)0.609
    Clopidogrel, n (%)195 (99.0)378 (99.7)0.271
    Statin, n (%)188 (95.4)366 (96.6)0.499
    RAS blocker, n (%)105 (53.3)205 (54.1)0.387
    Beta blocker, n (%)93 (47.2)194 (51.2)0.381
    Calcium channel blocker, n (%)56 (28.4)112 (29.6)0.847
Clinical presentation
    Stable angina, n (%)56 (28.4)92 (24.3)0.600
    Unstable angina, n (%)83 (42.1)187 (49.3)
    NSTEMI, n (%)26 (13.2)45 (11.9)
    STEMI, n (%)32 (16.2)55 (14.5)

SF-OCT: optical coherence tomography detected suboptimal findings; CVA: cerebrovascular accident; PCI: percutaneous coronary intervention; RAS: renin–angiotensin system; NSTEMI: non–ST-segment elevation myocardial infarction; STEMI: ST-segment elevation myocardial infarction

Table 2

Angiographic and procedural data.

VariableNo SF-OCT (197 lesions)SF-OCT (379 lesions)P-value
Vessel, n (%)
    LAD119 (60.4)251 (66.2)0.098
    LCX16 (8.1)39 (10.3)
    RCA56 (28.4)86 (22.7)
    Left main5 (2.5)3 (0.8)
Stent type, n (%)
    Biolimus A9-eluting stent93 (47.2)179 (47.2)0.404
    Everolimus-eluting stent98 (49.7)195 (51.5)
    Sirolimus-eluting stent1 (0.5)2 (0.5)
    Bare-metal stent5 (2.5)3 (0.8)
Stent diameter (mm)3.27 ± 0.432.98 ± 0.42<0.001
Stent length (mm)20.53 ± 7.2023.24 ± 7.56<0.001
Adjuvant procedure
    Adjuvant dilatation, n (%)37 (18.8)95 (25.1)0.095
    Adjuvant balloon diameter (mm)3.55 ± 0.493.39 ± 0.610.137
    Adjuvant balloon length (mm)11.14 ± 4.3912.03 ± 3.990.262
Quantitative Coronary Analysis
Baseline
    RD (mm)3.43 ± 0.433.00 ± 0.610.013
    MLD (mm)0.78 ± 0.450.64 ± 0.380.254
    Diameter stenosis (%)67.99 ± 27.7864.84 ± 31.860.718
    Lesion length (mm)19.5 ± 8.522.2 ± 10.40.287
Post-procedure
    RD (mm)3.63 ± 0.463.34 ± 0.580.092
    MLD (mm)3.35 ± 0.443.00 ± 0.460.015
    Diameter stenosis (%)5.45 ± 4.016.65 ± 5.350.402
OCT Quantitative Coronary Analysis
    MSA (mm2)6.85 ± 2.085.27 ± 2.02<0.001
    Proximal RA (mm2)9.38 ± 2.918.06 ± 3.23<0.001
    Distal RA (mm2)8.11 ± 2.536.61 ± 2.65<0.001
    Expansion rate (%)78.95 ± 9.2472.53 ± 11.06<0.001
    Underexpansion, n (%)101 (51.3)293 (77.3)<0.001

SF-OCT: optical coherence tomography detected suboptimal findings; LAD: left anterior descending artery; LCX: left circumflex artery; RCA: right coronary artery; RD: reference diameter; MLD: minimal lumen diameter; OCT: optical coherence tomography; RA: reference area; MSA: minimal stent area

Study flow chart.

PCI: percutaneous coronary intervention DES: drug-eluting stent; OCT: optical coherence tomography; CAG: coronary angiography; SF-OCT: optical coherence tomography detected suboptimal findings. SF-OCT: optical coherence tomography detected suboptimal findings; CVA: cerebrovascular accident; PCI: percutaneous coronary intervention; RAS: renin–angiotensin system; NSTEMI: non–ST-segment elevation myocardial infarction; STEMI: ST-segment elevation myocardial infarction SF-OCT: optical coherence tomography detected suboptimal findings; LAD: left anterior descending artery; LCX: left circumflex artery; RCA: right coronary artery; RD: reference diameter; MLD: minimal lumen diameter; OCT: optical coherence tomography; RA: reference area; MSA: minimal stent area

SF-OCTs after DES implantation

The prevalence of SF-OCTs was 65.8% (379/576). The incidences of the individual SF-OCT were as follows: 185 malapposition (32.1%), 182 small MSA (31.6%), 72 ED (12.5%), 44 thrombi (7.6%), and 39 TP (6.8%) (Fig 3A). Of 379 lesions with SF-OCTs, 256 (67.5%) had one SF-OCT, 100 (26.4%) had two, 23 (6.1%) had three and none had four or five (Fig 3B). Adjuvant procedures were performed after index PCI in 132 cases (22.9%) by physician’s preference, mainly by visualized significant underexpansion or malapposition in angiography or OCT. After the adjuvant procedures, 37 patients were assigned to the No SF-OCT group but 95 patients still showed SF-OCT despite of the adjuvant procedure, thereby assigned to the SF-OCT group (Fig 2). Table 3 shows the quantitative assessment of each individual SF-OCT. Average diameter of thrombus was 1.62 mm and most of the cases showed 1 thrombus in OCT finding. In malapposition, average maximal depth was 469 μm with average malapposition area of 2.07 mm2 and length of 2.43 mm. In cases of tissue protrusion, average of the diameter was 1.35mm and average area was 1.03mm2. Proximal edge dissection (n = 40) was more frequent than distal (n = 32), with similar findings of maximal flap opening, maximal flap length, longitudinal flap length and arc of dissection. The intra-observer κ coefficient for SF-OCT was 0.96, and the interobserver κ coefficient was 0.93. The logistic univariate model showed that diabetes mellitus, stent diameter, stent length, proximal reference area, distal reference area, underexpansion and adjuvant procedure were correlated with SF-OCT. The logistic multivariate model identified stent diameter (OR 0.212; 95% CI 0.136–0.328; P<0.001), and underexpansion (OR 3.244; 95% CI 2.197–4.789; P<0.001) as independent predictors of SF-OCT (Table 4).
Fig 3

Incidences of suboptimal findings detected by optical coherence tomography (OCT).

(A) Incidence of each component. (B) Cumulative incidence of suboptimal optical coherence tomography-detected findings.

Table 3

Quantification data of suboptimal optical coherence tomography-detected findings.

VariablesAverageQ1MedianQ3
Thrombus (n = 116)Number (n)1.27111
Longitudinal length (mm)0.790.480.731
Diameter (mm)1.20.861.161.55
Area (mm2)0.560.290.510.68
Significant thrombus (n = 44)Number (n)1.05111
Longitudinal length (mm)1.040.701.001.40
Diameter (mm)1.621.311.601.97
Area (mm2)0.860.590.741.10
Malapposition (n = 188)Maximal depth (μm)465310410578
Area (mm2)2.051.181.712.51
Length (mm)2.421.32.23.18
Significant malapposition (n = 185)Maximal depth (μm)469315410580
Area (mm2)2.071.211.722.53
Length (mm)2.431.302.203.15
Tissue protrusion (n = 263)Length (mm)0.890.620.841.1
Area (mm2)0.370.160.250.42
Significant tissue protrusion (n = 39)Length (mm)1.351.051.301.67
Area (mm2)1.030.560.691.02
Edge dissection (n = 100)Maximal flap opening (mm)0.360.20.310.46
Maximal flap length (mm)0.80.380.71.02
Longitudinal flap length (mm)1.730.931.52.3
Arc (°)2814.822.339.5
Proximal (n = 55)Maximal flap opening (mm)0.40.20.330.52
Maximal flap length (mm)0.880.470.731.07
Longitudinal flap length (mm)1.710.71.22.3
Arc (°)25.214.220.331.9
Distal (n = 45)Maximal flap opening (mm)0.320.20.290.41
Maximal flap length (mm)0.690.320.581.01
Longitudinal flap length (mm)1.740.251.82.35
Arc (°)33.318.431.648.3
Significant edge dissection (n = 72)Maximal flap opening (mm)0.440.270.370.52
Maximal flap length (mm)0.900.450.801.23
Longitudinal flap length (mm)1.871.001.602.58
Arc (°)29.114.223.943.0
Proximal (n = 40)Maximal flap opening (mm)0.490.280.380.71
Maximal flap length (mm)1.000.570.881.32
Longitudinal flap length (mm)1.950.801.352.83
Arc (°)27.614.221.936.9
Distal (n = 32)Maximal flap opening (mm)0.380.270.330.48
Maximal flap length (mm)0.780.400.691.18
Longitudinal flap length (mm)1.781.301.802.30
Arc (°)32.315.334.947.8
Small MSA (n = 182)--3.803.443.944.22

Q1: first quartile; Q3: third quartile; MSA: minimal stent area

Table 4

Predictors of suboptimal findings detected by optical coherence tomography.

VariableUnivariate analysisMultivariate analysis
OR95% CIP-valueOR95% CIP-value
LowHighLowHigh
Age1.0060.9911.0220.414
Male sex0.8040.5441.1890.274
Unstable angina (vs. stable angina)1.2690.8431.9080.253
MI (vs. stable angina)0.9750.6291.5110.910
Diabetes mellitus1.4491.0032.0950.048
Hypertension1.3940.9821.9780.063
LDL cholesterol0.9970.9931.0000.067
HDL cholesterol0.9910.9791.0020.118
Triglyceride1.0000.9981.0010.816
Peak CK-MB1.0000.9981.0010.620
Stent diameter0.2120.1390.324<0.0010.2120.1360.328<0.001
Stent length1.0521.0271.079<0.001
Pre-procedural RD0.1850.00217.8890.469
Pre-procedural MLD0.0000.000491.8300.183
Pre-procedural DS0.7250.4081.2870.272
Pre-procedural lesion length3.216.99910.3530.050
Post-procedural RD0.9780.0146.6700.104
Post-procedural MLD0.0000.00047.6100.068
Post-procedural DS0.2020.0281.4790.115
Proximal RA0.8790.8320.929<0.001
Distal RA0.8120.7580.870<0.001
Underexpansion3.2382.2404.681<0.0013.2442.1974.789<0.001
Adjuvant procedure1.4470.9442.216<0.001

OR: odds ratio; CI: confidence interval; MI: myocardial infarction; LDL: low-density lipoprotein; HDL: high-density lipoprotein; RD: reference diameter; MLD: minimal lumen diameter; DS: diameter stenosis; RA: reference area

Incidences of suboptimal findings detected by optical coherence tomography (OCT).

(A) Incidence of each component. (B) Cumulative incidence of suboptimal optical coherence tomography-detected findings. Q1: first quartile; Q3: third quartile; MSA: minimal stent area OR: odds ratio; CI: confidence interval; MI: myocardial infarction; LDL: low-density lipoprotein; HDL: high-density lipoprotein; RD: reference diameter; MLD: minimal lumen diameter; DS: diameter stenosis; RA: reference area

Clinical outcomes

Clinical outcomes on the basis of SF-OCT or No SF-OCT are shown in Table 5. During follow-up (median duration 21.5 months, interquartile range of 15.0–30.0 months), the incidence of MACE was numerically lower in the No SF-OCT group but there was no statistical significance between the two groups (3.0% vs. 5.0%, HR 1.601; 95% CI 0.639–4.011; P = 0.315) (Fig 4). Cox regression analysis was performed to evaluate the predictors of MACE. Overall, neither suboptimal OCT-detected findings nor the components showed significance in univariate analysis (Table 6). In the SF-OCT group, neither each components of SF-OCTs nor the severity of SF-OCTs based on quantitative measurements had statistical power for predicting MACE. Clinical results comparing no adjuvant procedure after OCT (n = 444) versus adjuvant procedures (n = 132) were significantly not different (4.1% vs. 5.3%, HR 1.249; 95% CI 0.521–2.996; P = 0.618).
Table 5

Clinical outcomes.

VariableNo SF-OCT (197 patients)SF-OCT (379 patients)P-value
Major cardiovascular adverse event6 (3.0)19 (5.0)0.272
Non-fatal myocardial infarction00-
All cause death1 (0.5)2 (0.5)0.975
TLR4 (2.0)13 (3.4)0.346
TVR5 (2.5)17 (4.5)0.247
Non-TLR/TVR5 (2.5)8 (2.1)0.743
Stent thrombosis1 (0.5)1 (0.3)0.637

SF-OCT: optical coherence tomography detected suboptimal findings; TLR: target lesion revascularization; TVR: target vessel revascularization

Fig 4

Survival free of major adverse cardiovascular events according to optical coherence tomography-detected poststent optimal versus suboptimal findings.

Table 6

Predictors of major adverse cardiovascular events.

VariableUnivariate analysis
HR95% CIP-value
LowHigh
Age0.9820.9491.0170.314
Male sex1.0760.4302.6950.875
Unstable angina (vs. stable angina)0.7170.2821.8190.483
MI (vs. stable angina)0.9450.3592.4870.908
Body mass index0.9720.8241.1460.734
Diabetes mellitus1.1610.5212.5880.715
Hypertension0.6810.3111.4940.338
Current smoker2.0000.8644.6290.106
Ex-smoker0.9100.2503.3070.886
Peak CK-MB0.9980.9921.0030.420
hs-CRP1.0070.9921.0220.374
LDL cholesterol1.0000.9931.0080.942
RAS blocker1.0140.4592.2390.973
Beta blocker0.5890.2631.3150.196
Stent diameter0.9370.3822.3010.887
Stent length0.9890.9371.0450.705
MSA1.0000.8321.2020.998
Underexpansion0.7080.3181.5750.397
Proximal RA0.9710.8531.1060.661
Distal RA0.9890.8521.1480.885
Expansion rate1.0140.9781.0520.438
SF-OCT1.6010.6394.0110.315
Thrombus6.7950.68267.6990.102
Malapposition1.1270.4982.5520.774
Tissue protrusion0.0370.000101.3820.414
Edge dissection0.5680.1342.4110.443
Small MSA1.2240.5412.7700.628
Thrombus diameter2.2790.47111.0290.306
Thrombus area3.8450.50229.4850.195
Malapposition depth0.9970.9921.0010.178
Malapposition area0.7780.4101.4770.443
Malapposition length1.0350.7261.4770.848
Tissue protrusion length1.3460.3065.9230.694
Tissue protrusion area0.1220.0028.3090.328
Edge dissection location0.6710.0617.4120.745
Edge dissection maximal flap opening3.9970.117136.5580.442
Edge dissection maximal flap length0.8590.1066.9710.887
Edge dissection longitudinal flap length1.5500.9082.6460.108
Edge dissection arc0.9820.8621.1180.780

HR: hazard ratio; CI: confidence interval; MI: myocardial infarction; hs-CRP: high sensitive C-reactive protein; LDL: low-density lipoprotein; RAS: renin-angiotensin system; MSA: minimal stent area; RA: reference area; SF-OCT: suboptimal findings detected by optical coherence tomography

SF-OCT: optical coherence tomography detected suboptimal findings; TLR: target lesion revascularization; TVR: target vessel revascularization HR: hazard ratio; CI: confidence interval; MI: myocardial infarction; hs-CRP: high sensitive C-reactive protein; LDL: low-density lipoprotein; RAS: renin-angiotensin system; MSA: minimal stent area; RA: reference area; SF-OCT: suboptimal findings detected by optical coherence tomography

Discussion

The main findings of the present study are as follows: (1) SF-OCTs are highly prevalent (65.8%, 379/576 cases) after DES implantation; (2) no SF-OCTs translated into clinical events over the follow-up (median 21.5 months); and (3) independent predictors of SF-OCT were stent diameter and underexpansion. Although several studies have reported that most SF-OCTs after DES implantation resolve spontaneously over a short-term follow-up period, these studies had small sample sizes and did not systematically deal with relevant clinical events [1, 4, 5, 13]. The important differences between the current study and the previous studies are: (1) the present study had the longest follow-up period; (2) the present study compared the clinical outcomes between the SF-OCT group and the No SF-OCT groups; (3) the present study quantitatively analyzed the severity of each component of SF-OCT and the relationship of each component with clinical events; and (4) the present study determined independent factors predictive of suboptimal OCT-detected findings. Additionally, the present study has an advantage over previous studies because defining SF-OCT means that these findings do not correlate with clinical events, which can reduce unnecessary additional interventions after PCI. However, our findings do not signify that intravascular imaging is unnecessary. Stent optimization is a crucial factor for future event prevention and intravascular imaging plays a significant role for correction of suboptimal findings. Even though adjuvant intervention was done after index procedure, underexpansion rate was high, post-procedural MLD was significantly smaller and high rate of SF-OCT was still observed. Since procedure was performed on operator’s discretion, further management of suboptimal findings was not strict and stent optimization criteria was not firmly determined. Nonetheless, data in current study suggest the concept of safety zone that if the measurements of SF-OCTs are found to similar to our data.

Malapposition

The incidence of OCT-defined stent malapposition was 32.1% (185/576 lesion) in our study. In studies using intravascular ultrasound (IVUS) for analysis, the incidence of malapposition immediately after stent implantation was approximately 10% [14, 15]. Studies using OCT for analysis have reported higher incidences of post-procedural stent malapposition than studies using IVUS, ranging from 22.2% to more than 50% [1–3, 5]. Since OCT has higher resolution than IVUS, it is superior for detecting malapposition. Our findings also support this high frequency of malapposition. However, this high frequency was not associated with clinical events, a finding similar to those of other previous studies using IVUS and OCT that had follow-up periods less than or approximately one year [1–3, 5, 14, 15]. Previous studies have reported that stent malapposition might be related to late stent thrombosis in both bare-metal stents and DES [16-18]. However, in researches which compared maximal malapposition depth between cases of late stent thrombosis and control showed significantly large scale of malapposition (1400 to 1800 μm) compared to our study (average 469 μm) [17, 18]. Considering the very low incidence of stent thrombosis, the potential impact of malapposition on stent thrombosis can be difficult to evaluate properly. At least, our study can suggest modest degree of malapposition may be left untouched without further correction.

Thrombus or tissue protrusion

The incidence of OCT-defined significant thrombus was 7.6% (44/576 cases), TP was 6.8% (39/576 cases) and any significant thrombus or TP was 14.2% (82/576 cases) in the present study. The rate of thrombus was lower than that in previous reports (37–51%) [1-3]. However, residual thrombus after DES implantation did not translate into clinical events in the current study. Quantitative thrombus analysis did not reveal any correlations between clinical events and the number of thrombi (Table 6). Only one TVR case showed a thrombus, which was not particularly severe (diameter in the fourth quartile, area in the third quartile). After stent implantation, thrombi are believed to form as a result of procedural problems, such as extended time required for stent location, insufficient heparinization during PCI, or squeezed remnants of pre-existing thrombi after stent implantation [1]. The incidence of TP is quite different from those reported in other OCT studies (81–97.5%) [1–4, 13, 19]. However, not all studies had a high prevalence: e.g., Kume et al. reported a TP incidence of 51.3% [5]. This discordance could be explained by inter-observer bias, or by procedural problems such as high stent balloon pressure or post-balloon pressure. Although quite different frequencies have been observed, most of the OCT studies investigating suboptimal findings concluded that TP appears to be a benign phenomenon. Our data also support these previous results. Soeda et al. reported that irregular protrusion was a powerful predictor of device-oriented clinical endpoints and target lesion revascularization [2]. As regards the definition of TP in this study, our data defined most irregular protrusion as thrombus and only smooth protrusion as TP. However, our study failed to demonstrate that either of the findings was related with clinical outcomes. Through quantification, we found that the average maximal length was 0.87 mm, and the average maximal area was 0.36 mm2. These data are similar to those reported in previous studies [4, 5, 13]. Considering incidence and clinical impact of significant criteria of thrombus or TP, it seems concordance comparing previous studies, which failed to prove significance of these findings [3, 6]. Despite limited sample size and lack of quantification in previous studies, natural course of thrombus and TP were mostly resolved, which can be suggestive explanation of no impact on clinical outcome [1, 4, 5].

Edge dissection

The incidence of OCT-defined significant ED was 12.5% (72/576 cases) in our study. Compared with other studies, in which the incidence of presence varied widely from 20% to 37.8% and significant criteria of 12% to 14%, our study reported similar frequency [1–3, 6, 8, 20]. Stent ED defined by IVUS or conventional angiography is considered to be associated with increased short-term and mid-term incidences of MACE and stent thrombosis [19, 21–23]. However, we did not find a significant correlation between OCT-defined ED and clinical outcomes after 2 years. Moreover, quantified dissection severity had no clinical impact. The different clinical outcomes might be because of the aim of our study, which was to investigate suboptimal findings in OCT that are apparently normal in angiography. In dissections seen only by OCT that are minor and non-flow-limiting, spontaneous healing might have a benign course and minimal correlation with clinical outcome. Our quantification efforts revealed that the average maximal flap opening was 0.44 mm, the maximal flap length was 0.90 mm, and the average ED longitudinal length was 1.87 mm. In previous studies, the average maximal flap length ranged from 0.7 to 1.0 mm, longitudinal flap length 2.04±1.60mm and the average maximal flap opening was 0.39±0.34 mm [4, 8, 13, 20]. These findings are consistent with those of the present study. These measurement ranges had no clear relationship with any of the clinical outcomes examined.

Small minimal stent area

Small MSA occupied majority portion among SF-OCTs of 31.6% (182/576 cases). Prevalence was similar compared to previous studies in CLI-OPCI II trial and Soeda et al., which was 23.4% and 41.2% respectively [2, 3]. Previous IVUS as well as OCT studies have shown that small MSA and inadequate lumen area are associated with major clinical outcomes [2, 3, 24, 25]. Although frequency of small MSA did not differ in this study, it was not associated with clinical events like other studies. Reasons of this discrepancy can be considered by dominant use of new generation stent and high rate of dual antiplatelet therapy, as well as limited number of patients enrolled in current study.

Study limitations

The major limitation of this study is the absence of follow-up OCT data to assess the natural healing course of SF-OCTs. However, the most important issue regarding these findings is whether they have a significant clinical impact that requires additional initial procedures. Another concern is this was a non-randomized retrospective study based on low event rates, relatively limited sample size and modest follow-up period to clarify the clinical outcome, raising the possibility of selection bias and therefore underpowered to determine the benefits of correcting SF-OCTs. Estimating predicted power of suboptimal findings in clinical events were performed with power analysis of 0.553, which this study may be underpowered for its primary endpoint. Since the data collection of clinical outcomes was based on retrospective chart review, there is a chance of underreporting. The dataset used in the study is not recent and therefore results can be underestimated. However, data in current study suggest the concept of safety zone since certain findings in OCT to or not to intervene is still yet controversial and our data can at least show that modest degree can be tolerated. Only the maximum depth and maximum malapposition area were measured, and the entire stent was not evaluated. Finally, all measurements were performed manually, meaning that a certain degree of manual error might be present. Larger studies with a longer follow-up duration are needed to confirm the relationships between clinical events and SF-OCTs.

Conclusions

The presence of angiographically insignificant SF-OCTs (ED, TP, malapposition, thrombus and small MSA) and their severity were not associated with clinical outcomes in this study. (XLSX) Click here for additional data file. 10 Aug 2020 PONE-D-20-16120 Clinical impact of angiographically insignificant suboptimal poststent findings detected by optical coherence tomography after drug-eluting stent implantation PLOS ONE Dear Dr. Yu, Thank you for submitting your manuscript to PLOS ONE. After careful consideration, we feel that it has merit but does not fully meet PLOS ONE’s publication criteria as it currently stands. Therefore, we invite you to submit a revised version of the manuscript that addresses the points raised during the review process. Generally, the reviewers found the paper favorable, however there are some issues that need to be addressed. Please address each item specifically and outline these changes in revised text and in a point by point fashion in the response letter. Please submit your revised manuscript by Sep 24 2020 11:59PM. If you will need more time than this to complete your revisions, please reply to this message or contact the journal office at plosone@plos.org. When you're ready to submit your revision, log on to https://www.editorialmanager.com/pone/ and select the 'Submissions Needing Revision' folder to locate your manuscript file. Please include the following items when submitting your revised manuscript: A rebuttal letter that responds to each point raised by the academic editor and reviewer(s). You should upload this letter as a separate file labeled 'Response to Reviewers'. A marked-up copy of your manuscript that highlights changes made to the original version. You should upload this as a separate file labeled 'Revised Manuscript with Track Changes'. An unmarked version of your revised paper without tracked changes. You should upload this as a separate file labeled 'Manuscript'. If you would like to make changes to your financial disclosure, please include your updated statement in your cover letter. Guidelines for resubmitting your figure files are available below the reviewer comments at the end of this letter. If applicable, we recommend that you deposit your laboratory protocols in protocols.io to enhance the reproducibility of your results. Protocols.io assigns your protocol its own identifier (DOI) so that it can be cited independently in the future. For instructions see: http://journals.plos.org/plosone/s/submission-guidelines#loc-laboratory-protocols We look forward to receiving your revised manuscript. Kind regards, Jay Widmer Academic Editor PLOS ONE Journal Requirements: When submitting your revision, we need you to address these additional requirements. 1. Please ensure that your manuscript meets PLOS ONE's style requirements, including those for file naming. The PLOS ONE style templates can be found at https://journals.plos.org/plosone/s/file?id=wjVg/PLOSOne_formatting_sample_main_body.pdf and https://journals.plos.org/plosone/s/file?id=ba62/PLOSOne_formatting_sample_title_authors_affiliations.pdf 2. Thank you for stating the following in the Acknowledgments Section of your manuscript: "This study was supported by Wonkwang University in 2018." We note that you have provided funding information that is not currently declared in your Funding Statement. However, funding information should not appear in the Acknowledgments section or other areas of your manuscript. We will only publish funding information present in the Funding Statement section of the online submission form. Please remove any funding-related text from the manuscript and let us know how you would like to update your Funding Statement. Currently, your Funding Statement reads as follows: "The authors received no specific funding for this work." [Note: HTML markup is below. Please do not edit.] Reviewers' comments: Reviewer's Responses to Questions Comments to the Author 1. Is the manuscript technically sound, and do the data support the conclusions? The manuscript must describe a technically sound piece of scientific research with data that supports the conclusions. Experiments must have been conducted rigorously, with appropriate controls, replication, and sample sizes. The conclusions must be drawn appropriately based on the data presented. Reviewer #1: Yes Reviewer #2: Partly ********** 2. Has the statistical analysis been performed appropriately and rigorously? Reviewer #1: I Don't Know Reviewer #2: Yes ********** 3. Have the authors made all data underlying the findings in their manuscript fully available? The PLOS Data policy requires authors to make all data underlying the findings described in their manuscript fully available without restriction, with rare exception (please refer to the Data Availability Statement in the manuscript PDF file). The data should be provided as part of the manuscript or its supporting information, or deposited to a public repository. For example, in addition to summary statistics, the data points behind means, medians and variance measures should be available. If there are restrictions on publicly sharing data—e.g. participant privacy or use of data from a third party—those must be specified. Reviewer #1: Yes Reviewer #2: No ********** 4. Is the manuscript presented in an intelligible fashion and written in standard English? PLOS ONE does not copyedit accepted manuscripts, so the language in submitted articles must be clear, correct, and unambiguous. Any typographical or grammatical errors should be corrected at revision, so please note any specific errors here. Reviewer #1: Yes Reviewer #2: Yes ********** 5. Review Comments to the Author Please use the space provided to explain your answers to the questions above. You may also include additional comments for the author, including concerns about dual publication, research ethics, or publication ethics. (Please upload your review as an attachment if it exceeds 20,000 characters) Reviewer #1: Interesting and important question to investigate. Study is reasonably designed. Size is modest, although it could be underpowered. Evidence of this includes a HR for the primary endpoint of 1.6 (0.6-4), suggesting a sizable effect could be present (in a larger sample). Also suggestion of under-powering is evident in the fact that statistically significant differences in MSA and other such measures (reference vessel size, etc) did not translate into difference in restenosis/TVR. Are we to conclude that final stent size does not impact restenosis? Questions that arise include: Why were all patients treated between Jan 2011 - May 2013, and the analysis just now being performed? Was the database cut off, and we have no more info, or was there some reason that this time period was chosen? The data is getting to be fairly old, and of course, techniques evolve which could impact how we apply findings. I'd like to understand a bit better what the operators knew about and did with the OCT results. They seemed to have them available, and yet sometimes they were still considered "suboptimal." Was this because the operator mis-interpreted the OCT images, had no option because there was no solution to the issue, or was it simply an issue that what the operator deemed insignificant or suboptimal was different from the later, agreed upon definition. Nonetheless, it points out a significant weak point in the study, this is not a randomized comparison of one strategy vs. another, rather it is a comparison of a stent procedure with optimal results vs. a procedure with suboptimal results, but one in which the operator deemed the results worthy to accept. That is an important distinction. I like and appreciate the suggestion of a "safety zone" in optimizing PCI results. The study contradicts what the majority of studies (and meta-analyses) have shown, that some of these suboptimal results do negatively impact outcomes. That being said, it is likely that as the author purport, some degree of suboptimal results can be tolerated without compromising clinical results. How much can be tolerated, is a fascinating question. Reviewer #2: This study sought to compare clinical outcomes may differ between patients with SF (suboptimal findings) OCT and without SF-OCTs after DES implantation. With this aim 576 with final OCT after DES were divided into SF-OCT group (n=379, 379 lesions) and No SF-OCT group (n=197, 197 lesions). The study population had no significant abnormal finding in final angiography. Among 379 patients with SF-OCT, 32.4% had multiple SF-OCTs. Malapposition (32.1%) was the most frequent, followed by small MSA (31.6%), edge dissection (12.5%), thrombus (7.6%) and tissue protrusion (6.8%). These features were also quantitatively analyzed. The SF-OCT group showed smaller ST diameter and longer length, and lower expansion rate. The incidence of MACE did not differ between the two groups (3% vs 5%, p=0.310). This is an interesting study suggesting that suboptimal OCT findings have no clinical consequences in patients with good angiographic results. Some issues, however, need to be addressed: 1. Sample size remains a major problem in this study (unpowered) as clinical events in patients with good angiographic results after DES are low. Please provide an estimate of the predicted power of the study and acknowledge that the study was unpowered for its primary endpoint. 2. Study criteria should be clarified. 2) Patients who underwent sequential OCT immediately after DES implantation or after adjuvant procedures; 3) No significant abnormal finding in final coronary angiography. Adjuvant procedures after OCT examination were performed in 132 patients (132 lesions). Were clinical results poorer in patients that required optimization of the stent after the initial OCT? 3. Information of clinical outcomes was collected by the retrospective review of the chart. Was this performed blindly to angiographic data and OCT findings? 4. Angiographic MLD after the procedure was smaller in the SF-OCT group. This is a classical variable associated with poorer long-term clinical and angiographic outcomes. Again a lack of power may explain the lack of clinical differences between the groups. 5. The logistic multivariate model identified stent diameter MSA and underexpansion as independent predictors of SF-OCT. The variable MSA by OCT is a factor already included in the SF-OCT list. 6. Only clear abnormal findings were factors were considered “significant” SF-OCTs. Some abnormal findings were considered SF-OCT. However cualitative crietria were arbitrary. Perhaps a different cut-offs would yield different results. Please address. Among them it is included a small MSA defined as in-stent minimum area <4.5 mm2 MSA has been found to predict clinical outcomes (mainly TLR and stent thrombosis) in many previous IVUS and OCT studies. Please address. Again the study may be simply unpowered in this regard. 7. Please describe with further details the QCA analysis and the system used. The reference diameter, minimal luminal diameter, percentage of stenosis, and lesion length were evaluated. Were measurements performed also after DES implantation? Any of the findings described in the SF OCT group related to poorer angiographic findings by QCA? ********** 6. PLOS authors have the option to publish the peer review history of their article (what does this mean?). If published, this will include your full peer review and any attached files. If you choose “no”, your identity will remain anonymous but your review may still be made public. Do you want your identity to be public for this peer review? For information about this choice, including consent withdrawal, please see our Privacy Policy. Reviewer #1: Yes: Timothy A. Mixon MD, Daniel Fronk MD. Reviewer #2: No [NOTE: If reviewer comments were submitted as an attachment file, they will be attached to this email and accessible via the submission site. Please log into your account, locate the manuscript record, and check for the action link "View Attachments". If this link does not appear, there are no attachment files.] While revising your submission, please upload your figure files to the Preflight Analysis and Conversion Engine (PACE) digital diagnostic tool, https://pacev2.apexcovantage.com/. PACE helps ensure that figures meet PLOS requirements. To use PACE, you must first register as a user. Registration is free. Then, login and navigate to the UPLOAD tab, where you will find detailed instructions on how to use the tool. If you encounter any issues or have any questions when using PACE, please email PLOS at figures@plos.org. Please note that Supporting Information files do not need this step. 18 Sep 2020 Ref: Manuscript Number PONE-D-20-16120 Clinical impact of angiographically insignificant suboptimal poststent findings detected by optical coherence tomography after drug-eluting stent implantation Answers to the editor’s and reviewers’ comments We thank the editor and the reviewers for their thoughtful and constructive comments on our study. Several issues were raised by the editor and reviewers, and we have addressed these issues point-by-point and enclosed our comments. Indeed, we realized that the comments were insightful, constructive, and helpful in strengthening our manuscript. We have used red color font to indicate the revised portions of our manuscript for the reviewers. We hope that the editor and the reviewers would be satisfied with our responses and find the revised manuscript suitable for publication in PLOS ONE. ---------------------------------------------------------------------------------------------------------------- Response to Reviewer #1: Comment #1. Interesting and important question to investigate. Study is reasonably designed. Size is modest, although it could be underpowered. Evidence of this includes a HR for the primary endpoint of 1.6 (0.6-4), suggesting a sizable effect could be present (in a larger sample). Also suggestion of under-powering is evident in the fact that statistically significant differences in MSA and other such measures (reference vessel size, etc) did not translate into difference in restenosis/TVR. Are we to conclude that final stent size does not impact restenosis? Answer: Thank you for your valuable comment we should address. As the reviewer pointed out, the small sample size is one of the important limitations of this study. However, what differentiates this study from previous studies is that the present study was conducted on patients who showed only suboptimal findings confined to OCT, where no definite abnormal finding was observed on coronary angiography. Therefore, the subjects of this study have a lower frequency of event occurrence in clinical outcomes than other studies. As a result, it is estimated that small MSA or underexpansion, which have been known as risk factors of the adverse clinical outcomes, did not translate into difference in clinical outcomes in this study. In addition, the fact that this study included patients who had only single vessel to intervene may be an additional explanation for the results that did not show significant difference in clinical outcomes between the two study groups. As the reviewer points out, in general, the final stent size is considered a factor that affects restenosis. However, it would be reasonable to interpret that a slight difference in stent size does not directly lead to a difference in clinical outcomes in patients with simple lesions and no definite angiographic abnormal findings. We believe that the conclusions of this study should be reconfirmed in future large-scale studies. The limitations of this study on small sample size are specified in the Limitation section. In page 25 line 9-14: Another concern is this was a non-randomized retrospective study based on low event rates, relatively limited sample size and modest follow-up period to clarify the clinical outcome, raising the possibility of selection bias and therefore underpowered to determine the benefits of correcting SF-OCTs. Estimating predicted power of suboptimal findings in clinical events were performed with power analysis of 0.553, which this study may be underpowered for its primary endpoint. In page 25 line 21-22: Larger studies with a longer follow-up duration are needed to confirm the relationships between clinical events and SF-OCTs. Comment #2. Why were all patients treated between Jan 2011 - May 2013, and the analysis just now being performed? Was the database cut off, and we have no more info, or was there some reason that this time period was chosen? The data is getting to be fairly old, and of course, techniques evolve which could impact how we apply findings. Answer: Thank you for your valid comments. We agree that the period of dataset is not recent. The first announcement of present study date was made in TCT 2014, and then TCT 2015 and TCT 2016. JACC Volume 64, Issue 11 Supplement, September 2014 (DOI: 10.1016/j.jacc.2014.07.426) JACC Volume 66, Issue 15 Supplement, October 2015 (DOI: 10.1016/j.jacc.2015.08.976) JACC Volume 68, Issue 18 Supplement, November 2016 (DOI: 10.1016/j.jacc.2016.09.697) Since then, considerable time has been spent due to repeated intravascular image analysis and statistical analysis to increase the precision of the data. The point you have mentioned has been added in the Limitation section. In page 25 line 15 to 16: The dataset used in the study is not recent and therefore results can be underestimated. Comment #3: I'd like to understand a bit better what the operators knew about and did with the OCT results. They seemed to have them available, and yet sometimes they were still considered "suboptimal." Was this because the operator mis-interpreted the OCT images, had no option because there was no solution to the issue, or was it simply an issue that what the operator deemed insignificant or suboptimal was different from the later, agreed upon definition. Nonetheless, it points out a significant weak point in the study, this is not a randomized comparison of one strategy vs. another, rather it is a comparison of a stent procedure with optimal results vs. a procedure with suboptimal results, but one in which the operator deemed the results worthy to accept. That is an important distinction. Answer: Thank you for providing these insights. Since procedure was performed on operator’s personal discretion, further management of suboptimal findings was not strict and stent optimization criteria was not firmly determined. Brief OCT analysis was performed in on-site and concurrent analysis in core-lab was not done. In addition, because the resolution of OCT is very high, there are often cases in real world practice where there is a sub-optimal finding on OCT that did not show any specific findings on coronary angiography. In these cases, there are many cases of contemplating whether it is beneficial to apply an additional procedure to correct suboptimal findings on OCT without specific findings on coronary angiography. This is because, by applying an additional procedure, we may get a chance to correct the suboptimal finding on the OCT, but on the contrary, there is also a risk of complications such as dissection or need for additional stent implantation by performing additional procedures to correct suboptimal findings of OCT. We believe this study is valuable in that it can provide some modest additional information to these common situations that can be encountered in many medical centers using OCT. The point you have mentioned has been added in the manuscript. In page 21 line 18 to 22: Even though adjuvant intervention was done after index procedure, underexpansion rate was high, post-procedural MLD was significantly smaller and high rate of SF-OCT was still observed. Since procedure was performed on operator’s discretion, further management of suboptimal findings was not strict and stent optimization criteria was not firmly determined. Comment #4: I like and appreciate the suggestion of a "safety zone" in optimizing PCI results. The study contradicts what the majority of studies (and meta-analyses) have shown, that some of these suboptimal results do negatively impact outcomes. That being said, it is likely that as the author purport, some degree of suboptimal results can be tolerated without compromising clinical results. How much can be tolerated, is a fascinating question. Answer: Thank you for your suggestion and we agree with your assessment. Even though under expansion rate was very high, our data did not show association with clinical outcome, unlike previous studies about suboptimal findings after PCI. In guidelines and results of large randomized studies tell us about importance of intravascular imaging for stent optimization, however, not all procedures undergo intravascular imaging in real world for several reasons (e.g., financial problem, lack of time, operator’s personal discretion, and impossible optimization due to lesion characteristics). Even though intravascular imaging was done, on-site precise measuring and analysis for optimization is quite burden. Certain findings in OCT to or not to intervene is still yet controversial and our data can at least show that modest degree can be tolerated. However, although this study provided the concept of "safety zone" in optimizing PCI results, it is difficult to accurately define what the size of the safety zone is only with the results of this study. Based on the results of this study, a large-scale OCT study should be conducted to clarify the boundaries of the safety zone. Thank you again for your valuable comments. In page 25 line 16 to 18: However, data in current study suggest the concept of safety zone since certain findings in OCT to or not to intervene is still yet controversial and our data can at least show that modest degree can be tolerated. ---------------------------------------------------------------------------------------------------------------- Response to Reviewer #2: Comment #1: Sample size remains a major problem in this study (unpowered) as clinical events in patients with good angiographic results after DES are low. Please provide an estimate of the predicted power of the study and acknowledge that the study was unpowered for its primary endpoint. Answer: Thank you for your sharp comments. Since present study was based on retrospective registry, estimating sample size was not possibly calculated. Although the follow-up period of this study was relatively long compared to other OCT studies, the small number of patients with clinical events due to the nature of the study population demographics (had only single vessel to intervene, no definite abnormal finding in final angiography) is an important limitation of this study. Estimating predicted power of suboptimal findings in clinical events were performed with power analysis of 0.553, which this study may be underpowered for its primary endpoint as you mentioned. The point you have mentioned has been added in the Limitation section. In page 25 line 12 to 14: Estimating predicted power of suboptimal findings in clinical events were performed with power analysis of 0.553, which this study may be underpowered for its primary endpoint. Comment #2: Study criteria should be clarified. 2) Patients who underwent sequential OCT immediately after DES implantation or after adjuvant procedures; 3) No significant abnormal finding in final coronary angiography. Adjuvant procedures after OCT examination were performed in 132 patients (132 lesions). Were clinical results poorer in patients that required optimization of the stent after the initial OCT? Answer: Thank you for providing the insight. Clinical results comparing no adjuvant procedure (n=444) versus adjuvant procedures (n=132) were significantly not different (4.1% vs. 5.3%, HR 1.249; 95% CI 0.521-2.996; P = 0.618). However, the lack of clinical difference between these two groups may be due to the fact that only patients with no definite abnormal findings in the final coronary angiography after additional procedures were included in this study. Therefore, it should not be interpreted that the implementation of the additional procedure does not affect clinical outcomes. This finding is addressed in the manuscript. In page 18 line 6 to 8: Clinical results comparing no adjuvant procedure (n=444) versus adjuvant procedures (n=132) were significantly not different (4.1% vs. 5.3%, HR 1.249; 95% CI 0.521-2.996; P = 0.618). Comment #3: Information of clinical outcomes was collected by the retrospective review of the chart. Was this performed blindly to angiographic data and OCT findings? Answer: Thanks for your kind advice. Collecting data of clinical information and outcomes was performed blindly to angiographic data and OCT findings. The point you mentioned has been added in the manuscript. In page 5 line 23 to 24: Collecting data of clinical information and outcomes was performed blindly to angiographic data and OCT findings. Comment #4: Angiographic MLD after the procedure was smaller in the SF-OCT group. This is a classical variable associated with poorer long-term clinical and angiographic outcomes. Again a lack of power may explain the lack of clinical differences between the groups. Answer: Thank you for your comment and we agree with your assessment. Even though underexpansion rate was high and angiographic MLD after the procedure was significantly smaller in the SF-OCT group, our data did not show association with clinical outcome. As you mentioned above, small sample size and lack of power can be the reason of this finding. However, what differentiates this study from previous studies is that the present study was conducted on patients who showed only suboptimal findings confined to OCT, where no definite abnormal finding was observed on coronary angiography. Therefore, the subjects of this study had a lower frequency of event occurrence in clinical outcomes than other studies. As a result, it is estimated that small MLD or MSA, which were known as risk factors of the adverse clinical outcomes, did not translate into difference in clinical outcomes in this study. In addition, the fact that this study included patients who had only single vessel to intervene may be an additional explanation for the results that did not show significant difference in clinical outcomes between the two study groups. It would be reasonable to interpret that slight differences in MLD (MLD 3.35 ± 0.44 vs. 3.00 ± 0.46mm, p=0.015), which are guaranteed to be extended to some extent (mean value of MLD in both group≥3.0mm), may not directly lead to a difference in clinical outcomes in patients with simple lesions and no definite angiographic abnormal findings. This has been pointed out in the manuscript. In page 21 line 18-20: Even though adjuvant intervention was done after index procedure, underexpansion rate was high, post-procedural MLD was significantly smaller and high rate of SF-OCT was still observed. Comment #5: The logistic multivariate model identified stent diameter MSA and underexpansion as independent predictors of SF-OCT. The variable MSA by OCT is a factor already included in the SF-OCT list. Answer: Thanks for your sharp point. The variable MSA is deleted and analyzed again in logistic multivariate model of independent predictors for SF-OCT (Table 4). In page 13 line 20 to page 14 line 2: The logistic univariate model showed that diabetes mellitus, stent diameter, stent length, proximal reference area, distal reference area, underexpansion and adjuvant procedure were correlated with SF-OCT. The logistic multivariate model identified stent diameter (OR 0.212; 95% CI 0.136-0.328; P<0.001), and underexpansion (OR 3.244; 95% CI 2.197-4.789; P<0.001) as independent predictors of SF-OCT (Table 4). In page 16 to 17 Table 4: Variable Univariate analysis Multivariate analysis OR 95% CI P-value OR 95% CI P-value Low High Low High Age 1.006 0.991 1.022 0.414 Male sex 0.804 0.544 1.189 0.274 Unstable angina (vs. stable angina) 1.269 0.843 1.908 0.253 MI (vs. stable angina) 0.975 0.629 1.511 0.910 Diabetes mellitus 1.449 1.003 2.095 0.048 Hypertension 1.394 0.982 1.978 0.063 LDL cholesterol 0.997 0.993 1.000 0.067 HDL cholesterol 0.991 0.979 1.002 0.118 Triglyceride 1.000 0.998 1.001 0.816 Peak CK-MB 1.000 0.998 1.001 0.620 Stent diameter 0.212 0.139 0.324 <0.001 0.212 0.136 0.328 <0.001 Stent length 1.052 1.027 1.079 <0.001 Pre-procedural RD 0.185 0.002 17.889 0.469 Pre-procedural MLD 0.000 0.000 491.830 0.183 Pre-procedural DS 0.725 0.408 1.287 0.272 Pre-procedural lesion length 3.216 .999 10.353 0.050 Post-procedural RD 0.978 0.014 6.670 0.104 Post-procedural MLD 0.000 0.000 47.610 0.068 Post-procedural DS 0.202 0.028 1.479 0.115 Proximal RA 0.879 0.832 0.929 <0.001 Distal RA 0.812 0.758 0.870 <0.001 Underexpansion 3.238 2.240 4.681 <0.001 3.244 2.197 4.789 <0.001 Adjuvant procedure 1.447 0.944 2.216 <0.001 Comment #6: Only clear abnormal findings were factors were considered “significant” SF-OCTs. Some abnormal findings were considered SF-OCT. However qualitative criteria were arbitrary. Perhaps a different cut-offs would yield different results. Please address. Among them it is included a small MSA defined as in-stent minimum area <4.5 mm2. MSA has been found to predict clinical outcomes (mainly TLR and stent thrombosis) in many previous IVUS and OCT studies. Please address. Again the study may be simply unpowered in this regard. Answer: You have asked an important question that we should address. Measurement of SF-OCTs and “significant” SF-OCTs are separately addressed in the quantification data of SF-OCT (Table 3). All statistical analysis was performed on basis of “significant” SF-OCTs. Average (3.80), median (3.94), Q1 (3.44) and Q3 (4.22) value of MSA in small MSA group is added in the quantification data of SF-OCT (Table 3). For the reason why small MSA did not affect clinical outcome in this study, it would be appreciated if you refer to the answer to Comment #4. In page 14 to 16 Table 3: Variables Average Q1 Median Q3 Thrombus (n=116) Number (n) 1.27 1 1 1 Longitudinal length (mm) 0.79 0.48 0.73 1 Diameter (mm) 1.2 0.86 1.16 1.55 Area (mm2) 0.56 0.29 0.51 0.68 Significant Thrombus (n=44) Number (n) 1.05 1 1 1 Longitudinal length (mm) 1.04 0.70 1.00 1.40 Diameter (mm) 1.62 1.31 1.60 1.97 Area (mm2) 0.86 0.59 0.74 1.10 Malapposition (n=188) Maximal depth (µm) 465 310 410 578 Area (mm2) 2.05 1.18 1.71 2.51 Length (mm) 2.42 1.3 2.2 3.18 Significant Malapposition (n=185) Maximal depth (µm) 469 315 410 580 Area (mm2) 2.07 1.21 1.72 2.53 Length (mm) 2.43 1.30 2.20 3.15 Tissue protrusion (n=263) Length (mm) 0.89 0.62 0.84 1.1 Area (mm2) 0.37 0.16 0.25 0.42 Significant Tissue protrusion (n=39) Length (mm) 1.35 1.05 1.30 1.67 Area (mm2) 1.03 0.56 0.69 1.02 Edge dissection (n=100) Maximal flap opening (mm) 0.36 0.2 0.31 0.46 Maximal flap length (mm) 0.8 0.38 0.7 1.02 Longitudinal flap length (mm) 1.73 0.93 1.5 2.3 Arc (°) 28 14.8 22.3 39.5 Proximal (n=55) Maximal flap opening (mm) 0.4 0.2 0.33 0.52 Maximal flap length (mm) 0.88 0.47 0.73 1.07 Longitudinal flap length (mm) 1.71 0.7 1.2 2.3 Arc (°) 25.2 14.2 20.3 31.9 Distal (n=45) Maximal flap opening (mm) 0.32 0.2 0.29 0.41 Maximal flap length (mm) 0.69 0.32 0.58 1.01 Longitudinal flap length (mm) 1.74 0.25 1.8 2.35 Arc (°) 33.3 18.4 31.6 48.3 Significant Edge dissection (n=72) Maximal flap opening (mm) 0.44 0.27 0.37 0.52 Maximal flap length (mm) 0.90 0.45 0.80 1.23 Longitudinal flap length (mm) 1.87 1.00 1.60 2.58 Arc (°) 29.1 14.2 23.9 43.0 Proximal (n=40) Maximal flap opening (mm) 0.49 0.28 0.38 0.71 Maximal flap length (mm) 1.00 0.57 0.88 1.32 Longitudinal flap length (mm) 1.95 0.80 1.35 2.83 Arc (°) 27.6 14.2 21.9 36.9 Distal (n=32) Maximal flap opening (mm) 0.38 0.27 0.33 0.48 Maximal flap length (mm) 0.78 0.40 0.69 1.18 Longitudinal flap length (mm) 1.78 1.30 1.80 2.30 Arc (°) 32.3 15.3 34.9 47.8 Small MSA (n=182) - - 3.80 3.44 3.94 4.22 Comment #7: Please describe with further details the QCA analysis and the system used. The reference diameter, minimal luminal diameter, percentage of stenosis, and lesion length were evaluated. Were measurements performed also after DES implantation? Any of the findings described in the SF OCT group related to poorer angiographic findings by QCA? Answer: Thank you very much for your valuable comments. A detailed explanation of QCA analysis is provided in the Methods section. Measurements of QCA analysis after DES implantation is presented in angiographic and procedural data (Table 2). Measurements of QCA analysis were analyzed in logistic multivariate model of independent predictors for SF-OCT, and none of those measurements were related to SF-OCTs. These findings are address in the predictors of SF-OCTs (Table 4). In page 6 line 5-10: Coronary angiograms were analyzed using a computer-based Telecardiology system, version 2.02 (Medcon Inc., Tel Aviv, Israel) by three radiologic technicians who were blinded to the study purpose. The reference diameter, minimal luminal diameter (MLD), percentage of stenosis, and lesion length were evaluated from diastolic frames using guided catheter magnification calibration in a single, matched view with a computerized quantitative analyzer using a caliper. In page 16 to 17 Table 4: Variable Univariate analysis Multivariate analysis OR 95% CI P-value OR 95% CI P-value Low High Low High Pre-procedural RD 0.185 0.002 17.889 0.469 Pre-procedural MLD 0.000 0.000 491.830 0.183 Pre-procedural DS 0.725 0.408 1.287 0.272 Pre-procedural lesion length 3.216 .999 10.353 0.050 Post-procedural RD 0.978 0.014 6.670 0.104 Post-procedural MLD 0.000 0.000 47.610 0.068 Post-procedural DS 0.202 0.028 1.479 0.115 Submitted filename: PlosOne_AAF_Response to Reviewers_200915_R1.docx Click here for additional data file. 5 Oct 2020 Clinical impact of angiographically insignificant suboptimal poststent findings detected by optical coherence tomography after drug-eluting stent implantation PONE-D-20-16120R1 Dear Dr. Yu, We’re pleased to inform you that your manuscript has been judged scientifically suitable for publication and will be formally accepted for publication once it meets all outstanding technical requirements. Within one week, you’ll receive an e-mail detailing the required amendments. When these have been addressed, you’ll receive a formal acceptance letter and your manuscript will be scheduled for publication. An invoice for payment will follow shortly after the formal acceptance. To ensure an efficient process, please log into Editorial Manager at http://www.editorialmanager.com/pone/, click the 'Update My Information' link at the top of the page, and double check that your user information is up-to-date. If you have any billing related questions, please contact our Author Billing department directly at authorbilling@plos.org. If your institution or institutions have a press office, please notify them about your upcoming paper to help maximize its impact. If they’ll be preparing press materials, please inform our press team as soon as possible -- no later than 48 hours after receiving the formal acceptance. Your manuscript will remain under strict press embargo until 2 pm Eastern Time on the date of publication. For more information, please contact onepress@plos.org. Kind regards, Jay Widmer Academic Editor PLOS ONE Reviewer's Responses to Questions Comments to the Author 1. If the authors have adequately addressed your comments raised in a previous round of review and you feel that this manuscript is now acceptable for publication, you may indicate that here to bypass the “Comments to the Author” section, enter your conflict of interest statement in the “Confidential to Editor” section, and submit your "Accept" recommendation. Reviewer #1: All comments have been addressed Reviewer #2: All comments have been addressed ********** 2. Is the manuscript technically sound, and do the data support the conclusions? The manuscript must describe a technically sound piece of scientific research with data that supports the conclusions. Experiments must have been conducted rigorously, with appropriate controls, replication, and sample sizes. The conclusions must be drawn appropriately based on the data presented. Reviewer #1: Yes Reviewer #2: Partly ********** 3. Has the statistical analysis been performed appropriately and rigorously? Reviewer #1: Yes Reviewer #2: Yes ********** 4. Have the authors made all data underlying the findings in their manuscript fully available? The PLOS Data policy requires authors to make all data underlying the findings described in their manuscript fully available without restriction, with rare exception (please refer to the Data Availability Statement in the manuscript PDF file). The data should be provided as part of the manuscript or its supporting information, or deposited to a public repository. For example, in addition to summary statistics, the data points behind means, medians and variance measures should be available. If there are restrictions on publicly sharing data—e.g. participant privacy or use of data from a third party—those must be specified. Reviewer #1: Yes Reviewer #2: Yes ********** 5. Is the manuscript presented in an intelligible fashion and written in standard English? PLOS ONE does not copyedit accepted manuscripts, so the language in submitted articles must be clear, correct, and unambiguous. Any typographical or grammatical errors should be corrected at revision, so please note any specific errors here. Reviewer #1: No Reviewer #2: Yes ********** 6. Review Comments to the Author Please use the space provided to explain your answers to the questions above. You may also include additional comments for the author, including concerns about dual publication, research ethics, or publication ethics. (Please upload your review as an attachment if it exceeds 20,000 characters) Reviewer #1: Thank you for addressing my concerns and questions. Some minor alterations are needed which are mainly language issues, including: Page 5, line 24: "was performed in a blinded fashion in regards to angiographic data and OCt..." P 18, line 7: "change to "not significantly different" P 21, line 18: "Even though adjuvant intervention was performed after the index procedure, the rate of under expansion was high...and a high rate of SF-OCT was still observed" p21, line 20-24: "Since the procedure was performed at the operator's discretion, further management of suboptimal findings was not based on strict protocol, and stent optimization criteria were not universally followed. Reviewer #2: (No Response) ********** 7. PLOS authors have the option to publish the peer review history of their article (what does this mean?). If published, this will include your full peer review and any attached files. If you choose “no”, your identity will remain anonymous but your review may still be made public. Do you want your identity to be public for this peer review? For information about this choice, including consent withdrawal, please see our Privacy Policy. Reviewer #1: No Reviewer #2: Yes: Fernando Alfonso MD 8 Oct 2020 PONE-D-20-16120R1 Clinical impact of angiographically insignificant suboptimal poststent findings detected by optical coherence tomography after drug-eluting stent implantation Dear Dr. Yu: I'm pleased to inform you that your manuscript has been deemed suitable for publication in PLOS ONE. Congratulations! Your manuscript is now with our production department. If your institution or institutions have a press office, please let them know about your upcoming paper now to help maximize its impact. If they'll be preparing press materials, please inform our press team within the next 48 hours. Your manuscript will remain under strict press embargo until 2 pm Eastern Time on the date of publication. For more information please contact onepress@plos.org. If we can help with anything else, please email us at plosone@plos.org. Thank you for submitting your work to PLOS ONE and supporting open access. Kind regards, PLOS ONE Editorial Office Staff on behalf of Dr. Jay Widmer Academic Editor PLOS ONE
  25 in total

1.  Natural history of intravascular ultrasound-detected edge dissections from coronary stent deployment.

Authors:  S J Sheris; M R Canos; N J Weissman
Journal:  Am Heart J       Date:  2000-01       Impact factor: 4.749

2.  Evaluation of intracoronary stenting by intravascular optical coherence tomography.

Authors:  B E Bouma; G J Tearney; H Yabushita; M Shishkov; C R Kauffman; D DeJoseph Gauthier; B D MacNeill; S L Houser; H T Aretz; E F Halpern; I-K Jang
Journal:  Heart       Date:  2003-03       Impact factor: 5.994

Review 3.  Coronary artery dissection and perforation complicating percutaneous coronary intervention.

Authors:  Jason H Rogers; John M Lasala
Journal:  J Invasive Cardiol       Date:  2004-09       Impact factor: 2.022

4.  Stent underexpansion and residual reference segment stenosis are related to stent thrombosis after sirolimus-eluting stent implantation: an intravascular ultrasound study.

Authors:  Kenichi Fujii; Stéphane G Carlier; Gary S Mintz; Yi-ming Yang; Issam Moussa; Giora Weisz; George Dangas; Roxana Mehran; Alexandra J Lansky; Edward M Kreps; Michael Collins; Gregg W Stone; Jeffrey W Moses; Martin B Leon
Journal:  J Am Coll Cardiol       Date:  2005-04-05       Impact factor: 24.094

5.  Late incomplete stent apposition after sirolimus-eluting stent implantation: a serial intravascular ultrasound analysis.

Authors:  Junya Ako; Yoshihiro Morino; Yasuhiro Honda; Ali Hassan; Shinjo Sonoda; Paul G Yock; Martin B Leon; Jeffrey W Moses; Heidi N Bonneau; Peter J Fitzgerald
Journal:  J Am Coll Cardiol       Date:  2005-09-20       Impact factor: 24.094

6.  Neointimal coverage of sirolimus-eluting stents at 6-month follow-up: evaluated by optical coherence tomography.

Authors:  Daisuke Matsumoto; Junya Shite; Toshiro Shinke; Hiromasa Otake; Yusuke Tanino; Daisuke Ogasawara; Takahiro Sawada; Oscar Luis Paredes; Ken-ichi Hirata; Mitsuhiro Yokoyama
Journal:  Eur Heart J       Date:  2006-11-29       Impact factor: 29.983

7.  Optical coherence tomography assessment of the acute effects of stent implantation on the vessel wall: a systematic quantitative approach.

Authors:  N Gonzalo; P W Serruys; T Okamura; Z J Shen; Y Onuma; H M Garcia-Garcia; G Sarno; C Schultz; R J van Geuns; J Ligthart; E Regar
Journal:  Heart       Date:  2009-08-10       Impact factor: 5.994

8.  Predictors and outcomes of stent thrombosis: an intravascular ultrasound registry.

Authors:  Neal G Uren; S P Schwarzacher; J A Metz; D P Lee; Y Honda; A C Yeung; P J Fitzgerald; P G Yock
Journal:  Eur Heart J       Date:  2002-01       Impact factor: 29.983

9.  Incidence, predictors, morphological characteristics, and clinical outcomes of stent edge dissections detected by optical coherence tomography.

Authors:  Daniel Chamié; Hiram G Bezerra; Guilherme F Attizzani; Hirosada Yamamoto; Tomoaki Kanaya; Gregory T Stefano; Yusuke Fujino; Emile Mehanna; Wei Wang; Ahmad Abdul-Aziz; Matthew Dias; Daniel I Simon; Marco A Costa
Journal:  JACC Cardiovasc Interv       Date:  2013-07-17       Impact factor: 11.195

10.  Local determinants of thrombus formation following sirolimus-eluting stent implantation assessed by optical coherence tomography.

Authors:  Hiromasa Otake; Junya Shite; Junya Ako; Toshiro Shinke; Yusuke Tanino; Daisuke Ogasawara; Takahiro Sawada; Naoki Miyoshi; Hiroki Kato; Bon-Kwon Koo; Yasuhiro Honda; Peter J Fitzgerald; Ken-ichi Hirata
Journal:  JACC Cardiovasc Interv       Date:  2009-05       Impact factor: 11.195

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.